Abstract
In recent years, a method for designing radiotherapy boluses using 3D printing technology has been established in the West Pomeranian Oncology Centre in Szczecin, Poland. The aim of the present study was to investigate whether the ionising radiation used in radiotherapy affects the physical properties of the printing material. Particularly, the purpose of this study was to determine the effect of a 60 Gy X-ray radiation dose on the hardness and dimensions of 3D-printed boluses. Four cuboids were printed on a Zortrax M200 printer with acrylonitrile–butadiene–styrene (ABS) polymer. All printed samples were exposed to 60 Gy of X-ray radiation delivered by a medical accelerator. After irradiation, changes in the hardness (using Vickers test) and dimensions of the prints were measured. The therapeutic X-ray dose had a minimal effect on the dimensions of the printed samples, resulting in a maximum contraction of only 0.4%. Changes of the hardness were not statistically significant. In conclusion, regarding the radiotherapy planning process, the application of this therapeutic X-ray dose does not significantly influence the hardness and dimensions of ABS-printed boluses.
Avoid common mistakes on your manuscript.
Introduction
Boluses are tissue-equivalent materials with specific shape that can be applied directly to the patient’s skin. They are widely used in radiotherapy to alter the dose distribution of ionising radiation, especially during irradiation of superficial neoplastic lesions. Boluses provide additional absorption and scattering of ionising radiation, resulting in a more homogenous dose delivered to the area being treated (Hogstrom and Almond 2006; Łukowiak et al. 2017, 2016; Yan et al. 2018; Stone et al. 2003; Khan 2003).
In Poland, most boluses are created manually using paraffin wax to fit the patient’s surface. Other materials available in some centres are polymeric gels or ultrasound transmission gels. The present study concentrates on 3D-printed boluses filled with paraffin or water. In recent years, a method for the production of radiotherapy boluses using 3D printing technology has been introduced in the West Pomeranian Oncology Centre in Szczecin, and Acrylonitrile–butadiene–styrene terpolymer (ABS) has been selected as a bolus material. ABS-printed boluses were more precise compared with paraffin wax boluses that were manually created, and this helped to improve treatment planning (Łukowiak et al. 2017, 2016). The use of 3D printing allows for the production of customised phantoms with a shape designed to fit the patient’s therapeutic needs and anatomy. The 3D-printed boluses, when filled with water or paraffin wax, possess properties similar to those of irradiated biological tissues. Because of the method and the material used to produce these boluses, however, there is growing concern that irradiation can cause them to contract or change their mechanical properties.
ABS is a thermoplastic polymer commonly used in 3D printing, as it can be moulded after heat treatment. Moreover, this material slightly changes its size while solidifying (Hakimian and Sulong 2012; Rahimi et al. 2014; Sreedharan and Jeevanantham 2018). So far it remained an open question whether the ionising radiation used in radiotherapy affects the physical properties of such 3D-printed boluses. Consequently, the objective of the present study was to assess the suitability of ABS for bolus printing by measuring the changes in hardness and dimensions of 3D-printed ABS samples after exposure to 60 Gy X-rays. Changes in the hardness of a material can result in a greater tendency to deformations, and even small changes may negatively impact dose distribution in irradiated tissues.
Material and methods
Material hardness was evaluated using Vickers test, also referred to as the microhardness method. In this test, an indenter (diamond right square pyramid with a dihedral angle of 136°) is pressed onto the flat surface of the material with a specific force for a definite time (Fig. 1). The size of the resulting impression is measured with a microscope and used to determine the hardness of the material (Herrmann 2011; Chen et al. 2011; Shahdada et al. 2007; Franco et al. 2004; Asikuzun et al. 2012). The hardness number is determined by the ratio:
where F is the applied force and S is the surface area of the resulting indentation, which can be determined by the formula:
where d is the average length of the diagonal left by the indenter. Hence:
The samples printed for the microhardness test were designed as hollow of cuboids using the Fusion 360 software. The file was exported to stereolithography format, which is compatible with the 3D printer’s software, and then printed with ABS polymer on a Zortrax M200 printer working on FDM technology. The parameters used to print were: nozzle temperature—260 °C, bed temperature—75 °C, layer thickness—0.14 mm, infill pattern—2 shells. Supports have not been applied.
Four identical samples were printed using this procedure and named I, II, R1 and R2. R1 and R2 were reference samples. The samples were stored at room temperature until further use.
Prints I and R1 were used for microhardness testing. Before testing, the surface of the samples was smoothed using sandpapers with 1200, 2400 and 4000 grit sizes. Using an unprepared print to measure microhardness is not accurate, as its surface is not smooth enough. The hardness of samples I and R1 was measured 30 times each, using a DURAMIN-40 AC3 hardness tester (Struers, Germany) with a load of 10 g applied for 10 s each time.
The dimensions (x and y) of samples II and R2 were measured 30 times each, using an electronic calliper with accuracy 0.02 mm and resolution 0.01 mm.
Prints I and II were irradiated with a single dose of 60 Gy, delivered with a Siemens ARTISTE 3 linear accelerator emitting X-rays at 6 MV. The measurements of microhardness and dimensions were repeated immediately after and 45 days after exposure, since the average duration of radiation treatment is 45 days for all prints.
The distribution of the data was assessed with the Shapiro–Wilk test. For normally distributed data, the statistical significance of the differences between means was evaluated using Student’s t-test. For non-normally distributed data, the Mann–Whitney U test was used (Statistica 13). Differences between the means were considered statistically significant at p < 0.05.
To check the dose distribution within the boluses, two randomly chosen treatment plans were analysed. Using the treatment planning system (TPS), boluses were planned, one with the volume of 51.2 cm3 and another one with the volume of 146.9 cm3. Then, a bolus shrink of 1 mm was simulated. The differences in dose distribution between the planed and simulated boluses were investigated. The mean, maximum and minimum dose for a planned Clinical Target Volume (CTV) were checked.
Results
Tables 1, 2 and 3 present the statistics of the measured data. The changes in hardness (samples I and R1) were not statistically significant. The changes in dimensions for sample II were statistically significant only after 45 days (0.3% for x and 0.4% for y dimension). The measurements for reference sample R2 for dimension x were not statistically significant and for dimension y, the change was statistically significant only after 45 days after exposure, and was only 0.1% (0.03 mm). The p values for the comparison of the means are presented in Table 4. The results of the dose distribution comparison, for planned and simulated shrunken boluses are collected in Table 5.
Discussion
This is a preliminary study, and further research will focus on performing a greater number of measurements using different irradiation parameters (type of radiation, dose, or beam energy). Based on the data obtained in the present study, it is suggested that even those changes in material hardness and dimensions that were statistically significant, are irrelevant to radiotherapy. Based on the results obtained, maximum shrinkage of the investigated boluses was 0.4% in all dimensions (Tables 2 and 3). The analysis of the planned TPS boluses shows that the maximum shrinkage, for a bolus with a length of 150 mm, should be 0.6 mm. However, a shrinkage of 1 mm, which is much greater than it should be, was simulated in the present study, because the TPS did not allow for the simulation of a smaller shrinkage. Even for a shrinkage of 1 mm, no significant differences in dose distribution were observed. For these changes to be relevant regarding treatment planning, the size of the print would have to be approximately 100 mm. Moreover, the changes in dimensions were observed 45 days after printing, not immediately after radiation exposure. Therefore, it could be that the observed reduction in size was not induced by radiation absorption, but represented just the natural contraction of the material. It can thus be suggested that only newly printed boluses should be used for treatment. However, more research is needed in this regard before a final conclusion can be drawn.
The ABS polymer is nontoxic and biologically inert with a density of 1.05 g/cm3 which is close to the density of soft tissue (1.05 g/cm3). Moreover, its atomic composition is similar to that of the human body. For these reasons, ABS seems to be an ideal material for bolus printing. The main disadvantages of using ABS are the fumes produced when printing and its non-biodegradability. Consequently, its use requires a proper ventilation system and generates a large amount of waste. Further work needs to be done to test other materials that could be used for printing boluses.
In conclusion, it is stated that ABS seems to be an appropriate material for creating 3D-printed boluses to be used in radiotherapy.
Data availability
Data will be available from authors upon request.
References
Asikuzun E, Ozturk O, Cetinkara HA, Yildirim G, Varilci A, Yılmazlar M, Terzioglu C (2012) Vickers hardness measurements and some physical properties of Pr2O3 doped Bi-2212 superconductors. J Mat Sci: Mat Electronics 23(5):1001–1010. https://doi.org/10.1007/s10854-011-0537-0
Chen XQ, Niu H, Li D, Li Y (2011) Modeling hardness of polycrystalline materials and bulk metallic glasses. Intermetallics 19(9):1275–1281. https://doi.org/10.1016/j.intermet.2011.03.026
Franco AR Jr, Pintaúde G, Sinatora A, Pinedo CE, Tschiptschin AP (2004) The use of a vickers indenter in depth sensing indentation for measuring elastic modulus and vickers hardness. Mat Res 7(3):483–491. https://doi.org/10.1590/S1516-14392004000300018
Hakimian E, Sulong AB (2012) Analysis of warpage and shrinkage properties of injection-molded micro gears polymer composites using numerical simulations assisted by the Taguchi method. Mater Des 42:62–71. https://doi.org/10.1016/j.matdes.2012.04.058
Herrmann K (2011) Hardness measurement of plastic and elastomers. In: Hardness Testing: Principles and Applications. ASM International, pp 145–158
Hogstrom KR, Almond PR (2006) Review of electron beam therapy physics. Phys Med Biol 51:R455–R489. https://doi.org/10.1088/0031-9155/51/13/R25
Khan F (2003) Treatment verification. In: The physics of radiation therapy. 3rd ed. Phila-delphia: Lippincott Williams & Wilkins, pp 244–245
Łukowiak M, Boehlke M, Lewocki M, Kot W, Matias D, Piątek- Hnat M, ElFray M, Jezierska K, Podraza W (2016) Use of a 3D printer to create a bolus for patients undergoing tele-radiotherapy. Int J Radiat Res 14(4):287–295. https://doi.org/10.18869/acadpub.ijrr.14.4.287
Łukowiak M, Jezierska K, Boehlke M, Więcko M, Łukowiak A, Podraza W, Lewocki M, Masojć B, Falco M (2017) Utilization of a 3D printer to fabricate boluses used for electron therapy of skin lesions of the eye canthi. J Appl Clin Med Phys 18(1):76–81. https://doi.org/10.1002/acm2.12013
Rahimi M, Esfahanian M, Moradi M (2014) Effect of reprocessing on shrinkage and mechanical properties of ABS and investigating the proper blend of virgin and recycled ABS in injection molding. J Mater Process Technol 214(11):2359–2365. https://doi.org/10.1016/j.jmatprotec.2014.04.028
Shahdada SA, McCabea JF, Bullb S, Rusbya S, Wassella RW (2007) Hardness measured with traditional Vickers and Martens hardness methods. Dent Mater 23(9):1079–1085. https://doi.org/10.1016/j.dental.2006.10.001
Sreedharan J, Jeevanantham AK (2018) Analysis of shrinkages in abs injection molding parts for automobile applications. Mater Today-Proc 5(5). Part 2:12744–12749. https://doi.org/10.1016/j.matpr.2018.02.258
Stone HB, Coleman N, Anscher MS, McBride WH (2003) Effects of radiations on normal tissue: consequences and mechanism. Lancet Oncol 4:529–536. https://doi.org/10.1016/S1470-2045(03)01191-4
Yan H, Guo F, Zhu D, Stryker S, Trumpore S, Roberts K, Higgins S, Nath R, Chen Z, Liu Wu (2018) On the use of bolus for pacemaker dose measurement and reduction in radiation therapy. J Appl Clin Med Phys 19(1):125–131. https://doi.org/10.1002/acm2.12229
Acknowledgements
The authors would like to thank Iwona Żuk from the Department of Medical Physics of the Pomeranian Medical University in Szczecin and Marek Boehlke from the Department of Medical Physics of West Pomeranian Oncology Center, for their technical assistance.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Jezierska, K., Sękowska, A., Podraza, W. et al. The effect of ionising radiation on the physical properties of 3D-printed polymer boluses. Radiat Environ Biophys 60, 377–381 (2021). https://doi.org/10.1007/s00411-021-00892-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00411-021-00892-z