Abstract
Background
Creating multiple beam interference fringes on the retina using the "retina fringe projector" (RFP) provides us with a scale which can be used to determine the absolute size of objects at the fundus.
Methods
The procedure is tested on an model eye, the optical parameters of which are varied in order to simulate different refraction conditions. The RFP was used to determine the size of a known object at the model fundus; the result was then checked against the real value of the object. In a second series of measurements, the RFP fringes were included in fundus photos of 52 patients with varying levels of ametropia, taken at the University Eye Clinic in Jena using the FF450 fundus camera (manufacturer: Carl Zeiss Meditec AG) for diagnostic purposes. The distance between the fringes was determined using both the RFP and the Littmann method.
Results
The differences to the real value of the fundus size of a known object deviated on average by 1 to 2%, and always remained under 3% in the model eye. A very good correlation between RFP and Littmann methods results could be found in the clinical part of the study. The resulting deviations are below 2.5%.
Conclusion
Integration of a RFP in a fundus camera permits measurement of absolute values of the retina, regardless of the type of observation or documentation. Knowing the real size of fundus objects holds a potential in many clinical classification scales of fundus objects such as drusen or vessels, or in optimizing photodynamic therapy spot size measurement.
Similar content being viewed by others
References
Littmann H (1982) Zur Bestimmung der wahren Größe eines Objektes auf dem Hintergrund des lebenden Auges. Klin Monatsbl Augenheilk 180:286–289, Littmann I
Littmann H (1988) Zur Bestimmung der wahren Größe eines Objektes auf dem Hintergrund des lebenden Auges. Klin Monatsbl Augenheilk 192:66–67, Littmann II
Holden AL, Fitzke FW (1988) Image size in the fundus: structural evidence for wide-field retinal magnification factor. Brit J Ophthalmol 21:228–230
Bennett AG (1988) A method of determining the equivalent powers of the eye and its crystalline lens without resort to phakometry. Ophthal Physiol Opt 8:53–59
Bengtsson B, Krakau CET (1992) Correction of optic disk measurements on fundus photographs. Graefes Arch Clin Exp Ophthalmol 230:24–28
Bennett AG, Rudnicka AR, Edgar DF (1994) Improvements on Littmann`s method of determining the size of retinal features by fundus photography. Graefes Arch Clin Exp Ophthalmol 232:361–367
Barr DB (1995) An appraisal of the accuracy of Littmann`s method of determining the real dimensions of a retinal object. Acta Ophthalmol Scand Suppl 216:1–12
Garway-Heath DF, Rudnicka AR, Lowe T, Koster PJ, Fitzke FW, Hitchings RA (1998) Measurement of optic disc size: equivalence of methods to correct for ocular magnification. Brit J Ophthalmol 82:643–649
Le Grand Y (1935) Sur la mesure de l'acuité oculaire au moyen de franges d`interférence. CR Seance Acad Sci (Paris) 200:490
Campbell FW, Green DG (1965) Optical and retinal factors affecting visual resolution. J Physiol London 181:576
Rassow B, Wolf D (1973) Erfahrungen mit dem Laserinterferenzstreifentest bei der Messung des retinalen Auflösungsvermögens. Albrecht Von Graefes Arch Klin Exp Ophthalmol 187:61–66
Rassow B, Wolf D (1977) Die Messung der “retinalen Sehschärfe“ mit dem Laserinterferenzgerät als klinische Routinemethode. Adv Ophthalmol 34:116–142
Kennedy SJ, Schwartz B, Takamoto T, Eu JKT (1983) Interference fringe scale for absolute ocular fundus measurement. Invest Opthalmol Vis Sci 24:169–174
Baumbach P (1986) Zur absoluten Längenmessung am Augenhintergrund. Diplomarbeit im Fachbereich Physik, Universität Hamburg 1986
Baumbach P, Rassow B, Wesemann W (1989) Absolute ocular fundus dimensions measured by multiple-beam interference fringes. Invest Ophthalmol Vis Sci 30:2314–2319
Baumbach P, Rassow B (1986) Augenuntersuchungsgerät zur Messung topografischer Daten des Augenhintergrunds. Deutsches Patentamt DE 362 1893 A1 vom 01.07.1986
Rudnicka AR, Edgar DF, Bennett AG (1992) Construction of a model eye and its applications. Ophthalmic Physiol Opt 12:485–490
Dawczynski J, Mohr T, Koschmieder I, Rassow B, A.K. Schubert, Ach T, Strobel J (2008) Measurement of the Real Size of Fundus Objects With the Retina-Stripe-Projector (RSP). 2008 Annual Meeting, Eyes on Innovation, April 27-May 1, 2008, Fort Lauderdale/Florida, Poster 1438
Author information
Authors and Affiliations
Corresponding author
Additional information
Conflict of interest
None of the authors has any financial interest in any product or method mentioned in the text.
Rights and permissions
About this article
Cite this article
Dawczynski, J., Gebhardt, M., Mohr, T. et al. Determination of absolute size of fundus objects. Graefes Arch Clin Exp Ophthalmol 249, 381–387 (2011). https://doi.org/10.1007/s00417-010-1498-3
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00417-010-1498-3