Abstract
Purpose
The purpose of this study was to evaluate the 5-year real-world results of Supera stent implantation in below the knee prosthetic bypasses (BKPBs). All the procedures were performed because of a history of recurrent thrombosis of the graft and significant stenotic kinking of the prosthesis during knee flexion. A Supera stent was implanted to prevent the next potential BKPB thrombosis.
Materials and Methods
Fourteen patients were included in this single-center, retrospective observational cohort study. All patients underwent Supera stent implantation in infrainguinal prosthetic bypass between 2012 and 2017, due to a history of recurrent thrombosis and kinking of the prosthetic bypass.
Results
Prior to Supera stent implantation procedure, all the patients had more than one episode of acute limb ischemia caused by thrombosis of the BKPB. The median number of BKPB thromboses prior to Supera stent implantation was 3 and ranged from 2 to 6. Technical success was achieved in all cases. Primary patency rates at 12, 24, 36 and 60 months were 71.4%, 57.1%, 57.1% and 14.3%, respectively. Secondary patency rates at 12, 24, 36 and 60 months were 78.6%, 64.3%, 64.3% and 35.7%, respectively. One stent fracture was reported during 60-month follow-up. Major amputation was performed in 6 patients in 5-year follow-up.
Conclusion
Supera stent in treatment of recurrent thrombosis of BKBP is a safe procedure with acceptable mid-term results. However, larger and comparable prospective studies are needed for broader analysis of this procedure.
Graphical Abstract
Avoid common mistakes on your manuscript.
Introduction
Failed infrainguinal bypass is associated with a poor limb prognosis, especially in cases of procedures that were initially performed for limb salvage [1]. Graft thrombosis in most cases is provoked by inadequate inflow, poor runoff vessels, significant anastomotic stenosis, or hypercoagulable state [2]. One of the reasons for reduced patency rates is stenotic kinking of the graft during knee flexion [3, 4]. Stenotic kinking is an underdiagnosed cause of graft occlusion. Diagnosis should be made by measuring the ankle brachial index (ABI) in flexion and extension of the knee joint, followed by duplex ultrasound and angiography in two perpendicular projections with the knee in flexion and extension. Treatment of stenotic kinking of the below the knee prosthetic bypass (BKPB) was previously limited to surgical revision with implantation of a new prosthetic graft [3, 5].
The Supera stent (Abbott Vascular, Santa Clara, Calif, USA) is a self-expanding, braided, interwoven nitinol stent with a helical construction. The Supera stent has high radial force and is therefore able to withstand increased compression and biomechanical stress [6,7,8].
The aim of this study was to evaluate the 5-year results of Supera stent implantation in BKPB due to recurrent graft thrombosis and stenotic kinking of the prosthesis.
Materials and Methods
This is a single-center, observational cohort study with a retrospective evaluation of the acquired data. All patients underwent Supera stent implantation in an infrainguinal polytetrafluoroethylene (PTFE) prosthetic bypass between 2012 and 2017. Inclusion criteria were: implantation of the Supera stent to the femoropopliteal below the knee prosthetic bypass (BKPB), history of recurrent graft thrombosis, proper outflow and inflow from the prosthesis, and stenotic kinking during knee flexion, and no other possible explanation for recurrent graft thrombosis. Exclusion criteria were implantation of the Supera stent in femoropopliteal bypass due to anastomotic stenosis. The study was conducted in accordance with the Helsinki Declaration. Institutional bioethics committee approval was obtained.
Procedural Details
Prior to Supera stent implantation procedure, all the patients had more than one episode of thrombosis of the BKPB. Episodes of graft thrombosis were successfully treated with targeted catheter-directed thrombolysis prior to the Supera stent implantation. In some cases, thrombolysis was combined with an additional procedure. The Supera stent was implanted to prevent recurrent graft thrombosis. This procedure was scheduled for the next hospitalization. Patients were eligible for Supera stent implantation based on the following criteria: history of recurrent graft thrombosis, proper outflow and inflow from the prosthesis as demonstrated by Duplex ultrasound and computed tomography angiography (CTA), hemodynamically significant stenotic kinking during knee flexion as demonstrated by duplex ultrasound and digital subtraction angiography (DSA), and no other possible explanation for recurrent graft thrombosis. Duplex ultrasound and CTA were performed before the procedure to evaluate the inflow and outflow of the BKPB. BKPB were patented at the time of stent implantation. DSA was performed in 90° flexion and full extension of the knee joint. If angiography revealed a stenotic kinking during knee flexion, a Supera stent was implanted at the site of the stenosis. After implantation, angiography was performed with the knee joint in 90° flexion and full extension. The diameter of the Supera stent used was not in a 1:1 ratio to the diameter of the graft. Based on the previous multiple episodes of graft thrombosis, it was suspected that mural thrombi were present in the graft. To match the Supera stent diameter to the graft lumen in a 1:1 ratio with mural thrombi, a 5.5-mm Supera was implanted in the 6-mm graft and a 7.5-mm Supera was implanted in the 8 mm graft.
Definitions
The definition of major adverse cardiovascular events (MACE) included: acute coronary syndrome, stroke (either ischemic or hemorrhagic stroke), cardiovascular death and all-cause death. Major adverse limb events (MALE) were defined as either major amputation of the revascularized limb or reintervention of the revascularized segment. Primary patency was defined as no need for target limb revascularization based on clinical indication or stenosis on follow-up duplex ultrasound. Secondary patency was defined as the cumulative patency rate including reintervention after occlusion. Technical success was defined as the achievement of vessel patency with less than 30% residual stenosis.
Follow-up
Complications during the 30-day post-procedure period were evaluated. Patients were reevaluated at clinic visits 1 month, 6 months, 12 months, and then every 12 months after the procedure. At the follow-up visits, patients underwent physical examination and duplex ultrasonography. Duplex ultrasound was performed with 90° of knee joint flexion and full knee joint extension. All patients were prescribed dual antiplatelet therapy (aspirin + clopidogrel) for at least 6 months.
Statistical Analysis
Continuous variables were reported as mean ± standard deviation or as median (range). Categorical variables were reported as counts (percentages). Kaplan–Meier analysis was used to determine patency rates. Statistical significance was confirmed with a p value < 0.05. Statistical analysis was performed with Statsoft Statistica 13.3 software (TIBCO Software Inc., Palo Alto, CA, USA, 2017).
Results
Fourteen patients were included in the study. Seven BKPBs (50%) were performed with Miller cuff modification of the distal anastomosis. Patients characteristics presenting symptoms and indications for initial BKPB surgery are summarized in Table 1.
The median time from BKPB surgery to the analyzed Supera stent implantation procedure was 12 months. The median number of thromboses of the BKPB prior to Supera stent implantation was 3 and ranged from 2 to 6. The time interval between episodes of BKPB thrombosis ranged from 1 to 29 months with a median of 4 months. The time interval between recurrent thrombosis and Supera stent implantation ranged from 1 to 2 months, and no reocclusion was observed during the waiting time. Images of BKPB thrombosis before and after treatment are presented in Fig. 1.
Stents characteristics are shown in Table 2. In all cases, only one stent was implanted. The median stent diameter was 5.5 mm. Technical success of Supera stent implantation was achieved in 100% of the procedures. Supera stent implantation in stenotic kinking of BKPB is illustrated in Fig. 2.
No MACE or MALE were observed during the 30-day follow-up period. However, 1 minor complication was noted: access site hematoma.
The median follow-up was 51 months and ranged from 7 to 108 months. However, 6 patients were lost to follow-up before 60 months. Two patients were lost to follow-up in the first year, three patients were lost to follow-up between 12 and 24 months, and one patient was lost to follow-up after 48 months. Four individuals lost to follow-up were patients who had undergone above-the-knee amputation. Primary patency rates at 12, 24, 36 and 60 months were 71.4%, 57.1%, 57.1% and 14.3%, respectively (Fig. 3). Secondary patency rates at 12, 24, 36 and 60 months were, respectively, 78.6%, 64.3%, 64.3% and 35.7%, respectively (Fig. 4). Episodes of Supera stent and graft thrombosis were treated with catheter-directed thrombolysis. Four cases of Supera stent restenosis were successfully treated with PTA. One stent fracture was observed during 60 months of follow-up. The stent fracture occurred 5 months after stent implantation, which resulted in BKPB thrombosis and above the knee amputation. All six amputations were associated with Supera stent and BKPB thrombosis and failure. Four amputations were above the knee amputations (AKA), and two were below the knee amputations (BKA).
Discussion
The feasibility of Supera stent for bypass graft salvage has not been analyzed in the existing studies except for one case report [9]. The results of this small series suggest that Supera stent implantation should be considered an acceptable option for the treatment of stenotic kinking of BKPB. In follow-up, eight patients did not experience BKPB thrombosis during the first 24 months after Supera stent implantation. However, recurrent thrombosis occurred at short intervals prior to stent implantation. Nevertheless, limb salvage rate was 57.1% and number of secondary events was relatively high. Those aspects question the results of this study and may suggest the need for more precise selection of patients. Four patients experienced BKPB thrombosis in the first year after Supera stent implantation, suggesting that stenotic kinking was not the cause of BKPB thrombosis. Implantation of the Supera stent in the BKPB results in straightening of the bypass. This could potentially lead to kinking at the different levels and tension at anastomoses. Graft kinking at the different levels or tension at anastomoses may be the reason for early graft thrombosis after Supera stent implantation. Hematological disorders may also be the reason for early thrombosis of the BKPB graft after Supera stent implantation. Progression of arteriosclerotic disease may lead to graft thrombosis in mid- or long-term observation.
Intimal hyperplasia at the distal anastomosis is the most common etiology for the femoropopliteal bypass failure [10]. However, stenotic kinking of the prosthetic vascular graft has also been shown to be a significant risk factor for BKPB failure [3, 11]. Kinking of a BKPB during knee flexion may provoke thrombosis by two mechanisms. First, knee flexion may occlude the graft and, if prolonged, lead to thrombosis. The second mechanism relates to turbulence, jet phenomena and in the long-term neointimal proliferations provoked by kinking of the vascular graft. These lesions may be tears or ulcerations of the neointimal lining that promote local thrombosis or hyperplasic neointimal stenosis of the graft [11,12,13].
Stenotic kinking of the BKPB during knee flexion is a rare phenomenon, and most of the cases are treated surgically. Most of the literature focuses on the treatment of anastomotic stenosis [14,15,16,17,18]. Fujimura et al. analyzed the results of treatment of previously occluded above-the-knee infrainguinal bypasses by Viabahn stent-graft implantation [19]. Presumably, other vasculomimetic stents could be used instead of Supera stent to achieve similar results. The optimal time interval between Supera stent implantation for stenotic kinking of the graft and recurrent thrombosis was not determined in this study. Considering the short time intervals between episodes of recurrent graft thrombosis, we suggest stent implantation directly after treatment of BKPB thrombosis.
The main limitations of the study are: small cohort of patients, one type of stent used, retrospective design of the study and lack of randomization and comparison with different treatment methods.
Conclusions
Implantation of the Supera stent to prevent thrombosis of the BKPB is a feasible option in cases of stenotic kinking of the prosthesis during knee flexion. Supera stent in the treatment of recurrent BKPB thrombosis is a safe procedure with acceptable mid-term results. Larger and comparable prospective studies are needed for a more comprehensive analysis of this procedure.
References
Baldwin ZK, Pearce BJ, Curi MA, Desai TR, McKinsey JF, Bassiouny HS, et al. Limb salvage after infrainguinal bypass graft failure. J Vasc Surg. 2004;39:951–7.
Singh N, Sidawy AN, DeZee KJ, Neville RF, Akbari C, Henderson W. Factors associated with early failure of infrainguinal lower extremity arterial bypass. J Vasc Surg. 2008;47:556–61.
Chakfe N, Jahn C, Nicolini P, Kretz JG, Edah-Tally S, Beaufigeau M, et al. The impact of knee joint flexion on infrainguinal vascular grafts: an angiographic study. Eur J Vasc Endovasc Surg. 1997;13:23–30.
LeVeen RF, Batterman HJ, Billman LA, Gatti JE. Entrapment of below-the-knee bypass grafts. Cardiovasc Interv Radiol. 1986;9:100–2.
Biemans JMA, Van Den Berg JC, Moll FL. Beware the kinking bypass-a hidden cause of graft reocclusion. EJVES Extra. 2002;3(5):88–90.
Scheinert D, Grummt L, Piorkowski M, Sax J, Scheinert S, Ulrich M, et al. A novel self-expanding interwoven nitinol stent for complex femoropopliteal lesions: 24-month results of the SUPERA SFA registry. J Endovasc Ther. 2011;18:745–52.
Scheinert D, Werner M, Scheinert S, Paetzold A, Banning-Eichenseer U, Piorkowski M, et al. Treatment of complex atherosclerotic popliteal artery disease with a new self-expanding interwoven nitinol stent: 12-month results of the leipzig supera popliteal artery stent registry. JACC Cardiovasc Interv. 2013;6:65–71.
Salamaga S, Stępak H, Żołyński M, Kaczmarek J, Błaszyk M, Stanišić MG, et al. Three-year real-world outcomes of interwoven nitinol supera stent implantation in long and complex femoropopliteal lesions. J Clin Med. 2023;12:4869.
Chan YC, Cheng SW, Cheung GC. A midterm analysis of patients who received femoropopliteal helical interwoven nitinol stents. J Vasc Surg. 2020;71:2048–55.
Monahan TS, Owens CD. Risk factors for lower-extremity vein graft failure. Semin Vasc Surg. 2009;22:216–26.
Taylor RS, McFarland RJ, Cox MI. An investigation into the causes of failure of PTFE grafts. Eur J Vasc Surg. 1987;1:335–43.
Geiger G, Hoevels J, Storz L, Bayer HP. Vascular grafts in below-knee femoro-popliteal bypass. A comparative study. J Cardiovasc Surg (Torino). 1984;25:523–9.
Zempo N, Esato K, O-Hara M, Fujioka K, Kuga T, Takenaka H. Is the preferential use of polytetrafluoroethylene grafts for below-knee femoropopliteal bypass justified? Int Surg. 1993;78:162–5.
Ali H, Elbadawy A, Saleh M, Hasaballah A. Balloon angioplasty for revision of failing lower extremity bypass grafts. J Vasc Surg. 2015;62:93–100.
Schneider PA, Caps MT, Nelken N. Infrainguinal vein graft stenosis: cutting balloon angioplasty as the first-line treatment of choice. J Vasc Surg. 2008;47:960–6.
Linni K, Ugurluoglu A, Aspalter M, Hitzl W, Hölzenbein T. Paclitaxel-coated versus plain balloon angioplasty in the treatment of infrainguinal vein bypass stenosis. J Vasc Surg. 2016;63:391–8.
Mathur K, Ayyappan MK, Hodson J, Hopkins J, Duddy MJ, Tiwari A, et al. Stenting as a bail-out option after failed percutaneous transluminal angioplasty in infrainguinal vein bypass grafts. Vascular. 2016;24:383–9.
Hicks CW, Stonko DP, Deery SE, Kavousi Y, Black JH, Lum YW, et al. Drug-eluting stents are associated with superior mid-term outcomes for the treatment of infrainguinal bypass graft stenoses. Ann Vasc Surg. 2022;87:13–20.
Fujimura N, Obara H, Iwasa K, Hattori T, Yamamoto H, Watada S, et al. Preliminary experience of viabahn stent graft inside the occluded prosthetic bypass graft for the treatment of above knee femoropopliteal bypass occlusion. Cardiovasc Interv Radiol. 2020;43:223–30.
Funding
This study was not supported by any funding.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical Approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. For this type of study, formal consent is not required. The study was accepted by institutional bioethical committee.
Informed Consent
For this type of study, informed consent is not required.
Consent for Publication
For this type of study, consent for publication is not required.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Salamaga, S., Stanišić, MG., Stępak, H. et al. Five-Year Experience of Interwoven Self-Expanding Stent Implantation in Stenotic Kinking of Below the Knee Prosthetic Bypasses. Cardiovasc Intervent Radiol 47, 670–677 (2024). https://doi.org/10.1007/s00270-024-03728-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00270-024-03728-7