Abstract
Purpose
Concerns remain about potential increased wear with dual mobility cups related to the multiple articulations involved in this specific design of implant. This finite element analysis study aimed to compare polyethylene (PE) wear between dual mobility cup and conventional acetabular component, and between the use of conventional ultra-high molecular weight PE (UHMWPE) and highly cross-linked PE (XPLE).
Methods
Patient-specific finite element modeling was developed for 15 patients undergoing primary total hip arthroplasty (THA). Five acetabular components were 3D modeled and compared in THA constructs replicating existing implants: a dual mobility cup with a 22.2-mm-diameter femoral head against UHMWPE or XLPE (DM22PE or DM22XL), a conventional cup with a 22.2-mm-diameter femoral head against UHMWPE (SD22PE) and a conventional cup with a 32-mm-diameter femoral head against UHMWPE or XLPE (SD32PE or SD32XL).
Results
DM22PE produced 4.6 times and 5.1 times more volumetric wear than SD32XL and DM22XL (p < 0.0001, Cohen’s d = 6.97 and 7.11; respectively). However, even if significant, the differences in volumetric wear between DM22XL and SD32XL as well as between DM22PE and SD22PE or SD32PE were small according to their effect size (p < 0.0001, Cohen’s |d|= 0.48 to 0.65) and could be therefore considered as clinically negligible.
Conclusion
When using XLPE instead of UHMWPE, dual mobility cup with a 22.2-mm-diameter femoral head produced a similar amount of volumetric wear than conventional acetabular component with a 32-mm-diameter femoral head against XLPE. Therefore, XLPE is advocated in dual mobility cup to improve its wear performance.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Although dual mobility cups demonstrate a marked increase in use to achieve hip stability during primary and revision total hip arthroplasty (THA), concerns remain about potential increased wear related to the multiple articulations involved in this specific design of implant [1,2,3,4]. Indeed, the polyethylene (PE) mobile component is involved in three prosthetic articulations [5,6,7]. At the small articulation, the concave inner bearing surface articulates with the femoral head and behaves as a low-friction bearing [5,6,7]. At the large articulation, the convex outer bearing surface articulates against the metal shell and behaves as a large effective PE-head bearing that increases the jump distance to dislocation [5,6,7]. The impingement-free range of motion within a dual mobility cup is also increased compared to a conventional bearing due to the third articulation that engages movement of the PE mobile component at the large articulation upon femoral neck contact onto the chamfer [5,6,7]. Consequently, the convex outer bearing surface of the PE mobile component was supposed to introduce an additional source of PE wear [4]. However, previous retrieval and biomechanical studies demonstrated that motion and wear within dual mobility cup predominate at the small articulation in vivo [6,7,8,9,10,11]. Nevertheless, their major limitations are related to the fact that retrieval studies are non-comparative and evaluate wear performance on explants potentially issued from THA with non-optimal functioning in vivo and/or PE damage to the mobile component that could occur at the time of revision [9,10,11]. In addition, biomechanical studies using hip simulator with gravimetric measurement usually do not consider parameters such as patient’s level of activity and hip anatomy, body mass index (BMI) or variation in implant positioning related to surgeon’s experience [6,7,8]. Another limitation of these ex vivo studies is related to their ability to determine the exact parts played in wear by the small and large articulations [6,7,8,9,10,11].
Therefore, there is a need for a better understanding, quantification and prediction of wear of dual mobility cup. In previous studies [12,13,14], we developed patient-specific modeling of the hip using finite element analysis (FEA) to simulate the in vivo biomechanical functioning of THA that enables simulation and evaluation of wear of dual mobility and conventional cups. This FEA study aimed to evaluate and compare PE wear between dual mobility and conventional cups, as well as between the use of conventional ultra-high molecular weight PE (UHMWPE) or highly cross-linked PE (XPLE).
Material and methods
Patients
Patient-specific finite element modeling was developed for 15 patients (eight men, mean age = 50 ± 14 years, mean BMI = 30 ± 4 kg/m2). The inclusion criteria were patients undergoing primary THA for hip osteoarthritis at our institution and having a preoperative computed-tomography (CT) scan of the entire pelvis and hip with a sufficient resolution to build computer models. The only exclusion criteria (to avoid bias) were to balance the gender ratio and to account for a large range of age and anatomic variability, especially regarding the femoral offset. These patient-specific models were described with more details in previous studies [12,13,14]. For each patient, 3D geometric models of the pelvis and femur were developed by segmentation of preoperative CT scan. Then, THA implants were replicated in these models. A generic musculoskeletal model was adapted to each patient anatomy. This model was used to predict hip kinematics and joint reaction force during level walking that were used as boundary conditions to predict wear of PE components in a FEA model (Fig. 1). Patient’s informed consent and Institutional Review Board approval were obtained before initiating this study (CER-VD#2013).
THA implants
For each virtual patient, five acetabular components with a 52-mm external diameter were 3D modeled and compared in THA constructs replicating existing implants (Symbol®, Dedienne santé, Mauguio, France): a dual mobility cup with a 22.2-mm-diameter femoral head against UHMWPE (DM22PE), a conventional cup with a 22.2-mm-diameter femoral head against UHMWPE (SD22PE), a conventional cup with a 32-mm-diameter femoral head against UHMWPE (SD32PE), a conventional cup with a 32-mm-diameter femoral head against XLPE (SD32XL) and a dual mobility cup with a 22.2-m- diameter femoral head against XLPE (DM22XL) (Fig. 1). A same design of femoral stem (Symbol®) was used with a simulated cobalt-chromium head.
Musculoskeletal and finite element model
A generic (inverse dynamics) OpenSim lower limb musculoskeletal model (Model Gait 2392) was used to predict hip kinematics and joint reaction force during level walking [15] (Figs. 1 and 2). This model was adapted to the hip anatomy of each patient (i.e. rotation centre and origin of the 3 gluteus muscles) that was estimated from the pre-operative CT scan as well as to their height and weight [14]. The rotation centre of the operated hip was evaluated by fitting a sphere into the acetabulum surface. Then, eleven level walking cycles of the gait model were simulated and averaged to define a single level walking cycle per patient. A Matlab script (MathWorks Inc, Natick, MA) was developed to automatically adapt the generic model using the OpenSim Application Programming Interface.
A finite element model was developed in Abaqus (3DS Dassault Systèmes, Vélizy-Villacoublay, France) for each THA implant [16]. The metal shell was placed first with 40° of inclination and 15° of anteversion with respect to the pelvic anatomical landmarks. For UHMWPE components, the elasto-plastic material properties were elastic modulus of 500 MPa, Poisson’s ratio of 0.45 and plastic yield stress of 16 MPa [17]. For annealed XLPE components, the elasto-plastic material properties were elastic modulus of 1000 MPa, Poisson’s ratio of 0.45 and plastic yield stress of 20 MPa [18]. Bone and metal structures were assumed rigid and modeled as analytical surfaces while PE was modeled as linear elastic. Linear hexahedral elements were then used for modeling the PE components. For both conventional and dual mobility cups, a contact was considered between the PE component and the femoral head. The PE mobile component was allowed to move freely with respect to the metal shell in dual mobility cup. Contrarily, the PE component was fully constrained as an insert into the metal shell in conventional cup. We used the same coefficient of friction (µ = 0.01) for all metal-PE interfaces [19]. The boundary conditions of the finite element models were defined to reproduce the kinematics and joint reaction force predicted by the musculoskeletal model during the entire walking gait cycle. The variable rotation of the femoral head was imposed by a connector element in Abaqus, while the varying joint reaction force vector was applied in parallel on the femoral head center. The metal shells were fixed. The same procedure was performed for each of the 15 patients and five cups.
Wear model
The Archard wear law was used to predict PE wear from contact pressure and sliding distance [20]. For UHMWPE, the wear coefficient was: 10.656 × 10−7 mm3 N−1 m−1 [21]. For annealed XLPE, 20% of the UHMWPE wear coefficient was considered [22]. For dual mobility cup, wear of the convex outer (“large articulation”) and concave inner (“small articulation”) bearing surfaces of the PE mobile component was considered. Pressure and sliding of the femoral head onto the PE component were predicted by the FEA model. A Python routine function was developed to define linear wear on each finite element node of the PE surface according to the Archard wear equation, and extrapolated to 1.0 million cycles (mc) of level walking gait that corresponds approximately to one year of normal activity [23]. Testing to 1.0 mc aimed to imitate the in vivo steady-state wear [6, 7, 23, 24]. The steady-state wear of PE bearing is mathematically linear as a function of cycle count, making wear highly predictable over 1.0 mc [6, 7, 23, 24]. Volumetric wear was derived from linear wear and corresponded to the difference between the initial and final volume of the PE component. Linear and volumetric wear values were reported in mm and mm3, respectively. A mesh convergence analysis was performed with the volumetric wear of DM22PE. We evaluated five mesh refinements to reach a variability lower than 1.5%, using an average mesh size of 0.75 mm and resulting in about 400,000° of freedom. The model was indirectly validated by comparison with similar numerical and experimental studies on comparable SD22PE designs [21].
Statistical analysis
Descriptive statistics were reported as mean ± standard deviation (SD). Normality of all variables was tested using Shapiro–Wilk test. DM22PE was considered as the control group for linear and volumetric wear comparisons. Comparison between two quantitative variables was performed using two-sided paired t tests and reported with mean ± SD and effect size (Cohen’s d) of the difference. The 95% confidence interval (95% CI) was reported for the difference and effect size. Relationship between two quantitative variables was assessed using the coefficient of determination (R2). Statistical analyses were performed with R 4.0 software (cran.r-project.org) with a level of significance set at p < 0.05.
Results
Volumetric and linear wear rates of DM22PE, SD22PE, SD32PE, SD32XL and DM22XL are reported in Table 1 and Fig. 3.
Volumetric wear
DM22PE produced 1.1 times more volumetric wear than SD22PE, 4.6 times more volumetric wear than SD32XL and 5.1 times more volumetric wear than DM22XL (p < 0.0001). However, DM22PE produced 0.9 times less volumetric wear than SD32PE (p < 0.0001). In addition, DM22XL produced 0.9 times less volumetric wear than SD32XL (difference = − 0.44 ± 0.22 mm3, 95% CI = [− 0.56, − 0.31]; d = − 0.65, 95% CI = [− 1.39, 0.09]; p < 0.0001). However, even if significant, the differences in volumetric wear between DM22PE and SD22PE or SD32PE and between DM22XL and SD32XLwere small according to their effect size and could be therefore considered as clinically negligible (Fig. 4).
Regarding the repartition of wear during the simulated gait cycle, the volumetric wear of the convex outer bearing surface at the large articulation was 0.13 ± 0.03 mm3 for DM22PE and 0.01 ± 0.003 mm3 for DM22XL, whereas the volumetric wear of the concave inner bearing surface at the small articulation was 23.0 ± 3.6 mm3 and 4.5 ± 0.7 mm3, respectively (Fig. 5).
For both DM22PE and DM22XL, the volumetric wear was positively correlated to joint reaction force (R2 = 0.656 and 0.640; p = 0.0003 and 0.0004), but not to BMI (R2 = 0.005 and 0.004; p = 0.81 and 0.83) (Fig. 6).
Linear wear
The linear wear of DM22PE was significantly higher than of SD22PE, SD32PE, SD32XL and DM22XL (p < 0.0001). In addition, the linear wear of DM22XLPE was significantly higher than of SD32XL (difference = 0.009 ± 0.003 mm, 95% CI = [0.007, 0.010]; d = 2.47, 95%-CI = [1.49, 3.42]; p < 0.0001).
Discussion
Although reports on the long-term wear in vivo of recent designs of dual mobility cup are sparse in literature, previous clinical and laboratory studies suggest that wear of dual mobility cup compared favorably with conventional PE bearings with large heads routinely used in THA [1,2,3]. For most of the modern designs of dual mobility cup, the mobile component is made of UHMWPE with survival rates reported up to 93% at ten years, 84% at 15 years and 74% at 22 years in primary THA [5, 25, 26]. As contemporary THA patients are more active and functionally demanding than patients evaluated in historical series, the use of XLPE for the mobile component was introduced in dual mobility cup to increase wear performance [27]. A recent analysis of the American Joint Replacement Registry further reported that patients undergoing primary THA under the age of 50 years demonstrated the highest rates of dual mobility cup utilization [2]. However, Boyer et al. [26] reported that patients of this specific age group presented higher risk of dual mobility cup revision for wear and aseptic loosening at long-term follow-up compared to patients over 60 years of age. Therefore, this current trend in the use of dual mobility cup emphasizes the need for a better understanding, quantification and prediction of wear in order to optimize both indication of this bearing and material selection for the PE mobile component. The most important finding of this study was that dual mobility cup with a mobile component made of UHMWPE produced significantly higher volumetric wear than conventional cup with a 32-mm-diameter head articulating against XLPE or a dual mobility cup with a mobile component made of XLPE. Importantly, a similar amount of volumetric wear could be expected between dual mobility cup with a mobile component made of XLPE and conventional cup with a 32-mm-diameter head articulating against XLPE.
To our knowledge, this study based on patient-specific finite element modeling is the first to evaluate and compare wear between dual mobility and conventional cups with PE components made of UHMWPE or XLPE. Several attempts at measuring PE wear of the outer convex and inner concave bearing surfaces of mobile components were performed through clinical, retrieval or hip joint simulator studies [3, 6,7,8,9,10,11]. However, no standardized method of quantifying wear of dual mobility cup has been reported in literature. Regarding the ex vivo studies, the quantification of wear for PE bearings could be overestimated in retrieval studies while hip simulator studies would tend to underestimate it [24, 28]. In retrieval studies, Adam et al. and Geringer et al. [10, 11] both demonstrated a volumetric wear rate of 54 mm3/year in dual mobility cup mobile components made of UHMWPE that was higher than the rate of 24 mm3 we observed in our numerical study after a simulated one year of level walking. Comparable to our results, Saikko et al. [8] reported, in a comparative hip joint simulator study, a similar amount of volumetric wear between dual mobility and conventional cups with PE components made of UHMWPE. Moreover, Saikko et al. [8] demonstrated that machining marks were almost intact at the convex outer bearing surface of mobile components suggesting that motion at the large articulation was minimal. Similarly, in a retrieval study evaluating PE damage and wear lesions to the bearing surfaces of mobile components, D’Apuzzo et al. [9] showed that, although occurring at both bearing surfaces, motion and wear within a dual mobility cup predominate at the small articulation. In our study, motion and volumetric wear predominated at the small articulation for both DM22PE and DM22XL. Importantly, the ratio of volumetric wear of the convex outer bearing surface to the concave inner bearing surface was 1/187 for DM22PE and 1/353 for DM22XL. Therefore, contrarily to the concern raised by Deckard et al. [4] about potential increased wear due to the large articulation, the convex outer bearing surface of the PE mobile component did not represent a significant additional source of wear with dual mobility cup. Besides these laboratory studies, Boyer et al. [29] reported that wear is tridimensional in vivo as the PE mobile component is intended to move freely around three axes. Therefore, measurement of the linear penetration rate on conventional two-dimensional radiographs is deemed not to be effective for determining wear of dual mobility cup [29]. Therefore, some authors proposed the use of 3D radiostereometric analysis (RSA) to measure wear in vivo though no correlation with explant analysis was performed to confirm the accuracy of this technique [30]. In our study, for both DM22PE and DM22XL, the linear wear reported was below the 0.1 mm/year threshold that was proposed by Dumbleton et al. [31] to increase the risk of osteolysis. In addition, the rate of linear wear predicted in our study with DM22XL was equivalent to the rate measured by Laende et al. [30] with 3D RSA technique to evaluate dual mobility cup with XLPE mobile component in vivo. This same linear wear rate of 0.020 mm/year was within the expected range of wear for conventional cup with large head against XLPE [30, 32]. Moreover, similarly to this 3D RSA study, we found that wear of dual mobility cup was independent from patient BMI for both DM22PE and DM22XL suggesting that dual mobility cup could be used in high BMI patients without increased risk of wear compared to conventional acetabular components.
This numerical study presented with some limitations. Wear varies from patient to patient, even in the case of identical implants, being influenced by numerous factors such as patient’s activity, and quality of bone, muscle and hip reconstruction, which is difficult to reproduce and control ex vivo. In the present study, we tried to account for this variability using patient-specific modeling issued from 15 different patients. In addition, only level walking was replicated, instead of more complex and demanding movements of daily living activities such as stair negotiation or running. This generic level walking is certainly the most important limitation of our study. In addition, we are obviously limited by a purely computational study that cannot reproduce the entire complexity of the post-operative reality. We might certainly have predicted higher wear at the convex outer surface of the dual mobility mobile components. We might also have observed more wear at the third articulation, which was virtual in the present study. Finally, this numerical study only replicated 1.0 mc of level walking, corresponding approximately to one year of normal activity after THA. Extending this postoperative time would have required more complex modeling. Besides, we considered here a constant wear coefficient, while it has been reported to be pressure dependent and multidirectional [33, 34]. However, for sake of simplicity, we assumed that it would not affect this comparative study. We thus believe that these limitations would not have changed our main conclusions.
Conclusion
This numerical study based on patient-specific finite element modeling demonstrated that, when using XLPE instead of UHMWPE, dual mobility cup with a 22.2-mm-diameter femoral head produced a similar amount of volumetric wear than conventional cup with a 32-mm-diameter femoral head articulating against XLPE. Therefore, the use of XLPE is advocated in dual mobility cup to improve its wear performance especially in young, active and high functional demand patients.
Data availability
All the data and material are saved in an anonymized repository file folder and available upon request.
References
Darrith B, Courtney PM, Della Valle CJ (2018) Outcomes of dual mobility components in total hip arthroplasty: a systematic review of the literature. Bone Joint J 100-B:11–19. https://doi.org/10.1302/0301-620X.100B1.BJJ-2017-0462.R1
Heckmann N, Weitzman DS, Jaffri H et al (2020) Trends in the use of dual mobility bearings in hip arthroplasty. Bone Joint J 102-B(7_Supple_B):27–32. https://doi.org/10.1302/0301-620X.102B7.BJJ-2019-1669.R1
Laura AD, Hothi H, Battisti C et al (2017) Wear of dual-mobility cups: a review article. Int Orthop 41:625–633. https://doi.org/10.1007/s00264-016-3326-9
Deckard ER, Azzam KA, Meneghini RM (2018) Contemporary dual mobility head penetration at five years: concern for the additional convex bearing surface? J Arthroplasty 33:S280–S284. https://doi.org/10.1016/j.arth.2018.02.061
Vielpeau C, Lebel B, Ardouin L et al (2011) The dual mobility socket concept: experience with 668 cases. Int Orthop 35:225–230. https://doi.org/10.1007/s00264-010-1156-8
Loving L, Herrera L, Banerjee S et al (2015) Dual mobility bearings withstand loading from steeper cup-inclinations without substantial wear. J Orthop Res 33:398–404. https://doi.org/10.1002/jor.22774
Loving L, Lee RK, Herrera L et al (2013) Wear performance evaluation of a contemporary dual mobility hip bearing using multiple hip simulator testing conditions. J Arthroplasty 28:1041–1046. https://doi.org/10.1016/j.arth.2012.09.011
Saikko V, Shen M (2010) Wear comparison between a dual mobility total hip prosthesis and a typical modular design using a hip joint simulator. Wear 268:617–621. https://doi.org/10.1016/j.wear.2009.10.011
D’Apuzzo MR, Koch CN, Esposito CI et al (2016) Assessment of damage on a dual mobility acetabular system. J Arthroplasty 31:1828–1835. https://doi.org/10.1016/j.arth.2016.01.039
Adam P, Farizon F, Fessy MH (2014) Dual mobility retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants. Orthop Traumatol Surg Res 100:85–91. https://doi.org/10.1016/j.otsr.2013.12.011
Geringer J, Boyer B, Farizon F (2011) Understanding the dual mobility concept for total hip arthroplasty. Investigations on a multiscale analysis-highlighting the role of arthrofibrosis. Wear 271:2379–2385. https://doi.org/10.1016/j.wear.2011.02.027
Terrier A, Levrero Florencio F, Rüdiger HA (2014) Benefit of cup medialization in total hip arthroplasty is associated with femoral anatomy. Clin Orthop Relat Res 472:3159–3165. https://doi.org/10.1007/s11999-014-3787-3
Rüdiger HA, Parvex V, Terrier A (2016) Impact of the femoral head position on moment arms in total hip arthroplasty: a parametric finite element study. J Arthroplasty 31:715–720. https://doi.org/10.1016/j.arth.2015.09.044
Rüdiger HA, Guillemin M, Latypova A, Terrier A (2017) Effect of changes of femoral offset on abductor and joint reaction forces in total hip arthroplasty. Arch Orthop Trauma Surg 137:1579–1585. https://doi.org/10.1007/s00402-017-2788-6
Delp SL, Loan JP, Hoy MG et al (1990) An interactive graphics-based model of the lower extremity to study orthopaedic surgical procedures. IEEE Trans Biomed Eng 37:757–767. https://doi.org/10.1109/10.102791
Terrier A, Latypova A, Guillemin M et al (2017) Dual mobility cups provide biomechanical advantages in situations at risk for dislocation: a finite element analysis. Int Orthop 41:551–556. https://doi.org/10.1007/s00264-016-3368-z
Teoh SH, Chan WH, Thampuran R (2002) An elasto-plastic finite element model for polyethylene wear in total hip arthroplasty. J Biomech 35:323–330. https://doi.org/10.1016/s0021-9290(01)00215-9
Pruitt LA (2005) Deformation, yielding, fracture and fatigue behavior of conventional and highly cross-linked ultra high molecular weight polyethylene. Biomaterials 26:905–915. https://doi.org/10.1016/j.biomaterials.2004.03.022
Saikko VO (1996) A three-axis hip joint simulator for wear and friction studies on total hip prostheses. Proc Inst Mech Eng H 210:175–185. https://doi.org/10.1243/PIME_PROC_1996_210_410_02
Archard JF, Hirst W (1956) The wear of metals under unlubricated conditions. Proc R Soc Lond A 236:397–410. https://doi.org/10.1098/rspa.1956.0144
Maxian TA, Brown TD, Pedersen DR, Callaghan JJ (1996) Adaptive finite element modeling of long-term polyethylene wear in total hip arthroplasty. J Orthop Res 14:668–675. https://doi.org/10.1002/jor.1100140424
Dumbleton JH, Wang A, Sutton K, Manley MT (2009) Chapter 14—Highly crosslinked and annealed UHMWPE. In: Kurtz SM, editor. UHMWPE biomaterials handbook (second edition), Academic Press, pp 205–219. https://doi.org/10.1016/B978-0-12-374721-1.00014-6
Schmalzried TP, Szuszczewicz ES, Northfield MR et al (1998) Quantitative assessment of walking activity after total hip or knee replacement. J Bone Joint Surg Am 80:54–59
Affatato S, Spinelli M, Zavalloni M et al (2008) Tribology and total hip joint replacement: current concepts in mechanical simulation. Med Eng Phys 30:1305–1317. https://doi.org/10.1016/j.medengphy.2008.07.006
Combes A, Migaud H, Girard J et al (2013) Low rate of dislocation of dual-mobility cups in primary total hip arthroplasty. Clin Orthop Relat Res 471:3891–3900. https://doi.org/10.1007/s11999-013-2929-3
Boyer B, Philippot R, Geringer J, Farizon F (2012) Primary total hip arthroplasty with dual mobility socket to prevent dislocation: a 22-year follow-up of 240 hips. Int Orthop 36:511–518. https://doi.org/10.1007/s00264-011-1289-4
Epinette JA, Harwin SF, Rowan FE et al (2017) Early experience with dual mobility acetabular systems featuring highly cross-linked polyethylene liners for primary hip arthroplasty in patients under fifty five years of age: an international multi-centre preliminary study. Int Orthop 41:543–550. https://doi.org/10.1007/s00264-016-3367-0
Sychterz CJ, Engh CA Jr, Young AM et al (2000) Comparison of in vivo wear between polyethylene liners articulating with ceramic and cobalt-chrome femoral heads. J Bone Joint Surg Br 82:948–951. https://doi.org/10.1302/0301-620x.82b7.9885
Boyer B, Neri T, Di Iorio A et al (2017) The linear penetration rate is not relevant for evaluating wear of dual mobility cups: an explant study. Int Orthop 41:599–603. https://doi.org/10.1007/s00264-017-3406-5
Laende EK, Richardson CG, Dunbar MJ (2020) Migration and wear of a dual mobility acetabular construct at 3 years measured by radiostereometric analysis. J Arthroplasty 35:1109–1116. https://doi.org/10.1016/j.arth.2019.11.010
Dumbleton JH, Manley MT, Edidin AA (2002) A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty 17:649–661. https://doi.org/10.1054/arth.2002.33664
Callary SA, Solomon LB, Holubowycz OT et al (2015) Wear of highly crosslinked polyethylene acetabular components. Acta Orthop 86:159–168. https://doi.org/10.3109/17453674.2014.972890
Kang L, Galvin AL, Brown TD, Jin Z, Fisher J (2008) Quantification of the effect of cross-shear on the wear of conventional and highly cross-linked UHMWPE. J Biomech 41:340–346. https://doi.org/10.1016/j.jbiomech.2007.09.005
Saikko V (2006) Effect of contact pressure on wear and friction of ultra-high molecular weight polyethylene in multidirectional sliding. Proc Inst Mech Eng H 220:723–731. https://doi.org/10.1243/09544119JEIM146
Funding
Open access funding provided by University of Lausanne. This study was supported by a research grant from the “Lausanne Orthopedic Research Foundation” (Not-for-Profit foundation).
Author information
Authors and Affiliations
Contributions
All the authors 1/have made substantial contributions to conception and design, or acquisition of data, or analysis and interpretation of data, 2/ have been involved in drafting the manuscript or revising it critically for important intellectual content, 3/ have given final approval of the version to be published, and 4/ agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.
Corresponding author
Ethics declarations
Ethical approval
Patient’s informed consent and Institutional Review Board approval were obtained before initiating this study (CER-VD#2013).
Conflict of interest
JW receives royalties from Dedienne santé and serves as paid consultant for Stryker Orthopaedics and Lima corporate. AA serves as paid consultant for Stryker Orthopaedics. ES, MB and AT have no conflict of interest to disclose.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Wegrzyn, J., Antoniadis, A., Sarshari, E. et al. Polyethylene wear of dual mobility cups: a comparative analysis based on patient-specific finite element modeling. International Orthopaedics (SICOT) 46, 779–787 (2022). https://doi.org/10.1007/s00264-022-05305-4