Purpose
Revision of failed total hip arthroplasty with severe acetabular bone loss represents a rare but challenging problem. Anatomic reconstruction with biological defect augmentation as solid bone transplants or cancellous bone restores bone stock while providing good component stability. The objective of this study was to present short- to mid-term results of revision total hip arthroplasty with a modular ring system in a consecutive series of patients with severe acetabular bone loss.
Methods
We retrospectively reviewed 39 consecutive patients (39 hips) with severe acetabular bone loss (Paprosky type 3 A and 3 B) reconstructed with revision total hip arthroplasty using the cementless modular ring system MRS-Titan®. The MRS-Titan® consists of independent ring systems that are adapted modularly to the defect situation. Combined with acetabular defect reconstruction by using cancellous bone in impaction grafting technique, the MRS-Titan® system offers a cement-free revision system that enables defect-adapted customization to individual anatomic matters, bridging of the acetabular bone defect and reconstruction of the centre of rotation. Initial stability of the implant was obtained by screw fixation. Harris hip score and sequential radiographs were used to evaluate clinical and radiographic results.
Results
At an average follow up of 31 months (range 12–69 months) 33 of 39 (84.6 %) of the implants were considered radiographically stable without signs of acetabular migration of more than 2 mm in the horizontal or vertical direction, implant rotation or screw breakage. Complications included six implant failures (15 %). We recorded one aseptic loosening of the MRS prosthesis after 20 months. Five patients (12.8 %) had to undergo revision surgery because of periprosthetic joint infection. Two patients had a dislocation postoperatively. The mean Harris hip score improved from 27 ± 14 points preoperatively to 76 ± 15 points at the time of last follow up.
Conclusion
By achieving stable implant fixation and providing potential for biological fixation, treatment of severe acetabular bone loss in revision total hip arthroplasty with the modular ring system MRS-Titan provides a durable solution with good clinical and radiographic results and allows us to accomplish major goals of reconstruction.
Similar content being viewed by others
References
Dreinhofer KE, Dieppe P, Sturmer T, Grober-Gratz D, Floren M, Gunther KP, Puhl W, Brenner H (2006) Indications for total hip replacement: comparison of assessments of orthopaedic surgeons and referring physicians. Ann Rheum Dis 65:1346–1350. doi:10.1136/ard.2005.047811
Malchau H, Herberts P, Eisler T, Garellick G, Soderman P (2002) The Swedish Total Hip Replacement Register. J Bone Joint Surg Am 84-A(Suppl 2):2–20
Sakai T, Ohzono K, Nishii T, Takao M, Miki H, Nakamura N, Sugano N (2013) Modular acetabular reconstructive cup in acetabular revision total hip arthroplasty at a minimum ten year follow-up. Int Orthop 37:605–610. doi:10.1007/s00264-013-1818-4
Wirtz DC, Niethard FU (1997) Etiology, diagnosis and therapy of aseptic hip prosthesis loosening—a status assessment. Z Orthop Grenzgeb 135:270–280. doi:10.1055/s-2008-1039388
Gravius S, Pagenstert G, Weber O, Kraska N, Rohrig H, Wirtz DC (2009) Acetabular defect reconstruction in revision surgery of the hip. Autologous, homologous or metal? Orthopade 38:729–740. doi:10.1007/s00132-009-1428-4
Gravius S, Randau T, Wirtz DC (2011) What can be done when hip prostheses fail? New trends in revision endoprosthetics. Orthopade 40:1084–1094. doi:10.1007/s00132-011-1844-0
Kim DH, Cho SH, Jeong ST, Park HB, Hwang SC, Park JS (2010) Restoration of the center of rotation in revision total hip arthroplasty. J Arthroplast 25:1041–1046. doi:10.1016/j.arth.2009.07.023
Pulido L, Rachala SR, Cabanela ME (2011) Cementless acetabular revision: past, present, and future revision total hip arthroplasty: the acetabular side using cementless implants. Int Orthop 35:289–298. doi:10.1007/s00264-010-1198-y
Chen XD, Waddell JP, Morton J, Schemitsch EH (2005) Isolated acetabular revision after total hip arthroplasty: results at 5–9 years of follow-up. Int Orthop 29:277–280. doi:10.1007/s00264-005-0674-2
Patil N, Hwang K, Goodman SB (2012) Cancellous impaction bone grafting of acetabular defects in complex primary and revision total hip arthroplasty. Orthopedics 35:e306–e312. doi:10.3928/01477447-20120222-24
Dunlop DG, Brewster NT, Madabhushi SP, Usmani AS, Pankaj P, Howie CR (2003) Techniques to improve the shear strength of impacted bone graft: the effect of particle size and washing of the graft. J Bone Joint Surg Am 85-A:639–646
Beckmann NA, Weiss S, Klotz MC, Gondan M, Jaeger S, Bitsch RG (2014) Loosening after acetabular revision: comparison of trabecular metal and reinforcement rings A systematic review. J Arthroplasty 29:229–235. doi:10.1016/j.arth.2013.04.035
Wimmer MD, Randau TM, Deml MC, Ascherl R, Noth U, Forst R, Gravius N, Wirtz D, Gravius S (2013) Impaction grafting in the femur in cementless modular revision total hip arthroplasty: a descriptive outcome analysis of 243 cases with the MRP-TITAN revision implant. BMC Musculoskelet Disord 14:19. doi:10.1186/1471-2474-14-19
Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty An end-result study using a new method of result evaluation. J Bone Joint Surg Am 51:737–755
Charlson ME, Pompei P, Ales KL, MacKenzie CR (1987) A new method of classifying prognostic comorbidity in longitudinal studies: development and validation. J Chronic Dis 40:373–383
Charlson ME, Sax FL, MacKenzie CR, Braham RL, Fields SD, Douglas RG Jr (1987) Morbidity during hospitalization: can we predict it? J Chronic Dis 40:705–712
Charnley J (1961) Arthroplasty of the hip. A new operation. Lancet 1:1129–1132
Sink EL, Leunig M, Zaltz I, Gilbert JC, Clohisy J, Academic Network for Conservational Hip Outcomes Research G (2012) Reliability of a complication classification system for orthopaedic surgery. Clin Orthop Relat Res 470:2220–2226. doi:10.1007/s11999-012-2343-2
Paprosky WG, Sporer SS, Murphy BP (2007) Addressing severe bone deficiency: what a cage will not do. J Arthroplast 22:111–115. doi:10.1016/j.arth.2007.01.018
DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res 121:20–32
Peters CL, Miller M, Erickson J, Hall P, Samuelson K (2004) Acetabular revision with a modular anti-protrusio acetabular component. J Arthroplast 19:67–72
Bonnomet F, Vanhille W, Lefebvre Y, Clavert P, Gicquel P, Kempf JF (2001) Failure of acetabular cups fixed with cement and thick embedded wire mesh. Rev Chir Orthop Reparatrice Appar Mot 87:544–555
Azuma T, Yasuda H, Okagaki K, Sakai K (1994) Compressed allograft chips for acetabular reconstruction in revision hip arthroplasty. J Bone Joint Surg Br 76:740–744
Morsi E, Garbuz D, Stockley I, Catre M, Gross AE (1996) Total hip replacement in dysplastic hips using femoral head shelf autografts. Clin Orthop Relat Res 324:164–168
Brooker AF, Bowerman JW, Robinson RA, Riley LH Jr (1973) Ectopic ossification following total hip replacement incidence and a method of classification. J Bone Joint Surg Am 55:1629–1632
Choi HR, Anderson D, Foster S, Beal M, Lee JA, Barr C, Malchau H, McCarthy J, Kwon YM (2013) Acetabular cup positioning in revision total hip arthroplasty with Paprosky type III acetabular defects: Martell radiographic analysis. Int Orthop 37:1905–1910. doi:10.1007/s00264-013-2008-0
Bonnomet F, Clavert P, Gicquel P, Lefebvre Y, Kempf JF (2001) Reconstruction by graft and reinforcement device in severe aseptic acetabular loosening: 10 years survivorship analysis. Rev Chir Orthop Reparatrice Appar Mot 87:135–146
Friedrich MJ, Gravius S, Schmolders J, Wimmer MD, Wirtz DC (2014) Biological acetabular defect reconstruction in revision hip arthroplasty using impaction bone grafting and an acetabular reconstruction ring. Oper Orthop Traumatol 26:126–140. doi:10.1007/s00064-013-0270-3
Berry DJ, Muller ME (1992) Revision arthroplasty using an anti-protrusio cage for massive acetabular bone deficiency. J Bone Joint Surg Br 74:711–715
Rosson J, Schatzker J (1992) The use of reinforcement rings to reconstruct deficient acetabula. J Bone Joint Surg Br 74:716–720
Zehntner MK, Ganz R (1994) Midterm results (5.5-10 years) of acetabular allograft reconstruction with the acetabular reinforcement ring during total hip revision. J Arthroplast 9:469–479
Bohm P, Banzhaf S (1999) Acetabular revision with allograft bone. 103 revisions with 3 reconstruction alternatives, followed for 0.3–13 years. Acta Orthop Scand 70:240–249
van der Linde M, Tonino A (2001) Acetabular revision with impacted grafting and a reinforcement ring: 42 patients followed for a mean of 10 years. Acta Orthop Scand 72:221–227. doi:10.1080/00016470152846510
Siebenrock KA, Trochsler M, Sadri H, Ganz R (2001) Hooked roof cup in revision of difficult loose hip prosthesis cups. Results after a minimum of 10 years. Orthopade 30:273–279
Winter E, Piert M, Volkmann R, Maurer F, Eingartner C, Weise K, Weller S (2001) Allogeneic cancellous bone graft and a Burch-Schneider ring for acetabular reconstruction in revision hip arthroplasty. J Bone Joint Surg Am 83-A:862–867
Eggli S, Muller C, Ganz R (2002) Revision surgery in pelvic discontinuity: an analysis of seven patients. Clin Orthop Relat Res 398:136–145
Gerber A, Pisan M, Zurakowski D, Isler B (2003) Ganz reinforcement ring for reconstruction of acetabular defects in revision total hip arthroplasty. J Bone Joint Surg Am 85-A:2358–2364
Capone A, Setzu V, Ennas F, Civinini R, Gusso MI (2004) Ganz reinforcement rings in acetabular revision: indications and medium-term results. La Chirurgia degli organi di movimento 89:107–117
Goodman S, Saastamoinen H, Shasha N, Gross A (2004) Complications of ilioischial reconstruction rings in revision total hip arthroplasty. J Arthroplast 19:436–446
Schlegel UJ, Bitsch RG, Pritsch M, Clauss M, Mau H, Breusch SJ (2006) Mueller reinforcement rings in acetabular revision: outcome in 164 hips followed for 2–17 years. Acta Orthop 77:234–241. doi:10.1080/17453670610045966
Acknowledgments
We would like to thank Professor Dr. Peter Thümler for his cordial permission to allow us to make use of his visual material.
Author information
Authors and Affiliations
Corresponding author
Additional information
Authors Jan Schmolders and Max J. Friedrich contributed equally to this work.
Rights and permissions
About this article
Cite this article
Schmolders, J., Friedrich, M.J., Michel, R.D. et al. Acetabular defect reconstruction in revision hip arthroplasty with a modular revision system and biological defect augmentation. International Orthopaedics (SICOT) 39, 623–630 (2015). https://doi.org/10.1007/s00264-014-2533-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00264-014-2533-5