Abstract
Objectives
This study investigated the progression of pubofemoral distance (PFD) with age and assessed the correlation between PFD and late acetabular index (AI) measurements.
Methods
This prospective observational study was conducted between January 2017 and December 2021. We enrolled 223 newborns who underwent the first, second, and third hip ultrasounds, and pelvis radiograph at a mean age of 18.6 days, 3.1 months, 5.2 months, and 6.8 months, respectively. The difference between PFD measured at serial ultrasounds and the correlation with AI were analyzed.
Results
The PFD increased significantly (p < 0.001) at serial measurements. The mean PFD at the first, second, and third ultrasounds were 3.3 (2.0–5.7), 4.3 (2.9–7.2), and 5.1 (3.3–8.0) mm, respectively. The PFD at three ultrasounds were all significantly (p < 0.001) and positively correlated with AI, with the Pearson correlation coefficients being 0.658, 0.696, and 0.753 for the first, second, and third ultrasounds, respectively. Using AI as reference, the diagnostic ability of PFD was calculated by the areas under the receiver operating characteristic curve, which was 0.845, 0.902, and 0.938 for the first, second, and third PFD, respectively. For the first, second, and third ultrasounds, PFD cutoff values of ≥ 3.9, ≥ 5.0, and ≥ 5.7 mm, respectively, yielded the greatest sensitivity and specificity in predicting late abnormal AI.
Conclusion
The PFD naturally progresses with age and is positively correlated with AI. The PFD has potential for predicting residual dysplasia. However, the cutoff for abnormal PFD values may require adjustment according to the patient’s age.
Key Points
• The pubofemoral distance measured in hip ultrasonography naturally increases as the infant’s hips mature.
• The early pubofemoral distance demonstrates a positive correlation with late acetabular index measurements.
• The pubofemoral distance may help physicians predict abnormal acetabular index. However, the cutoff for abnormal pubofemoral distance values may require adjustment according to patient’s age.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Developmental dysplasia of the hip (DDH) is an abnormal development of the osseous and soft tissue of the hip joint, leading to insufficient coverage of the femoral head. DDH has varying clinical manifestations and typically presents as mild dysplasia, subluxation, or frank dislocation. The reported incidence of DDH is 3–5 per 1000 newborns [1, 2]. The standard screening strategy for DDH in infants before the age of 4 months involves universal physical examination for hip stability followed by selective hip ultrasonography for newborns with either abnormal physical examination findings or risk factors [2, 3]. The Graf method and the femoral head coverage (FHC) ratio are the most extensively used ultrasonography parameters for DDH screening; both tools provide quantitative results and accurate diagnoses [4,5,6]. After age of 4 months, femoral head ossification starts and the acetabular index (AI) measured from anteroposterior pelvic radiography can be used to evaluate acetabular morphology.
Couture et al and Tréguier et al have recently developed a simple ultrasound method based on the pubofemoral distance (PFD), which may be adopted as an adjunctive tool for DDH screening [7, 8]. The PFD can be an accessional ultrasound parameter in addition to the Graf method and FHC ratio to serve as a fast, low-cost, highly sensitive, and specific screening tool [7,8,9,10,11]. Moreover, the PFD can be used to assist Graf method and FHC ratio to evaluate the effectiveness of hip reduction and to monitor patients underwent Pavlik harness treatment with high intra- and inter-observer reliability [12, 13].
Early ultrasound findings may predict late AI measurements. Studies have extensively explored the correlation between ultrasound and AI measurements for DDH screening; nevertheless, most of such studies have focused on ultrasound screening performed by the Graf method and the FHC ratio [14,15,16,17,18,19,20]. Evidence regarding the correlation between the PFD and AI is relatively scarce. Whether an increase in PFD is associated with an increase in AI is uncertain. In addition, similar to the alpha angle in the Graf method and the FHC ratio, the PFD may differ with age. Although studies using the Graf method and the FHC ratio for screening have adequately documented spontaneous improvements in hip morphology, the natural history of the PFD is less extensively discussed and remains unclear [21,22,23,24].
Accordingly, the primary objective of this study was to investigate the progression of the PFD with age in consecutive infants. The secondary objective was to assess the correlation between the PFD and late AI measurements.
Materials and methods
This prospective observational study was conducted from January 1, 2017, to December 31, 2021. We included all newborns who were delivered at our institution or were referred to our hospital for DDH. Newborns with neurogenic dislocation, syndromic dislocation, or a history of Pavlik harness treatment were excluded. The study was performed in accordance with the ethical principles set out in the 1964 Declaration of Helsinki and was approved by the institutional review board at our institution (IRB number 2017–06-010AC).
A neonatologist performed the Barlow and Ortolani maneuvers on all newborns to screen for hip stability [25, 26]. Hip status was categorized according to the Barlow and Ortolani test results into three groups: stable hips (involving a fixed position of the hip center during examination), subluxatable hips (involving any laxity of the hip center without dislocation), and dislocatable or dislocated hips (involving complete displacement of the hip center from the acetabulum). Hips that were considered to be subluxatable required confirmation from two senior pediatric orthopedic surgeons. In addition to physical examination findings, risk factors such as the female sex, twin pregnancy, firstborn birth order, breech presentation, or a family history of DDH were documented [2].
All newborns were divided into three groups according to their Barlow and Ortolani test results and risk factor status: group 1, comprising newborns who had stable hips but had DDH risk factors; group 2, comprising newborns with subluxatable, dislocatable, or dislocated hips regardless of the presence of risk factors; and group 3, comprising newborns with stable hips and no risk factors (Fig. 1).
After informed consent was obtained from the parents, the newborns in groups 1 and 2 were subjected to serial ultrasound examinations of the bilateral hips. The first ultrasound examination was performed immediately after the hip physical examination and risk factor survey in the newborn nursery. The second and third ultrasound examinations, conducted at the outpatient clinic, monitored hip morphology at the age of approximately 3 and 5 months.
A pediatric orthopedic surgeon certified in musculoskeletal ultrasound examination performed the hip sonography using a 7.5-MHz linear transducer (LOGIQ e ultrasound, GE Healthcare) positioned perpendicular to the hip joint, with no tilting during the examination. Each of the patients was placed in the lateral decubitus position, with the hips slightly flexed, adducted, and internally rotated throughout the ultrasound examination [27]. The criteria for the selection of qualified hip ultrasound images were as follows: inclusion of the lower iliac margin at the triradiate cartilage, the labrum, the deepest point of the acetabulum, the pubis, and the chondroosseous border of the proximal femur [27]. For each indicated participant, the most representative image out of three repeated image acquisitions at the bilateral hips was selected for ultrasound interpretation.
The PFD was measured as the shortest distance between the pubic bone and the medial margin of the femoral epiphysis (Fig. 2) [7, 8]. Couture and Tréguier performed their ultrasound examination in the supine position when they introduced the PFD; nevertheless, subsequent studies have demonstrated that the PFD measured in the lateral decubitus position can also yield high sensitivity and specificity in detecting hip instability [10]. At our institution, ultrasound examinations are performed in the lateral decubitus position in accordance with Graf’s guidelines for hip sonogram positioning [27].
An anteroposterior plain film radiograph of the pelvis was conducted for each of the patients at the age of 6 months in order to monitor their hip development. The AI is defined as the angle between the Hilgenreiner line and a line drawn from the triradiate epiphysis to the lateral edge of the acetabulum (Fig. 3). Although most current studies establish the normal AI range based on various ages, it should be noted that there is evidence that suggests the female may have higher AI than male before age of 7 years [28]. The mean AI for females is approximately 3° higher than for males when measured at age of 6 months to 1 year [28]. Based on the definition for normal AI range from most literatures, we used the mean AI for infants at the same age to determine abnormal AI regardless of patient’s sex in this study. An AI value exceeding 23.5° at 6 months of age is considered abnormal [29].
A senior pediatric orthopedic surgeon measured all PFD values in this study. A pediatric orthopedic fellow who was blinded to the ultrasound findings measured all AI values. To ensure consistency of measurements for the two observers, one observer measured the PFD for anonymized sonograms captured from 50 hips, and the other observer measured the AI for anonymized radiograph from the same image set. The same observers, blinded to the patients’ information and their previous results, repeated the image interpretation for the same 50 hips 3 weeks later. The results showed high consistency between two measurements for two observers, with the intraclass correlation coefficient being 0.95 (95% confidence interval [CI]: 0.91–0.99) for the PFD and 0.93 (95% CI: 0.90–0.96) for the AI measurement.
The images from patients who completed three ultrasound examinations and the pelvic plain films captured at 6 months were analyzed. The primary outcome measure was the difference between the PFD values measured from three hip ultrasound images. The secondary outcome measure was the correlation of the PFD values measured from three hip ultrasounds with the AI measured at the age of 6 months, and the cutoff for abnormal PFD values using AI as the reference standard.
Statistical analysis
A paired-sample t-test was used to compare the PFD values measured on three consecutive hip ultrasound images. A two-sample t-test was used to compare the PFD values between the male and the female groups. The Pearson correlation coefficient was calculated to determine the correlation between the PFD and AI measurements. A receiver operating characteristic (ROC) curve analysis was performed to investigate the diagnostic ability and cutoff for abnormal PFD values, with the AI serving as the reference standard. Statistical significance was defined as p < 0.05. All statistical analyses were performed using SPSS (version 22; IBM Corp.).
Results
From January 2017 to December 2021, a total of 4637 newborns delivered at our institution and 88 infants referred to our institution underwent Barlow and Ortolani tests and a risk factor survey for DDH screening. A total of 1652 newborns who had stable hips but had risk factors and 495 newborns with abnormal physical examination findings underwent the first ultrasound examination for the bilateral hips (Fig. 1). We excluded 51 patients for the following reasons: receiving Pavlik harness treatment (n = 44), having syndromic dislocation (n = 3), and having neurogenic dislocation (n = 4). Moreover, 1423 and 845 patients subsequently underwent the second and third hip ultrasound examinations, respectively, and 223 patients finally underwent anteroposterior pelvic radiography at the age of 6 months (Fig. 1). A total of 223 patients completed three serial hip ultrasounds and the pelvis plain film were included in our final analysis. The mean ages at the first, second, and third ultrasound examinations, and pelvis plain film were 18.6 (0–22) days, 3.1 (2.5–4.2) months, 5.2 (4.2–6.1) months, and 6.8 (5.8–7.6) months, respectively. The characteristics of enrolled patients are summarized in Table 1.
Regarding the PFD measured on three serial hip ultrasounds, the mean PFD increased significantly (p < 0.001) between each measurement (Fig. 4). The mean PFD of the first, second, and third ultrasounds were 3.3 (2.0–5.7), 4.3 (2.9–7.2), and 5.1 (3.3–8.0) mm, respectively (Table 2). The trend of naturally increased PFD was observed in both boys and girls; however, the mean PFD were not significantly different between sex in three serial ultrasounds (3.1 vs. 3.4 mm, p = 0.31; 4.3 vs. 4.2 mm, p = 0.47; 5.0 vs. 5.2 mm, p = 0.48, Table 2).
The PFD values measured on three consecutive hip ultrasounds all exhibited a significant positive correlation with the AI measured at the age of 6 months. The Pearson correlation coefficients for the PFD measured on the first, second, and third hip ultrasounds were 0.658 (p < 0.001), 0.696 (p < 0.001), and 0.753 (p < 0.001), respectively. The correlation between the measured PFD values and AI is illustrated in Fig. 5.
Using the AI at the age of 6 months as the reference standard, the diagnostic ability of the PFD was assessed by using ROC curve analysis (Fig. 6a). The PFD values from three ultrasound exams all exhibited high diagnostic performance in predicting AI. The areas under the ROC curve (AUC) for the PFD measured on the first, second, and third hip ultrasounds were 0.845 (95% CI: 0.775–0.915), 0.902 (95% CI: 0.839–0.965), and 0.938 (95% CI: 0.888–0.988), respectively.
The ROC curve analysis was further performed for boys and girls (Fig. 6b, c). The PFD value from three hip ultrasounds demonstrated high diagnostic ability to predict abnormal AI in both genders. For the male, the AUC for the PFD measured on the first, second, and third ultrasounds were 0.795 (95% CI: 0.670–0.920), 0.857 (95% CI: 0.734–0.979), and 0.905 (95% CI: 0.808–1.0), respectively. For the female, the AUC for the PFD measured on the first, second, and third ultrasounds were 0.883 (95% CI: 0.808–0.957), 0.935 (95% CI: 0.875–0.994), and 0.963 (95% CI: 0.915–1.0), respectively.
The cutoff value of PFD to exhibit greatest sensitivity and specificity to detect abnormal AI measurement based on age and sex distribution is listed in Table 2.
Discussion
In this study, the PFD significantly increased with age in newborns. PFD values measured on three serial hip ultrasounds all exhibited a significant positive correlation with AI measured at the age of 6 months, and the early PFD value may be served as an effective tool for predicting late abnormal AI measurements.
Studies have extensively reported that ultrasound parameters used for DDH screening improved with infant hip maturation [21, 22, 24]. However, most of such studies have focused on changes in ultrasound features used in the Graf method and FHC ratio. Although most studies have demonstrated that the alpha angle and FHC ratio increased naturally with age, few studies have considered the natural history of the PFD. Ban et al examined 240 infants aged 0–12 months who had Graf type Ia or Ib hips. They reported that the PFD increased linearly with age [23]. However, their study design involved performing ultrasound examinations for different infants with various ages. Their results suggested that older infants may have higher PFD values, but they did not reveal the natural progression of the PFD or the trend of the PFD with age among the infants.
On the basis of our results, we suggest that PFD values could be affected by the age at ultrasound acquisition. Because the PFD may naturally increase with age, PFD values observed at different ages may account for the various cutoff values of abnormal PFD measurements in previous studies. Tréguier et al defined the cutoff for abnormal PFD values to be 6 mm from patients with varying ages, ranging from < 28 to > 90 days [8]. Teixeria et al proposed that the cutoff for abnormal PFD values should be 4.9 mm when measured in hip flexion, 4.6 mm when measured in a neutral position, or asymmetry > 1.5 mm to contralateral hip [11]. Husum et al suggested that the cutoff for abnormal PFD values should be > 4.4 mm when measured in the lateral decubitus position [10]. Both studies conducted by Teixeria et al and Husum et al have established thresholds for abnormal PFD values in infants aged approximately 1 month. Motta et al reported a mean PFD of 3.5 mm for dysplastic hips in infants aged from 3 days to 6 months [9]. The heterogeneity of patient age in the aforementioned studies may explain the inconsistency for the cutoff values of abnormal PFD. Moreover, whether the patient’s gender can affect PFD interpretation is unknown from current literatures. From our results, we did not observe statistical difference of PFD between sex from three serial ultrasounds. However, the cutoff value for abnormal PFD was slightly different between boys and girls in the ultrasounds performed at various age. The PFD distribution in different gender may require more evidence by research with larger sample scale and more long-term follow-up.
Studies on the correlation between PFD and AI measurements are limited in the previous research. As demonstrated in the literature, the alpha angle and FHC ratio are negatively correlated with AI measurements; that is, patients with a lower alpha angle or FHC ratio are more likely to have a higher AI [14,15,16,17,18,19,20]. Nonetheless, previous studies have not clarified whether the same correlation can be observed between PFD and AI measurements. Previous research has defined abnormal PFD values by using the Graf method as the reference standard [8,9,10,11]; hence, evidence regarding the association between PFD and AI measurements is extremely limited in the literature. According to our review of current evidence, no large-scale study or quantitative evidence confirming the correlation between PFD and AI measurements was available prior to the present study.
The main strength of this study is that it provides information about the natural progression of the PFD and its correlation with late AI measurements in consecutive infants. The PFD is a relatively novel ultrasound parameter for DDH screening compared with the Graf method and the FHC ratio. Hence, according to our review of the literature, our study may be the first to further discover the natural history of the PFD and its association with radiographic findings. Our results suggest that normal PFD values may vary depending on the timing of observation and that the PFD has the potential to predict residual dysplasia. Further research with a larger sample size and longer follow-up period can more clearly identify the precise cutoff for abnormal PFD values at various ages.
Our study has several limitations. First, this is a study with a relatively small sample size and a short follow-up period. Second, lack of routine dynamic ultrasound examinations for DDH screening may have affected the PFD interpretation. Finally, we did not calculate the thickness of the pulvinar or pubic cartilage separately when measuring the PFD; one study suggested that both factors should be considered when interpreting PFD results [30].
In conclusion, the PFD naturally progresses with age. The PFD not only demonstrates a significant positive correlation with AI measurements but also has the potential for predicting residual dysplasia. The patient’s age may need to be considered when determining the cutoff for abnormal PFD values. In conjunction with Graf method and FHC ratio, the PFD measurement can help current ultrasound method for DDH screening.
Abbreviations
- AI:
-
Acetabular index
- DDH:
-
Developmental dysplasia of the hip
- FHC:
-
Femoral head coverage
- PFD:
-
Pubofemoral distance
References
Dezateux C, Rosendahl K (2007) Developmental dysplasia of the hip. Lancet 369:1541–1552
Mulpuri K, Song KM, Goldberg MJ, Sevarino K (2015) Detection and nonoperative management of pediatric developmental dysplasia of the hip in infants up to six months of age. J Am Acad Orthop Surg 23:202–205
Kotlarsky P, Haber R, Bialik V, Eidelman M (2015) Developmental dysplasia of the hip: what has changed in the last 20 years? World J Orthop 6:886–901
Omeroğlu H (2014) Use of ultrasonography in developmental dysplasia of the hip. J Child Orthop 8:105–113
Harcke HT, Pruszczynski B (2017) Hip ultrasound for developmental dysplasia: the 50% rule. Pediatr Radiol 47:817–821
American Institute of Ultrasound in Medicine; American College of Radiology (2013) AIUM practice guideline for the performance of an ultrasound examination for detection and assessment of developmental dysplasia of the hip. J Ultrasound Med 32:1307–1317
Couture A, Baud C, Prodhomme O, Saguintaah M, Veyrac C (2011) Ultrasound of the neonatal hip: initial evaluation and follow-up. J Radiol 92:142–165
Tréguier C, Chapuis M, Branger B et al (2013) Pubo-femoral distance: an easy sonographic screening test to avoid late diagnosis of developmental dysplasia of the hip. Eur Radiol 23:836–844
Motta GGB, Chiovatto ARS, Chiovatto ED et al (2022) Measurement of pubofemoral distance in the diagnosis of developmental dysplasia of the hip: sensitivity and specificity. J Ultrasound Med 41:1205–1212
Husum HC, Hellfritzsch MB, Hardgrib N, Møller-Madsen B, Rahbek O (2019) Suggestion for new 4.4 mm pubo-femoral distance cut-off value for hip instability in lateral position during DDH screening. Acta Orthop 90:88–93
Teixeira SR, Dalto VF, Maranho DA, Zoghbi-Neto OS, Volpon JB, Nogueira-Barbosa MH (2015) Comparison between Graf method and pubo-femoral distance in neutral and flexion positions to diagnose developmental dysplasia of the hip. Eur J Radiol 84:301–306
Luan Q, Teng J, Shi M et al (2022) Use of ultrasonography to evaluate early outcomes of reduction in developmental dysplasia of the hip. Pediatr Radiol 52:1521–1527
Maranho DA, Donati FN, Dalto VF, Nogueira-Barbosa MH (2017) The pubo-femoral distance decreases with Pavlik harness treatment for developmental dysplasia of the hip in newborns. Skeletal Radiol 46:1201–1207
Morin C, Harcke HT, MacEwen GD (1985) The infant hip: real-time US assessment of acetabular development. Radiology 157:673–677
Terjesen T, Rundén TO, Tangerud A (1989) Ultrasonography and radiography of the hip in infants. Acta Orthop Scand 60:651–660
Dornacher D, Cakir B, Reichel H, Nelitz M (2010) Early radiological outcome of ultrasound monitoring in infants with developmental dysplasia of the hips. J Pediatr Orthop B 19:27–31
Pillai A, Joseph J, McAuley A, Bramley D (2011) Diagnostic accuracy of static Graf technique of ultrasound evaluation of infant hips for developmental dysplasia. Arch Orthop Trauma Surg 131:53–58
Copuroglu C, Ozcan M, Aykac B, Tuncer B, Saridogan K (2011) Reliability of ultrasonographic measurements in suspected patients of developmental dysplasia of the hip and correlation with the acetabular index. Indian J Orthop 45:553–557
Atalar H, Dogruel H, Selek H, Tasbas BA, Bicimoglu A, Gunay C (2013) A comparison of ultrasonography and radiography in the management of infants with suspected developmental dysplasia of the hip. Acta Orthop Belg 79:524–529
Spaans AJ, Beek FJA, Uiterwaal C, Pruijs JEH, Sakkers RJ (2019) Correlation between ultrasonic and radiographic imaging of developmental dysplasia of the hip. J Child Orthop 13:155–160
Castelein RM, Sauter AJ, de Vlieger M, van Linge B (1992) Natural history of ultrasound hip abnormalities in clinically normal newborns. J Pediatr Orthop 12:423–427
Schwend RM, Shaw BA, Segal LS (2014) Evaluation and treatment of developmental hip dysplasia in the newborn and infant. Pediatr Clin North Am 61:1095–1107
Ban Y, Luan Q, Shi M et al (2021) Establishing reference values for the pubofemoral distance in normal infant medial hips by ultrasound. Acta Radiol 62:551–556
Riad JP, Cundy P, Gent RJ, Piotto L, Morris L, Hirte C (2005) Longitudinal study of normal hip development by ultrasound. J Pediatr Orthop 25:5–9
Barlow TG (1962) Early diagnosis and treatment of congenital dislocation of the hip. J Bone Joint Surg Br 44(2):292–301
Ortolani MJP (1937) Un segno poco noto e sua importanza per la diagnosi precose di prelussazione congenita dell’anca. Pediatria 45:129–136
Graf R (1980) The diagnosis of congenital hip-joint dislocation by the ultrasonic Combound treatment. Arch Orthop Trauma Surg 97:117–133
Akel I, Songür M, Karahan S, Yilmaz G, Demirkıran HG, Tümer Y (2013) Acetabular index values in healthy Turkish children between 6 months and 8 years of age: a cross-sectional radiological study. Acta Orthop Traumatol Turc 47(1):38–42
Noordin S, Umer M, Hafeez K, Nawaz H (2010) Developmental dysplasia of the hip. Orthop Rev (Pavia) 2:e19
Printemps C, Cousin I, Le Lez SS et al (2021) Pulvinar and pubic cartilage measurements to refine universal ultrasound screening for developmental dysplasia of the hip: Data from 1896 infant hips. Eur J Radiol 139:109727
Acknowledgements
The authors acknowledge the statistical support of the Research Center of Biostatistics, Taipei Veterans General Hospital.
Funding
The authors state that this work has not received any funding.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Guarantor
The scientific guarantor of this publication is Yu-Ping Su, MD, PhD, the corresponding author of this study.
Conflict of interest
The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.
Statistics and biometry
The authors acknowledge the statistical support of the Research Center of Biostatistics, Taipei Veterans General Hospital.
Informed consent
Written informed consent was obtained from the legal guardian of participants in this study.
Ethical approval
Approval from Institutional Review Board of authors’ institute was obtained.
Methodology
• Prospective
• Observational
• Performed at one institution
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Chang, WC., Hsu, KH. & Su, YP. Natural progression of the pubofemoral distance with age and its correlation with future acetabular index. Eur Radiol 33, 6473–6481 (2023). https://doi.org/10.1007/s00330-023-09579-z
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-023-09579-z