Experimental study of bone-implant contact with a parabolic acetabular component (Hofer-Imhof)

  • Original article
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Incomplete bone contact with threaded acetabular components is usually attributable to the tapered thread design and sclerotic bone, and might be the reason for early loosening. The self-tapping flat threads of the Hofer-Imhof cup allow accurate insertion without incarceration, even in sclerotic bone. To demonstrate the effectiveness of this threaded cup design based on digitized measurements of bone contact area, six Hofer-Imhof cups were implanted into three human cadavers. The bone contact area at the threads was digitally analyzed on cadaver-explanted cups cut into horizontal and coronal sections. In cups fixed with ideal form fit, bone contact was 31.6% (range 8–55%) at the threads and 28.5% (range 0–49%) between the threads on average. In cups with incomplete bone contact secondary to insufficient reaming, the bone contact area averaged 27.4% (range 10–44%) at the threads. In sclerotic bone, the highest bone contact at the threads was 69%. The threads were the first part of the cup to achieve primary bone contact. Factors contributing to poor contact were incomplete reaming, sections which passed through the incisura of the acetabulum, and incomplete contact with the medial aspect of the acetabulum. The aim was to achieve complete bone contact. Because of the higher contact areas observed in sclerotic bone, some sclerotic bone should remain after reaming.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: 10 February 1999

Rights and permissions

Reprints and permissions

About this article

Cite this article

Effenberger, H., Böhm, G., Huber, M. et al. Experimental study of bone-implant contact with a parabolic acetabular component (Hofer-Imhof). Arch Orth Traum Surg 120, 160–165 (2000). https://doi.org/10.1007/s004020050035

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s004020050035

Keywords

Navigation