Abstract
Objectives
Our study comprised a single-center retrospective in vitro correlation between spectral properties, namely ρ/Z values, derived from scanning blood samples using dual-energy computed tomography (DECT) with the corresponding laboratory hemoglobin/hematocrit (Hb/Hct) levels and assessed the potential in anemia-detection.
Methods
DECT of 813 patient blood samples from 465 women and 348 men was conducted using a standardized scan protocol. Electron density relative to water (ρ or rho), effective atomic number (Zeff), and CT attenuation (Hounsfield unit) were measured.
Results
Positive correlation with the Hb/Hct was shown for ρ (r-values 0.37–0.49) and attenuation (r-values 0.59–0.83) while no correlation was observed for Zeff (r-values −0.04 to 0.08). Significant differences in attenuation and ρ values were detected for blood samples with and without anemia in both genders (p value < 0.001) with area under the curve ranging from 0.7 to 0.95. Depending on the respective CT parameters, various cutoff values for CT-based anemia detection could be determined.
Conclusion
In summary, our study investigated the correlation between DECT measurements and Hb/Hct levels, emphasizing novel aspects of ρ and Zeff values. Assuming that quantitative changes in the number of hemoglobin proteins might alter the mean Zeff values, the results of our study show that there is no measurable correlation on the atomic level using DECT. We established a positive in vitro correlation between Hb/Hct values and ρ. Nevertheless, attenuation emerged as the most strongly correlated parameter with identifiable cutoff values, highlighting its preference for CT-based anemia detection.
Clinical relevance statement
By scanning multiple blood samples with dual-energy CT scans and comparing the measurements with standard laboratory blood tests, we were able to underscore the potential of CT-based anemia detection and its advantages in clinical practice.
Key Points
-
Prior in vivo studies have found a correlation between aortic blood pool and measured hemoglobin and hematocrit.
-
Hemoglobin and hematocrit correlated with electron density relative to water and attenuation but not Zeff.
-
Dual-energy CT has the potential for additional clinical benefits, such as CT-based anemia detection.
Graphical Abstract
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Attenuation measurements in single energy computed tomography (SECT) for tissue identification and discrimination are recognized to possess substantial technical limitations. The overlap in attenuation of various tissue types is attributed to the similarity in linear attenuation coefficients, primarily influenced by two physical effects: photoelectric absorption and Compton scattering [1]. Photoelectric absorption refers to the energy of X-ray photons that interact with the tightly bound electrons of the inner shell, specifically, the K-shell. These interactions lead to the absorption of X-ray photons while electrons are ejected. Also, modifications of the scan protocol, especially changes in tube voltage and therefore photon energy, can significantly affect the values of attenuation. Accordingly, a valid comparison is only possible under standardized conditions [2,3,4,5].
The assessment of attenuation in SECT is comparable with the estimation of electron density relative to water (ρ or rho) using dual-energy CT (DECT) [6]. In addition to the attenuation application of DECT, it can provide the measurement of the effective atomic number (Zeff) which allows for an optimization of tissue discrimination on the atomic level. The atomic number Z corresponds to the number of protons in the nucleus, according to the atomic number of the periodic table of the elements. As the K-shell binding energy is proportional to the Z of an element, the photoelectric effect is proportional to the Z3 of the scanned material [7]. This measurement is possible since DECT uses two different X-ray energy spectra [1, 6, 8,9,10,11,12,13,14]. Modifications of the scan protocol in DECT also have an impact on ρ/Z estimations, which must be taken into account, but have a minor effect on the tissue discrimination [1, 15].
The idea of correlating hemoglobin (Hb) values with computed tomography (CT) findings has been discussed in literature before. The objective was mainly to quantify possible anemia and to improve treatment, for example in oncological patients. The literature describes some subjective findings like the aortic ring sign or the visualization of the interventricular septum in unenhanced CTs which suggest the presence of anemia [16, 17]. A linear correlation between the in vivo attenuation measured from the aortic blood pool and the laboratory-determined blood values like Hb and hematocrit (Hct) was found before in the literature [18,19,20,21,22,23,24,25,26].
These studies always depended on attenuation and additionally, these values were always measured in non-standardized conditions under the inclusion of both, native and contrast-enhanced examinations, as well as changing body configurations or scan protocols. The aim of our study was to correlate the Hb and Hct values with ρ/Z values derived under controlled in vitro conditions using spectral properties of DECT as an additive to attenuation values.
Materials and methods
This retrospective study was approved by the local institutional Ethics Committee (BASEC-Nr. 2023-00368). Written informed consent was waived by the local Ethics Committee.
Blood samples
A conservative effect size assumption was made (R² = 0.02) due to possible measurement errors from blood sedimentation. The power analysis resulted in a sample size of 813.
Our central laboratory has established a standardized collection, transportation, analysis, and storage (at 4 °C) of blood samples. The samples, contained in hematology tubes with ethylene diamine tetraacetic acid, Becton Dickinson Vacutainer ref. 368861 to prevent clotting, were promptly transported to the laboratory for clinical analysis. Blood samples were stored for 2 weeks, allowing for subsequent analysis if attending physicians required additional information. This storage period facilitated a retrospective use of the blood samples for our scanning. Samples from the past two days were provided anonymized by our central laboratory.
Blood values
The anonymized Hb and Hct values were provided by our laboratory after the completion of scans and post-processing. Identification was achieved through the combination of internal laboratory and study-specific identification numbers. The presence of anemia was evaluated by Hb values below 13 g/dL for men and 12 g/dL for women, according to the cutoff values established by the World Health Organization (WHO) [27].
Blood scanning
All blood samples provided by the central laboratory were already suitable for laboratory analysis, allowing us to include them, without the need for applying exclusion criteria. To ensure reliable identification of post-processed, anonymized samples, a standardized plastic grid arrangement was employed (Fig. 1). Since our main CT device, a dual-source DECT, was not applicable to this study, the DECT scans of the blood samples were conducted using a single source dual-energy CT system SOMATOM x.ceed (Siemens Healthineers). Two scan modes, dual-spiral (ds DECT) and twinbeam (tb DECT), were employed. Adhering to established literature we adopted a standardized scan protocol [1, 15, 19, 28,29,30,31], deactivating all automatic voltage/ tube current adaptations (refer to Table 1 for technical specifications). Immediately prior to each scanning, the blood samples underwent 20 gentle agitation cycles to counteract sedimentation as advised by the laboratory physician (Fig. 2).
Post-processing and measurements of attenuation, ρ, and Z values
All images were assessed using Syngo.via® software (Siemens Healthcare GmbH 2009-2021, VB60) for multimodality reading. Images were reconstructed using 3 mm slice thickness, employing a soft reconstruction kernel (Qr40) with a standardized 512 matrix. The specific DECT parameters (ρ and Zeff) were elicited using the “rho/Z” application profile from Syngo.via with adapted subcategory for ds and tb acquisition. A trained reader with 2.5 years of cross-sectional imaging expertise, managed subsequent processing of the scanned blood samples under consensus reading with a senior consultant radiologist (18 years of experience). After axial alignment, freehand dual-energy region of interest (ROI) was employed for the measurement of ρ or rho, Zeff, and attenuation (Hounsfield unit [HU]). To prevent artifacts, meticulous attention was given to the proper placement of the ROI at the tube’s central axis, avoiding wall contact (Fig. 3). For precise measurements ROIs were placed with two different orientations, and mean values were used for further evaluation.
Statistics
Hb and Hct values served as the independent variables, while attenuation, ρ, and Z measurements were designated as dependent variables. Descriptive statistics were used for reporting the DECT values. T-test for independent samples, ROC-analysis, calculation of Cohen’s kappa (k), and intraclass correlation coefficient (ICC, class 3), as well as linear regression analysis, were performed in JASP Team (2023; version 0.17.3) and Microsoft Excel® (Microsoft Office Professional Plus 2019).
Results
Blood samples
A study cohort of 813 samples, comprising 348 men and 465 women, spanning an age range of 17–98 years for men and 17–97 years for women were assessed. The mean age was 65.58 ± 17.38 years for men and 59.47 ± 20.4 years for women respectively.
Blood values
Examinations were conducted on 183 blood samples from men with laboratory-confirmed anemia (mean Hblab = 10.37 ± 1.74 g/dL) and 165 samples without confirmed anemia (mean Hblab = 14.65 ± 1.09 g/dL). Similarly, 244 blood samples from anemic women (mean Hblab = 9.99 ± 1.46 g/dL), and 221 without anemia (mean Hblab = 13.37 ± 1.03 g/dL) were included in the analyses.
Assessment of ρ/Z, attenuation, and blood values
Reliability tests (n = 50) demonstrated excellent intrareader reliability with high ICC (3,1) values of 0.93 (95% CI: 0.87–0.96) for ρ, 0.92 (95% CI: 0.86–0.95) for Z and 0.97 (95% CI: 0.94–0.98) for attenuation. A comparative analysis of CT measurements between dsDECT and tbDECT revealed significant differences in attenuation, ρ, and Z-measurements (p values < 0.0001) for both scan modes.
Positive correlations with Hb and hematocrit values with correlation coefficients (r-values) ranging from 0.59 to 0.83 for attenuation and 0.37 to 0.49 for ρ were found for both genders. Z values exhibited no significant correlation in either tb or ds CT (r-values −0.04 to 0.08). Detailed results are summarized in Table 2. The highest correlation coefficient was identified for the combination of attenuation and tbDECT, demonstrated in Fig. 4a for men and Fig. 4b for women.
Cutoff-determination for CT-based anemia detection
Due to the missing correlation of Z, no cutoff analysis was determined for this value.
Comparison of blood samples with and without laboratory-confirmed anemia revealed significant differences (p value < 0.001) in both genders for ρ and attenuation values (refer to Fig. 5a–d). Receiver operating characteristic (ROC) analysis was performed for ρ and attenuation (refer to Fig. 6a–b). The area under the curve ranges from 0.7 up to 0.95 for men and from 0.7 to 0.92 for women, depending on the acquisition technique and measured parameter. The optimal gender-specific cutoff values for ρ (Cρ) range from 39 to 56. For attenuation (Ca) cutoff values from 37 to 44 were determined. Detailed results, along with associated sensitivity and specificity for both DECT scan modes (ds and tb) are listed in Table 3 [32].
Discussion
Anemia is a hematologic condition characterized by a reduction in the number of circulating red blood cells (RBC) or their Hb content, diagnosed through estimation of Hb concentration usually in the laboratory [33, 34]. Anemia detection in CT datasets has the potential to identify clinically silent anemia, prompting further laboratory and clinical investigations.
We assumed a correlation with the atomic number Zeff. If Zeff is measured in a material consisting of only one element, it reflects the atomic number. For mixed materials, Zeff is determined by the respective elements and their number of electrons, resulting in an averaged value that cannot be assigned to a specific atomic number. In DECT, Zeff is determined using a coefficient that is calculated from previous calibration data, including ρ measurements, which, in turn, was considered within the utilized application profiles [14]. Zeff reflects the blood composition on an atomic level. Hb, an iron-containing oxygen-transport protein composed of four heme molecules and one globin protein, is present in erythrocytes up to about 96% of a red blood cell’s dry weight [35,36,37]. Even if quantitative changes in the number of Hb proteins might alter the mean Zeff values, the results of our study show no positive or negative correlation measurable using DECT (refer to Table 2).
It appears that anemia has no significant or discernible effect on the overall elemental blood composition on an atomic level measurable in CT, despite changes in RBC concentration. Overall, Zeff seems to be a good parameter for determining single elements or differentiating selective materials, but has limitations in complex liquid materials like blood [6, 12, 13, 38].
By using two energy spectra, the electron density relative to water (ρ, rho), comparable to the attenuation measurements in SECT, can be determined in DECT as well [15]. To our knowledge, we used ρ for the first time in the literature for a linear regression analysis between Hb and CT measurements. Therefore, we demonstrated a positive correlation between Hb and ρ (refer to Table 2). As the solid blood components mainly consist of RBC, and the hematocrit indicates the percentage of cellular blood components, an additional positive correlation between ρ and hematocrit was observed, as expected.
While ρ showed a positive correlation with the Hb/Hct value, within the scope of our analyses, the strongest correlation was observed for attenuation (refer to Table 2). This was evident in both tbDECT and dsDECT with overall better results for tb acquisition (refer to Fig. 4). Better correlation for attenuation may be because ρ measurements are based on calibration scans and calculations from calibration coefficients that depend on low- and high-energy DECT images [14].
Consequently, numerous studies have investigated the relationship between CT measurements and Hb/Hct levels before, establishing a positive correlation between the blood and attenuation values [16, 19, 21,22,23,24]. However, we observed some important limitations for some of them. First, we noticed variations in the scan protocols that may significantly affect attenuation measurements in SECT [4, 5]. Title et al, for example, used two different CT scanners with different tube voltages (120 kV and 140 kV) for their study population, and Chaudhry et al reported a peak tube voltage of a maximum of 120 kV [19, 22]. Jung et al and Zhou et al used care dose implementations like automatic tube current, with tube current values ranging from 30 milliampere seconds (mAs) up to 500 mAs [20, 21]. Also, physical patient variation like different body weights, influencing care dose implementations, was unavoidable with the in vivo study designs [26]. Additionally, the time delay between the CT examination and the laboratory determination of Hb/Hct is an important limitation, since the blood values are not stable parameters in living patients. Depending on the study, time latency was reported to range from hours, as in the study from Title et al, up to 7 days, as Zopfs et al used as inclusion criteria [21, 23, 24, 26].
Our approach with an in vitro study eliminates these influencing factors at once. On the one hand, DECT measurements are known to be more valid an show less susceptibility to changes in the scan protocol [1, 15]. On the other hand, we overcome the confounding factor of changing scan parameters through a standardized scan protocol (refer to Table 1). The standardized scan setting allowed us to exclude patient-related confounding factors. The fact that the Hb level is relatively stable in properly stored blood samples leads to a theoretical time delay between laboratory evaluation and a CT scan of zero [39, 40].
Nevertheless, the in vitro setting resulted in the limitation that, despite thorough agitation of the blood samples, a minor influence of the sedimentation rate cannot be ruled out. The study was conducted using one of the latest single-source CT devices available on the market, despite dual-source DECT devices representing the current state-of-the-art. It should be noted that this decision may slightly limit the comparison with clinical practice. A comparison between tb and ds acquisition techniques as possible; however, despite the standardized scan protocol, it should be noted that the tube voltage differed between the two scan modes, and therefore, the comparison must be critically evaluated. The superior results obtained with tbDECT, as mentioned before, remain inconclusively explained, but, besides the differed utilized tube voltage, a plausible reason may lie in the scanning procedure. While tbDECT requires a single scan to generate low- and high-energy images, dsDECT necessitates two scans for the same purpose. Although only a few seconds elapse between these scans, this latency may noticeably affect erythrocyte sedimentation within the measurement results. Another relevant limitation of our in vitro study is the grid arrangement we utilized, which proved to be the most practical for tilting the numerous blood samples and preventing blood sedimentation. The resultant lack of dense material surrounding the blood samples and the absence of beam hardening effects may restrict the comparability with in vivo studies.
Some studies have determined cutoff values for anemia detection using attenuation values as a predictor [21, 26]. Our aim with this in vitro study was to complement existing in vivo studies and provide a scientifically grounded foundation for future research in this area. Nevertheless, our statistical analysis indicates good differentiation between blood samples with and without anemia within the use of CT measurements (refer to Figs. 5 and 6). We identified several gender-specific ρ and attenuation cutoff values for ds and tb DECT. Given the better correlation between attenuation and Hb values, observed in our study while using DECT, which has also been shown in further studies with virtually non-contrast CT images or photon-counting CTs, attenuation seems to be the preferred parameter for computed tomographic anemia detection [25, 41]. But this, of course, does not replace the comprehensive laboratory analysis and anemia diagnostics and only can be used as a screening parameter.
In summary, our study investigated the correlation between DECT measurements and blood parameters, emphasizing novel aspects of ρ and Zeff values. The use of DECT offers less variation based on scan protocols and scanning blood samples addresses previous time latency issues. If quantitative changes in the Hb count might alter the mean Zeff values, the results of our study show that there is no measurable correlation using DECT. However, we conducted a linear regression analysis between Hb/Hct and ρ for the first time, demonstrating a positive correlation and indicating the potential of this parameter in further in vivo studies. Nevertheless, attenuation emerged as the most strongly correlated parameter with identifiable cutoff values, especially in tb DECT, highlighting its preference for CT-based anemia detection.
Abbreviations
- CT:
-
Computed tomography
- DECT:
-
Dual-energy CT
- ds:
-
Dual-spiral
- Hb:
-
Hemoglobin
- Hct:
-
Hematocrit
- HU:
-
Hounsfield unit
- mAs:
-
Milliampere seconds
- RBC:
-
Red blood cells
- ROC:
-
Receiver operating characteristic
- ROI:
-
Region of interest
- SECT:
-
Single-energy CT
- tb:
-
Twinbeam
- Zeff :
-
Effective atomic number
- ρ or rho:
-
Electron density relativ to water
References
Chatzaraki V, Bolsi A, Kubik-Huch RA et al (2022) Influence of radiation dose, photon energy, and reconstruction kernel on rho/Z analysis in spectral computer tomography: a phantom study. In Vivo 36:678–686
Shafiq-Ul-Hassan M, Zhang GG, Hunt DC et al (2018) Accounting for reconstruction kernel-induced variability in CT radiomic features using noise power spectra. J Med Imaging (Bellingham) 5:011013/1–01101314
Vernuccio F, Cannella R, Bartolotta TV, Galia M, Tang A, Brancatelli G (2021) Advances in liver US, CT, and MRI: moving toward the future. Eur Radiol Exp 5:1–16
Levi C, Gray JE, McCullough EC, Hattery RR (1982) The unreliability of CT numbers as absolute values. AJR Am J Roentgenol 139:443–447
Yuan R, Mayo JR, Hogg JC et al (2007) The effects of radiation dose and CT manufacturer on measurements of lung densitometry. Chest 132:617–623
Landry G, Reniers B, Granton PV et al (2011) Extracting atomic numbers and electron densities from a dual source dual energy CT scanner: experiments and a simulation model. Radiother Oncol 100:375–379
Marin D, Boll DT, Mileto A, Nelson RC (2014) State of the art: dual-energy CT of the abdomen. Radiology 271:327–342
Rutherford RA, Pullan BR, Isherwood I (1976) Measurement of effective atomic number and electron density using an EMI scanner. Neuroradiology 11:15–21
Petersilka M, Bruder H, Krauss B, Stierstorfer K, Flohr TG (2008) Technical principles of dual source CT. Eur J Radiol 68:362–368
Fornaro J, Leschka S, Hibbeln D et al (2011) Dual- and multi-energy CT: approach to functional imaging. Insights Imaging 2:149–159
Hünemohr N, Krauss B, Tremmel C, Ackermann B, Jäkel O, Greilich S (2014) Experimental verification of ion stopping power prediction from dual energy CT data in tissue surrogates. Phys Med Biol 59:83–96
Hünemohr N, Paganetti H, Greilich S, Jäkel O, Seco J (2014) Tissue decomposition from dual energy CT data for MC based dose calculation in particle therapy. Med Phys 41:061714/1–06171414
Tatsugami F, Higaki T, Kiguchi M et al (2014) Measurement of electron density and effective atomic number by dual-energy scan using a 320-detector computed tomography scanner with raw data-based analysis: a phantom study. J Comput Assist Tomogr 38:824–827
Bharati A, Mandal SR, Gupta AK et al (2019) Development of a method to determine electron density and effective atomic number of high atomic number solid materials using dual-energy computed tomography. J Med Phys 44:49–56
Chatzaraki V, Born C, Kubik-Huch RA, Froehlich JM, Thali MJ, Niemann T (2021) Influence of radiation dose and reconstruction kernel on fat fraction analysis in dual-energy CT: a phantom study. In Vivo 35:3147–3155
Corcoran HL, Cook DE, Proto AV (1988) Diagnosis of anemia on computed tomography scans of the thorax. J Comput Tomogr 12:116–121
Desaint P, Bernard C, Contou D (2022) Diagnosing anemia with chest CT-scan: the aortic ring sign. Am J Med Sci 363:e17
Black DF, Rad AE, Gray LA, Campeau NG, Kallmes DF (2011) Cerebral venous sinus density on noncontrast CT correlates with hematocrit. AJNR Am J Neuroradiol 32:1354–1357
Chaudhry AA, Gul M, Chaudhry A, Sheikh M, Dunkin J (2015) Quantitative evaluation of noncontrast computed tomography of the head for assessment of anemia. J Comput Assist Tomogr 39:842–848
Jung C, Groth M, Bley TA et al (2012) Assessment of anemia during CT pulmonary angiography. Eur J Radiol 81:4196–4202
Zhou Q-Q, Yu Y-S, Chen Y-C et al (2018) Optimal threshold for the diagnosis of anemia severity on unenhanced thoracic CT: a preliminary study. Eur J Radiol 108:236–241
Title RS, Harper K, Nelson E, Evans T, Tello R (2005) Observer performance in assessing anemia on thoracic CT. AJR Am J Roentgenol 185:1240–1244
Lan H, Nishihara S, Nishitani H (2010) Accuracy of computed tomography attenuation measurements for diagnosing anemia. Jpn J Radiol 28:53–57
Zopfs D, Rinneburger M, Pinto Dos Santos D et al (2021) Evaluating anemia using contrast-enhanced spectral detector CT of the chest in a large cohort of 522 patients. Eur Radiol 31:4350–4357
Decker JA, Huber A, Senel F et al (2022) Anemia detection by hemoglobin quantification on contrast-enhanced photon-counting CT data sets. Radiology 305:650–652
Kamel EM, Rizzo E, Duchosal MA et al (2008) Radiological profile of anemia on unenhanced MDCT of the thorax. Eur Radiol 18:1863–1868
World Health Organization (2011) Haemoglobin concentrations for the diagnosis of anaemia and assessment of severity. World Health Organization. https://iris.who.int/handle/10665/85839
Mavridis S, El-Gedaily M, Kubik-Huch RA et al (2023) Assessment of image quality in chest CT with precision matrix and increased framing rate using single source CT: a phantom study. In Vivo 37:99–105
Obmann MM, Kelsch V, Cosentino A, Hofmann V, Boll DT, Benz MR (2019) Interscanner and intrascanner comparison of virtual unenhanced attenuation values derived from twin beam dual-energy and dual-source, dual-energy computed tomography. Invest Radiol 54:1–6
Euler A, Obmann MM, Szucs-Farkas Z et al (2018) Comparison of image quality and radiation dose between split-filter dual-energy images and single-energy images in single-source abdominal CT. Eur Radiol 28:3405–3412
Pourvaziri A, Parakh A, Cao J et al (2022) Comparison of four dual-energy CT scanner technologies for determining renal stone composition: a phantom approach. Radiology 304:580–589
Unal I (2017) Defining an optimal cut-point value in ROC analysis: an alternative approach. Comput Math Methods Med 2017:1–14
Chaparro CM, Suchdev PS (2019) Anemia epidemiology, pathophysiology, and etiology in low- and middle-income countries. Ann N Y Acad Sci 1450:15–31
Kuhn, Diederich V, Keller TCS L et al (2017) Red blood cell function and dysfunction: redox regulation, nitric oxide metabolism, anemia. Antioxid Redox Signal 26:718–742
Hall WD, Hurst JW, Walker HK (1990) Hemoglobin and hematocrit. Butterworths, Boston
Huehns ER, Shooter EM (1965) Human haemoglobins. J Med Genet 2:48–90
Weed RI, Reed CF, Berg G (1963) Is hemoglobin an essential structural component of human erythrocyte membranes? J Clin Invest 42:581–588
Nanako O, Shuhei S, Kentaro I et al (2017) Evaluation of urinary stone composition and differentiation between urinary stones and phleboliths using single-source dual-energy computed tomography. Acta Med Okayama 71:91–96
Heil W, Heins M, Withold W (1995) Storage of serum or whole blood samples? Effects of time and temperature on 22 serum analytes. Eur J Clin Chem Clin Biochem 1995:231–238
Guder WG, Narayanan S, Wisser H, Zawta B (2008) Samples: from the patient to the laboratory: the impact of preanalytical variables on the quality of laboratory results. Wiley-VCH, Weinheim
Decker JA, Huber A, Senel F et al (2023) Anemia detection and quantification in contrast-enhanced CT scans on a novel photon-counting detector CT: a prospective proof-of-concept study. Eur J Radiol 166:110967. 1–7
Acknowledgements
Special thanks to Hans-Ruedi Schmid, for generously supplying the blood samples.
Funding
The authors state that this work has not received any funding. Open access funding provided by University of Zurich.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Guarantor
The scientific guarantor of this publication is the corresponding author Tilo Niemann.
Conflict of interest
Bastian Schulz, MD, received a scientific grant from Guerbet AG, Zurich, Switzerland. This company had no role in the design of the study, in the collection, analyses, or interpretation of data, in the writing of the manuscript, or in the decision to publish the results. The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.
Statistics and biometry
No complex statistical methods were necessary for this paper.
Informed consent
Written informed consent was waived by the local Ethics Committee “Ethikkomission Nordwest-und Zentralschweiz”.
Ethical approval
This retrospective study was performed in line with the principles of the Declaration of Helsinki and under the legal guidelines of the “Bundesgesetz über die Forschung am Menschen (Humanforschungsgesetz, HFG). The questionnaire and methodology for this retrospective study were approved by the local Ethics Committee “Ethikkomission Nordwest- und Zentralschweiz” (22.03.2023, BASEC-Nr. 2023-00368).
Study subjects or cohorts overlap
The authors are not aware of any overlap in study subjects or cohorts.
Methodology
-
Single-center retrospective in vitro correlation study.
Additional information
Publisher’s Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Schulz, B., Euler, A., Schmid, HR. et al. In vitro blood sample assessment: investigating correlation of laboratory hemoglobin and spectral properties of dual-energy CT measurements (ρ/Z). Eur Radiol (2024). https://doi.org/10.1007/s00330-024-10820-6
Received:
Revised:
Accepted:
Published:
DOI: https://doi.org/10.1007/s00330-024-10820-6