Skip to main content

Advertisement

Log in

Effect of corneal astigmatism on intraocular pressure measurement using ocular response analyzer and Goldmann applanation tonometer

  • Glaucoma
  • Published:
Graefe's Archive for Clinical and Experimental Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

To assess the effect of corneal astigmatism on intraocular pressure (IOP) measurements using an Ocular Response Analyzer (ORA) and a Goldmann applanation tonometer (GAT).

Methods

We prospectively examined 59 normal eyes of 59 healthy volunteers (18 men, 41 women; age, mean ± standard deviation, 40.5 ± 14.2 years; age range, 19–68 years). We quantitatively assessed the values of corneal-compensated intraocular pressure (IOPcc) and Goldmann-correlated intraocular pressure (IOPG) using an ORA (Reichert Ophthalmic Instruments). We also measured the IOP using a GAT (GAT-IOP). The amount of corneal astigmatism was assessed with an autokeratometer. We carried out these measurements three times, and the mean value obtained was used for statistical analysis.

Results

The mean IOPcc, IOPG, and GAT-IOP were 14.7 ± 2.6, 14.0 ± 2.8, and 14.2 ± 1.7 mmHg respectively. The mean corneal astigmatism was 0.94 ± 0.55 D. We found no significant correlation between IOPcc and corneal astigmatism (Pearson’s correlation coefficient r = −0.04, p = 0.79), or between IOPG and corneal astigmatism (r = 0.09, p = 0.52). However, we found a weak, but significant, correlation between GAT-IOP and corneal astigmatism (Pearson’s correlation coefficient r = 0.34, p = 0.009).

Conclusions

Both IOPcc and IOPG measured with ORA were less affected by the amount of corneal astigmatism, and the GAT-IOP readings were significantly higher in eyes with greater corneal astigmatism, suggesting that IOPcc as well as IOPG may be helpful for accurate IOP measurements in eyes with some corneal astigmatism.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

  1. Goldmann H, Schmidt T (1965) On applanation tonography. Ophthalmologica 150:65–75

    Article  CAS  PubMed  Google Scholar 

  2. Whitacre MM, Stein RA, Hassanein K (1993) The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 115:592–596

    CAS  PubMed  Google Scholar 

  3. Doughty MJ, Zaman ML (2000) Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 44:367–408

    Article  CAS  PubMed  Google Scholar 

  4. Feltgen N, Leifert D, Funk J (2001) Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol 85:85–87

    Article  CAS  PubMed  Google Scholar 

  5. Brandt JD, Beiser JA, Kass MA, Gordon MO (2001) Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS). Ophthalmology 108:1779–1788

    Article  CAS  PubMed  Google Scholar 

  6. Shimmyo M, Ross AJ, Moy A, Mostafavi R (2003) Intraocular pressure, Goldmann applanation tension, corneal thickness, and corneal curvature in Caucasians, Asians, Hispanics, and African-Americans. Am J Ophthalmol 136:603–613

    Article  PubMed  Google Scholar 

  7. Gunvant P, Baskaran M, Vijaya L et al (2004) Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer. Br J Ophthalmol 88:518–522

    Article  CAS  PubMed  Google Scholar 

  8. Broman AT, Congdon NG, Bandeen-Roche K, Quigley HA (2007) Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of intraocular pressure. J Glaucoma 16:581–588

    Article  PubMed  Google Scholar 

  9. Whitacre MM, Stein R (1993) Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 38:1–30

    Article  CAS  PubMed  Google Scholar 

  10. Holladay JT, Allison ME, Prager TC (1983) Goldmann applanation tonometry in patients with regular corneal astigmatism. Am J Ophthalmol 96:90–93

    CAS  PubMed  Google Scholar 

  11. Mark HH, Mark TL (2003) Corneal astigmatism in applanation tonometry. Eye 17:617–618

    Article  CAS  PubMed  Google Scholar 

  12. Rask G, Behndig A (2006) Effects of corneal thickness, curvature, astigmatism and direction of gaze on Goldmann applanation tonometry readings. Ophthalmic Res 38:49–55

    Article  PubMed  Google Scholar 

  13. Luce DA (2005) Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg 31:156–162

    Article  PubMed  Google Scholar 

  14. AlMubrad TM, Ogbuehi KC (2008) The effect of repeated applanation on subsequent IOP measurements. Clin Exp Optom 91:524–529

    Article  PubMed  Google Scholar 

  15. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310

    CAS  PubMed  Google Scholar 

  16. Medeiros FA, Weinreb RN (2006) Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer. J Glaucoma 15:364–370

    Article  PubMed  Google Scholar 

  17. Kotecha A, Elsheikh A, Roberts CR, Zhu H, Garway-Heath DF (2006) Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci 47:5337–5347

    Article  PubMed  Google Scholar 

  18. Lam A, Chen D, Chiu R, Chui WS (2007) Comparison of IOP measurements between ORA and GAT in normal Chinese. Optom Vis Sci 84:909–914

    Article  PubMed  Google Scholar 

  19. Martinez-de-la-Casa JM, Garcia-Feijoo J, Fernandez-Vidal A, Mendez-Hernandez C, Garcia-Sanchez J (2006) Ocular response analyzer versus Goldmann applanation tonometry for intraocular pressure measurements. Invest Ophthalmol Vis Sci 47:4410–4414

    Article  PubMed  Google Scholar 

  20. Kaufmann C, Bachmann LM, Thiel MA (2004) Comparison of dynamic contour tonometry with Goldmann applanation tonometry. Invest Ophthalmol Vis Sci 45:3118–3121

    Article  PubMed  Google Scholar 

  21. Saleh TA, Adams M, McDermott B, Claridge KG, Ewings P (2006) Effects of central corneal thickness and corneal curvature on the intraocular pressure measurement by Goldmann applanation tonometer and ocular blood flow pneumatonometer. Clin Experiment Ophthalmol 34:516–520

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Kazutaka Kamiya.

Additional information

Conflict of interest

The authors have no financial or proprietary interest in any materials mentioned herein.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Hagishima, M., Kamiya, K., Fujimura, F. et al. Effect of corneal astigmatism on intraocular pressure measurement using ocular response analyzer and Goldmann applanation tonometer. Graefes Arch Clin Exp Ophthalmol 248, 257–262 (2010). https://doi.org/10.1007/s00417-009-1202-7

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00417-009-1202-7

Keywords

Navigation