Abstract
Purpose
This study investigated whether an interaural delay, e.g. caused by the processing latency of a hearing device, can affect sensitivity to interaural level differences (ILDs) in normal hearing subjects or cochlear implant (CI) users with contralateral normal hearing (SSD-CI).
Methods
Sensitivity to ILD was measured in 10 SSD-CI subjects and in 24 normal hearing subjects. The stimulus was a noise burst presented via headphones and via a direct cable connection (CI). ILD sensitivity was measured for different interaural delays in the range induced by hearing devices. ILD sensitivity was correlated with results obtained in a sound localization task using seven loudspeakers in the frontal horizontal plane.
Results
In the normal hearing subjects the sensitivity to interaural level differences deteriorated significantly with increasing interaural delays. In the CI group, no significant effect of interaural delays on ILD sensitivity was found. The NH subjects were significantly more sensitive to ILDs. The mean localization error in the CI group was 10.8° higher than in the normal hearing group. No correlation between sound localization ability and ILD sensitivity was found.
Conclusion
Interaural delays influence the perception of ILDs. For normal hearing subjects a significant decrement in sensitivity to ILD was measured. The effect could not be confirmed in the tested SSD-CI group, probably due to a small subject group with large variations. The temporal matching of the two sides may be beneficial for ILD processing and thus sound localization for CI patients. However, further studies are needed for verification.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Patients supplied with either one or two cochlear implants (CIs) have limited access to monaural and binaural auditory cues compared to normal hearing (NH) listeners and, therefore, difficulties in localizing sounds. However, CI users with bilateral CIs or unilateral CI using a hearing aid on the contralateral ear (bimodal) are also able to localize sounds, but less precisely than NH or CI users with single-sided deafness (SSD) [1,2,3].
The most relevant cues for sound localization are interaural time differences (ITD) and interaural level differences (ILD). Sensitivities to ITDs and ILDs are highly correlated and part of a complex neural processing [4, 5]. Deteriorated ITD and ILD sensitivity occurs even with mild hearing loss [5, 6]. In NH subjects, low frequency ITD information dominates when available [4, 7], whereas CI users primarily use ILD information for sound localization [1, 8, 9]. In 2018, Prejban et al. described that SSD-CI subjects were able to lateralize sounds with ILDs presented via headphones [10].
In SSD-CI or bimodally fitted CI users, the different auditory inputs of both ears (acoustic vs. electric stimulation) diminish binaural processing [11]. For example, a hearing aid and a CI have different processing times until the auditory nerve is stimulated. In a study by Zirn and coworkers, it was shown that the time period from signal presentation to stimulation of the auditory nerve with a CI (manufacturer MED-EL, Innsbruck, Austria) is comparable to that of an NH ear for stimulation frequencies around 1 kHz [12]. Higher frequencies arrive earlier in the NH ear than with electric stimulation and lower frequencies arrive later in the NH ear, due to the mechanics of the traveling wave effect in the cochlea. Processing delays of other CI manufacturers are frequency-independent and in the range of 9 to 12.5 ms [13] and most hearing aids have processing delays between 3 and 11 ms [14]. Recent studies showed that sound localization can be improved by minimizing the interaural delays between electrical CI stimulation and acoustic stimulation by a hearing aid [15, 16]. The best results in sound localization of speech-shaped signals were achieved in SSD-CI subjects with a small additional delay of about 1 ms on the CI side [17].
However, it is still unclear, which binaural cue is improved by the temporal matching of the two sides. One hypothesis is that the adjusted temporal signals on both sides allow processing of more synchronized ITD information. However, since access to ITD cues is only limited in CI systems, one other hypothesis is that the better timing is beneficial for more accurate temporal integration of ILD cues. The effect of interaural delay on sensitivity to ILD cues has not been adequately investigated so far, even in NH subjects.
The aim of this study was to measure to which extend interaural processing delays affect ILD sensitivity in NH subjects or experienced SSD-CI subjects. For this purpose, the sensitivity to ILDs was determined and compared for different interaural delays. In addition, sound localization abilities were measured in both subject groups and results were compared with the results of ILD sensitivity.
Materials and methods
Subjects
10 SSD-CI users (mean age: 51 years, 6 female and 4 male) and 24 NH subjects (mean age: 28 years, 16 female and 8 male) participated in this study.
Only SSD-CI subjects with a mean contralateral pure-tone audiogram thresholds (pure-tone average PTA-4 of 0.5, 1, 2 and 4 kHz) better or equal 25 dB HL and without contralateral hearing aid were included in the study. The mean PTA-4 was 14.3 ± 5.5 dB HL. Individual pure-tone audiograms are visualized in Fig. 1. The demographic data of the SSD-CI subjects is shown in Table 1.
All SSD-CI participants received implants from the same manufacturer (MED-EL, Innsbruck, Austria) with Flex28 electrode arrays and used the stimulation strategy encoding temporal fine structure in the four most apical electrodes FS4. All subjects used behind the ear processors (OPUS2 or SONNET) and had at least one year of experience with their CI. The mean monosyllable speech perception score was 42.5% at 65 dB sound pressure level (Freiburg monosyllable score, FMS [18]). Measurements were made in a free field condition with contralateral masking (insert earphones, broadband noise of 60 dB SPL).
All NH subjects had a PTA-4 better than 15 dB HL and hearing thresholds lower than 30 dB HL for all octave frequencies between 250 Hz and 8 kHz. The individual interaural difference in PTA-4 was 10 dB HL or less. All SSD-CI subjects suffered from postlingual sensorineural hearing loss prior to CI implantation. In six subjects, deafness was caused by one or more sudden hearing losses, two suffered from Menière’s disease, one from head trauma and one from otitis media affecting the inner ear.
The study was approved by the local institutional review board (No. 213/16). All subjects gave their written consent and the SSD-CI subjects received financial compensation for participating in this study. The tests were conducted in accordance with the Code of Ethics of the World Medical Association (Declaration of Helsinki) for experiments with humans.
Loudness balancing for SSD-CI subjects
Prior to the ILD and localization task, the SSD-CI subjects had to perform an interaural loudness-balancing to match the perceived loudness in the CI ear with the contralateral ear. To ensure sufficient dynamic range, the volume level of the CI processor in the subject’s everyday program was set to at least 85% in the Maestro fitting software (MED-EL, Innsbruck, Austria). If the currently used volume level was below 85%, the most comfortable level (MCL) was lowered globally and the volume level was increased afterwards.
The stimuli in the CI ear were presented via auxiliary input of the sound processor driven by a headphone amp (Lake People, Konstanz, Germany) equipped with an analog volume potentiometer. The stimuli in the contralateral ear were presented using a headphone (Sennheiser HDA 200, Wedemark, Germany) driven by the integrated headphone-amp of the used audio interface and digital to analog converter RME Fireface UC (Audio AG, Haimhausen, Germany) and calibrated to a fixed sound pressure level of 65 dB SPL. All tasks were implemented with software MATLAB R2010a (MathWorks, Natick, United States).
The used stimulus was a broadband noise burst (bandpass filtered between 100 Hz and 15 kHz) with a duration of 500 ms and a time ramp of 20 ms. The stimulus was presented twice to the left ear and then twice to the right ear. Pauses in between stimuli were 200 ms. The subjects were asked whether the perceived loudness were the same on both sides. If one side was perceived louder, level was adjusted on the CI-side by the experimenter and the test was repeated unlimited times until a loudness considered as equal on both sides was reached. Afterwards, the loudness balance was rechecked with the same stimulus alternating between the left and right ear to obtain the final setting and potentially corrected again in level by the subject with the remote control until a loudness considered as equal was reached.
ILD sensitivity (just noticeable difference)
All tests were performed in an anechoic chamber (size: 4.1 × 2.6 × 2.1 m, reverberation time RT60: 0.05 s). The just noticeable differences (JND) of ILDs were measured using a two alternative forced choice method (2-AFC) with a 1up-2down-procedure to estimate the threshold of 70.7% correct [19]. The task was realized with the psylab-toolbox (version 2.8, Jade Hochschule Oldenburg, Germany, [20]). The initial ILD was 15 dB with a step size of 6 dB. The step size was halved after each reversal and the minimum step size was 0.75 dB. The test stopped after six reversals with minimum step size. A reversal is defined as local minimum or local maximum. The mean of the local minimum and maximum values of the JND task with minimum step size was defined as JND-ILD.
Two consecutive pairs of sound stimuli were presented dichotically to both ears. One pair had fixed ILDs and one pair had adaptively changing ILDs per trial. The order of the pairs was randomized. The subjects were asked to indicate the change in perceived sound position between the two signals (left–right or right-left). A level roving of ± 5 dB was implemented to avoid any monaural intensity cues.
To keep the perceived binaural loudness for each tested ILD constant, the level adjustments for each ear was based on the level calculation of an ideal panpot of a mixing desk [21]. ILD cues were presented as an asymmetrical level separation of the ILD information by increasing the level on one side by a smaller amount and decreasing the level of the other side by a larger amount. In the group of NH subjects, the signal was always lateralized towards the left ear, i.e. signals with ILDs had a higher level on the left side than on the right side. For the SSD-CI group the signal was always lateralized towards the normal hearing ear.
Each subject performed 8 trials with different test stimuli. Two different reference ILDs were tested (0 dB and 10 dB) with 4 different interaural delays each. For the reference ILD of 10 dB, the level on the left ear (or normal hearing ear in SSD-CI subjects) was increased by 5 dB, whereas the right ear (or CI in SSD-CI subjects) was decreased by 5 dB. To realize the interaural delay, the left ear in NH and the normal-hearing ear in SSD-CI subjects was delayed. For the NH subjects delays of 0, 5, 10 and 15 ms were tested to cover the range of processing delays potentially introduced by hearing aids or CIs [14]. The SSD-CI subjects were tested with delays of -5, 0, 5 and 10 ms, whereas 5 ms refers to a delay of the CI ear. Each subject completed an initial training trial with a reference ILD of 0 dB and no interaural delay.
Sound localization test
The sound localization was tested for seven frontal loudspeaker positions within ± 60° (0°, ± 21°, ± 42°, ± 59°). Each position was tested five times in random order. The sound stimuli were five pulsed broadband noise bursts (length 30 ms, 3 ms ramping, pause 70 ms), presented at 65 dB SPL with a level roving of ± 6 dB. The subjects were asked to fixate an initial active LED at 0° during stimulus presentation to avoid cues from head movements. The subjects indicated the perceived position by changing the active LED of a frontal LED array with a rotary encoder. The loudspeakers were hidden behind an acoustically transparent curtain to avoid visual cues. No feedback was given to the subjects during the test. Prior to testing, a training session was conducted to assure the correct handling. The setup is further described in earlier studies [22]. One SSD-CI subject (P10) and two NH subjects did not perform the sound localization task and are therefore excluded from the following analysis.
The localization error was defined as the difference between the perceived and the presented angle. To obtain a single value for localization accuracy, the mean localization error (MLE) was calculated as the median of the localization errors per angle. Additionally, the interquartile ranges per angle were calculated to estimate the uncertainty of sound localization.
Statistics
Boxplots and median values were used for descriptive analyses throughout the manuscript. Nonparametric tests were utilized for statistical analyses of ILD sensitivity and the sound localization accuracy. The Mann Whitney U-test was applied for pairwise group comparisons between SSD-CI subjects and NH. Statistical differences of JND-ILDs within the subject groups were determined with the Friedman test. Correlations were tested via Spearman rank correlation. A p-value < 0.05 was considered as significant and a Bonferroni correction of the p-value was performed in case of multiple pairwise comparison. IBM SPSS Statistics 27 (IBM, Armonik, New York) was used for the analysis.
Results
JND-ILD
All subjects were able to perform the test and to perceive a lateralized sound impression when ILD cues were presented. The JND-ILDs (reference ILD: 0 dB) for both subject groups are plotted in Fig. 2. The median JND-ILDs in the SSD-CI group varied from 4.9 dB (no delay) to 6.2 dB (10 ms delay). No significant effect of interaural delay on JND-ILDs was found in this subject group (Z = 1.788, p = 0.618). The median JND-ILD in the NH group varied between 1.1 dB (no delay) and 3.2 dB (15 ms delay). The JND-ILDs were significantly affected by an interaural delay at a reference ILD of 0 dB (Z = 40.190, p < 0.001). In a paired post-hoc comparison, significant differences between JND-ILDs were found between 0 and 10 ms (Z = -3.410, p = 0.004), 0 and 15 ms (Z = -5.981, p < 0.001) and between 5 and 15 ms (Z = -4.416, p < 0.001).
The NH subjects performed significantly better in all conditions (0 ms: Z = 4.202, p < 0.001; 5 ms: Z = 3.559, p < 0.001; 10 ms: Z = 3.576, p < 0.001). A few SSD-CI subjects yielded ILD sensitivities, which were comparable to NH subjects.
The results for a reference ILD of 10 dB are visualized in Fig. 3. The median JND-ILDs in the SSD-CI group varied from 6.7 dB (10 ms delay) to 11.4 dB (5 ms delay). No impact of interaural delay on JND-ILDs was found in this subject group (Z = 6.455, p = 0.091). The median JND-ILD in the NH group varied between 1.6 dB (no delay) and 2.7 dB (10 ms delay). The JND-ILDs were significantly affected by an interaural delay at a reference ILD of 10 dB (Z = 7.826, p = 0.050). The JND-ILD with no delay was significantly lower than the JND-ILD for 15 ms delay (Z = -2.739, p = 0.037). The NH subjects performed significantly better in all conditions (0 ms: Z = 3.406, p < 0.001; 5 ms: Z = 4.027, p < 0.001; 10 ms: Z = 3.479, p < 0.001).
JND-ILDs compared to a reference ILD of 0 dB increased in both subject groups. For the NH subjects, the increase was largest but non-significant for 0 and 5 ms (0 ms: Z = − 2.227, p = 0.104; 5 ms: Z = − 2.435, p = 0.06), for the SSD-CI subjects the increase in JND-ILD was significant for 5 ms (5 ms: Z = − 2.803, p = 0.02). It should be noted that interquartile ranges in the SSD-CI group considerably increased for the reference ILD of 10 dB compared to 0 dB, especially for a delay of -5 ms.
For all tested interaural delays, no correlation between age and ILD sensitivity was found in both subject groups.
Sound localization test
Boxplots of the localization error are shown in Fig. 4. The median localization error was 12.6° for the SSD-CI subjects (range: 1.6–18.7°) and for the NH-subjects the median localization error was 1.8° (range: 0.8–5.2°). The NH subjects performed significantly better than the SSD-CI subjects (Z = 3.375, p < 0.001). In both subject groups, no correlation was found between the results of the sound localization task and the JND-ILD task with a reference ILD of 0 dB (SSD-CI: rs = − 0.200, NH: rs = − 0.150) and with a reference ILD of 10 dB (SSD-CI: rs = − 0.503, NH: rs = 0.297).
Interquartile ranges (IQR) of localization task were used to analyze the uncertainty of sound localization in the different subject groups. IQR were separately analyzed for more central and smaller (± 21° and 0°) and more lateral and larger (± 42° and ± 59°) angles of sound incidence (Fig. 5). While no difference in IQR was detectable in the NH subjects, the SSD-CI subjects had a more reliable sound localization for the larger angles than for the smaller angles (NH: Z = -1.057, p = 0.290; SSD-CI: Z = − 1.886, p = 0.059).
Discussion
Sensitivity to interaural level differences was measured in 10 SSD-CI subjects and 24 NH subjects for broadband noise pulses presented via headphones and via a direct cable connection into the CI. The impact of interaural delays on ILD sensitivity was investigated for two different reference ILDs. Results were correlated with localization accuracy for frontally presented signals (broadband noise).
NH subjects showed decreasing ILD sensitivity with increasing interaural delay for both reference ILDs. In the SSD-CI group, no significant impact of interaural delay on ILD sensitivity was found, possibly due to the high individual variability in the results. Another aspect could be that generally worse ILD sensitivity in this subject group hinder the detection of the (even in the NH relatively small) JND differences induced by interaural delays. The median JND-ILD for SSD-CI subjects varied between 4.7 and 11.3 dB and was significantly higher than for the NH group. For a reference ILD of 0 dB, some of the SSD-CI subjects were able to achieve ILD sensitivities in the range of NH.
This is the first study to investigate the impact of interaural processing latencies on ILD sensitivity in NH and SSD-CI subjects.
ILD sensitivity
Although several studies investigating ILD sensitivity in NH subjects were published, there is currently nearly no data available discussing ILD sensitivity in CI users. The results obtained in experiments measuring JNDs for ILD sensitivity are highly dependent on the experimental design (e.g. type of stimulus, test procedure, etc.). Therefore, results of different studies are only comparable to a limited extend. In the present study, a short broadband stimulus was used, whereas some other studies with NH subjects used narrowband noise signals with different center frequencies [23]. In some other studies an AFC method was used [24], whereas in different studies subjects were asked to estimate the lateralized side for signals with defined ILDs [10, 23].
In a study by Ochi et al. in 2014 [24], sensitivity thresholds for ILDs and ITDs were measured in NH subjects for low- and high-frequency signals (duration 400 ms). The JND-ILD ranged from 1 to 4 dB for the low-frequency signals and from 1 to 5 dB for the high-frequency signals [24]. Spencer and colleagues also measured JND-ILDs in NH subjects for low- and high-frequency signals and obtained mean JND-ILDs of 1.9 dB [25], similar to Bernstein and colleagues with a JND-ILD of 2 dB [6]. However, Bernstein et al. used signals with center frequencies of 500 Hz or 4 kHz rather than broadband noise signals. Although the experimental design in our study was completely different, the median JND-ILD in this study of 1.1 dB for NH subjects obtained in our study is in line with previous studies. Since broadband noise was used in our study, this may explain the slightly better ILD sensitivity compared to the results of other studies.
Only a limited number of studies investigated on the ILD sensitivity in SSD-CI subjects so far. In a study published by Prejban and colleagues in 2018 [10] lateralization abilities of SSD-CI subjects based on ILDs were measured for ILDs between 0 and 24 dB with a step size of 4 dB. Since some of the SSD-CI subjects analyzed in their study were able to lateralize signals with ILDs of 4 dB, a sensitivity threshold in the range of 4 dB can be assumed, which is in the range of the results of our study obtained for a reference ILD of 0 dB without interaural delay.
In a study by Francart and colleagues in 2008, JND-ILD was estimated in bimodally fitted subjects by measuring psychometric functions [26]. The estimated median JND-ILD was 2 dB and, thus, better than the results of SSD-CI subjects found in our study with or without interaural delay. However, they stimulated the CI and the hearing aid directly (after conducting a pitch matching procedure) and measured the loudness growth function by loudness balancing both sides for different levels.
The SSD-CI subjects performed significantly worse than NH in all ILD tasks. One possible explanation is the reduced auditory input with a CI combined with a diminished binaural integration of the auditory information. It should also be considered that the acoustic and the electric inputs had different and unmatched dynamic ranges. With an adapted or enhanced electrical dynamic range, as performed by Gajecki and Nogueira 2021, the ILD sensitivity could be improved for CI subjects [27]. Also, different loudness growth functions for acoustic and electric stimulation might hinder better ILD sensitivity in SSD-CI subjects [26]. Another aspect that could account for different results of NH and SSD-CI subjects was the age difference in both subjects groups. The SSD-CI group was significantly older than the NH subject group. Several studies reported declining sensitivity to binaural cues with age [28, 29], but they focused mainly on the temporal integration processing. An early study by Herman et al. in 1977 assumed that age had no effect on level-based lateralization cues, but only on temporal cues [30]. However, because of the small subject groups and age ranges the results of our study do not allow us to draw conclusions about age effects on JND-ILDs.
Impact of processing delays
While interaural delays decreased ILD sensitivity in NH subjects, no significant effects were found in SSD-CI subjects. NH results are in line with a study from 2015 from Kong et al., where detrimental effect of interaural delays on binaural processing was found [31]. Angermeier and colleagues reported recently that an additional interaural delay deteriorates speech perception in noise and decreases spatial release from masking [32]. They measured speech perception in noise with varying interaural delays from 0 to 7 ms and were obtained in NH subjects.
To the authors’ knowledge, the impact of interaural delays on ILD sensitivity in CI subjects was not investigated so far. Some previous studies measured the effects of interaural delays on sound localization accuracy in SSD-CI users [17, 33]. Zirn and coworkers [33] delayed the CI to compensate for the hearing aid delay in bimodally fitted subjects, whereas Seebacher et al. [17] investigated whether a small additional delay of the CI (delay values below 4 ms) could improve the localization accuracy. In both studies, the temporal matching of both sides resulted in improved localization accuracy. Another study found no benefit in sound localization by matching the CI to the hearing aid delay [2]. In our study no significant correlation was found between JND-ILD and sound localization errors. However, in our study a large range of processing delays with rather large step size was tested in the JND task.
In our study, no significant impact of the interaural delay on the ILD sensitivity was found for the SSD-CI subjects. However, trend towards decreased ILD sensitivity for higher interaural delays were shown. Potential factors that may have reduced an effect were the small sample of subjects and the large inter-individual variances in JND-ILD results. It is also possible that the high compression of CI stimulation minimizes sensitivity to ILDs. Based on the frequency-specific delays in CI measured by Zirn and coworkers, it may be relevant to compensate the interaural delay in a frequency-specific manner [11, 12].
Correlation to localization accuracy
No correlation was found between JND-ILD and localization accuracy in both subject groups. In the NH group, this might be explained by the very good (i.e. low) JND-ILD for all subjects. The differences in ILD between the seven tested loudspeakers is higher than measured JND-ILD. Furthermore, NH subjects are relying predominantly on ITD cues, which are also present in the broadband noise of the localization task.
In the SSD-CI group, ILD information is shown to be dominant for sound localization since high-pass and broadband signals can be localized more reliably than low pass signals [1]. Due to the individual variability in JND-ILDs, a correlation with sound localization ability could be assumed. However, no correlation was found. A different explanation could be the measurement method used for sound localization. The results of sound localization are highly dependent on the experimental task and the used method of the analysis of the results. In our study, the mean localization error was used which is a single measure for sound localization averaged over all tested angles of sound incidence. A different measure is the root mean square (RMS) value. In a study reporting the RMS value, localization abilities in bilateral CI users were described with a mean RMS value of 20.4° [9]. Furthermore, since the difference of angular placement of tested loudspeakers was quite high, high ILD cues were available in the localization task. In the SSD-CI group it was shown in the analysis of interquartile ranges of the localization results, that these were more certain in the identification of the position of (i.e. higher ILDs) of sound incidence compared to lower ILDs. For a potential correlation with ILD sensitivity, it would be beneficial to test for a minimum audible angle instead of localization ability for different loudspeaker positions.
Another potential explanation for the lack of correlation is the small number of SSD-CI subjects in this study. Further studies are necessary to investigate the relation between sound localization and ILD sensitivity in CI users.
Conclusion
A detrimental effect of interaural delay on ILD sensitivity was found in NH. The effect could not be confirmed in the tested SSD-CI group. Based on this result, the beneficial effects of binaural temporal matching shown in other studies could not be explained by improved ILD sensitivity.
Data availability
Data available on request from the authors.
References
Dorman MF, Zeitler D, Cook SJ et al (2015) Interaural level difference cues determine sound source localization by single-sided deaf patients fit with a cochlear implant. Audiol Neurootol 20:183–188. https://doi.org/10.1159/000375394
Zaleski-King A, Goupell MJ, Barac-Cikoja D et al (2019) Bimodal cochlear implant listeners’ ability to perceive minimal audible angle differences. J Am Acad Audiol 30:659–671. https://doi.org/10.3766/jaaa.17012
Dorman MF, Loiselle LH, Cook SJ et al (2016) Sound source localization by normal-hearing listeners, hearing-impaired listeners and cochlear implant listeners. Audiol Neurootol 21:127–131. https://doi.org/10.1159/000444740
Wightman FL, Kistler DJ (1992) The dominant role of low-frequency interaural time differences in sound localization. J Acoust Soc Am 91:1648–1661. https://doi.org/10.1121/1.402445
Koehnke J, Culotta CP, Hawley ML et al (1995) Effects of reference interaural time and intensity differences on binaural performance in listeners with normal and impaired hearing. Ear Hear 16:331–353. https://doi.org/10.1097/00003446-199508000-00001
Bernstein LR, Trahiotis C (2019) No more than “slight” hearing loss and degradations in binaural processing. J Acoust Soc Am 145:2094. https://doi.org/10.1121/1.5096652
Macpherson EA, Middlebrooks JC (2002) Listener weighting of cues for lateral angle: the duplex theory of sound localization revisited. J Acoust Soc Am 111:2219–2236. https://doi.org/10.1121/1.1471898
Dirks C, Nelson P, Sladen DP et al. (2019) Mechanisms of localization and speech perception with colocated and spatially separated noise and speech maskers under single-sided deafness with a cochlear implant. Ear Hear:1293–1306. https://doi.org/10.1097/AUD.0000000000000708
Dorman MF, Loiselle L, Stohl J et al (2014) Interaural level differences and sound source localization for bilateral cochlear implant patients. Ear Hear 35:633–640. https://doi.org/10.1097/AUD.0000000000000057
Prejban DA, Hamzavi J-S, Arnoldner C et al (2018) Single sided deaf cochlear implant users in the difficult listening situation: speech perception and subjective benefit. Otol Neurotol 39:e803–e809. https://doi.org/10.1097/MAO.0000000000001963
Pieper SH, Hamze N, Brill S et al. (2022) Considerations for fitting cochlear implants bimodally and to the single-sided deaf. Trends Hear 26. https://doi.org/10.1177/23312165221108259
Zirn S, Arndt S, Aschendorff A et al (2015) Interaural stimulation timing in single sided deaf cochlear implant users. Hear Res. https://doi.org/10.1016/j.heares.2015.08.010
Wess J, Brungart DS, Bernstein JG et al (2017) The effect of interaural mismatches on contralateral unmasking with single-sided vocoders. Ear Hear 38:374–386. https://doi.org/10.1097/AUD.0000000000000374
Dillon H, Keidser G, OʼBrien A et al. (2003) Sound quality comparisons of advanced hearing aids. The Hearing Journal 56:30. https://doi.org/10.1097/01.HJ.0000293908.50552.34
Zirn S, Angermeier JM, Wesarg T (2018) Impact of the interaural stimulation timing mismatch on localization performance in bimodal HA/CI users. J Acoust Soc Am 143:1939. https://doi.org/10.1121/1.5036343
Angermeier J, Hemmert W, Zirn S (2021) Sound localization bias and error in bimodal listeners improve instantaneously when the device delay mismatch is reduced. Trends Hear 25:23312165211016164. https://doi.org/10.1177/23312165211016165
Seebacher J, Franke-Trieger A, Weichbold V et al (2018) Improved interaural timing of acoustic nerve stimulation affects sound localization in single-sided deaf cochlear implant users. Hear Res 371:19–27. https://doi.org/10.1016/j.heares.2018.10.015
Hahlbrock K-H (1953) Über Sprachaudiometrie und neue Wörterteste (Speech audiometry and new word-tests). Arch Ohren Nasen Kehlkopfheilkd 162:394–431. https://doi.org/10.1007/BF02105664
Levitt H (1971) Transformed up-down methods in psychoacoustics. J Acoust Soc Am 49:467–477. https://doi.org/10.1121/1.1912375
Hansen M Lehre und Ausbildung in Psychoakustik mit psylab: freie Software für psychoakustische Experimente (Teaching and training in psychoacoustics with psylab: free software for psychoacoustic experiments). In: DAGA 2006, Braunschweig, pp 591–592
Eberhard Sengpiel—sengpielaudio Das theoretische ideale Panpot—Werteberechnung (The theoretical ideal panpot—value calculation). http://www.sengpielaudio.com/DasTheoretischeIdealePanpot-Werte.pdf. Accessed 12 July 2018.
Körtje M, Eichenauer A, Stöver T et al (2022) Impact of reverberation on speech perception and sound localization accuracy in cochlear implant users with single-sided deafness. Otol Neurotol 43:e30–e37. https://doi.org/10.1097/MAO.0000000000003335
Brown AD, Tollin DJ (2021) Effects of interaural decoherence on sensitivity to interaural level differences across frequency. J Acoust Soc Am 149:4630. https://doi.org/10.1121/10.0005123
Ochi A, Yamasoba T, Furukawa S (2014) Factors that account for inter-individual variability of lateralization performance revealed by correlations of performance among multiple psychoacoustical tasks. Front Neurosci 8:27. https://doi.org/10.3389/fnins.2014.00027
Spencer NJ, Hawley ML, Colburn HS (2016) Relating interaural difference sensitivities for several parameters measured in normal-hearing and hearing-impaired listeners. J Acoust Soc Am 140:1783. https://doi.org/10.1121/1.4962444
Francart T, Brokx J, Wouters J (2008) Sensitivity to interaural level difference and loudness growth with bilateral bimodal stimulation. Audiol Neurootol 13:309–319. https://doi.org/10.1159/000124279
Gajecki T, Nogueira W (2021) Enhancement of interaural level differences for bilateral cochlear implant users. Hear Res 409:108313. https://doi.org/10.1016/j.heares.2021.108313
Ross B, Fujioka T, Tremblay KL et al (2007) Aging in binaural hearing begins in mid-life: evidence from cortical auditory-evoked responses to changes in interaural phase. J Neurosci 27:11172–11178. https://doi.org/10.1523/JNEUROSCI.1813-07.2007
Füllgrabe C (2013) Age-dependent changes in temporal-fine-structure processing in the absence of peripheral hearing loss. Am J Audiol 22:313–315. https://doi.org/10.1044/1059-0889(2013/12-0070)
Herman GE, Warren LR, Wagener JW (1977) Auditory lateralization: age differences in sensitivity to dichotic tima and amplitude cues. J Gerontol:187–191. https://doi.org/10.1093/geronj/32.2.187
Kong L, Xie Z, Lu L et al. (2015) Similar impacts of the interaural delay and interaural correlation on binaural gap detection. PLOS One 10:e0126342. https://doi.org/10.1371/journal.pone.0126342
Angermeier J, Hemmert W, Zirn S (2022) Measuring and modeling cue dependent spatial release from masking in the presence of typical delays in the treatment of hearing loss. Trends Hear 26:23312165221094200. https://doi.org/10.1177/23312165221094202
Zirn S, Angermeier J, Arndt S et al (2019) Reducing the device delay mismatch can improve sound localization in bimodal cochlear implant or hearing-aid users. Trends Hear 23:2331216519843876. https://doi.org/10.1177/2331216519843876
Acknowledgements
The authors would like to thank Mara Wolter for conducting several of the measurements.
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors disclose no conflicts of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.