Abstract
Background
Liner malseating is well described in ceramic-on-ceramic total hip arthroplasties (THAs). However, limited information is known on this complication among dual-mobility articulations. As such, this systematic review analyzed liner malseating in dual-mobility THAs concerning prevalence, clinical implications, and associated risk factors.
Methods
A PRISMA criteria-based systematic review was performed, and PubMed, Web of Science, MEDLINE, and Cochrane used as data bases. All original studies from 1980 to 2022 were considered eligible for inclusion, and Methodological Index for Nonrandomized Studies (MINORS) used for quality assessment.
Results
In total, five retrospective cohort studies with 2330 patients (2673 dual-mobility THAs) were included. Mean age was 66.9 years, mean BMI was 29.8 kg/m2, and 35% of patients were female. Rates of malseating ranged from 0.15% to 5.8%, with a total of 53 malseated liners identified throughout all studies (1.98%). Based on THA manufacturer, malseating occurred in 48 Stryker (1.96%) and 5 Biomet Zimmer (2.14%) THAs. Mean clinical follow-up was 2.2 years (mean range, 1.3 to 6.4 years). Except one patient reporting of pain at 2 years, no revision or negative clinical implication was noted in any of the malseated liners, including normal ranged metal ions measured in four cases. A smaller acetabular component size was identified as a statistically significant risk factor for malseating in one study. Mean MINORS score was 9.8.
Conclusions
Liner malseating is a rare finding in patients undergoing THAs with dual-mobility articulations. While prelim results demonstrate no negative clinical consequences to date, existing studies are limited, refer to short-term outcomes only, and do not prospectively follow-up affected patients.
Level of evidence: IV.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Background
Total Hip Arthroplasty (THA) is widely considered the surgery of the century [1] with number of primary THAs projected to increase by 71% annually through 2030, in the United States alone [2]. However, dislocation remains one of the most common complications with up to 2% of THAs experiencing prosthetic hip dislocation within the first postoperative year [3]. The dual-mobility articulation, developed in 1976 by French physician Giles Bosquet, has become increasingly popular in recent decades [4], allowing for reduced rates of dislocation compared to conventional designs, while demonstrating implant survivorship rates of over 90% at midterm [5].
Nonetheless, dual-mobility designs are not without complications, including intraprosthetic dislocation, metallosis, liner fracture, and liner malseating [6]. The latter has been described in detail in the context of ceramic-on-ceramic THAs with malseating rates ranging over 20% [7], while resulting in significant clinical complications [8]. However, limited is known on liner malseating among dual-mobility THAs, as well as potential clinical implications [6]. As such, we analyzed liner malseating in dual-mobility THAs concerning overall malseating rate, patient characteristics, clinical and functional implications, as well as associated risk factors in the first systematic review to date.
Patients and methods
The systematic review was performed on the basis of the PRISMA criteria (Preferred Reporting Items for Systematic Reviews and Meta-analyses) [9], and PubMed, Web of Science, Ovid Medline, and Cochrane used as data bases. Search criteria were set as: “(liner malseating OR dual-mobility liner malseating OR liner malseating hip arthroplasty)”. Inclusion criteria were: (1) Patients treated with a dual-mobility THA, (2) between 1980 and 2022, (3) and evaluation for liner malseating. Exclusion criteria were: (1) non-dual-mobility THAs, (2) experimental studies, and (3) non-English full texts. The search was performed by two independent reviewers (DK, HB). Duplicated search results were removed, and the remaining articles analyzed based on title, and if considered eligible as full text.
The quality assessment was performed using the Methodological Index for Nonrandomized Studies (MINORS) score independently applied by the same two reviewers, and a final score between 0 and 16 reached by consensus [10]. In addition, localization, publication year, study type, and level of evidence based on Ackley et al. [11] were analyzed for the purpose of quality and potential bias assessment. Outcome parameters included number and characteristics of dual-mobility THAs, number of patients and their baseline demographics (sex, age, BMI), definition and radiographic evaluation of liner malseating, clinical and functional outcomes among affected patients, as well as factors associated with malseating. Results were descriptively summarized as means and ranges in case of continuous variables, as well as percentages and absolute numbers for categorical variables.
Results
A total of 39 studies were identified based on PubMed (n = 17), Web of Science (n = 13), Ovid MEDLINE (n = 8), and Cochrane (n = 1) search (Fig. 1). After removal of duplicates, 17 articles were screened for study inclusion, and 12 records considered to be possibly eligible based on their title. After exclusion of 7 studies on non-dual-mobility THAs, 5 retrospective cohort studies, published between 2019 and 2021, were included in the final analysis (Table 1) [6, 12,13,14,15]. Except one multicenter study, all were single institution cohorts, and all conducted at major university-based US institutions between 2010 and 2020. Mean MINORS score was 9.8 (range, 8 to 11), and no study fulfilled level of evidence III or above.
There were a total of 2330 patients (range, 219 to 1322) treated with 2673 dual-mobility THAs (2440 Stryker shells, 233 Zimmer Biomet shells). Mean age was 66.9 years (range, 65.8 to 68), mean BMI was 29.8 kg/m2 (range, 28.3 to 31 kg/m2), and 35% of patients were female (based on studies with available sex differentiation only).
In all but one study [14], linear malseating was evaluated by at least 2 independent reviewers, and at least a third reviewer used for confirmation and/or consensus. Patients without both an AP and cross table lateral radiograph were excluded in all studies. Radiographic follow-up was précised by 3 studies only, with two using a minimum of 6 weeks [6, 15], one reporting of a mean follow-up of 6 and 10 months for Stryker and Zimmer Biomet implants, respectively [13]. Liner malseating was defined in 4 studies, with all proposing a similar definition of gap, asymmetry, angulation, or disruption between liner and acetabular component [6, 12, 13, 15]. In total, 53 malseatings were identified among the 2673 THAs (1.98%), with the rates among studies ranging from 0.15% to 5.8% [6, 14]. Referred to implant type used, 48 malseatings occurred with Stryker (1.96%; Fig. 2) and 5 with Biomet Zimmer shells (2.14%; Fig. 3). Differentiation based on sex was performed in 2 studies, with 35 malseatings occurring in females, 9 in men [6, 12].
Clinical follow-up was described inconsistently among studies with 3 reporting of a mean follow-up (overall mean 2.2 years) [12, 13, 15], 4 including a minimum follow-up (overall mean 1.27 years) [12,13,14,15], and another 4 reporting of a maximum follow-up (overall mean 6.36 years) [6, 12, 13, 15]. Except one patient reporting of pain at 2 years [6], no clinical consequence was noted in the group of malseated liners, including no revision or negatively impacted functionality. Among patients with malseated liners, 4 patients had a metal ion measurement, all of them within a normal range [6, 12, 14]. Functionality was evaluated in 2 studies, with one demonstrating no differences between malseated and non-malseated liners [12].
Three studies performed a statistical analysis to identify factors associated with liner malseating. Guntin et al. [12] identified a component size of 50 mm or less to be associated with malseating. Likewise, Romero et al. [6] reported of a smaller cup size, as well as females, and a lower BMI as statistically significant risk factors, although no significance was noted in the course of a multivariable logistic regression analysis. Finally, Siljander et al. [13] could not show a significant difference between Stryker and Zimmer Biomet liners.
Discussion
This is the first systematic review on liner malseating in patients undergoing dual-mobility THAs. We found a low rate of malseated liners (1.98%) in 2330 patients based on 5 different studies. Importantly, no revision or other negative clinical impact was noted in any of the aforementioned patients at short-term.
Liner malseating is not a new phenomenon and has previously been described with ceramic-on-ceramic liners. Of note, these prior reports demonstrated a substantially higher rate of malseating, ranging from 7.2% up to 25% at short-term [7, 16,17,18,19]. An explanation on this discrepancy is difficult, as the malseating mechanism itself is not understood in its entirety. Inferior interposition of soft tissue [6], deformation of acetabular components with under-reaming [20], and prominent screw heads are among the most common mechanisms discussed [15]. As such, the aforementioned discrepancy might be attributable to surgical factors, as suggested by Salem et al. [14]. Moreover, the authors believe low interobserver and intraobserver reliability [12], as well as studies limited to high-volume university centers only, to offer a potential explanation on differences in malseating rates between dual mobility and ceramic-on-ceramic THAs.
Malseating goes beyond a pure radiographic finding. Prior investigations on malseated ceramic-on-ceramic liners identified significant complications, including liner dissociation, metal fretting, implant interface motion, complete liner dislocation, liner fracture and penetration of the femoral head through the acetabular shell [8, 21,22,23,24]. In our analysis none of these complications were noted, although we acknowledge inconsistent and short-term follow-ups. Moreover, Romero et al. [6] identified earlier fretting onset compared with well-seated liners in the experimental arm of their study, suggesting possible long-term effects not analyzed in current investigations.
Factors associated with malseating are essential, as they might allow for risk factor modification before undergoing THA. In the course of our investigations, one study reported of smaller sized cups as a risk factor for malseating [12]. This is in contrast to findings on malseating in ceramic-on-ceramic THAs that identified higher age, reduced preoperative flexion, and THA for osteoarthritis, but not implant size, as significant risk factors [7]. No further factors, including sex, preoperative BMI, or implant design, were noted in our review, although analysis was limited by low event rates of malseated liners throughout.
This article had a number of limitations. Foremost included studies were retrospective cohort studies with none reaching level of evidence III or above. This was also reflected in a moderate quality assessed by two independent reviewers using the MINORS score. In addition, studies reported outcomes in one country only, including 3 studies from the same clinic (Hospital for Special Surgery) [6, 14, 15], limiting generalizability. Moreover, both radiographic and clinical follow-up were defined inconsistently throughout studies, and affected patients not followed up in detail. In specific, metal ions were only available in 4 cases, and functionality not compared to unaffected hips in all studies.
In conclusion, this systematic review found liner malseating to be a rare finding in dual-mobility THAs. While no direct clinical consequences were noted in any patient, a short-term clinical and inconsistent radiographic follow-up must be acknowledged. Detailed follow-up of affected liners at mid-term, including multiple radiographic controls, metal ion measurements, and functional assessments are necessary before drawing final conclusions.
References
Learmonth ID, Young C, Rorabeck C (2007) The operation of the century: total hip replacement. Lancet 370(9597):1508–1519. https://doi.org/10.1016/S0140-6736(07)60457-7
Sloan M, Premkumar A, Sheth NP (2018) Projected volume of primary total joint arthroplasty in the U.S., 2014 to 2030. J Bone Joint Surg Am 100(17):1455–1460
Ali Khan MA, Brakenbury PH, Reynolds IS (1981) Dislocation following total hip replacement. J Bone Joint Surg Br 63-B(2):214–218. https://doi.org/10.1302/0301-620X.63B2.7217144
Caton JH, Ferreira A (2017) Dual-mobility cup: a new French revolution. Int Orthop 41(3):433–437. https://doi.org/10.1007/s00264-017-3420-7
Harwin SF, Sodhi N, Ehiorobo J, Khlopas A, Sultan AA, Mont MA (2019) Outcomes of dual mobility acetabular cups in total hip arthroplasty patients. Surg Technol Int 34:367–370
Romero J, Wach A, Silberberg S et al (2020) 2020 Otto Aufranc Award: Malseating of modular dual mobility liners. Bone Joint J 102-B(7_Supple_B):20–26. https://doi.org/10.1302/0301-620X.102B7.BJJ-2019-1633.R1
Higuchi Y, Seki T, Takegami Y, Osawa Y, Kusano T, Ishiguro N (2019) What factors predict ceramic liner malseating after ceramic-on-ceramic total hip arthroplasty? Orthop Traumatol Surg Res 105(3):453–459. https://doi.org/10.1016/j.otsr.2019.01.009
Higuchi Y, Hasegawa Y, Komatsu D, Seki T, Ishiguro N (2017) Incidence of ceramic liner malseating after ceramic-on-ceramic total hip arthroplasty associated with osteolysis: A 5- to 15-year follow-up study. J Arthroplasty 32(5):1641–1646. https://doi.org/10.1016/j.arth.2016.11.044
Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, Shamseer L, Tetzlaff JM, Akl EA, Brennan SE, Chou R, Glanville J, Grimshaw JM, Hróbjartsson A, Lalu MM, Li T, Loder EW, Mayo-Wilson E, McDonald S, McGuinness LA, Moher D (2021) The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. Syst Rev 10(1):89. https://doi.org/10.1186/s13643-021-01626-4
Slim K, Nini E, Forestier D, Kwiatkowski F, Panis Y, Chipponi J (2003) Methodological index for non-randomized studies (minors): development and validation of a new instrument. ANZ J Surg 73(9):712–716. https://doi.org/10.1046/j.1445-2197.2003.02748.x
Ackley BJ, Swan BA, Ladwig G, Tucker S (2008) Evidence-based nursing care guidelines: medical-surgical interventions. Mosby Elsevier, St. Louis, p 7
Guntin J, Plummer D, Della Valle C, DeBenedetti A, Nam D (2021) Malseating of modular dual mobility liners. Bone Jt Open 2(10):858–864. https://doi.org/10.1302/2633-1462.210.BJO-2021-0124.R1
Siljander MP, Gausden EB, Wooster BM et al (2022) Liner malseating is rare with two modular dual-mobility designs. Bone Joint J 104-B(5):598–603. https://doi.org/10.1302/0301-620X.104B5.BJJ-2021-1734.R1
Salem HS, Harwin SF, Westrich GH, Delanois RE, Mont MA (2020) Dual mobility acetabular systems for total hip arthroplasty: a multicenter study and technique report. Surg Technol Int 37:356–360
Chalmers BP, Dubin J, Westrich GH (2020) Modular dual-mobility liner malseating: a radiographic analysis. Arthroplast Today 6(4):699–703. https://doi.org/10.1016/j.artd.2020.07.034
Fernández-Fernández R, Oñorbe-San Francisco F, Gil-Garay E (2021) Long-Term Outcomes of a Titanium-Encased Ceramic Liner Total Hip Arthroplasty (15 to 21Year Results). J Arthroplasty 36(11):3697–3702. https://doi.org/10.1016/j.arth.2021.06.016
Langdown AJ, Pickard RJ, Hobbs CM, Clarke HJ, Dalton DJ, Grover ML (2007) Incomplete seating of the liner with the Trident acetabular system: a cause for concern? J Bone Joint Surg Br 89(3):291–295. https://doi.org/10.1302/0301-620X.89B3.18473
Carvajal Alba JA, Schiffman ED, Scully SP, Parvataneni HK (2010) Incomplete seating of a metal-backed alumina liner in ceramic-on-ceramic total hip arthroplasty. Orthopedics 33(1):15. https://doi.org/10.3928/01477447-20091124-11
Miller AN, Su EP, Bostrom MP, Nestor BJ, Padgett DE (2009) Incidence of ceramic liner malseating in Trident acetabular shell. Clin Orthop Relat Res 467(6):1552–1556. https://doi.org/10.1007/s11999-009-0770-5
Small SR, Meding JB, Oja JW et al (2017) Shell design and reaming technique affect deformation in mobile-bearing total hip arthroplasty acetabular components. Proc Inst Mech Eng H 231(8):691–698. https://doi.org/10.1177/0954411917701952
Yamamoto K, Shishido T, Tateiwa T, Katori Y, Masaoka T, Imakiire A, Clarke I (2004) Failure of ceramic THR with liner dislocation: a case report. Acta Orthop Scand 75:500–502
Ha YC, Kim SY, Kim HJ, Yoo JJ, Koo KH (2007) Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty. Clin Orthop Relat Res 458:106–110
Hasegawa M, Sudo A, Hirata H, Uchida A (2003) Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup. J Arthroplasty 18:658–661
Simon JA, Dayan AJ, Ergas E, Stuchin SA, Di Cesare PE (1998) Catastrophic failure of the acetabular component in a ceramic-polyethylene bearing total hip arthroplasty. J Arthroplasty 13(1):108–113. https://doi.org/10.1016/s0883-5403(98)90085-7
Funding
Open Access funding enabled and organized by Projekt DEAL. The authors declare that no funds, grants, or other support were received during the preparation of this manuscript.
Author information
Authors and Affiliations
Contributions
Literature analysis (DK, SA, HB, DA, PK, OA), review of MINORS criteria (DK, OA), writing (DK, OA, SA), internal review (DK, HB, DA, PK).
Corresponding author
Ethics declarations
Conflict of interest
The authors have no relevant financial or non-financial interests to disclose.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Karczewski, D., Andronic, O., Akgün, D. et al. No clinical consequence of liner malseating in dual-mobility THAs at short term: a systematic review. Arch Orthop Trauma Surg 143, 6453–6459 (2023). https://doi.org/10.1007/s00402-023-04855-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-023-04855-5