Abstract
Background
The aim of this work is to provide the currently missing evidence that may allow an update of the Paediatric Dosage Card provided by the European Association of Nuclear Medicine (EANM) for conventional PET/CT systems.
Methods
In a total of 2082 consecutive [18F]FDG-PET scans performed within the EuroNet-PHL-C2 trial, the administered [18F]FDG activity was compared to the activity recommended by the EANM Paediatric Dosage Card. None of these scans had been rejected beforehand by the reference nuclear medicine panel of the trial because of poor image quality. For detailed quality assessment, a subset of 91 [18F]FDG-PET scans, all performed in different patients at staging, was selected according to pre-defined criteria, which (a) included only patients who had received substantially lower activities than those recommended by the EANM Paediatric Dosage Card, and (b) included as wide a range of different PET systems and imaging parameters as possible to ensure that the conclusions drawn in this work are as generally valid as possible. The image quality of the subset was evaluated visually by two independent readers using a quality scoring system as well as analytically based on a volume-of-interest analysis in 244 lesions and the healthy liver. Finally, recommendations for an update of the EANM Paediatric Dosage Card were derived based on the available data.
Results
The activity recommended by the EANM Paediatric Dosage Card was undercut by a median of 99.4 MBq in 1960 [18F]FDG-PET scans and exceeded by a median of 15.1 MBq in 119 scans. In the subset analysis (n = 91), all image data were visually classified as clinically useful. In addition, only a very weak correlation (r = 0.06) between activity reduction and tumour-to-background ratio was found. Due to the intended heterogeneity of the dataset, the noise could not be analysed statistically sound as the high range of different imaging variables resulted in very small subsets. Finally, a suggestion for an update of the EANM Paediatric Dosage Card was developed, based on the analysis presented, resulting in a mean activity reduction by 39%.
Conclusion
The results of this work allow for a conservative update of the EANM Paediatric Dosage Card for [18F]FDG-PET/CT scans performed with conventional PET/CT systems.
Graphical abstract
Avoid common mistakes on your manuscript.
Introduction
The choice of radiopharmaceutical activities in molecular imaging of paediatric patients always involves a trade-off between adequate image quality and radiation exposure. Therefore, the EANM Paediatric Dosage Card and the North American consensus guidelines for radiopharmaceutical activities in paediatrics were developed in 2007 and 2011, respectively, providing recommendations on activities for different radiotracers [1,2,3]. Despite several updates [4,5,6], these recommendations have repeatedly been criticised by nuclear medicine specialists, as, in their view, the [18F]FDG activities were set too high [7]. Because no evidence could be presented to support their view that lower [18F]FDG activities result in good image quality, the recommendations were left unchanged. Based on the original [18F]FDG-PET data, collected within a large international multicentre trial using [18F]FDG-PET for staging and restaging of Hodgkin lymphoma patients [7], we aimed at providing the missing evidence and at updating the EANM Paediatric Dosage Card for conventional PET/CT systems.
Methods
The present analysis is based on 2082 consecutive [18F]FDG-PET scans from the EuroNet-PHL-C2 trial (EudraCT number: 2012–004053-88; Clinical Trials.gov Identifier: NCT02684708), in which more than 2800 children, adolescents and young adults with Hodgkin’s lymphoma were treated between 2015 and 2021. After [18F]FDG-PET imaging at one of the more than 200 paediatric oncology centres located in Europe, Israel, Australia and New Zealand, reference reading was centrally performed at Leipzig University Hospital by experienced nuclear medicine physicians [8]. As the study protocol did not define mandatory criteria on how to perform the [18F]FDG-PET scans, the image data closely reflect the clinical reality in various PET centres. More information on the EuroNet-PHL-C2 trial can be found in [9] and in the supplement.
To select a subset of data suitable for the planned analysis, the injected activities recommended by the EANM Paediatric Dosage Card were determined for all 2082 scans based on administered [18F]FDG activity and patient weight. Then, a subset of 91 scans was selected, which met the following criteria:
-
(1)
At least 70 MBq activity reduction compared to the EANM Paediatric Dosage Card recommendation.
-
(2)
Image acquisition for initial staging as lymphoma lesions respond rapidly to chemotherapy. Thus, in [18F]FDG-PET images for response assessment minimal uptake due to favourable response to chemotherapy or due to reduced administered [18F]FDG activity cannot be distinguished.
-
(3)
Broadest possible spectrum of different PET/CT systems. However, if criteria 1 and 2 resulted in more than five scans performed on equivalent systems, all further scans on those systems were skipped.
For further analysis, the percentage activity reduction ∆A% for this subgroup was calculated as:
In addition, mean and standard deviation of the standardised uptake value (SUV) were determined in sphere volumes of interest (VOIs) in “tumour” (1 ml volume around up to 3 lesions) and “background” (30 ml volume in healthy liver tissue). These were used to calculate the tumour-to-background ratio (TBR) as the quotient of mean SUV in the lesion VOI and the liver background VOI. As quantitative measure for the overall noise in the image, the coefficient of variation (CoV) was calculated as the quotient of standard deviation and mean SUV in the liver background VOI.
Lastly, a dedicated image quality assessment was performed by two experienced nuclear medicine physicians based on a visual quality score (QSV). It was defined by the homogeneity of the liver uptake and the visual impression of tumour-to-background in neck and mediastinum and has a value range between 0 (not suitable for clinical reporting) and 9 (excellent image quality).
More information on the data included in the analysis and the determination of TBR, CoV and QSV are provided in the supplement (including Supplemental Table 1). All post-processing and statistics were performed in RStudio 2022.07.2.
Results
Characteristics of all 2082 consecutive [18F]FDG-PET scans
Of the 2082 [18F]FDG-PET scans, 996 had been performed for initial staging (47.8%), 893 were early response (42.9%) and 193 were late response scans (9.3%). The underlying patient population had a wide range of weights (median 55 kg, range 10–140 kg). The median patient age was 15 years (range 2–24 years), with 15.3% younger or equal to 10 years, 82.4% between 11 and 18 years, and 2.3% between 18 and 24 years. In 1960 (94.1%) of the 2082 PET scans, the administered activity was below the recommended activity (median 99.4 MBq, range 0.4–279.8 MBq). In 119 (5.7%) PET scans, the recommended activity was exceeded by a median of 15.1 MBq (range 0.5–126.0 MBq). In the remaining three PET scans, the recommended activity had been administered. Detailed information on the patient characteristics including histograms of patient weight, patient age and administered/recommended activities can be found in the supplement (including Supplemental Fig. 1).
Analysis of the selected subset of 91 [18F]FDG-PET scans
Characteristics of the subset
The characteristics of the subset (n = 91) are depicted in Fig. 1. The activity recommended by the EANM Paediatric Dosage Card had been reduced by a median of 45% (range 14–68%, Supplemental Table 2).
Homogeneity of the dataset
PET/CT data from four different vendors (CPS, General Electric, Philips and Siemens) were included in the analysis. While attenuation, scatter and decay correction had been applied to all data, other features such as time-of-flight reconstruction or post-reconstruction filtering were only partially used. An overview of the PET/CT imaging parameters potentially affecting image quality can be found in Supplemental Table 3.
Assessment of the image quality based on visual assessment (QSV)
All QSV values were ≥ 3 and thus found to be suitable for sufficient reporting. A histogram of the assigned QSV values divided among the four manufacturers is depicted in Supplemental Fig. 3. No statistically significant differences in QSV were found between Siemens and GE systems (t(37) = 0.85, p = 0.40), nor between Siemens and Philips systems (t(32) = 1.24, p = 0.22). As only six scans performed on CPS systems were available, these data were not included in the statistical analysis. To give a visual impression of the image quality, Fig. 2 shows two examples for a QSV of 4.0 and 7.5, respectively.
Noise analysis based on the coefficient of variation (CoV)
A noise analysis was performed based on the CoV in the 30 ml VOI (mean volume 30.2 ± 1.1 ml) of the healthy liver. Due to the wide range of influencing factors such as frame duration, voxel volume and activity-per-weight in relation to the size of the subset, however, the noise could not be adequately analysed by a univariate analysis. Instead, the visual quality assessment by two independent readers must suffice at this point, clearly indicating that all scans were appropriate for clinical evaluation. Further details on the influence of the individual factors can be found in Supplemental Fig. 4.
Investigation of the dependence of the tumour-to-background ratio (TBR) on the administered activity
A total of 244 lesions were analysed, resulting in a median TBR of 5.6 (range 0.8–13.7). Although the hypothesis of correlation could not be completely rejected (p = 0.352), the correlation was found to be very weak (Fig. 3, r = 0.06), inferring an almost negligible advantage of higher activities.
Proposed correction for the EANM Paediatric Dosage Card
An upper limit of 3.7 MBq/kg, as given in the North American consensus guideline of the SNMMI [3], covers most of the data in the subset. A linear regression analysis between body weight and administered activity (slope, 3.24 ± 0.29 MBq/kg, intercept, − 9.16 ± 15.68 MBq, Supplemental Fig. 5) arrives at a similar conservative upper limit of 3.53 MBq/kg (slope plus one standard deviation). Consequently, this value represents an adequate basis for an update of the EANM Paediatric Dosage Card [1, 4]. To be consistent with the previous versions of the EANM Paediatric Dosage Card and the methodology described in [1, 2, 4, 6, 10], the reduced values for conventional [18F]FDG-PET/CT given in Table 1 are proposed for a frame duration of 3 min per bed position.
A mean ratio of 0.61 ± 0.07 was found between a 3.7 MBq/kg activity regimen and the activities recommended by the existing EANM Paediatric Dosage Card. Consequently, the baseline value of 25.9 MBq provided by the dosage card for [18F]FDG torso was reduced by this factor. Table 2 shows the recommended activities and effective doses, taken from ICRP128 [11] for [18F]FDG, in comparison to an activity of 3.7 MBq/kg. In order to be consistent with the previous values, and due to the lack of data for effective doses according to the weighting factors of ICRP 103 [12], the tissue weighting factors of ICRP 60 were applied [13].
This suggested update of the EANM Paediatric Dosage Card, based on the analysis presented in this work, results in a mean activity reduction by 39%.
Figure 4 shows the proposed activities to be administered based on the current EANM Paediatric Dosage Card [4], the recommended update developed in this work and the corresponding values of a 3.7 MBq/kg regimen as a function of body weight. To put these results in the context of the overall study, data from all 2082 PET/CT examinations are plotted.
The activity values proposed by this work are consistent with the recommendations by Dickson et al. [10] for PET/MR systems with higher sensitivity.
Discussion
The present analysis comprises a subset of 2082 PET/CT scans from the EuroNet-PHL-C2 trial, a large multicentric clinical trial in a paediatric population. All 2082 PET/CT scans, 1960 of which were below the activity recommended by the EANM Paediatric Dosage Card by a median of 99.4 MBq, were evaluated by an independent evaluation board; none of the scans was classified as not eligible, suggesting good adherence to imaging guidelines [14] and showing that the activity recommendations of the latest version of the EANM Paediatric Dosage Card [4] can be reduced. To ensure the highest possible relevance of the present retrospective analysis, a subset of 91 patients was selected according to predefined criteria based on the highest absolute activity reduction, which should correspond to the poorest image quality.
For this analysis, we decided to include patients with an activity reduction based on absolute values instead of a relative reduction of the administered activity. When applying this criterion, only a very small part of the lightweight patients are included. However, this choice is justified as, with increasing weight, more attenuation and scatter are observed in the patients because of an increase of the soft tissue proportion (above all muscle mass/fat), with the consequence that an impairment of the image quality is to be expected. In addition, there is a larger distribution volume in this patient group, leading to a lower concentration of the radiopharmaceutical per body section in heavier patients. Both factors might lead to a much stronger decrease of the image quality in heavier patients compared to younger patients. Consequently, we chose to select PET/CT scans based on the highest absolute activity reduction, which should correspond to the poorest image quality.
Retrospective PET/CT data were accessible in DICOM format. Because the amount of acquisition and reconstruction information available in the respective DICOM headers was very different for the manufacturers considered, there was a wide variation in the number of parameters available. Consequently, some of the datasets could not be clearly assigned certain acquisition and reconstruction parameters, which would have been imperative for assessing the image quality in relation to the measurement parameters. As an example, only 45% of the datasets contained information on iterations, subsets and postfilter applied during reconstruction – parameters with major influence on the noise characteristics in the image. While the patient weight, which is needed for calculation of the standardised uptake value, SUV, had been entered for all patients, the patient height was only available for 59% of the patients, making it difficult to include the patients’ physique in the analysis. Due to these limitations in combination with the limited data size of n = 91, a detailed univariate analysis of individual parameters was not feasible, as each of the subgroups becomes too small to deliver statistically meaningful results. On a positive note, however, the data cover a wide range of possible clinical situations, making them ideal for deriving universally applicable recommendations. Therefore, all available data were included in our weight-based activity analysis.
The consistently good quality of the datasets, which can be seen in Fig. 4, allows two important observations: First, it shows that the activities proposed by the present version of the EANM Paediatric Dosage Card [4], and even the significantly reduced administered activities, exceed the amount of activity required for acceptable image quality. Second, it shows an unavoidable limitation of this study. To reliably explore the lower limit of data quality, a considerable proportion of the patients would have to be administered with lower activities (resulting in a lower image quality), which would, however, entail the risk of clinically unevaluable data, and would thus be ethically unacceptable. Although the data used in this work allow us to reduce the recommendation for the activity to be administered according to the EANM Paediatric Dosage Card, we do not necessarily reach an absolute lower limit.
In the last years, two publications studied, in single-centre settings with modern time-of-flight PET/CT systems (Siemens mCT), how the activity to be administered can be reduced in comparison to the EANM Paediatric Dosage Card [15, 16]. While both studies propose higher activity reductions than what is proposed in this work, one should note that both studies were based on state-of-the-art PET/CT equipment in combination with optimised acquisition and reconstruction settings. In contrast, our analysis is based on PET/CT images from a wide range of PET/CT systems from different manufacturers with different acquisition and reconstruction settings, resulting in more conservative, but also more universally applicable reduction recommendations.
It is important to note that these dosage recommendations should be taken in context of “good practice” for nuclear medicine and do not substitute for national and international legal or regulatory provisions. In addition, these recommendations represent a conservative limit needed to obtain a baseline quality necessary for diagnostic analysis of the image data. Lower activities might be conceivable if the PET/CT system including site-specific imaging and reconstruction parameters affecting image quality (e.g. frame duration or reconstruction parameters) suit the needs of the clinic with respect to their specific equipment, clinical preference and the particular needs of their patients.
Conclusion
Based on our results, we suggest amended values for activities to be administered based on the EANM Paediatric Dosage Card for [18F]FDG-PET/CT scans with conventional PET/CT systems.
References
Lassmann M, Biassoni L, Monsieurs M, Franzius C, Jacobs F. The new EANM paediatric dosage card. Eur J Nucl Med Mol Imaging. 2007;34:796–8. https://doi.org/10.1007/s00259-007-0370-0.
Lassmann M, Biassoni L, Monsieurs M, Franzius C. The new EANM paediatric dosage card: additional notes with respect to F-18. Eur J Nucl Med Mol Imaging. 2008;35:1666–8. https://doi.org/10.1007/s00259-008-0799-9.
Gelfand MJ, Parisi MT, Treves ST, Pediatric Nuclear Medicine Dose Reduction W. Pediatric radiopharmaceutical administered doses: 2010 North American consensus guidelines. J Nucl Med. 2011;52:318–22. https://doi.org/10.2967/jnumed.110.084327.
Lassmann M, Treves ST. Pediatric radiopharmaceutical administration: harmonization of the 2007 EANM Paediatric Dosage Card (version 1.5.2008) and the 2010 North American Consensus guideline. Eur J Nucl Med Mol Imaging. 2014;41:1636. https://doi.org/10.1007/s00259-014-2731-9.
Treves ST, Gelfand MJ, Fahey FH, Parisi MT. 2016 update of the North American consensus guidelines for pediatric administered radiopharmaceutical activities. J Nucl Med. 2016;57:15N-N18.
Machado JS, Beykan S, Herrmann K, Lassmann M. Recommended administered activities for (68)Ga-labelled peptides in paediatric nuclear medicine. Eur J Nucl Med Mol Imaging. 2016;43:2036–9. https://doi.org/10.1007/s00259-015-3289-x.
Holm S, Borgwardt L, Loft A, Graff J, Law I, Hojgaard L. Paediatric doses–a critical appraisal of the EANM paediatric dosage card. Eur J Nucl Med Mol Imaging. 2007;34:1713–8. https://doi.org/10.1007/s00259-007-0508-0.
Kurch L, Mauz-Korholz C, Bertling S, Wallinder M, Kaminska M, Marwede D, et al. The EuroNet paediatric hodgkin network - modern imaging data management for real time central review in multicentre trials. Klin Padiatr. 2013;225:357–61. https://doi.org/10.1055/s-0033-1354416.
EuroNet-PHL-C2. https://www.gpoh.de/studienportal/abgeschlossene_studien_register/euronet_phl_c2/index_ger.html. Last accessed 25 May 2023.
Dickson J, Eberlein U, Lassmann M. The effect of modern PET technology and techniques on the EANM paediatric dosage card. Eur J Nucl Med Mol Imaging. 2022;49:1964–9. https://doi.org/10.1007/s00259-021-05635-2.
ICRP. ICRP Publication 128. Radiation dose to patients from radiopharmaceuticals: a compendium of current information related to frequently used substances. Ann ICRP. 2015;44(2S). https://doi.org/10.1177/0146645314558019.
ICRP. Publication 103: the 2007 recommendations of the International Commission of Radiological Protection. Ann ICRP. 2007;37 (2–4). https://www.icrp.org/publication.asp?id=ICRP%20Publication%20103.
ICRP. Publication 60: 1990 recommendations of the International Commission on Radiological Protection. Ann ICRP. 1991;21(1–3). https://www.icrp.org/publication.asp?id=icrp%20publication%2060.
Vali R, Alessio A, Balza R, Borgwardt L, Bar-Sever Z, Czachowski M, et al. SNMMI procedure standard/EANM practice guideline on pediatric (18)F-FDG PET/CT for oncology 1.0. J Nucl Med. 2021;62:99–110. https://doi.org/10.2967/jnumed.120.254110.
Kertesz H, Beyer T, London K, Saleh H, Chung D, Rausch I, et al. Reducing radiation exposure to paediatric patients undergoing [18F]FDG-PET/CT imaging. Mol Imaging Biol. 2021;23:775–86. https://doi.org/10.1007/s11307-021-01601-4.
Cox CPW, van Assema DME, Verburg FA, Brabander T, Konijnenberg M, Segbers M. A dedicated paediatric [(18)F]FDG PET/CT dosage regimen. EJNMMI Res. 2021;11:65. https://doi.org/10.1186/s13550-021-00812-8.
Acknowledgements
We would like to thank all paediatric oncology centres and nuclear medicine departments that participated in the EuroNet-PHL-C2 study and provided data for this study. Our special thanks go to the patients, the clinical board and the national chairpersons of the EuroNet-PHL-C2 trial.
Funding
Open Access funding enabled and organized by Projekt DEAL. The work was in part funded by the Mitteldeutsche Kinderkrebsforschung – Stiftung für Forschung und Heilung, Leipzig, Germany.
Author information
Authors and Affiliations
Contributions
All authors contributed to the study conception and design. Data collection was performed by LK, AP, UR, CMK, RK, TWG and DK. Data analysis and data interpretation were done by JTG, UE, ML and LK. The first draft of the manuscript was written by JTG, LK and ML. All authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
All patients and/or their guardians provided written informed consent to participate in the EuroNet-PHL-C2 trial. This also included informed consent to provide anonymised data for scientific side projects like our analysis. The Clinical Board of the EuroNet-PHL-C2 study approved this project. An ethics approval was applied to the local ethics committee of the University of Leipzig (065/19-ek). No ethical concerns were expressed. All procedures in connection with this work were performed in accordance with the principles of the 1964 Declaration of Helsinki and its later amendments or comparable ethical standards.
Competing interests
M Lassmann has received institutional grants by IPSEN Pharma, Nordic Nanovector and Novartis. No other potential conflicts of interest relevant to this article exist.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Tran-Gia, J., Eberlein, U., Lassmann, M. et al. Analysis of image data from the EuroNet PHL-C2 trial indicates a potential reduction in injected F-18 FDG activities in children: a proposal to update the EANM Paediatric Dosage Card. Eur J Nucl Med Mol Imaging 51, 405–411 (2024). https://doi.org/10.1007/s00259-023-06396-w
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00259-023-06396-w