Effects of Regional and General Anesthesia on the Therapeutic Outcome of Benign Thyroid Nodules Treated with High Intensity Focused Ultrasound (HIFU)

  • Original Scientific Report
  • Published:
World Journal of Surgery Aims and scope Submit manuscript

Abstract

Background

The study investigated whether anesthesia performed during high-intensity-focused-ultrasound treatment (HIFU) of benign thyroid nodules influenced the therapy outcome, based on volume reduction and the amount of energy delivered.

Methods

Thirty patients with benign thyroid nodules were treated with HIFU under general or regional anesthesia at two centers from 2014 to 2019. During HIFU, a therapeutic ultrasound probe, EchoPulse (Teraclion, Malakoff, France), heats the focus to 80–90 degrees Celsius. Nodal volumes were measured by ultrasound before and 3 months after therapy. For statistical analysis, the total population was divided into two groups according to the anesthesia performed. In a retrospective long-term multicenter study, volume reduction and the energy delivered were analyzed using the Wilcoxon signed-rank test and the Mann–Whitney test.

Results

At three months follow-up, the total study population had an average volume reduction of 39.26% (range 4.03–91.16%, p < 0.001, n = 30), the general anesthesia group of 47.46% (range 13.64–91.16%, p = 0.001, n = 15) and the regional anesthesia group of 31.06% (range 4.03—68.63%, p = 0.001, n = 15). Under regional anesthesia a median energy of 3.16 kJ/cm3 (range: 0.96 – 8.2 kJ/cm3) and under general anesthesia a median energy of 0.88 kJ/cm3 (range: 0.18 – 1.63 kJ/cm3) were delivered. All results were significant with p < 0.05. The complication rate was 6.67%.

Conclusion

HIFU is an effective method to treat benign thyroid nodules. Comparing anesthesia methods, volume reduction is higher in patients treated under general anesthesia and less energy has to be delivered under general anesthesia.

Trial registration number

2020–1728-evBO.

Agency

Ethik-Kommission bei der Landesäztekammer Hessen.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Papini E, Monpeyssen H, Frasoldati A et al (2020) European thyroid association clinical practice guideline for the use of image-guided ablation in benign thyroid nodules. Eur Thyroid J 9:172–185

    Article  Google Scholar 

  2. Happel C, Kranert WT, Bockisch B et al (2016) 131I and 99m Tc-uptake in focal thyroid autonomies development in Germany since the 1980s. Nuklearmedizin 55:236–241

    Article  Google Scholar 

  3. Durante C, Grani G, Lamartina L et al (2018) The diagnosis and management of thyroid nodules. JAMA 319:914

    Article  Google Scholar 

  4. Giovanella L, Avram AM, Iakovou I et al (2019) EANM practice guideline/SNMMI procedure standard for RAIU and thyroid scintigraphy. Eur J Nucl Med Mol Imaging 46:2514–2525

    Article  CAS  Google Scholar 

  5. Lang BHH, Wu ALH (2017) High intensity focused ultrasound (HIFU) ablation of benign thyroid nodules - a systematic review. J Ther Ultrasound 5:11

    Article  Google Scholar 

  6. Dietlein M, Grünwald F, Schmidt M et al (2016) Radioiodtherapie bei benignen Schilddrüsenerkrankungen (Version 5). Nuklearmedizin 55:213–220

    Article  Google Scholar 

  7. Federführende Fachgesellschaft:Deutsche Gesellschaft für Allgemein- und Viszeralchirurgie (DGAV) (2015) Operative Therapie benigner Schilddrüsenerkrankungen. AWMF online

  8. Korkusuz H, Gröner D, Racynski N et al (2018) Thermal ablation of thyroid nodules: are radiofrequency ablation, microwave ablation and high intensity focused ultrasound equally safe and effective methods? Eur Radiol 28:929–935

    Article  Google Scholar 

  9. Sennert M, Happel C, Korkusuz H et al (2018) Further investigation on high-intensity focused ultrasound (HIFU) treatment for thyroid nodules. Acad Radiol 2:88–94

    Article  Google Scholar 

  10. Korkusuz H, Fehre N, Sennert M et al (2015) Volume reduction of benign thyroid nodules 3 months after a single treatment with high-intensity focused ultrasound (HIFU). J Ther Ultrasound 3:4

    Article  Google Scholar 

  11. Korkusuz H, Sennert M, Fehre N et al (2015) Localized thyroid tissue ablation by high intensity focused ultrasound: volume reduction, effects on thyroid function and immune response. Fortschr Röntgenstr 187:1011–1015

    Article  CAS  Google Scholar 

  12. Korkusuz H (2016) Thermoablation von Schilddrüsenknoten. Nuklearmediziner 39:227–231

    Article  Google Scholar 

  13. Hill CR, Haar GR (1995) Review article: high intensity focused ultrasound–potential for cancer treatment. Br J Radiol 68:1296–1303

    Article  CAS  Google Scholar 

  14. Wu F, Wang ZB, Cao YD et al (2003) A randomised clinical trial of high-intensity focused ultrasound ablation for the treatment of patients with localised breast cancer. Br J Cancer 89:2227–2233

    Article  CAS  Google Scholar 

  15. Yang R, Sanghvi NT, Rescorla FJ et al (1993) Liver cancer ablation with extracorporeal high-intensity focused ultrasound. Eur Urol 23:17–22

    Article  Google Scholar 

  16. Chung SR, Baek JH, Suh CH et al (2020) Efficacy and safety of high-intensity focused ultrasound (HIFU) for treating benign thyroid nodules: a systematic review and meta-analysis. SAGE J 61:1636–1643

    Google Scholar 

  17. Trimboli P, Pelloni F, Bini F et al (2019) High-intensity focused ultrasound (HIFU) for benign thyroid nodules: 2-year follow-up results. Endocrine 65:312–317

    Article  CAS  Google Scholar 

  18. Prakash PS, Oh HB, Tan WB et al (2019) The efficacy and safety of high-intensity focused ultrasound (HIFU) therapy for benign thyroid nodules—a single center experience from Singapore. World J Surg 43:1957–1963. https://doi.org/10.1007/s00268-019-04976-2

    Article  PubMed  Google Scholar 

  19. Muller A, Petrusca L, Auboiroux V et al (2013) Management of respiratory motion in extracorporeal high-intensity focused ultrasound treatment in upper abdominal organs: current status and perspectives. Cardiovasc Intervent Radiol 36:1464–1476

    Article  CAS  Google Scholar 

  20. Diodato A, Cafarelli A, Schiappacasse A et al (2018) Motion compensation with skin contact control for high intensity focused ultrasound surgery in moving organs. Phys Med Biol 63:035017

    Article  CAS  Google Scholar 

  21. Bing C, Cheng B, Staruch R et al (2019) Breath-hold MR-HIFU hyperthermia: phantom and in vivo feasibility. Int J Hyperthermia 36:1083–1096

    Article  CAS  Google Scholar 

  22. Haar G, Coussios C (2007) High intensity focused ultrasound: Physical principles and devices. Int J Hyperthermia 23:89–104

    Article  Google Scholar 

  23. Izadifar Z, Izadifar Z, Chapman D et al (2020) An introduction to high intensity focused ultrasound: systematic review on principles, devices, and clinical applications. JCM 9:460

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Anne Fischer.

Ethics declarations

Conflict of interest

The authors declare that they had no conflict of interest.

Ethical standards

The study has been approved by the ethics committees and complies with the Declaration of Helsinki and local regulations.

Informed consent

Informed consent was obtained from all individual participants included in the study.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Vorländer, C., Fischer, A. & Korkusuz, H. Effects of Regional and General Anesthesia on the Therapeutic Outcome of Benign Thyroid Nodules Treated with High Intensity Focused Ultrasound (HIFU). World J Surg 46, 1076–1081 (2022). https://doi.org/10.1007/s00268-022-06447-7

Download citation

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00268-022-06447-7

Navigation