We were delighted to read the letter from P Bream, J Beecham-Chick and M Lessne in response to publication of our manuscript entitled ‘First Clinical Results of the Merit WRAPSODY™ Cell-Impermeable Endoprosthesis for Treatment of Access Circuit Stenosis in Haemodialysis Patients’ [1]. The respondents have suggested that there has been improper use of the endoprosthesis and make this comment based on the clinical example shown in Fig. 3 of the manuscript. They rightly highlight that stent placment has extended through the cephalic arch and into the mid subclavian vein but suggest that this could have been a complication rather than recognise how well placed the stent is to treat the stenosis and ensure in line flow. The respondents also go on to suggest that this case represents a poor example of vein preservation for the lifetime of the patient due to the potential ‘jailing’ of the basilic and axillary veins.
Whilst they are completely right to raise such concerns, I would like to outline the important complex clinical decision-making processes that we as clinicians went through and, more importantly, the backstory to the patient we have treated in this case as that may help the respondent’s and wider readers to realise that this was far from ‘a poor example of vein preservation’ but in fact a great example of considering the K-DOQI Kidney life plan recommendations [2].
The images in Fig. 3 were of a frail 77-year-old individual who had been on dialysis for 6 years and had no transplant option. This patient had a background of hypertensive nephropathy, severe COPD from lifelong smoking and limited exercise tolerance. There had been two previous access creation attempts before a left brachiocephalic AVF was established in 2015. This developed Cephalic Arch Stenosis in 2018 and was treated with Balloon Angioplasty on two occasions. In our institution, there is a weekly Multi-Disciplinary Team (MDT) meeting at which the access surgeon, nephrology team and radiologist discuss all images in the context of the patient to make an appropriate plan. The MDT felt that stenting for this case was appropriate, and the least invasive option given the severe frailties and limited prognosis. The WRAPSODY stent was placed in Sept 2019 and provided a patent circuit and ongoing good dialysis adequacy throughout the study period. This patient recently died from an exacerbation of COPD. It is our view that this was a case of an excellent MDT working together to decide on the ‘right access option at the right time in the right patient for all the right reasons’ [2].
As an MDT, we always take into consideration future AV access options for our patients and do recognise that this is critically important especially if any treatment or intervention of an existing circuit lesion could possibly prevent future new circuit creation. Such decision-making will, however, be done in the context of the patient and all possible renal replacement therapy options including transplantation. In some circumstance, preserving future options will not be ‘right’ or in the best interest of a specific patient’s ESKD life plan as this case highlighted.
References
Gilbert J, Rai J, Kingsmore D, Skousen J, Ptohis N. First clinical results of the merit WRAPSODYTM cell-impermeable endoprosthesis for treatment of access circuit stenosis in haemodialysis patients. Cardiovasc Intervent Radiol. 2021. https://doi.org/10.1007/s00270-021-02953-8.
Lok CE, Huber TS, Lee T, Shenoy S, Yevzlin AS, Abreo K, et al. KDOQI clinical practice guideline for vascular access: 2019 update. Am J Kidney Dis. 2020;75(4):S1-164.
Acknowledgements
The authors are grateful to the volunteers who participated in the study. Special thanks to the team of site personnel, including Dr. A Wigham, Dr. R Tapping, Dr. R Kasthuri, Dr. S Pagoni, Dr. G Dimitriadis, Dr. V Margellos, S Crosbie, and M Gavrila. We would also like to thank the team from Merit Medical as well as the data monitoring and clinical event committee.
Funding
Funding for the study was provided in part by the sponsor, Merit Medical Systems, Inc. Study devices were supplied by the sponsor. The sponsor was involved in establishing the design of the study; collection, analysis, and interpretation of the data; as well as the preparation of and decision to submit the manuscript.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
Mr. Gilbert reports personal fees from Merit Medical Systems and Bard BD, outside the submitted work. Mr. Kingsmore reports grant funding from WL Gore Ltd., outside the submitted work. Dr. Skousen is an employee of Merit Medical Systems. Dr. Ptohis and Dr. Rai have no conflicts to disclose.
Ethical Approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed Consent
Informed consent was obtained from all individual participants included in the study.
Consent for Publication
For this type of study, consent for publication is not required.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Gilbert, J., Rai, J., Kingsmore, D. et al. First Clinical Results of the Merit WRAPSODY™ Cell-Impermeable Endoprosthesis for Treatment of Access Circuit Stenosis in Haemodialysis Patients. Cardiovasc Intervent Radiol 44, 1916–1917 (2021). https://doi.org/10.1007/s00270-021-03001-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00270-021-03001-1