The past decade has brought unbelievable advances in CT technology. These have been reflected in a progressive widening of the radiation beam in the axial direction (from “fan beam” to “cone beam” geometry) and in a corresponding increase in the number of detector rings.

From the four detectors and 2 cm axial field of view (AFOV), acquired in half a second, that characterised the systems of a decade ago, we now have, with the present technology, 256/312 multi-row detectors that allow the acquisition of an 8/16 cm axial field in less than a third of a second. In addition, dual-source CT is offering a wealth of new possibilities.

These new systems, whose realisation demanded the solution of complex mechanics, electronics and information technology problems, have opened up a completely new field of medical investigation: the four-dimensional study of the heart and the vascular system.

The past few years have also seen evolutions in the field of PET/CT. Indeed, this combined modality has benefitted not only from improved CT technology, but also from gradual advances in the PET sphere—advances prompted by the introduction of new crystal detectors (such as LSO, GSO, LYSO), new iterative reconstruction algorithms, faster electronics, the “gating” technique (synchronisation of data acquisition with respiratory motion) and, especially, “time of flight” PET technology (TOF) [14].

Yet despite these significant improvements, the length of the detector cylinder, or length of the system’s sensitive volume (“field of view”), has remained essentially the same (15–18 cm). With this acquisition geometry, only a very small fraction of the radiation emitted at any given instant is acquired (geometric efficiency ≤0.2). Therefore, to obtain, for example, a “whole-body” study of about 80 cm in length (from the temples to the bladder of the patient), a series of six to eight static scans (“views”) need to be acquired. The adjacent single views, obtained with the movement of the bed, must, of course, also provide an appropriate region of overlap. Obviously, the acquisition time increases in proportion to the number of views. A “standard study” with the injection of 10 mCi of 18F-FDG takes about 3 min per view and, therefore, approximately 25 min per entire study. To this about a minute should be added for the CT scan. However, when used for whole-body studies in diagnostic oncology—this use accounts for approximately 90% of all PET scans—this procedure presents major limitations. The limited length of the sensitive volume not only imposes the need for multiple views, and thus prolongs the overall duration of the examination, it also (and this is particularly important) limits the maximum “line of response” (LOR) angle (with respect to the radial direction of the system), that is the line of flight of the pair of photons arising from a single annihilation event.

PET sensitivity increases as the square of the LOR angle (a fact whose importance became very clear when we moved from 2-D to 3-D acquisition). Therefore, leaving aside the intrinsic efficiency of the detectors and the electronics, the main cause of the system’s low global efficiency lies in the reduced value of the acceptance angle (between 2 and 7%, for a point source placed at the centre of the field of view). The drawbacks of this low global efficiency—low image quality, long scan times and the need for high radiopharmaceutical doses—translate into low productivity, high costs per single study and a substantial reduction in the range of possible applications of the technique. Long scan times also favour the generation of artefacts due to possible movements of the patient, especially in CT fusion studies. Similarly, the overlapping of subsequent views can easily give rise to artefacts in organs included in more than one field of view. An even more serious limitation is the lack of quantitative evaluations: the fact that different body regions are acquired at different times practically rules out the possibility of performing dynamic (tracer time course) studies.

New extended field of view geometry: the challenges

In short, PET systems with an extended AFOV (≥ 90 cm) would be the ideal solution to a wide range of problems. However, to be of practical value, these systems must solve the new problems, both from the system physics and cost point of view, raised by this geometry. As regards the first of these aspects, it must be appreciated that AFOV increases are accompanied by increases in the number not only of true coincidence events (being revealed), but also of spurious events (random or due to scattered radiation). In addition, when the angle at which the photon reaches the detector increases, so too does the probability that the photon will be intercepted by two or more crystals, making its location imprecise and reducing the quality of the image (parallax effect or “crosstalk”). This is, indeed, a major problem and one that is exacerbated by the highly asymmetric shape of crystal detectors, which is due to the need to achieve, simultaneously, high spatial resolution and high detection efficiency. This problem, which already afflicts standard length PET scanners, can be resolved by the use of detectors, possibly as a columnar structure, able to identify the depth within the crystal at which the interaction took place (depth of interaction, DOI). Electronic detection could, of course, prove to be a critical element given that it has to manage a photon flux that grows more than proportionately to the length of the AFOV. The major problem, however, is the cost of the system that, in PET, is largely ascribable (about 40–60%) to the volume of the crystal scintillators used.

First proposals and prototypes

Is there concrete evidence of an evolution of PET towards systems with longer fields of view? Siemens recently proposed a PET/CT system that increases the number of detector rings from three to four, extends the AFOV from 16 to 22 cm and implements TOF technology.

Developments of this kind can probably be considered part of the dawn of a new era, which began a decade ago with the study by Badawi et al. [5] on the effect of PET geometry on PET system performance. In this study, which took the characteristics of the GE Advance PET system as reference, the effects of increasing both AFOV (from 15 to 60 cm) and detector ring diameter (DRD) (from about 60 to 108 cm) were simulated on an anthropomorphic phantom with Monte Carlo code (1 m scan length, 10 mCi of FDG administered, waiting time of 1 h). At a constant DRD, an increase in AFOV from 15 to 60 cm allows a noise equivalent count (NEC) growing by a factor of about 7. In the same conditions, when AFOV increases between 15 and 30 cm, the scatter increases from 30 to 34% and then it remains practically at the same value up to 60 cm length. Tomograph performances decreased significantly with reduction of the DRD: the correlated scatter increase is responsible for this not so obvious result. The maximum count rate was close to 100 MHz (250 MHz assuming the case of isotopes with shorter half-life than 18F), a result considered difficult but not impossible to achieve with the technology of the time.

The first proposal for a truly “total-body” PET system (120–160 cm AFOV) came from an Italian physicist, Crosetto [6]. The first two prototypes that subsequently emerged can be seen to be moving in opposite directions in terms of philosophy and approach: the Knoxville-based Siemens Medical Solutions group [7, 8] developed a tomograph (P39 5H) based on large LSO “flat panels”: 53 cm in the axial direction and 37 cm in the transaxial direction, assembled to form five sides of a hexagon in constant rotation (30 rpm). Each panel consists of 70 block detectors and each block is a 12 × 12 array of 4 × 4 × 20 mm3 crystals. The sixth side contains the system for transmission measurements. This is a cost-effective geometry that makes it possible to reduce the number of photomultipliers by a factor of 4. The system has an axial sensitivity of 2%, measured with a 70 cm source (NU2-2001); the scatter fraction is 40% and the spatial resolution is approximately 5 mm. In clinical conditions, the system operates at a count rate lower than 5 MHz. Only two views (10 min each) are needed for a complete scan (80 cm).

In the study mentioned above [8], the prototype was compared directly with the HI-REZ in two patients. The prototype performs two 10-min views (total 20 min), 90 min after injection, reconstruction algorithm OP-OSEM3D, while the HI-REZ performs seven 2-min views (total 14 min) 140 min after injection, reconstruction algorithm not specified. A 3-D Gaussian filter is applied to both images. The first patient (60 kg) was administered 10.1 mCi of activity and the second (107 kg), who underwent a similar protocol, received 12.2 mCi. The prototype images showed lower resolution and higher noise than those produced by the commercial model, but the prototype (by definition) should have room for improvement.

The second prototype (SHR-92000), which has a 68.5 cm AFOV, was made by the Japanese group Hamamatsu [9]. The detection system consists of 12 layers of block detector rings stacked axially, each ring consisting of a circular array of 60 block detectors. Each block detector consists of a newly designed scintillation position-sensitive plate photomultiplier, which subtends a 16 × 8 BGO crystal array, each one measuring 2.9 × 6.3 × 20 mm3. To reduce the scatter fraction, metal septa are positioned between different adjacent rings (5-mm thick, 180-mm long). These septa do not appear in the image, since the patient bed, in performing an acquisition, continuously scans a distance equal to one pitch of the septa (57.6 mm). This is therefore an intermediate geometry, somewhere between 2-D and 3-D (2.5-D). The system has an axial sensitivity of 0.97% (NU2-2001) and a scatter fraction of 31.4%, and the NEC peak is 113.6 kcps (at 10.5 kBq/ml). No data regarding spatial resolution are available. In this case too, the full scan of the patient requires only two 10-min views: the clinical protocol provides for the injection of 3.4 mCi for a standard patient and a waiting time of 65 min. According to the authors, the resulting image showed very high sharpness, but “vignetting” due to the movement of the septa appears clearly detectable in the presented image.

New detectors: the challenges

The successful development of a PET system with an extended AFOV clearly depends very much on progress made in the field of detector technology. In this regard, the “resistive plate chamber” (RPC) technology, derived from high energy physics, is a particularly promising development. RPC detectors, which can be produced in large plates, have high time, spatial and energy resolution and are relatively inexpensive. They also allow evaluation of the DOI, thereby solving the parallax problem. The high time resolution also allows the implementation of TOF techniques. One limitation, however, is their rather low intrinsic sensitivity, which is resolved by creating, in the radial direction, a “sandwich” of several plates (60–120). Using this technology, researchers at the Coimbra University in Portugal [1013], after creating a high-resolution prototype for animal studies, are now working on a 2 m-long total-body prototype, at the end of which is mounted a small CT system. Assuming an acceptance angle of 30°, a length of 2 m and the presence of 120 plates, it is estimated that sensitivity can be increased, compared with the GE Advance system, by a factor of 20 (including a factor of 4.4 due to TOF). The estimated scatter fraction is 46% and the cost of the final system should not exceed that of a conventional PET system.

Cost or quality?

The possibilities offered by increasing the length of the field of view, well beyond that suggested by Badawi [5], were recently analysed by Eriksson et al. [14]. They evaluated the “sensitivity profile”, i.e. the response of detectors in the axial direction when a very long thin rectilinear source is placed on the axis. They evaluated the current Siemens models (HI-REZ: three rings of detectors, 16.4 cm field of view; and TruePoint: four rings of detectors, 22 cm field of view). Both systems use LSO crystals. With the first scanner, seven views are needed to cover a 70 cm axial field, whereas with the second, five views are sufficient. From the three-ring (16.4 cm field of view) to the four-ring system (22 cm field of view), sensitivity increases by a factor of about 2. This “gain” is reflected in a reduction of the time needed for each view (from 180 to 137 s) and in improved “image quality”. The time needed for the whole scanning procedure (product of time for each view by number of views) is reduced by a factor of about 2, corresponding to the increase in absolute sensitivity.

The authors also considered two hypothetical models with more detector rings (10 and 20 rings corresponding to a length of 55.6 and 111.6 cm, respectively), assuming their use with the current technology (and angle of acceptance). With the longer model, the whole scan (39 s) is accomplished with a single view, and the sensitivity is increased by a factor of 30 compared with the three-ring system and by a factor of 17 compared with the four-ring system. It is important to note that the performances of these systems are calculated without the use of TOF, which further increases the sensitivity by a factor of 2–4. With this latest technology and the 112 cm-long system, it can be assumed that a single scan would take just 8–20 s. This means that the entire examination could be carried out within the space of a breath-hold, thereby giving the added advantage of eliminating the significant blurring effect related to respiratory motion. If we consider that the intrinsic spatial resolution in these conditions is less than a millimetre, this can certainly be seen as a revolutionary development.

However, the financial aspect cannot be ignored. Considering, as mentioned earlier, that the cost of the crystals accounts for 40–60% of the total cost, a large field of view system would carry a price tag of around 5–6 million euros, compared with the (approximately) 2 million euro price tag of current scanners, and this is without considering the additional costs due to the considerable increase in technological complexity.

So should we be looking for cheaper solutions that do not have recourse to high energy physics technologies?

In a recent paper Eriksson et al. [15] evaluated the performance of a hypothetical system based on the same LSO crystals technology used in their previous study [14]; in this case, however, the crystal thickness was reduced (4.3 versus 20 mm). The 196 cm-long tomograph would allow the body to be scanned, with a single view, in 5.8 min, as opposed to the 7.5 min taken by the TruePoint TrueV system with its 22 cm-long axial field of view, five views and 20 mm-thick crystals. The “long” system has been shaped to provide approximately the same sensitivity as both the “short” system and an RPC system of the same length. The cost of this thin crystals system should not exceed the cost of current systems. Another “economical” solution, based on existing technology (BGO), is proposed by Wai-Hoi et al. [16]. This system, with its 100 cm AFOV, consists of 275 rings, each with 748 crystals. Each BGO crystal is 3.5 × 3.5 × 20 mm3 in size and the DRD is 82.5 cm. Compared with the GE Advance, this system reduces the examination time by a factor of 6 and the dose of by a factor of 4. The cost of producing the system, analysed in detail in the study, would be US $1,250,000, of which less than half (US $500,000) is due to the cost of the crystals (compared with US $2,000,000 using LYSO).

But can we now do without the benefits of TOF?

In a recent study, Poon et al. [17] developed a detailed simulation, based on anthropomorphic phantoms, of LSO tomographs up to 2 m long. With very fast electronics (several hundred MHz) a 2 m-long system should be able to increase sensitivity by a factor of 25–50, compared with the “reference” GE Discovery STE system.

At this point the question to be answered is: is it better to exploit the TOF technology to further increase the signal to noise ratio by a factor between 2 and 3 (creating, as seen above, a high-cost system) or, instead, to “settle for” a factor of 25–50 and spend the TOF benefits to concentrate on reducing the thickness of the crystals, thereby developing a relatively low-cost system.

Whichever option prevails, it seems likely that we will see, in the not too distant future, extended field of view PET scanners, with sensitivity increased by more than one order of magnitude.