Forte ceramic-on-delta ceramic cementless total hip arthroplasty: an 8- to 15-year follow-up study

  • Orthopaedic Surgery
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Introduction

Forte ceramic head on delta ceramic liner articulation showed satisfactory midterm results without ceramic-related complication. We aimed to investigate the clinical and radiological outcomes of cementless total hip arthroplasty (THA) with forte ceramic head on delta ceramic liner articulation.

Materials and methods

Overall, 107 patients (57 men, 50 women; 138 hips) who underwent cementless THA with forte ceramic head on delta ceramic liner articulation were enrolled. The mean follow-up duration was 11.6 years. For the clinical assessments, Harris hip score (HHS), Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), presence of thigh pain, and presence of squeaking were evaluated. Radiographs were assessed to search for osteolysis, stem subsidence, loosening of implants. Kaplan–Meier survival curves were evaluated.

Results

The mean HHS and WOMAC improved from 57.1 and 28.1 preoperatively to 81.4 and 13.1 at the final follow-up, respectively. Nine revisions (6.5%) were performed; 5 hips for stem loosening, 1 hip for ceramic liner fracture, 2 hips for periprosthetic fracture, and 1 hip for progressive osteolysis around cup and stem. Thirty-two patients (37 hips) complained squeaking, in which 4 cases (2.9%) were identified as ceramic-related noises. After a mean follow-up period of 11.6 years, 91% (95% CI 87.8–94.2) were free from revision of both femoral and acetabular components due to any reason.

Conclusions

Cementless THA with forte ceramic-on-delta ceramic articulation showed acceptable clinical and radiological results. Serial surveillance of these patients should be performed due to the possibility of cerami- related complications such as squeaking, osteolysis, and ceramic liner fracture.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

Data availability

The data analyzed during the current study are not publicly available due to the policy of the institutional review board of Asan Medical Center but are available from the corresponding author on reasonable request.

References

  1. Lau YJ, Sarmah S, Witt JD (2018) 3rd generation ceramic-on-ceramic cementless total hip arthroplasty: a minimum 10-year follow-up study. Hip Int 28:133–138. https://doi.org/10.1177/1120700018768617

    Article  PubMed  Google Scholar 

  2. Kim HS, Park JW, Ha JH, Lee YK, Ha YC, Koo KH (2022) Third-generation ceramic-on-ceramic total hip arthroplasty in patients with osteonecrosis of the femoral head: a 10- to 16-year follow-up study. J Bone Joint Surg Am 104:68–75. https://doi.org/10.2106/JBJS.20.00720

    Article  PubMed  Google Scholar 

  3. Byun JW, Yoon TR, Park KS, Seon JK (2012) Third-generation ceramic-on-ceramic total hip arthroplasty in patients younger than 30 years with osteonecrosis of femoral head. J Arthroplasty 27:1337–1343. https://doi.org/10.1016/j.arth.2011.07.004

    Article  PubMed  Google Scholar 

  4. Koo KH, Ha YC, Kim SY, Yoon KS, Min BW, Kim SR (2014) Revision of ceramic head fracture after third generation ceramic-on-ceramic total hip arthroplasty. J Arthroplasty 29:214–218. https://doi.org/10.1016/j.arth.2013.04.007

    Article  PubMed  Google Scholar 

  5. Howard DP, Wall PDH, Fernandez MA, Parsons H, Howard PW (2017) Ceramic-on-ceramic bearing fractures in total hip arthroplasty: an analysis of data from the National Joint Registry. Bone Joint J 99:1012–1019. https://doi.org/10.1302/0301-620X.99B8.BJJ-2017-0019.R1

    Article  PubMed  Google Scholar 

  6. Lombardi AV Jr, Berend KR, Seng BE, Clarke IC, Adams JB (2010) Delta ceramic-on-alumina ceramic articulation in primary THA: prospective, randomized FDA-IDE study and retrieval analysis. Clin Orthop Relat Res 468:367–374. https://doi.org/10.1007/s11999-009-1143-1149

    Article  PubMed  Google Scholar 

  7. Choy WS, Cha YH, Jeon CY, Lee KS, Kim HY (2018) A minimum ten years of follow-up of alumina head on delta liner total hip arthroplasty. J Arthroplasty 33:470–476. https://doi.org/10.1016/j.arth.2017.08.043

    Article  PubMed  Google Scholar 

  8. Al-Hajjar M, Carbone S, Jennings LM, Begand S, Oberbach T, Delfosse D et al (2017) Wear of composite ceramics in mixed-material combinations in total hip replacement under adverse edge loading conditions. J Biomed Mater Res B Appl Biomater 105:1361–1368. https://doi.org/10.1002/jbm.b.33671

    Article  CAS  PubMed  Google Scholar 

  9. Jarrett CA, Ranawat AS, Bruzzone M, Blum YC, Rodriguez JA, Ranawat CS (2009) The squeaking hip: a phenomenon of ceramic-on-ceramic total hip arthroplasty. J Bone Joint Surg Am 91:1344–1349. https://doi.org/10.2106/JBJS.F.00970

    Article  PubMed  Google Scholar 

  10. Gruen TA, McNeice GM, Amstutz HC (1979) “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res 141:17–27

    Google Scholar 

  11. DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res 121:20–32

    Google Scholar 

  12. Zicat B, Engh CA, Gokcen E (1995) Patterns of osteolysis around total hip components inserted with and without cement. J Bone Joint Surg Am 77:432–439

    Article  CAS  PubMed  Google Scholar 

  13. Engh CA, Massin P, Suthers KE (1990) Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res 257:107–128

    Article  Google Scholar 

  14. Udomkiat P, Wan Z, Dorr LD (2001) Comparison of preoperative radiographs and intraoperative findings of fixation of hemispheric porous-coated sockets. J Bone Joint Surg Am 83:1865–1870. https://doi.org/10.2106/00004623-200112000-00015

    Article  CAS  PubMed  Google Scholar 

  15. Kang BJ, Ha YC, Hwang SC, Lee YK, Koo KH (2014) Midterm results of large diameter Biolox forte ceramic head on delta ceramic liner articulation in total hip arthroplasty. J Arthroplasty 29:2412–2414. https://doi.org/10.1016/j.arth.2014.03.003

    Article  PubMed  Google Scholar 

  16. Castagnini F, Bordini B, Biondi F, Cosentino M, Stea S, Traina F (2020) Mixed ceramic combinations in primary total hip arthroplasty achieved reassuring mid-to-longterm outcomes. J Mater Sci Mater Med 31:56. https://doi.org/10.1007/s10856-020-06393-7

    Article  CAS  PubMed  Google Scholar 

  17. Toni A, Giardina F, Guerra G, Sudanese A, Montalti M, Stea S et al (2017) 3rd generation alumina-on-alumina in modular hip prosthesis: 13 to 18 years follow-up results. Hip Int 27:8–13. https://doi.org/10.5301/hipint.5000429

    Article  PubMed  Google Scholar 

  18. Elkins JM, Pedersen DR, Callaghan JJ, Brown TD (2013) Do obesity and/or stripe wear increase ceramic liner fracture risk? An XFEM analysis. Clin Orthop Relat Res 471:527–536. https://doi.org/10.1007/s11999-012-2562-6

    Article  PubMed  Google Scholar 

  19. Yoo JJ, Kim YM, Yoon KS, Koo KH, Song WS, Kim HJ (2005) Alumina-on-alumina total hip arthroplasty. A five-year minimum follow-up study. J Bone Joint Surg Am 87:530–535. https://doi.org/10.2106/JBJS.D.01753

    Article  PubMed  Google Scholar 

  20. Hamilton WG, McAuley JP, Blumenfeld TJ, Lesko JP, Himden SE, Dennis DA (2015) Midterm results of delta ceramic-on-ceramic total hip arthroplasty. J Arthroplasty 30:110–115. https://doi.org/10.1016/j.arth.2015.02.047

    Article  PubMed  Google Scholar 

  21. Kim SC, Lim YW, Jo WL, Park HW, Han SB, Kwon SY et al (2019) Fourth-generation ceramic-on-ceramic THA results in improvements in midterm outcomes compared to third-generation THA but does not resolve noise problems: a cohort study of a single-hip system. BMC Musculoskelet Disord 20:263. https://doi.org/10.1186/s12891-019-2641-x

    Article  PubMed  PubMed Central  Google Scholar 

  22. Hu D, Tie K, Yang X, Tan Y, Alaidaros M, Chen L (2015) Comparison of ceramic-on-ceramic to metal-on-polyethylene bearing surfaces in total hip arthroplasty: a meta-analysis of randomized controlled trials. J Orthop Surg Res 10:22. https://doi.org/10.1186/s13018-015-0163-2

    Article  PubMed  PubMed Central  Google Scholar 

  23. Kim MW, Kim SM, Chung YY (2018) Total hip arthroplasty using ceramic-on-ceramic bearing surfaces: long-term assessment of squeaking sounds. Hip Pelvis 30:18–22. https://doi.org/10.5371/hp.2018.30.1.18

    Article  PubMed  PubMed Central  Google Scholar 

  24. Gillespie JA, Kennedy JW, Patil SR, Meek DR (2016) Noise production in ceramic-on-ceramic total hip arthroplasty is associated with lower patient satisfaction and hip scores. J Orthop 13:282–284. https://doi.org/10.1016/j.jor.2016.06.018

    Article  PubMed  PubMed Central  Google Scholar 

  25. Kiyama T, Kinsey TL, Mahoney OM (2013) Can squeaking with ceramic-on-ceramic hip articulations in total hip arthroplasty be avoided? J Arthroplasty 28:1015–1020. https://doi.org/10.1016/j.arth.2012.10.014

    Article  PubMed  Google Scholar 

  26. Mai K, Verioti C, Ezzet KA, Copp SN, Walker RH, Colwell CW Jr (2010) Incidence of “squeaking” after ceramic-on-ceramic total hip arthroplasty. Clin Orthop Relat Res 468:413–417. https://doi.org/10.1007/s11999-009-1083-4

    Article  PubMed  Google Scholar 

  27. Tai SM, Munir S, Walker WL, Pearce SJ, Walter WK, Zicat BA (2015) Squeaking in large diameter ceramic-on-ceramic bearings in total hip arthroplasty. J Arthroplasty 30:282–285. https://doi.org/10.1016/j.arth.2014.09.010

    Article  PubMed  Google Scholar 

  28. Keurentjes JC, Kuipers RM, Wever DJ, Schreurs BW (2008) High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings. Clin Orthop Relat Res 466:1438–1443. https://doi.org/10.1007/s11999-008-0177-8

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  29. Baek SH, Kim SY (2008) Cementless total hip arthroplasty with alumina bearings in patients younger than fifty with femoral head osteonecrosis. J Bone Joint Surg Am 90:1314–1320. https://doi.org/10.2106/JBJS.G.00755

    Article  PubMed  Google Scholar 

  30. Baruffaldi F, Mecca R, Stea S, Beraudi A, Bordini B, Amabile M et al (2020) Squeaking and other noises in patients with ceramic-on-ceramic total hip arthroplasty. Hip Int 30:438–445. https://doi.org/10.1177/1120700019864233

    Article  PubMed  Google Scholar 

  31. Sexton SA, Yeung E, Jackson MP, Rajaratnam S, Martell JM, Walter WL et al (2011) The role of patient factors and implant position in squeaking of ceramic-on-ceramic total hip replacements. J Bone Joint Surg Br 93:439–442. https://doi.org/10.1302/0301-620X.93B4.25707

    Article  CAS  PubMed  Google Scholar 

Download references

Funding

The authors have no relevant financial or non-financial interests to disclose.

Author information

Authors and Affiliations

Authors

Contributions

All authors contributed to the study conception and design. PWY is the guarantor of submission and participated in study design, data collection, data analysis and data interpretation; J-KM participated in writing, data analysis, data interpretation, and literature search; SL and C-HK participated in data interpretation and figures; JYY participated in data collection and data analysis; SL participated in data analysis and figures; K-SL participated in data collection and data interpretation.

Corresponding author

Correspondence to Pil Whan Yoon.

Ethics declarations

Conflict of interest

The authors, their immediate families, and any research foundations with which they are affiliated have not received any benefits from any commercial entity related to the subject of this article.

Ethical approval

This retrospective study was performed following approval from the institutional review board of Asan Medical Center (No. 2022–1062). All methods were carried out in accordance with relevant guidelines.

Consent to participate

Informed consent was waived by the institutional review board of Asan Medical Center.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Moon, JK., Lee, S., Kim, CH. et al. Forte ceramic-on-delta ceramic cementless total hip arthroplasty: an 8- to 15-year follow-up study. Arch Orthop Trauma Surg 143, 5475–5483 (2023). https://doi.org/10.1007/s00402-023-04793-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00402-023-04793-2

Keywords

Navigation