High migration rate of two types of threaded acetabular cups

  • Original Article
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Introduction

Survivorship analysis was performed on 479 Link V-type threaded cups and on 110 modified V-type Bad Bramstedt cups (a modification of the Link V-type cup since 1993 with a smaller primary coil) to evaluate the effect of the modifications.

Materials and methods

Indication for hip arthroplasty with a Link V-type cup was an inflammatory arthritis in 310 patients, osteoarthritis in 138 patients and dysplastic hip joints in 31 patients. The average follow-up was 8.6±3.2 (range 3.0–15.2) years. The 110 modified V-type Bad Bramstedt cups had a mean follow-up period of 4.5±0.7 (range 3.0–5.9) years, including 49 with inflammatory arthritis, 49 with osteoarthritis and 12 with dysplastic hip joints. To evaluate the migration rate and radiolucent lines, radiographical examination according to the method of Nunn et al. and Delee and Charnley was performed on 264 Link V-type cups with a mean follow-up period of 8.2±2.7 years and 59 modified V-type Bad Bramstedt cups with a mean follow-up period of 4.6±0.7 years. Relevant parameters influencing cup migration were analysed.

Results

The cumulative survival rate of the former Link V-type acetabular cup was 94.5% after 5 years, 88.1% after 10 years and 70.2% after 15 years. The Bad Bramstedt cup showed a 5-year survivorship rate of 97.9%. Migration of more than 3 mm or tilting of the cup greater than 5 degrees was found in 73% of the former type and in 39% of the modified cup. Radiolucent lines greater than 2 mm and detectable in two zones appeared in 6.4% of the former Link V-type and in 1.7% of the Bad Bramstedt cup design. Among the influencing factors analysed, length of follow-up and primary cup positioning showed a significant correlation to cup migration.

Conclusion

The modification of the Link V-type cup showed no satisfactory improvement in cup migration. Therefore, both threaded cups were abandoned in favour of cementless press-fit cups.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2a,b
Fig. 3a–c
Fig. 4a,b
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Aigner C (1998) 10-year results of the corund-blasted Zweymüller threaded acetabular ring. Z Orthop 136:110–114

    CAS  PubMed  Google Scholar 

  2. Antolic V, Iglisc A, Herman S, Srakar F, Macek Leber A, Krajl Iglic V, Brajnik D, Stanic U (1994) Resultant hip joint force after total hip replacement. Acta Chir Orthop Traumatol Cech 61:42–44

    CAS  PubMed  Google Scholar 

  3. Baldursson H, Hansson LI, Olsson TH, Selvik G (1980) Migration of the acetabular socket after total hip replacement determined by roentgen stereophotogrammetry. Acta Orthop Scand 51:535–540

    CAS  PubMed  Google Scholar 

  4. Beckenbaugh RD, Ilstrup DM (1978) Total hip arthroplasty. A review of three hundred and thirty-three cases with long-term follow-up. J Bone Joint Surg Am 60:306–313

    CAS  PubMed  Google Scholar 

  5. Blum HJ, Noble PC, Tullos HS (1990) Migration and rotation of cementless acetabular cups: incidence, etiology and clinical significance. 57th Annual AAOS Meeting, New Orleans, Louisiana, February 8–13

    Google Scholar 

  6. Brujin JD, Seelen JL, Feenstra RM, Hansen BE, Bernoski FP (1995) Failure of the Mecring screw-ring acetabular component in total hip arthroplasty. A three to seven-year follow-up study. J Bone Joint Surg Am 77:760–766

    PubMed  Google Scholar 

  7. Capello WN, Colyer RA, Kernek CB, Carnahan JV, Hess JJ (1993) Failure of the Mecron screw-in ring. J Bone Joint Surg Br 75:835–836

    CAS  PubMed  Google Scholar 

  8. Carlsson AS, Gentz C-F, Sanzén L (1986) Socket loosening after hip arthroplasty. Radiographic observations in 241 cases up to 15 years. Acta Orthop Scand 57:97–100

    Google Scholar 

  9. Cordero-Ampuero J, Garcia-Cimbrelo E, Munuera L (1994) Fixation of cementless acetabular cups. A radiographic 4–8-year study of 102 porous-coated components. Acta Orthop Scand 65:263–266

    PubMed  Google Scholar 

  10. Cotterill P, Hunter GA, Tile M (1982) A radiographic analysis of 166 Charnley-Müller total hip arthroplasties. Clin Orthop 163:120–126

    PubMed  Google Scholar 

  11. Coudane H, Fery A, Sommelet J, Lacoste J, Leduc P, Gaucher A (1981) Aseptic loosening of cemented total arthroplasties of the hip in relation to positioning of the prosthesis: new utilization of the Tschuprow-Cramer statistical test. Acta Orthop Scand 52:201–205

    CAS  PubMed  Google Scholar 

  12. Delaunay CP, Kapandji AI (1998) Survivorship of rough-surfaced threaded acetabular cups. 382 consecutive primary Zweymüller cups followed for 0.2–12 years. Acta Orthop Scand 69:379–383

    PubMed  Google Scholar 

  13. DeLee JG, Charnley J (1976) Radiological demarcation of cementless sockets in total hip replacement. Clin Orthop 121:20–32

    PubMed  Google Scholar 

  14. Dickob M, Bleher J, Puhl W (1994) Standardisierte Pfannenwanderungsanalyse in der Hüftendoprothetik mittels digitaler Bildverarbeitung. Unfallchirurg 97:92–97

    CAS  PubMed  Google Scholar 

  15. Dominikus M, Wanivenhaus AH, Morscher M, Powell G, Krismer M, Wölfl G (1998) Different cup migration in rheumatoid arthritis and arthrosis. A radiographic analysis of 127 uncemented acetabular cups. Acta Orthop Scand 69:455–462

    CAS  PubMed  Google Scholar 

  16. Hamilton HW (1986) Prosthetic position in total hip replacement. Orthop Rev 15:294–300

    CAS  PubMed  Google Scholar 

  17. Havelin LI, Vollset SE, Engesaeter LB (1995) Revision for aseptic loosening of uncemented cups in 4352 primary total hip prostheses. A report form the Norwegian Arthorplasty Register. Acta Orthop Scand 66:494–500

    Google Scholar 

  18. Kaplan EL, Meier P (1958) Nonparametric estimation from incomplete observations. J Am Stat Assoc 53:457–481

    Google Scholar 

  19. Karachalios T, Hartofilakidis G, Zacharakis N, Tsekoura M (1993) A 12- to 18-year radiographic follow-up study of Charnley low-friction arthroplasty. The role of the centre of rotation. Clin Orthop 296:140–147

    PubMed  Google Scholar 

  20. Krismer M, Stoeckl B, Fischer M, Bauer R, Mayrhofer P, Ogan M (1996) Early migration predicts late aseptic failure of hip sockets. J Bone Joint Surg Br 78:422–426

    CAS  PubMed  Google Scholar 

  21. Krismer M, Tschupik JP, Bauer R, Mayrhofer P, Stöckl B, Fischer M, Biedermann R (1997) Einzel-Bild-Röntgen-Analyse (EBRA) zur Messung der Migration von Hüftendoprothesen. Orthopäde 26:229–236

    Google Scholar 

  22. Kutschera HP, Eyb R, Schartelmüller T, Toma C, Zweymüller K (1993) Das zementfreie Zweymüller Hüft-System. Ergebnisse einer 5-Jahres-Nachuntersuchung. Z Orthop 131:513–517

    CAS  PubMed  Google Scholar 

  23. Malchau H, Herberts P, Wang YX, Kärrholm J, Romanus B (1986) Long-term clinical and radiological results of the Lord total hip prosthesis. A prospective study. J Bone Joint Surg Br 78:884–891

    Article  Google Scholar 

  24. Manley MT, Capello WN, D’Antonio JA, Edidin AA, Geesink RGT (1998) Fixation of acetabular cups without cement in total hip arthroplasty. J Bone Joint Surg Am 80:1175–1185

    CAS  PubMed  Google Scholar 

  25. McBroom R, Müller M (1984) Aseptic loosening: fifteen years’ experience with the Müller total hip arthroplasty. J Bone Joint Surg Br 66:300–301

    Google Scholar 

  26. Mjöberg B (1991) Fixation and lossening of hip prostheses. A review. Acta Orthop Scand 62:500–508

    PubMed  Google Scholar 

  27. Mjöberg B, Selvik G, Hansson LI, Rosenqvist R, Önnerfält R (1986) Mechanical loosening of total hip prostheses. A radiographic and roentgen stereogrammatic study. J Bone Joint Surg Br 68:770–774

    PubMed  Google Scholar 

  28. Morscher E, Schmassmann A (1983) Failures of total hip arthroplasty and probable incidence of revision surgery in the future. Arch Orthop Trauma Surg 101:137–142

    CAS  PubMed  Google Scholar 

  29. Morscher EW (1992) Current status of acetabular fixation in primary total hip arthroplasty. Clin Orthop 274:172–193

    PubMed  Google Scholar 

  30. Nunn D, Freeman MAR, Hil PF, Evans SJW (1989) The measurement of migration of the acetabular component of hip prostheses. J Bone Joint Surg Br 71:629–631

    CAS  PubMed  Google Scholar 

  31. Önsten I, Bengner U, Besjakov J (1993) Socket migration after Charnley arthroplasty in rheumatoid arthritis and osteoarthritis. A roentgen stereophotogrammetric study. J Bone Joint Surg Br 75:677–680

    PubMed  Google Scholar 

  32. Önsten I, Akesson K, Obrant KJ (1995) Micromotion of the acetabular component and periactabular bone morphology. Clin Orthop 310:103–110

    PubMed  Google Scholar 

  33. Pagnano MW, Hanssen AD, Lewallen DG, Shaughnessy WJ (1996) The effect of superior placement of the acetabular component on the rate of loosening after total hip arthroplasty. J Bone Joint Surg Am 78:1004–1012

    CAS  PubMed  Google Scholar 

  34. Pellici PM, Wilson PD Jr, Sledge CB, Salvati EA, Ranawat CS, Poss R (1982) Revision total hip arthroplasty. Clin Orthop 170:34–41

    PubMed  Google Scholar 

  35. Rapperport DJ, Carter DR, Schurman DJ (1987) Contact finite element stress analysis of porous ingrowth acetabular cup implantation, ingrowth, and loosening. J Orthop Res 5:548–561

    CAS  PubMed  Google Scholar 

  36. Rosler J, Perka C (2000) The effect of anatomical positional relationships on kinetic parameters after total hip replacement. Int Orthop 24:23–27

    Article  PubMed  Google Scholar 

  37. Simank H-G, Brocal DRC, Reiser D, Thomsen M, Sabo D, Lukoschek M (1997) Middle-term results of threaded acetabular cups. High failure rates five years after surgery. J Bone Joint Surg Br 79:366–370

    Article  CAS  PubMed  Google Scholar 

  38. Snorrason F, Kärrholm J (1990) Primary migration of fully-threaded acetabular prostheses. A roentgen stereophotogrammetric analysis. J Bone Joint Surg Br 72:647–652

    CAS  PubMed  Google Scholar 

  39. Stauffer RN (1982) Ten-year follow-up study of total hip replacements. With particular reference to roentgenographic loosening of the components. J Bone Joint Surg Am 64:983–990

    CAS  PubMed  Google Scholar 

  40. Wilson-MacDonald J, Morscher E, Masar Z (1990) Cementless uncoated polyethylene acetabular components in total hip replacement. Review of five- to 10-year results. J Bone Joint Surg Br 72:423–430

    CAS  PubMed  Google Scholar 

  41. Wroblewski BM (1979) Wear of high-density polyethylene on bone and cartilage. J Bone Joint Surg Br 61:498–500

    Google Scholar 

  42. Wroblewski BM, Lynch M, Atkinson JR, Dowson D, Isaac GH (1987) External wear of the polyethylene socket in cemented total hip arthroplasty. J Bone Joint Surg Br 69:61–63

    CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Bernd Fink.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Fink, B., Protzen, M., Hansen-Algenstaedt, N. et al. High migration rate of two types of threaded acetabular cups. Arch Orthop Trauma Surg 124, 17–25 (2004). https://doi.org/10.1007/s00402-003-0592-y

Download citation

  • Received:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00402-003-0592-y

Keywords

Navigation