Introduction

Large diaphyseal bone defects can be the result of trauma, osteomyelitis and resection of bone tumors. These segmental defects are often reconstructed with structural cortical allografts, which provide initial mechanical strength [15, 17]. However, in-growth of these structural cortical bone grafts is limited and is therefore associated with long-term problems like nonunion and fatigue fractures. These complications result in survival rates of only 64–76% after 10 years [7, 9, 29]. In two case reports an alternative method of reconstruction was described using cylindrical metal cages filled with morsellized bone graft combined with intramedullary fixation [8, 21]. A study analyzing retrievals of failed spinal cage reconstructions reports only 30% viable bone in the cages [26]. This could be caused by stress shielding of the cage. Bone grafts incorporate without load due to a trauma response. However, in the long term, load seems to be another important factor in bone healing by stimulating bone remodeling and maintaining bone quantity and quality [10, 18]. If the load is absent the bone will degrade over time; but if the bone is loaded, studies with morsellized bone grafts in reconstructions have shown that it remodels into healthy bone [16, 22, 23]. It is our aim to reconstruct segmental diaphyseal defects with a cage that contains bone grafts which are mechanically loaded and hypnotized that clinical results can thereby be improved. The graft in the diaphyseal cage reconstruction of Cobos [8] could be loaded by removing the load-transfer to the cage and facilitate load-transfer directly to the morsellized bone graft (Fig. 1). This is achieved by allowing the cortical host bone to rest on the graft instead of the cage. The reconstruction could be stabilized with a dynamized intramedullary nail instead of a statically locked nail. Obviously the morsellized bone graft needs to provide stability, which can be achieved by impaction, and could subsequently carry a large portion of the dynamic in vivo load. The question of this study was if an impacted morsellized bone graft in a cage would be stable enough under dynamic loading conditions in an in vitro reconstruction. Deformation of the graft should not exceed 6–10 mm, which is the dynamic range of the slotted holes in commercial intramedullary nails, which is obviously limited to prevent leg length discrepancy. More specifically we assessed the effects of two different cage types, washing of the bone grafts (which could have an effect on the stability properties of the bone grafts), and mismatch between the fit of the circular cage and the irregular host-bone (containment) on the stability of the reconstruction.

Fig. 1
figure 1

Loading patterns in a static and dynamic reconstruction

Materials

We tested four groups with three variables; cage type, preparation of the graft (standard or washed) and bone-cage fit (Table 1).

Table 1 Variables of the test groups

Two types of cages were used in this study. The first cage was a prefabricated titanium Harms cage, commonly used for spinal applications but with potential application possibilities to reconstruct segmental bone defects (DePuy Motech, Warsaw, IN, USA). The height of the cage was 4.5 cm and it was ellipse-shaped (diameter of 22 × 28 mm) and had a pore size of 4 × 6 mm (Fig. 2). The second cage was a stainless steel mesh, commonly used for reconstructions in hip revision surgery (Stryker Howmedica X-Change system, Newbury, UK), which was cut to a height of 4.5 cm and folded into a cylinder with a diameter of 24 mm. This mesh was fixed with three cerclage wires (Fig. 3) and deformed by hand to a diameter of 22 × 28 mm. The mesh had a pore size of 3 × 3 mm.

Fig. 2
figure 2

Harms cage filled with bone graft in a contained and loaded condition

Fig. 3
figure 3

Stryker mesh cage

The morsellized bone graft used to fill the cages was harvested from a fresh frozen bovine distal femur. The distal femurs were sectioned longitudinally into two parts, and the cartilage and cortical bone were removed. Cancellous bone was nibbled by hand using a rongeur that produced large pure cancellous bone chips. The bone chips were subsequently processed in a Noviomagus bone mill (Spierings Medische Techniek, Nijmegen, The Netherlands) with the smallest rasping blade leading to a particle size of about 2 mm. In the group with the washed graft, the graft was washed using 2 l of saline with a pulse lavage jet (Stryker, Newbury, UK).

In the last group (Table 1) the graft was loaded with a plunger (cylinder with a 22 mm outer diameter and 14 mm inner diameter), which did not completely seal off the cage leaving some space for a possible escape of the graft under load between the plunger and the cage and between the plunger and the intramedullary nail of 10 mm (Fig. 4). This situation mimics clinical reality where a cage sometimes does not ideally fit around the host bone. In the other groups an ellipse-shaped plunger (22 × 28 mm outer diameter and 10 mm inner diameter) was used which sealed off the cage and nail leading to a full containment of the graft (Fig. 2).

Fig. 4
figure 4

Mismatch between Harms cage and cylindrical plunger leading to un-containment

Methods

The cages were filled in five layers with a total of 28 g morsellized bone graft. After each layer 10 standardized impactions were made with a mass of 1.5 kg (Fig. 5). The mass was dropped from a height of 35 cm. The last layer was followed by 40 impactions of the same weight. After the final impaction the remaining height of the graft layer that was inside the cage was approximately 35 mm. At the center of the grafted cage a 10 mm thick nail was located to mimic an intramedullary nail (Fig. 6). The graft in the cage was axially loaded using an MTS loading device (MTS Systems Corporation, Minneapolis, Minnesota, USA). An initial preload of 30 Newton was applied to pre-condition the reconstruction. Subsequently a dynamic load was applied with a maximum of 600 N for a period of 800 cycles and then raised to 1,200 N for another period of 800 cycles (frequency of 1 Hz). During loading, the displacement of the plunger into the graft material was recorded. In each group five cages (N = 5) were filled with bone graft and tested according to this protocol.

Fig. 5
figure 5

Impactor with mass of 1.5 kg

Fig. 6
figure 6

Impactor and Harms cage with a central nail

Statistical analysis

The ultimate subsidence values, at the end of testing the different groups, were statistically analyzed using ANOVA with a significance level of P = 0.05.

Results

Deformation of the graft material showed a typical visco-elastic/plastic response with a high deformation rate at the beginning of a loading period which leveled off with the number of loading cycles (Fig. 7). During loading, escape of fat through the cage openings was observed. There was no escape of bone graft through the openings of the mesh or the cage. Similarly, we observed no extrusion of bone graft particles at the locations of uncontained fitting of the cage (group 4).

Fig. 7
figure 7

Height of the graft column in mm under dynamic loaded condition

The cancellous graft had a compacted appearance after impacting and subsequent loading, and was difficult to remove out of the cage and mesh. The cage and mesh deformed showed negligible visible deformation. The cerclage wire around the mesh was strong enough to withstand the loading forces.

The Harms cage and mesh cage, which contained unwashed bone grafts were relatively stable with a maximal average subsidence of 2.90 mm (8.3% loss of volume) and 4.13 mm (11.8% loss of volume) for the Harms cage and mesh cage, respectively. Hence, both reconstructions provided an adequate stability of which the subsidence did not exceed the limit of 6 mm (Table 2). In terms of stability, the more rigid Harms cage was superior to the mesh cage with deformations that were significantly smaller at both load levels of 600 N (P = 0.004) and 1,200 N (P = 0.001).

Table 2 Deformation of the graft in mm under loading

Washing the bone grafts had surprisingly little effect on the stability of the reconstruction. Hence, no statistical effect of this parameter on the deformation values was found; at a load of 600 N (P = 0.981) and 1,200 N (P = 0.882).

The stability appeared to be susceptible to the quality of fit between the cage and the plunger. The uncontained unwashed bone graft in a Harms cage was not stable enough and exceeded the compression limit of 6 mm at a load of 1200 N (Table 2). The effect of containment of the unwashed graft in the Harms cage was significant at a load level of 600 N (P = 0.001) and 1,200 N (P < 0.001).

Discussion

Large segmental diaphyseal bone defects are commonly reconstructed using large massive cortical allografts. However, incorporation of these grafts is limited and therefore associated with long-term problems like nonunions and failure. Furthermore, they are expensive and carry the risk of transmitting diseases [20]. Therefore alternatives are required. The drawback is that only a small number of publications exist probably owing to the variety of patients with a different diagnosis and stages of diseases making comparison almost impossible [7, 9, 13, 20].

Direct postoperatively cortical grafts obviously have better mechanical properties as compared to morsellized bone graft. However, from a biological perspective the in-growth of a cancellous bone-graft is superior to a cortical bone-graft, which results in a higher incorporation and union rate, and therefore superior long-term characteristics can be expected [15, 17].

The reconstruction with a cage containing a loaded bone graft potentially may have adequate biologic features for the morsellized graft and sufficient initial mechanical features owing to a combination of impaction and containment in a cage/mesh [23]. Although the microstructure of the morsellized graft diminishes somewhat after impaction and subsequent loading, it is still less dense than the structural cortical graft which should provide for superior biological properties [6].

The application of structural grafts is also limited by the fact that it is often difficult to obtain an adequate fit between the graft and the host bone. An intimate host graft junction, which facilitates union, is virtually never achieved. The open structure of the cages facilitates in-growth and neovascularization in the large bone defect [5]. However, in-growth can be limited in oncological resections due to the lack of periosteum and adequate soft tissue coverage. A biological reconstruction in these cases especially with adjuvant chemo and radiotherapy is probably difficult to achieve. The intramedullary nail and the cages used (Harms and Stryker mesh) have a proven human application. Therefore, it is likely that the combined use of these materials will not trigger any unexpected adverse biological processes and are in this respect safe to use [8, 23, 25]. An incorporated metal mesh, however, can induce stress shielding in the long term. Resorbable meshes may reduce stress shielding but lack sufficient intrinsic stability to withstand the loads applied.

In this study it was shown that graft stability was influenced by containment. Although the graft was rather stable and forgives some un-containment, this study shows that the cage should fit around the bone optimally in order to maintain an adequate stability.

The stability of the graft may be explained by the fact that the chips of bone graft get entangled after impaction, which increases the inter-particle shear resistance [1, 4]. This is probably the explanation why no protrusion was observed of graft material through the pores in the cages even though the pore size of the cages exceeded the chip size.

The titanium Harms cage is more rigid than the Stryker mesh and consequently shows less deformation when loaded. This probably explains the difference in stability between the two groups. The Stryker mesh, however, can be folded for a custom fit leading to a better fit and consequently a better containment of the bone chips. From that perspective the mesh shows more promise from a surgical point of view, compared to the prefabricated Harms cage.

Washing the graft had little or even a slight negative effect on graft stability in this study. During compression of the unwashed graft fat and blood escaped through the cage openings and therefore adds no relevant extra stability. This is in contrast to studies in which cup stability is increased by graft washing [11, 27]. The difference between cup stability and graft stability in cages could be explained by the fact that containment of the bone chips is far better within the cages with axial loading as compared to the situation around a cup under shear-loaded conditions. Consequently, the stability of the reconstruction becomes less dependent on the inter-particle shear resistance, which is higher in the case of washed particles.

Unloaded grafts are more prone to resorption and could lead to nonunion and migration [26]. Loading of morsellized grafts could lead to instability. However, long-term reports of reconstructions with loaded morsellized graft around an implant do not show instability [16, 23]. Bone grafts in loaded conditions stimulate bone apposition, which could lead to superior results in our reconstruction with bone graft loading instead of load bypassing the graft [3, 24, 28]. To obtain continuous loading of the bone grafts, the deformation of the reconstruction should be less than the range of the dynamization distance of the intramedullary nail. Our goal was therefore that the graft layer would not deform more than 6 mm after loading. In most study groups this goal was achieved.

The load of 600–1,200 N was determined as twice the body weight of 60 kg. This weight of a small person is in accordance with the small cage size of 22–28 mm and should be a reasonable load magnitude for initial loading after reconstruction.

The impacted graft was stable enough if adequate containment could be achieved together with a cage with an adequate intrinsic stiffness. It is realized that stability requires firm impactions. If this is not achieved inferior results have been reported [2, 12, 19]. Relative to in vivo application of this technique, if necessary the dynamic sliding range of the nail could be enlarged to 1 cm especially in larger defects. This would improve the chances of long-term load transfer from the bone fragments onto the bone chips and should therefore have positive effects on the incorporation of the bone grafts.

Obviously, this study was simplified relative to the clinical reality. An important limitation of this study is the lack of in vivo resorption and incorporation of the bone graft. Resorption of morsellized bone graft is followed by replacement of new bone. This could lead to instability during the remodeling phase. This phenomenon needs to be assessed in animal experiments with large segmental defects. The bovine bone used in this study has different properties compared to human bone. The stiffness however is in the same range and is mainly influenced by impaction, particle size and water content [2, 14].

Another limitation of the study was the simplified loading configuration. In vivo loading also includes bending and torsion. However, it is expected that the intramedullary nail will transfer these loads, and that owing to the dynamization of the nail, predominantly compressive loads are transferred through the bone grafts. In that sense, the loading configurations seem a good approximation of what occurs in vivo. The indication of the reconstruction with the intramedullary nail and cage is limited to segmental defects in the meta/diaphyseal region of large bones.

We believe that this reconstruction is a promising alternative to the massive cortical grafts in the reconstruction of large diaphyseal defects, and its biological merits need to be assessed in animal experiments with large segmental defects.

Conclusion

Morsellized impacted bone graft in a cage is stable in dynamic loaded conditions in an in vitro reconstruction of a segmental diaphyseal bone defect. In terms of stability the method appears to be not very sensitive to the cage type and whether the bone chips are washed or not. However, containment of the cage with the bone segment is an important parameter and should be optimized during surgery. If this is achieved, we believe that this method of reconstruction could be a promising alternative for the reconstruction of large diaphyseal bone defects, and its biological behavior needs to be assessed in an animal model.