Determination of absolute size of fundus objects

  • Basic Science
  • Published:
Graefe's Archive for Clinical and Experimental Ophthalmology Aims and scope Submit manuscript

Abstract

Background

Creating multiple beam interference fringes on the retina using the "retina fringe projector" (RFP) provides us with a scale which can be used to determine the absolute size of objects at the fundus.

Methods

The procedure is tested on an model eye, the optical parameters of which are varied in order to simulate different refraction conditions. The RFP was used to determine the size of a known object at the model fundus; the result was then checked against the real value of the object. In a second series of measurements, the RFP fringes were included in fundus photos of 52 patients with varying levels of ametropia, taken at the University Eye Clinic in Jena using the FF450 fundus camera (manufacturer: Carl Zeiss Meditec AG) for diagnostic purposes. The distance between the fringes was determined using both the RFP and the Littmann method.

Results

The differences to the real value of the fundus size of a known object deviated on average by 1 to 2%, and always remained under 3% in the model eye. A very good correlation between RFP and Littmann methods results could be found in the clinical part of the study. The resulting deviations are below 2.5%.

Conclusion

Integration of a RFP in a fundus camera permits measurement of absolute values of the retina, regardless of the type of observation or documentation. Knowing the real size of fundus objects holds a potential in many clinical classification scales of fundus objects such as drusen or vessels, or in optimizing photodynamic therapy spot size measurement.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

References

  1. Littmann H (1982) Zur Bestimmung der wahren Größe eines Objektes auf dem Hintergrund des lebenden Auges. Klin Monatsbl Augenheilk 180:286–289, Littmann I

    Article  CAS  Google Scholar 

  2. Littmann H (1988) Zur Bestimmung der wahren Größe eines Objektes auf dem Hintergrund des lebenden Auges. Klin Monatsbl Augenheilk 192:66–67, Littmann II

    Article  CAS  Google Scholar 

  3. Holden AL, Fitzke FW (1988) Image size in the fundus: structural evidence for wide-field retinal magnification factor. Brit J Ophthalmol 21:228–230

    Article  Google Scholar 

  4. Bennett AG (1988) A method of determining the equivalent powers of the eye and its crystalline lens without resort to phakometry. Ophthal Physiol Opt 8:53–59

    Article  CAS  Google Scholar 

  5. Bengtsson B, Krakau CET (1992) Correction of optic disk measurements on fundus photographs. Graefes Arch Clin Exp Ophthalmol 230:24–28

    Article  PubMed  CAS  Google Scholar 

  6. Bennett AG, Rudnicka AR, Edgar DF (1994) Improvements on Littmann`s method of determining the size of retinal features by fundus photography. Graefes Arch Clin Exp Ophthalmol 232:361–367

    Article  PubMed  CAS  Google Scholar 

  7. Barr DB (1995) An appraisal of the accuracy of Littmann`s method of determining the real dimensions of a retinal object. Acta Ophthalmol Scand Suppl 216:1–12

    PubMed  Google Scholar 

  8. Garway-Heath DF, Rudnicka AR, Lowe T, Koster PJ, Fitzke FW, Hitchings RA (1998) Measurement of optic disc size: equivalence of methods to correct for ocular magnification. Brit J Ophthalmol 82:643–649

    Article  CAS  Google Scholar 

  9. Le Grand Y (1935) Sur la mesure de l'acuité oculaire au moyen de franges d`interférence. CR Seance Acad Sci (Paris) 200:490

    Google Scholar 

  10. Campbell FW, Green DG (1965) Optical and retinal factors affecting visual resolution. J Physiol London 181:576

    PubMed  CAS  Google Scholar 

  11. Rassow B, Wolf D (1973) Erfahrungen mit dem Laserinterferenzstreifentest bei der Messung des retinalen Auflösungsvermögens. Albrecht Von Graefes Arch Klin Exp Ophthalmol 187:61–66

    Article  PubMed  CAS  Google Scholar 

  12. Rassow B, Wolf D (1977) Die Messung der “retinalen Sehschärfe“ mit dem Laserinterferenzgerät als klinische Routinemethode. Adv Ophthalmol 34:116–142

    PubMed  CAS  Google Scholar 

  13. Kennedy SJ, Schwartz B, Takamoto T, Eu JKT (1983) Interference fringe scale for absolute ocular fundus measurement. Invest Opthalmol Vis Sci 24:169–174

    CAS  Google Scholar 

  14. Baumbach P (1986) Zur absoluten Längenmessung am Augenhintergrund. Diplomarbeit im Fachbereich Physik, Universität Hamburg 1986

  15. Baumbach P, Rassow B, Wesemann W (1989) Absolute ocular fundus dimensions measured by multiple-beam interference fringes. Invest Ophthalmol Vis Sci 30:2314–2319

    PubMed  CAS  Google Scholar 

  16. Baumbach P, Rassow B (1986) Augenuntersuchungsgerät zur Messung topografischer Daten des Augenhintergrunds. Deutsches Patentamt DE 362 1893 A1 vom 01.07.1986

  17. Rudnicka AR, Edgar DF, Bennett AG (1992) Construction of a model eye and its applications. Ophthalmic Physiol Opt 12:485–490

    Article  PubMed  CAS  Google Scholar 

  18. Dawczynski J, Mohr T, Koschmieder I, Rassow B, A.K. Schubert, Ach T, Strobel J (2008) Measurement of the Real Size of Fundus Objects With the Retina-Stripe-Projector (RSP). 2008 Annual Meeting, Eyes on Innovation, April 27-May 1, 2008, Fort Lauderdale/Florida, Poster 1438

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jens Dawczynski.

Additional information

Conflict of interest

None of the authors has any financial interest in any product or method mentioned in the text.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Dawczynski, J., Gebhardt, M., Mohr, T. et al. Determination of absolute size of fundus objects. Graefes Arch Clin Exp Ophthalmol 249, 381–387 (2011). https://doi.org/10.1007/s00417-010-1498-3

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00417-010-1498-3

Keywords

Navigation