Introduction

Boluses are tissue-equivalent materials with specific shape that can be applied directly to the patient’s skin. They are widely used in radiotherapy to alter the dose distribution of ionising radiation, especially during irradiation of superficial neoplastic lesions. Boluses provide additional absorption and scattering of ionising radiation, resulting in a more homogenous dose delivered to the area being treated (Hogstrom and Almond 2006; Łukowiak et al. 2017, 2016; Yan et al. 2018; Stone et al. 2003; Khan 2003).

In Poland, most boluses are created manually using paraffin wax to fit the patient’s surface. Other materials available in some centres are polymeric gels or ultrasound transmission gels. The present study concentrates on 3D-printed boluses filled with paraffin or water. In recent years, a method for the production of radiotherapy boluses using 3D printing technology has been introduced in the West Pomeranian Oncology Centre in Szczecin, and Acrylonitrile–butadiene–styrene terpolymer (ABS) has been selected as a bolus material. ABS-printed boluses were more precise compared with paraffin wax boluses that were manually created, and this helped to improve treatment planning (Łukowiak et al. 2017, 2016). The use of 3D printing allows for the production of customised phantoms with a shape designed to fit the patient’s therapeutic needs and anatomy. The 3D-printed boluses, when filled with water or paraffin wax, possess properties similar to those of irradiated biological tissues. Because of the method and the material used to produce these boluses, however, there is growing concern that irradiation can cause them to contract or change their mechanical properties.

ABS is a thermoplastic polymer commonly used in 3D printing, as it can be moulded after heat treatment. Moreover, this material slightly changes its size while solidifying (Hakimian and Sulong 2012; Rahimi et al. 2014; Sreedharan and Jeevanantham 2018). So far it remained an open question whether the ionising radiation used in radiotherapy affects the physical properties of such 3D-printed boluses. Consequently, the objective of the present study was to assess the suitability of ABS for bolus printing by measuring the changes in hardness and dimensions of 3D-printed ABS samples after exposure to 60 Gy X-rays. Changes in the hardness of a material can result in a greater tendency to deformations, and even small changes may negatively impact dose distribution in irradiated tissues.

Material and methods

Material hardness was evaluated using Vickers test, also referred to as the microhardness method. In this test, an indenter (diamond right square pyramid with a dihedral angle of 136°) is pressed onto the flat surface of the material with a specific force for a definite time (Fig. 1). The size of the resulting impression is measured with a microscope and used to determine the hardness of the material (Herrmann 2011; Chen et al. 2011; Shahdada et al. 2007; Franco et al. 2004; Asikuzun et al. 2012). The hardness number is determined by the ratio:

Fig. 1
figure 1

Schematic illustration of Vickers hardness test. d the average length of the diagonals d1 and d2 left by the indenter, F force applied to the indenter

$$HV= \frac{F}{S}\left[\frac{N}{{mm}^{2}}\right]$$
(1)

where F is the applied force and S is the surface area of the resulting indentation, which can be determined by the formula:

$$S=\frac{{d}^{2}}{1.854}\left[{\mathrm{mm}}^{2}\right]$$
(2)

where d is the average length of the diagonal left by the indenter. Hence:

$$HV=0.189 \frac{F}{{d}^{2}}\left[\frac{\mathrm{N}}{{\mathrm{mm}}^{2}}\right]$$
(3)

The samples printed for the microhardness test were designed as hollow of cuboids using the Fusion 360 software. The file was exported to stereolithography format, which is compatible with the 3D printer’s software, and then printed with ABS polymer on a Zortrax M200 printer working on FDM technology. The parameters used to print were: nozzle temperature—260 °C, bed temperature—75 °C, layer thickness—0.14 mm, infill pattern—2 shells. Supports have not been applied.

Four identical samples were printed using this procedure and named I, II, R1 and R2. R1 and R2 were reference samples. The samples were stored at room temperature until further use.

Prints I and R1 were used for microhardness testing. Before testing, the surface of the samples was smoothed using sandpapers with 1200, 2400 and 4000 grit sizes. Using an unprepared print to measure microhardness is not accurate, as its surface is not smooth enough. The hardness of samples I and R1 was measured 30 times each, using a DURAMIN-40 AC3 hardness tester (Struers, Germany) with a load of 10 g applied for 10 s each time.

The dimensions (x and y) of samples II and R2 were measured 30 times each, using an electronic calliper with accuracy 0.02 mm and resolution 0.01 mm.

Prints I and II were irradiated with a single dose of 60 Gy, delivered with a Siemens ARTISTE 3 linear accelerator emitting X-rays at 6 MV. The measurements of microhardness and dimensions were repeated immediately after and 45 days after exposure, since the average duration of radiation treatment is 45 days for all prints.

The distribution of the data was assessed with the Shapiro–Wilk test. For normally distributed data, the statistical significance of the differences between means was evaluated using Student’s t-test. For non-normally distributed data, the Mann–Whitney U test was used (Statistica 13). Differences between the means were considered statistically significant at p < 0.05.

To check the dose distribution within the boluses, two randomly chosen treatment plans were analysed. Using the treatment planning system (TPS), boluses were planned, one with the volume of 51.2 cm3 and another one with the volume of 146.9 cm3. Then, a bolus shrink of 1 mm was simulated. The differences in dose distribution between the planed and simulated boluses were investigated. The mean, maximum and minimum dose for a planned Clinical Target Volume (CTV) were checked.

Results

Tables 1, 2 and 3 present the statistics of the measured data. The changes in hardness (samples I and R1) were not statistically significant. The changes in dimensions for sample II were statistically significant only after 45 days (0.3% for x and 0.4% for y dimension). The measurements for reference sample R2 for dimension x were not statistically significant and for dimension y, the change was statistically significant only after 45 days after exposure, and was only 0.1% (0.03 mm). The p values for the comparison of the means are presented in Table 4. The results of the dose distribution comparison, for planned and simulated shrunken boluses are collected in Table 5.

Table 1 Vickers microhardness HV of samples Print I and reference sample Print R1, before irradiation and 1 and 45 days after irradiation with 60 Gy X-rays; SD standard deviation; ND normal distribution
Table 2 Changes in dimension x of samples Print II and reference sample Print R2, before irradiation and 1 and 45 days after irradiation with 60 Gy X-rays; SD standard deviation; ND normal distribution
Table 3 Changes in dimension y of samples Print II and reference sample Print R2, before irradiation and 1 and 45 days after irradiation with 60 Gy X-rays; SD standard deviation; ND normal distribution
Table 4 Differences in microhardness HV, dimension x and dimension y, for samples Print I and Print II, and reference samples Print R1 and Print R2; Differences between the means were considered statistically significant at p < 0.05
Table 5 Comparison of dose distributions, for two planned and simulated shrunken boluses, for a shrinkage value of 1 mm. ΔDmean, ΔDmax and ΔDmin are the absolute values of difference in mean, maximum and minimum dose between planned and simulated boluses

Discussion

This is a preliminary study, and further research will focus on performing a greater number of measurements using different irradiation parameters (type of radiation, dose, or beam energy). Based on the data obtained in the present study, it is suggested that even those changes in material hardness and dimensions that were statistically significant, are irrelevant to radiotherapy. Based on the results obtained, maximum shrinkage of the investigated boluses was 0.4% in all dimensions (Tables 2 and 3). The analysis of the planned TPS boluses shows that the maximum shrinkage, for a bolus with a length of 150 mm, should be 0.6 mm. However, a shrinkage of 1 mm, which is much greater than it should be, was simulated in the present study, because the TPS did not allow for the simulation of a smaller shrinkage. Even for a shrinkage of 1 mm, no significant differences in dose distribution were observed. For these changes to be relevant regarding treatment planning, the size of the print would have to be approximately 100 mm. Moreover, the changes in dimensions were observed 45 days after printing, not immediately after radiation exposure. Therefore, it could be that the observed reduction in size was not induced by radiation absorption, but represented just the natural contraction of the material. It can thus be suggested that only newly printed boluses should be used for treatment. However, more research is needed in this regard before a final conclusion can be drawn.

The ABS polymer is nontoxic and biologically inert with a density of 1.05 g/cm3 which is close to the density of soft tissue (1.05 g/cm3). Moreover, its atomic composition is similar to that of the human body. For these reasons, ABS seems to be an ideal material for bolus printing. The main disadvantages of using ABS are the fumes produced when printing and its non-biodegradability. Consequently, its use requires a proper ventilation system and generates a large amount of waste. Further work needs to be done to test other materials that could be used for printing boluses.

In conclusion, it is stated that ABS seems to be an appropriate material for creating 3D-printed boluses to be used in radiotherapy.