Introduction

In recent years, there has been growing use of X-ray fluoroscopy in clinical departments outside the main imaging departments (Rehani et al. 2010). Unfortunately, in these departments—at least in the gulf region—clinicians such as interventional cardiologists, electro-physiologists, vascular surgeons, urologists, orthopedic surgeons, gastroenterologists, to name a few, typically lack of training in radiation protection. This might pose some radiation-related risks to patients and staff. Recent studies have demonstrated eye lens opacities among 38 to 53% of main operators and 21 to 45% of nurses in cardiac catheterization (cath) labs (Rehani et al. 2011; Ciraj-Bjelac et al. 2010; 2012; Vano et al. 2010; 2016). These lens opacities have the potential to lead to cataracts later in life. The observation on the prevalence of eye lens opacities among cath lab staff prompted a number of research activities worldwide. For occupational exposure, the International Commission on Radiological Protection (ICRP) has recently reduced the annual dose limit to the eye lens by a factor of 7.5 from 150 to 20 mSv (Stewart et al. 2012). A recently published study indicates that 44% of interventional cardiologists would exceed the annual limit if eye protection tools are not utilized (Domienik-Andrzejewska et al. 2020).

Nearly 13.5 million interventional procedures were performed in the US in 2016, the number of fluoroscopy-guided cardiac intervention was 4.1 million (Mettler et al. 2020). In Kuwait, there are more than five thousand interventional procedures performed every year as per data of 2009 (Tsapaki et al. 2009). At the moment, no information is available on eye doses for staff working in interventional procedures in Kuwait, and there is a similar lack of information from neighboring countries in the region. With increasing emphasis on radiation-induced cataracts and reduction in threshold dose for the lens of the eye, there is a need to assess the radiation doses to staff working in interventional procedures in Kuwait. To meet this demand, the International Atomic Energy Agency (IAEA) has supported Kuwait through a technical cooperation (TC) project KUW9009 which aims at assessing radiation doses to eyes of medical and paramedical staff involved in interventional procedures in Kuwait. In addition, this TC project promotes optimization in radiological protection by providing the required scientific data and input for developing related guidelines for the healthcare personnel exposed to recurrent doses of ionizing radiation to the eye lenses.

Materials and methods

Under the framework of TC project KUW9009, a detailed work plan for dose measurements, data collection and analysis was developed. To that end, an IAEA expert mission was organised to evaluate the collected data and recommend protective measures for reducing eye radiation doses. The study was carried out in cardiac catheterization laboratories of three hospitals in Kuwait, namely, Chest Diseases Hospital, Al-Salam Hospital, and Mubarak Hospital. The study was approved by the ministry of health institutional ethical review board before the commencement of data collection. Data on technical parameters, workload, use of radiation protection devices, use of personal dosimeters and exposure history were collected. The study covered only percutaneous coronary intervention (PCI) procedures performed in the participating centers. Informed consent was obtained from all participants. Interventional cardiology procedures are usually performed by a team consisting of two cardiologists and a nurse/technician working in close proximity to the patient as shown in Fig. 1. The interventional cardiologist is the main operator who performs the procedure assisted by a second cardiologist and a nurse. During these cardiac procedures there is some variability in the work practice and position of the nurses compared to the positions of the cardiologists.

Fig. 1
figure 1

The usual positions of staff during percutaneous coronary intervention (PCI) procedures. The position of the ceiling suspended screen is also indicated but varies with the operator

The three sites included in the present study used a Philips BV Pulsera fluoroscopic unit with a typical range for the tube voltage of 40–110 kV. These units are subjected to regular quality control. Staff workload varies, but roughly 10–20 PCI procedures are performed weekly with a fluoroscopy time ranging between 8 and 12 min and a pulsed fluoroscopy set at 7.5 pulses per second and cine recording with a cine rate of 15 to 20 frames per second. The kerma-area-product (KAP) generated for each procedure using the KAP meter incorporated in the imaging system was also recorded. Ceiling suspended screens and table mounted lead curtains were used for radiation shielding. No protective lead spectacles were available. Simple statistical analysis was performed using Microsoft Excel (Microsoft Office Pro Plus 2019).

ED3 (Rotunda, USA) active personal extremity dosimeters were used for real-time dose monitoring. Currently, commercial electronic dosemeters for eye dose assessment are not available, and ED3 was the only electronic dosimeter available for the present study. The ED3 makes use of Silicon diodes to measure Hp(0.07) for a photon energy range from 60 keV up to 1.25 MeV with ± 30% angular response (Rotunda Scientific Technologies 2017). Until October 2019 there was no International Electrotechnical Commission (IEC) standard for pulsed X-ray testing and thus the calibration provided by the company was for gamma rays only. Two ED3 detectors were placed one each on the right and left side of the eyes using a head strap, or were clipped to the regular spectacles. The detectors, although calibrated for Hp(0.07), provided reasonable estimates of the doses delivered to the lens of the eyes, even though Hp(3) would be ideal (Behrens and Dietze 2010; Carinou et al. 2015; Vano et al. 2015 and Ginjaume et al. 2019). Specifically, the Hp(0.07) and the Hp(3) eye doses measured are comparable within 15% (Ciraj-Bjelac and Rehani 2014). The number of available ED3 dosimeters for this study was limited. Consequently, eye doses were measured separately for cardiologists and nurses in the three different hospitals leading to a total of 60 monitored procedures.

The eye lens doses were estimated by taking the median values of the occupational doses per procedure measured with the ED3 dosimeters in a sample of PCI procedures at the same hospital. The eye lens dosses were then normalized to patient doses by dividing measured Hp(0.07) to respective KAP values. Annual eye doses were estimated by multiplying the dose per procedure with the annual workload of PCI procedures. Finally, the results were compared to results from other countries.

Another objective of the KUW9009 project was to increase radiation protection awareness among interventional cardiology staff. Several training courses, seminars, webinars, and radiation safety expert site visits were organised. The key components of the implemented training actions were radiation protection, optimizing of work practice, eye lens doses monitoring and reduction. Training in radiation protection is the most desirable and successful action to reduce occupational doses (Vano et al. 2016). Additionally, for clinicians who were not interested in any technical details a simplified version of the training material was developed as a mobile app that provides a higher level of outreach than more detailed training material. The eye radiation dose app “ERD” developed as part of the present project includes a quick link to guidelines, training materials and an eye dose calculation tool. When clicking the “dose calculator” tab the app invites the user to select the type of procedure and simply add the annual workload, to get an estimate of the annual eye radiation dose. This is done by taking into account typical staff doses per procedure available at the IAEA radiation protection website (IAEA 2017). The results are displayed in a color-coded index to alert users when the level of eye lens radiation exposure is high compared to the ICRP dose limit. The app is freely available and can be downloaded at https://erd.com.kw.

Results

Table 1 shows the eye doses measured during PCI procedures in the investigated cardiac catheterization laboratories. For cardiologists the dose per procedure range from 18.5 to 30.3 µSv for the left eye and from 12.6 to 23.6 µSv for the right eye.

Table 1 Summary of measured eye doses per procedure in a sample of percutaneous coronary intervention (PCI) procedures at hospitals A, B, and C

The average fluoroscopy time, KAP, eye dose values and eye dose values normalized to the corresponding KAP during PCI procedures are presented in Table 2. The overall average value of KAP during the investigated PCI procedures for cardiologists was 82.6 Gy.cm2 (based on individual average values in three hospitals of 92.2, 76.9, 78.7). The ratio of main operator eye lens dose to KAP in the presence of ceiling suspended screen was 0.292 (range 0.235–0.329) µSv/Gy.cm2.

Table 2 Eye doses in the presence of ceiling suspended screen normalised to respective KAP for cardiologists

The estimated annual doses—based on a workload of 1,000 fluoroscopy guided PCI procedures performed yearly in Kuwait—are shown in Fig. 2. As a result, for the left eye lens dose a value of 30.3 mSv was estimated, while for the right eye lens dose a value of 23.6 mSv was estimated. Thus, the estimated annual eye dose received by cardiologists in Kuwait might exceed the dose limit recommended by ICRP and implemented in Kuwait.

Fig. 2
figure 2

Estimated annual eye lens doses, based on a typical annual workload of 1,000 percutaneous coronary intervention (PCI) procedures, for interventional cardiologists and nurses working at Chest Diseases Hospital compared to the annual dose limit of 20 mSv/y as recommended by ICRP and implemented in Kuwait

Table 3 compares eye doses reported in recent publications as compared to those obtained in the present study.

Table 3 Comparison of published data on average eye doses during percutaneous coronary intervention (PCI) procedures for interventional cardiologists

Discussion

This is the first study not only in Kuwait but also in the whole gulf region assessing eye doses to staff involved in percutaneous coronary interventions using active personal dosimeters. Eye lens doses were estimated based on typical workload of medical staff involved in those procedures. The results obtained indicate that doses to the lenses of the eyes of the cardiologists can potentially exceed the ICRP annual dose limit, which necessitates actions to reduce eye lens doses. Those actions could include wearing lead spectacles in addition to improving the placement of ceiling suspended screens, in addition to the optimization of techniques to reduce patient doses which will have direct implications on staff doses.

PCI procedures usually result in high-radiation exposure to patients and staff. The radiation dose to the lens of the eye varied widely in the present study which is not unexpected. This is mainly due to differences in staff skills, experience, radiation protection awareness and work environment. Table 1 show that the cardiologist dose during PCI procedures range between 18.5 and 30.3 µSv for the left eye and between 12.6 and 23.6 µSv for the right eye. When compared with other studies (Table 3) the present results are within the range observed by others and are rather in good agreement.

The results show that the doses to the left eye are higher than those to the right eye, which is expected as demonstrated by other investigators (Antic et al. 2013; Kim et al. 2008; Häusler et al. 2009). Lower doses to eye lens were found for nurses compared to cardiologists (Fig. 2), which is expected because nurses usually stand away from the X-ray beam at the patient foot end (Fig. 1). In contrast to the main operator, nurses have a greater degree of flexibility in terms of their location with respect to the patient, which can make a big difference in eye lens dose. In some situations, however, the nurses need to be very close to the patients during a cardiac procedure, which makes it difficult to protect them properly against the radiation. This could possibly explain why nurses in hospital B showed three times higher doses than those in the other hospitals (Table 1).

In interventional cardiology procedures the eye dose depends on many factors that usually vary during those procedures. The correlation between patient and staff eye doses has been studied in recent years (Antic et al. 2013). For example, it has been shown that typically 1 Gycm2 of KAP to the patient results in an estimated average dose to the lens of the eye of 10 μSv due to scatter dose at a C-arm (Leyton et al. 2016). By applying radiation protection measures, the eye dose due to scattered radiation can be reduced by up to 6%, at a C-arm (Vano et al 2016). Variations in fluoroscopy time and KAP were observed (Table 2), which were probably due to the complexity of the procedure and the experience of the cardiologist, because similar systems were used for all measurements. The eye lens doses to cardiologists in hospital A were higher than those in other hospitals. This is evidenced by the highest fluoroscopy time, which could be due to more complex procedures performed in hospital, a dedicated cardiac center, as compared to those performed in the other investigated hospitals. Variations of eye lens doses due to the influence of parameters like procedure complexity, technical parameters and proper positioning of ceiling suspended screens need further investigations.

Taking into account the typical annual workload of staff performing PCI procedures the estimated value of eye dose for interventional cardiologists can reach up to 30 mSv per year, highlighting the need for either additional lead spectacles or optimal placement of ceiling suspended screens and procedure parameters. Similar estimates made for nurses result in eye doses of about 3 mSv/year. For interventional cardiologists in Kuwait, doses to the lens of the eye thus exceed the annual eye dose limit.

Table 3 summarizes published data on eye doses during PCI procedures. The comparison demonstrates that the average dose per procedure for the main operator ranges from 10 μSv up to 300 μSv. This indicates that eye exposure levels associated with cardiology procedures vary in different countries. With current workloads and lack of proper radiation protective measures the annual dose limit is likely to be exceeded also in countries other than Kuwait.

Personal protective equipment is very effective in reducing eye doses, particularly when combining ceiling suspended screens with lead spectacles. The use of lead spectacles—with a thickness of 0.5 mm or 0.7 mm lead equivalence—has a small but meaningful effect on eye dose depending on photon energy, while other key parameters like beam projection, procedure geometry and correct use of available radiation protection equipment are more effective (Koukorava et al. 2014). Using ceiling suspended screens with lead spectacles can reduce eye lens dose by 83–90% (Matsubara et al. 2020), while using side shields or ‘wrap around’ eyewear can reduce the dose by 74% and 21% for the left and right eyes, respectively (Koukorava et al. 2014). When using leaded spectacles the impact of dosimeter position on the measured dose is great. The dosimeters should be very close to the eye and unshielded by the spectacles, to avoid underestimation of eye lens dose (Honorio da Silva et al. 2020). Eye dose calculation tools can be used to estimate the effectiveness of using radiation protection equipment and optimize the eye protection (Covens et al. 2018).

The lack of monitoring the occupational doses in interventional cardiology and the lack of the regular use of personal dosimeters and radiation protection tools has already been emphasized earlier in an international survey (Padovani et al. 2011). The situation is similar in Kuwait and in the gulf region. Indeed, there is a need to strengthen radiation protection awareness and occupational monitoring with emphasis to eye dose monitoring. For that reason, a mobile app “ERD” was developed as an awareness tool that can be used for optimisation of working practices and estimation of cumulative eye lens doses based on the number of performed procedures. The ERD app can be used for risk assessment and retrospective estimation of eye lens doses to health professionals who are likely to receive doses higher than 3/10th of the annual dose limit (Dauer et al. 2017).

The limitations of the present study include the following: one limitation is the lack of appropriate electronic active personal dosimeters for eye lens dose evaluation. The ED3 dosimeters, even though not specifically designed for eye lens dosimetry, is the only currently available electronic dosimeter that can be used for an estimation of eye lens doses. ED3 dosimeters are uncomfortable for the daily use during PCI procedures but are better than any passive dosimeters with head band. Also, the dose estimates in the present study do not take into account any other procedures the staff might be involved in. Even though the contribution of such other procedures to eye lens doses may be small, it will add up to the already high doses interventional cardiologists receive. Another limitation is the fact that the measurements were not performed simultaneously for cardiologist and nurses, due to the limited number of available dosimeters. In addition, the estimated workload used in the present study is based on an earlier publication from 2009 (Tsapaki et al. 2009); since then the number of PCI procedures has been increasing, and the current workload may be much higher. Finally, further studies on aspects identified in this study could not be pursued in view of the COVID-19 pandemic. For example, the current study would have benefited from a better assessment of the location of the nurse in different hospitals, because the preliminary information available did not allow to explain the high eye doses of nurses in hospital B as compared to those in the other two hospitals. Assessment of an appropriate placement of ceiling suspended screens is a difficult task. In addition, there is a lack of literature on this issue as most publications tend to just mention the importance of this aspect rather than assessing it in more detail.

Despite these limitations, it is clear that the present study has provided convincing evidence for the need for a better radiation protection of interventional cardiologists and associated staff in Kuwait.

Conclusions

In the present study, eye lens doses to interventional cardiology staff were measured for the first time in Kuwait and in gulf region as a whole. The doses per procedure for the left eye of the main operator when ceiling suspended screens were used ranged between 18.5 and 30.3 µSv for the left eye, and between 12.6 and 23.6 µSv for the right eye. The results show that the ICRP annual eye lens dose limit of 20 mSv can be exceeded for an annual workload of 1000 PCIs for interventional cardiologists. This demonstrates the need for additional radiation protection optimization measures. In addition, regular radiation protection training and awareness for interventional cardiology staff is needed to reduce eye lens doses.