Skip to main content
Log in

Retrievable IVC square stent filter: Experimental study

  • Laboratory Investigations
  • Published:
CardioVascular and Interventional Radiology Aims and scope Submit manuscript

Abstract

Purpose: In vitro and in vivo evaluation of a new retrievable, home-made, inferior vena cava (IVC) Square stent filter (SSF) with two trapping levels.

Methods: In vitro, the SSF was compared in a flow model with the stainless steel Greenfield filter (SGF) for embolitrapping efficiency by serially passing 300 emboli of 3 and 6 mm in diameter and 15–30 mm in length in each type of filter. Nine swine were used for the in vivo testing of the SSF for deployment and retrievability, emboli-trapping efficiency, stability, and self-centering ability and two were used (total of 11 swine) for testing repositioning and retrievability of the SSF at 2 weeks and for gross and histologic IVC changes at 2 months.

Results: In vitro, the SSF and SGF had similar efficiency in trapping large emboli but the SSF had significantly better efficiency than the SGF for trapping all sizes of emboli (91.7% vs 81%), medium size emboli (93% vs 80%), and small emboli (86% vs 69%). Efficiency decreased in both filters from the first to the fifth embolus in each series but was still significantly better for the SSF. With the SSF, 89% of emboli were caught at the primary and 11% at the secondary filtration level. In the nine animals used for acute studies, the SSF was easily placed in all 27 attempts, assumed a central position 26 times, and was easily retrieved in 21 of 22 attempts. One tilted filter needed additional manipulation for retrieval. During emboli injection in five swine, the SSF had 97.2% emboli-trapping efficiency and demonstrated good stability. In the two animals used for longer-term evaluation, the filters were easily retrieved 2 weeks after implantation. Histologic evaluation at 2 months showed neointimal proliferation around the SSF wires in contact with the IVC wall, which was otherwise normal.

Conclusion: The SSF is a promising filter. It is easy to place and retrieve, is stable after placement, and has high efficiency for trapping emboli. Promising results justify further experimental and eventual clinical studies with a commercially manufactured SSF.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Nakagawa N, Cragg AH, Smith TP, Castañeda F, Barnhart WH, De-Jong SC (1994) A retrievable nitinol vena cava filter: Experimental and initial clinical results. J Vasc Interv Radiol 5:507–512

    Article  PubMed  CAS  Google Scholar 

  2. Xian ZY, Roy S, Hosaka J, Kuroki K, Kvernebo K, Enge I, Laerum F (1995) In vitro evaluation of a new temporary venous filter: The Spring filter. Cardiovasc Intervent Radiol 18:315–320

    Article  PubMed  CAS  Google Scholar 

  3. Kuszyk BS, Venbrux AC, Samphilipo MA Jr, Magee CA, Olson JL, Osterman FA (1995) Subcutaneously tethered temporary filter: Pathologic effects in swine. J Vasc Interv Radiol 6:895–902

    Article  PubMed  CAS  Google Scholar 

  4. Cope C, Baum RA, Duszak RA Jr (1996) Temporary use of a Bird’s Nest filter during iliocaval thrombolysis. Radiology 198:765–767

    PubMed  CAS  Google Scholar 

  5. Bovyn G, Gory P, Reynaud P, Ricco JB (1997) The Tempofilter: A multicenter study of a new temporary caval filter implantable for up to six weeks. Ann Vasc Surg 11:520–528

    Article  PubMed  CAS  Google Scholar 

  6. Vos LD, Tielbeek AV, Born EP, Gooszen HC, Vroegindeweij D (1997) The Günther temporary inferior vena cava filter for short-term protection against pulmonary embolism. Cardiovasc Intervent Radiol 20:91–97

    Article  PubMed  CAS  Google Scholar 

  7. Lorch H, Zwaan M, Kulke C, Weiss HD (1998) In vitro studies of temporary vena cava filters. Cardiovasc Intervent Radiol 21:146–150

    Article  PubMed  CAS  Google Scholar 

  8. Zwaan M, Lorch H, Kulke C, Kagel C, Schweider G, Siemens HJ, Muller G, Eberhardt I, Wagner T, Weiss HD (1998) Clinical experience with temporary vena caval filters. J Vasc Interv Radiol 9:594–601

    Article  PubMed  CAS  Google Scholar 

  9. Hosaka J, Roy S, Kvernebo K, Enge I, Laerum F (1998) In vitro function of an adjustable temporary venous Spring filter: Comparison with the temporary RF02 filter and the permanent Greenfield vena cava filter. Acad Radiol 5:620–625

    Article  PubMed  CAS  Google Scholar 

  10. Millward SF (1998) Temporary and retrievable inferior vena cava filters: Current status. J Vasc Interv Radiol 9(3):381–387

    Article  PubMed  CAS  Google Scholar 

  11. Lund G, Rysavy JA, Salomonowitz E, Cragg AH, Kotula F, Castañeda-Zúñiga WR, Hunter DW, Coleman CC, Amplatz K (1984) A new vena caval filter for percutaneous placement and retrieval: Experimental study. Radiology 152(2):369–372

    PubMed  CAS  Google Scholar 

  12. Hunter DW, Lund G, Rysavy JA, Castañeda-Zú~niga W, Cardella JF, Young AT, Vladover Z, Amplatz K (1987) Retrieving the Amplatz retrievable vena cava filter. Cardiovasc Intervent Radiol 10:32–36

    Article  PubMed  CAS  Google Scholar 

  13. Teitelbaum GP, McGurrin M, Davies BL, Moritz MJ, Carabasi RA, Jarrell BE, English GW (1988) Insertion and recovery of a new retrievable vena caval filter.Work in progress. Invest Radiol 23(7):527–533

    Article  PubMed  CAS  Google Scholar 

  14. Neuerburg J, Günther RW, Rassmussen E, Vorwerk D, Tonn K, Handt S, Kupper W, Hansen JV (1993) New retrievable percutaneous vena cava filter: Experimental in vitro and in vivo evaluation. Cardiovasc Intervent Radiol 16:224–229

    Article  PubMed  CAS  Google Scholar 

  15. Irie T, Yamauchi T, Makita K, Kusano S (1995) Retrievable IVC filter: Preliminary in vitro and in vivo evaluation. J Vasc Interv Radiol 6:449–454

    Article  PubMed  CAS  Google Scholar 

  16. Vesely TM, Krysl J, Smith SR, Killion D, Hicks ME (1996) Preliminary investigation of the Irie inferior vena caval filter. J Vasc Interv Radiol 7:529–535

    Article  PubMed  CAS  Google Scholar 

  17. Dibie A, Musset D, Heissler M, Fournet JC, Palau R, Laborde F (1998) In vivo evaluation of a retrievable vena cava filter—the Dibie-Musset filter: Experimental results. Cardiovasc Intervent Radiol 21:151–157

    Article  PubMed  CAS  Google Scholar 

  18. Dorfman GS (1990) Percutaneous inferior vena caval filters. Radiology 174(2):987–992

    PubMed  CAS  Google Scholar 

  19. Grassi CJ (1991) Inferior vena caval filters: Analysis of five currently available devices. AJR 156:813–821

    PubMed  CAS  Google Scholar 

  20. Neuerburg J, Günther RW (1994) Developments in inferior vena cava filters: A European viewpoint. Semin Intervent Radiol 11(4):349–357

    Article  Google Scholar 

  21. Darcy MD, Cardella JF, Hunter DW, Smith TP, Castañeda-Zúñiga WR, Lund G, Amplatz K (1986) Experience with the Amplatz retrievable vena caval filter. Work in progress. Radiology 161(3):611–614

    PubMed  CAS  Google Scholar 

  22. Epstein DH, Darcy MD, Hunter DW, Coleman CC, Tadavarthy SM, Murray PD, Castañeda-Zúñiga WR, Amplatz K (1989) Experience with the Amplatz retrievable vena cava filter. Radiology 172:105–110

    PubMed  CAS  Google Scholar 

  23. McCowan TC, Ferris EJ, Carver DK, Baker ML (1990) Amplatz vena caval filter: Clinical experience in 30 patients. AJR 155(1):177–181

    PubMed  CAS  Google Scholar 

  24. Neuerburg JM, Günther RW, Vorwerk D, Dondelinger RF, Jager H, Lackner KJ, Schild HH, Plant GR, Joffre FG, Schneider PA, Janssen JHA (1997) Results of a multicenter study of the retrievable Tulip vena cava filter: Early clinical experience. Cardiovasc Intervent Radiol 20(1):10–16

    Article  PubMed  CAS  Google Scholar 

  25. Greenfield LJ, Proctor MC, James EA, Abrams GD, Moursi MM (1994) Staging of fixation and retrievability of Greenfield filters. J Vasc Surg 20(5):744–750

    PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Pavcnik, D., Uchida, B.T., Keller, F.S. et al. Retrievable IVC square stent filter: Experimental study. Cardiovasc Intervent Radiol 22, 239–245 (1999). https://doi.org/10.1007/s002709900374

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s002709900374

Key words

Navigation