Impact of post-dilatation with a focal expanding balloon for optimization of intracoronary stenting

  • ORIGINAL ARTICLE
  • Published:
Heart and Vessels Aims and scope Submit manuscript

Abstract

Optimal stenting frequently requires additional stent post-dilatation following initial stent deployment. Stent post-dilatation using a focal expanding balloon (FB) that grows 0.5 mm larger centrally may achieve a larger final stent lumen with fewer stent edge injuries as compared to use of a conventional unidiameter balloon (UB). In the present prospective study, of 128 stented lesions in 122 patients, 63 lesions had stents dilated with FB (Group F), while 65 lesions had stents dilated with UB (Group U). All balloons for stent post-dilatation were half-sized up to reference diameter by on-line quantitative coronary analysis. There were no differences in the reference diameter, pre-procedural minimal lumen diameter (MLD), balloon/artery ratio, and final balloon pressure between the two groups. Post-procedural MLD in Group F was significantly larger than that in Group U (3.03 ± 0.43 vs 2.80 ± 0.47 mm, P < 0.001). Stent edge injury occurred in 4 patients, and stent thrombosis in 2 patients in Group U, but not in Group F. Minimal lumen diameter at 6 months in Group F was significantly larger than that in Group U (2.05 ± 0.63 vs 1.82 ± 0.66 mm, P < 0.05), and incidence of restenosis was significantly lower in Group F than Group U (9% vs 22%, P < 0.05). By using a focal expanding balloon for stent optimization, a larger stent lumen can be obtained safely, and subsequent incidence of restenosis can be reduced.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. DL Fischman MB Leon DS Baim RA Schatz MP Savage I Penn K Detre L Veltri D Ricci M Nobuyoshi (1994) ArticleTitleA randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. Stent Restenosis Study Investigators N Engl J Med 331 496–501 Occurrence Handle8041414 Occurrence Handle10.1056/NEJM199408253310802 Occurrence Handle1:STN:280:DyaK2czhs1amsg%3D%3D

    Article  PubMed  CAS  Google Scholar 

  2. PW Serruys P de Jaegere F Kiemeneij C Macaya W Rustch G Heyndrickx H Emanuelsson J Marco V Legrand P Materne (1994) ArticleTitleA comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease. Benestent Study Group N Eng J Med 331 489–495 Occurrence Handle10.1056/NEJM199408253310801 Occurrence Handle1:STN:280:DyaK2czhs1anuw%3D%3D

    Article  CAS  Google Scholar 

  3. A Schömig A Kastrati H Mudra R Blasini H Scuhlen V Klauss G Richardt FJ Neumann (1994) ArticleTitleFour-year experience with Palmaz-Schatz stenting in coronary angioplasty complicated by dissection with threatened or present vessel closure Circulation 90 2716–2724 Occurrence Handle7994813

    PubMed  Google Scholar 

  4. A Schömig A Kastrati R Dietz B Rauch FJ Neumann HH Katus U Busch (1994) ArticleTitleEmergency coronary stenting for dissection during percutaneous transluminal coronary angioplasty: angiographic follow-up after stenting and after repeat angioplasty of the stented segment J Am Coll Cardiol 23 1053–1060 Occurrence Handle8144767 Occurrence Handle10.1016/0735-1097(94)90589-4

    Article  PubMed  Google Scholar 

  5. SL Goldberg A Colombo S Nakamura Y Almagor L Maiello JM Tobis (1994) ArticleTitleBenefit of ultrasound in the development of Palmaz-Schatz stent J Am Coll Cardiol 24 996–1003 Occurrence Handle7930236 Occurrence Handle1:STN:280:DyaK2M%2FhsFCmug%3D%3D Occurrence Handle10.1016/0735-1097(94)90861-3

    Article  PubMed  CAS  Google Scholar 

  6. KH Mak G Belli SG Ellis DJ Moliterno (1996) ArticleTitleSubacute stent thrombosis, evolving issues and current concepts J Am Coll Cardiol 27 494–503 Occurrence Handle8557927 Occurrence Handle10.1016/0735-1097(95)00466-1 Occurrence Handle1:STN:280:DyaK287lt1ehsA%3D%3D

    Article  PubMed  CAS  Google Scholar 

  7. I Moussa C Di Mario L Di Francesco B Reimers S Blengino A Colombo (1996) ArticleTitleSubacute stent thrombosis and the anticoagulation controversy, changes in drug therapy, operator technique, and the impact of the intravascular ultrasound Am J Cardiol 78 IssueIDSuppl 3A 13–17 Occurrence Handle8751841 Occurrence Handle10.1016/S0002-9149(96)00486-9 Occurrence Handle1:STN:280:DyaK28zlsVGgtg%3D%3D

    Article  PubMed  CAS  Google Scholar 

  8. GJ Karrilon MC Morice E Benveniste P Bunouf P Aubry S Cattan B Chevalier P Commeau A Cribier C Eiferman G Glollier Y Guerin M Henry T Lefevre B Livarek Y Lauvard J Marco S Makowski JP Monassier JM Pernes P Rouux C Spaulding G Zemour (1996) ArticleTitleIntracoronary stent implantation without ultrasound guidance and with replacement of conventional anticoagulation by antiplatelet therapy: 30-day clinical outcome of the French multicenter registry Circulation 94 1519–1527

    Google Scholar 

  9. RE Kuntz CM Gibson M Nobuyoshi DS Baim (1993) ArticleTitleGeneralized model of restenosis after conventional balloon angioplasty, stenting and directional atherectomy J Am Coll Cardiol 21 15–25 Occurrence Handle8417056 Occurrence Handle1:STN:280:DyaK3s7gslGjsw%3D%3D Occurrence Handle10.1016/0735-1097(93)90712-A

    Article  PubMed  CAS  Google Scholar 

  10. R Hoffman GS Mintz R Mehran AD Pichard KM Kent AF Satler JJ Popma W Hongsheng MB Leon (1998) ArticleTitleIntravascular ultrasound predictors of angiographical restenosis in lesions treated with Palmaz-Schatz stents J Am Coll Cardiol 31 43–39 Occurrence Handle10.1016/S0735-1097(97)00438-5

    Article  Google Scholar 

  11. GW Stone FG Goar JB Hodgson PJ Fitzgerald EL Alderman PG Yock J Coverdale H Sheehan TJ Linnemeier (1999) ArticleTitleAnalysis of the relation between stent implantation pressure and expansion Am J Cardiol 83 1397–1400 Occurrence Handle10235100 Occurrence Handle10.1016/S0002-9149(99)00104-6 Occurrence Handle1:STN:280:DyaK1M3kvFWkuw%3D%3D

    Article  PubMed  CAS  Google Scholar 

  12. JB Hodgson D Cumberland M Crocker (1995) ArticleTitlePreliminary results using a new “focal” balloon with an expanding central section J Am Coll Cardiol 25 346A Occurrence Handle10.1016/0735-1097(95)92891-8

    Article  Google Scholar 

  13. H Mudra F Werner E Regar V Klauss KH Henneke M Rothman C di Mario (1997) ArticleTitleOne balloon approach for optimized Palmaz-Schatz stent implantation, The MUSCAT trial Cathet Cardiovasc Diagn 42 130–136 Occurrence Handle9328693 Occurrence Handle10.1002/(SICI)1097-0304(199710)42:2<130::AID-CCD7>3.0.CO;2-E Occurrence Handle1:STN:280:DyaK2svmvFahsw%3D%3D

    Article  PubMed  CAS  Google Scholar 

  14. JB Hodgson (1997) ArticleTitleFocal Angioplasty, Theory and Clinical Application Cathet Cardiovasc Diagn 42 445–451 Occurrence Handle9408636 Occurrence Handle10.1002/(SICI)1097-0304(199712)42:4<445::AID-CCD27>3.0.CO;2-H Occurrence Handle1:STN:280:DyaK1c%2Fnt12gtg%3D%3D

    Article  PubMed  CAS  Google Scholar 

  15. PW Seruuys DP Foley RL Kirkeeide SB King (1993) ArticleTitleRestenosis revisited, insights provided by quantitative coronary angiography Am Heart J 126 1243–1267 Occurrence Handle10.1016/0002-8703(93)90689-7

    Article  Google Scholar 

  16. DP Foley PW Seruuys (1994) ArticleTitleRestenosis after percutaneous coronary intervention, the evolving angiographic perspectives Coron Art Dis 4 1129–1136

    Google Scholar 

  17. KJ Beatt PW Serruys PG Hugenholtz (1990) ArticleTitleRestenosis after coronary angioplasty, new standards for clinical studies J Am Coll Cardiol 15 491–498 Occurrence Handle2299089 Occurrence Handle1:STN:280:DyaK3c7jtFCqug%3D%3D Occurrence Handle10.1016/S0735-1097(10)80081-6

    Article  PubMed  CAS  Google Scholar 

  18. KJ Beatt HE Luijten PJ de Feyter M va de Brand PJ de Feyter JH Reiber PW Serruys (1988) ArticleTitleChanges in diameter of coronary artery segments adjacent to stenosis after percutaneous transluminal coronary angioplasty, failure of percent diameter stenosis measurement to reflect morphologic changes induced by balloon dilatation J Am Coll Cardiol 12 315–323 Occurrence Handle2969018 Occurrence Handle1:STN:280:DyaL1c3ntlyqsw%3D%3D Occurrence Handle10.1016/0735-1097(88)90400-7

    Article  PubMed  CAS  Google Scholar 

  19. DP Foley J Deckers AA van den Boss GR Heyndrickx GJ Larrman H Suryapranata F Zilstra HR Buller PW Serruys (1993) ArticleTitleIs there a need for repeat coronary angiography 24 h after successful balloon angioplasty to evaluate early luminal deterioration and facilitate quantitative analysis? Am J Cardiol 72 1341–1348 Occurrence Handle8256723 Occurrence Handle10.1016/0002-9149(93)90176-D Occurrence Handle1:STN:280:DyaK2c%2FnvV2jsg%3D%3D

    Article  PubMed  CAS  Google Scholar 

  20. WR Hermans DP Foley BJ Rensing PW Serruys (1994) ArticleTitleMorphologic changes during follow-up after successful percutaneous coronary balloon angioplasty, quantitative angiographic analysis in 778 lesions-further evidence for the restenosis paradox Am Heart J 127 483–494 Occurrence Handle8122593 Occurrence Handle10.1016/0002-8703(94)90654-8 Occurrence Handle1:STN:280:DyaK2c7mvVWiuw%3D%3D

    Article  PubMed  CAS  Google Scholar 

  21. RT Lee RD Kamm (1994) ArticleTitleVascular mechanism for cardiologist J Am Coll Cardiol 23 1289–1295 Occurrence Handle8176085 Occurrence Handle1:STN:280:DyaK2c3jtVyisg%3D%3D Occurrence Handle10.1016/0735-1097(94)90369-7

    Article  PubMed  CAS  Google Scholar 

  22. IK de ScheerDer K Wang EV Verbeken XR Zhou JH Piessens F van de Werf (1998) ArticleTitleStent deployment pressures defines the stent/vessel wall relationship and has important implications for early and late outcome J Invasive Cardiol 10 151–157 Occurrence Handle10762785

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Fumiaki Mori.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Mori, F., Tsurumi, Y., Hagiwara, N. et al. Impact of post-dilatation with a focal expanding balloon for optimization of intracoronary stenting. Heart Vessels 22, 152–157 (2007). https://doi.org/10.1007/s00380-006-0952-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00380-006-0952-y

Key words

Navigation