Haemodynamic effects of dual-chamber pacing versus ventricular pacing during a walk test in patients with depressed or normal left ventricular function

  • Original Article
  • Published:
European Journal of Nuclear Medicine and Molecular Imaging Aims and scope Submit manuscript

Abstract

Purpose

Dual-chamber rate-modulated pacing provides haemodynamic benefits compared with ventricular pacing at rest, but it is unclear whether this also holds true during physical exercise in patients with heart failure. This study assessed the haemodynamic response to a walk test during dual-chamber pacing and ventricular pacing in patients with depressed or normal left ventricular (LV) function.

Methods

Twelve patients with an LV ejection fraction <50% and 11 patients with an LV ejection fraction ≥50% underwent two randomised 6-min walk tests under dual-chamber rate-modulated pacing and ventricular pacing at a fixed rate of 70 beats/min. All patients had a dual-chamber pacemaker implanted for complete heart block. LV function was monitored by a radionuclide ambulatory system.

Results

In patients with depressed LV function, the change from dual-chamber pacing to ventricular pacing induced a decrease in end-systolic volume at the peak of the walk test (P<0.05), with no difference in end-diastolic volume. As a consequence, higher increases in LV ejection fraction (P<0.0001) and stroke volume (P<0.01) were observed during ventricular pacing. No difference in cardiac output was found between the two pacing modes. In patients with normal LV function, the change from dual-chamber pacing to ventricular pacing induced a significant decrease in cardiac output (P<0.005 at rest and P<0.05 at the peak of the walk test).

Conclusion

Compared with dual-chamber rate-modulated pacing, ventricular pacing improves cardiac function and does not affect cardiac output during physical activity in patients with depressed LV function, whereas it impairs cardiac output in those with normal function.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

Similar content being viewed by others

References

  1. Lamas GA, Orav EJ, Stambler BS, Ellenbogen KA, Sgarbossa EB, Huang SK, et al. Quality of life and clinical outcomes in elderly patients treated with ventricular pacing as compared with dual-chamber pacing. Pacemaker selection in the elderly investigators. N Engl J Med 1998;338:1097–104

    Article  CAS  PubMed  Google Scholar 

  2. Lamas GA, Lee KL, Sweeney MO, Silverman R, Leon A, Yee R, et al. Ventricular pacing or dual-chamber pacing for sinus-node dysfunction. N Engl J Med 2002;346:1854–62

    Article  PubMed  Google Scholar 

  3. Janosik DL, Labovitz AJ. Basic physiology of cardiac pacing. In: Ellenbogen KA, Kay GN, Wilkoff BL, editors. Clinical cardiac pacing. Philadelphia PA: WB Saunders; 1995. p. 367–98

    Google Scholar 

  4. Sulke N, Chambers J, Dritsas A, Sowton E. A randomised double-blind crossover comparison of four rate-responsive pacing modes. J Am Coll Cardiol 1991;17:696–706

    CAS  PubMed  Google Scholar 

  5. Sulke N, Dritsas A, Bostock J, Wells A, Morris R, Sowton E. “Subclinical” pacemaker syndrome: a randomised study of symptom-free patients with ventricular demand (VVI) pacemakers upgraded to dual-chamber devices. Br Heart J 1992;67:57–64

    CAS  PubMed  Google Scholar 

  6. Kass DA. Pathophysiology of physiologic cardiac pacing. JAMA 2002;288:3159–61

    Article  PubMed  Google Scholar 

  7. Kristensson BE, Arnman K, Ryden L. The haemodynamic importance of atrioventricular synchrony and rate increase at rest and during exercise. Eur Heart J 1985;6:773–78

    CAS  PubMed  Google Scholar 

  8. Ferro A, Duilio C, Santomauro M, Cuocolo A. Walk test at increased levels of heart rate in patients with dual-chamber pacemaker and with normal or depressed left ventricular function. Eur Heart J 2003;24:2123–32

    Article  PubMed  Google Scholar 

  9. Wilkoff BL, Cook JR, Epstein AE, Greene HL, Hallstrom AP, Hsia H, et al. Dual-chamber pacing or ventricular backup pacing in patients with an implanted defibrillator. The dual chamber and VVI implantable defibrillator (DAVID) trial. JAMA 2002;288:3115–23

    Article  PubMed  Google Scholar 

  10. Connolly SJ, Kerr CR, Gent M, Roberts RS, Yusuf S, Gillis AM, et al. Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes. N Engl J Med 2000;342:1385–91

    Article  CAS  PubMed  Google Scholar 

  11. Zaret BL, Jain D. Continuous monitoring of left ventricular function with miniaturized non-imaging detectors. In: Zaret BL, Beller GA, editors. Nuclear cardiology: state of the art and future directions. St. Louis: Mosby; 1999. p. 191–200

    Google Scholar 

  12. Tamaki N, Gill JB, Moore RH, Yasuda T, Boucher CA, Strauss HW. Cardiac response to daily activities and exercise in normal subjects assessed by an ambulatory ventricular function monitor. Am J Cardiol 1987;59:1164–69

    Article  CAS  PubMed  Google Scholar 

  13. Tamaki N, Yasuda T, Moore RH, Gill JB, Boucher CA, Hutter AM Jr, et al. Continuous monitoring of left ventricular function by ambulatory radionuclide detector in patients with coronary artery disease. J Am Coll Cardiol 1988;12:669–79

    CAS  PubMed  Google Scholar 

  14. Ferro A, Salvatore M, Cuocolo A. Radionuclide monitoring of left ventricular function after sublingual nifedipine administration at rest and during moderate physical activity. J Nucl Cardiol 2001;8:669–76

    Article  CAS  PubMed  Google Scholar 

  15. Poole-Wilson PA. The 6-minute walk. A simple test with clinical application. Eur Heart J 2000;21:507–8

    Article  CAS  PubMed  Google Scholar 

  16. Pace L, Cuocolo A, Nappi A, Nicolai E, Trimarco B, Salvatore M. Accuracy and repeatability of left ventricular systolic and diastolic function measurements using an ambulatory radionuclide monitor. Eur J Nucl Med 1992;19:800–6

    Article  CAS  PubMed  Google Scholar 

  17. Pace L, Cuocolo A, Stefano ML, Nappi A, Nicolai E, Imbriaco M, et al. Left ventricular systolic and diastolic function measurements using an ambulatory radionuclide monitor: effects of different time averaging on accuracy. J Nucl Med 1993;34:1602–6

    CAS  PubMed  Google Scholar 

  18. Imbriaco M, Cuocolo A, Pace L, Nappi A, Nicolai E, Maurea S, et al. Repeatability of haemodynamic responses to cardiac stimulations by ambulatory monitoring of left ventricular function. J Nucl Biol Med 1993;37:238–44

    Google Scholar 

  19. Heyndrickx GR, Vantrimpont PJ, Rousseau MF, Pouleur H. Effects of asynchrony on myocardial relaxation at rest and during exercise in conscious dogs. Am J Physiol 1988;254:H817–22

    CAS  PubMed  Google Scholar 

  20. Sweeney MO, Hellkamp AS, Ellenbogen KA, Greenspon AJ, Freedman RA, Lee KL, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation 2003;107:2932–7

    Article  PubMed  Google Scholar 

  21. Gallagher KP, Osakada G, Matsuzaki M, Kemper WS, Ross J Jr. Myocardial blood flow and function with critical coronary stenosis in exercising dogs. Am J Physiol 1982;12:H698–707

    Google Scholar 

  22. Ross J Jr. Myocardial perfusion-contraction matching. Implications for coronary heart disease and hibernation. Circulation 1991;83:1076–83

    PubMed  Google Scholar 

  23. Scholl JM, Chaitman BR, David PR, Dupras G, Brevers G, Val PG, et al. Exercise electrocardiography and myocardial scintigraphy in the serial evaluation of the results of percutaneous transluminal coronary angioplasty. Circulation 1982;66:380–90

    CAS  PubMed  Google Scholar 

  24. Broderick T, Sawada S, Armstrong WF, Ryan T, Dillon JC, Bourdillon PD, et al. Improvement in rest and exercise-induced wall motion abnormalities after coronary angioplasty: an exercise echocardiographic study. J Am Coll Cardiol 1990;15:591–9

    CAS  PubMed  Google Scholar 

  25. Lamas GA, Ellebogen KA. Evidence base for pacemaker mode selection. Circulation 2004;109:443–51

    Article  PubMed  Google Scholar 

  26. Rostagno C, Olivo G, Comeglio M, Boddi V, Banchelli M, Galanti G, et al. Prognostic value of 6-minute walk corridor test in patients with mild to moderate heart failure: comparison with other methods of functional evaluation. Eur J Heart Failure 2003;5:247–52

    Article  Google Scholar 

  27. Bittner V, Weiner DH, Yusuf S, Rogers WJ, McIntyre KM, Bangdiwala SI, et al. Prediction of mortality and morbidity with a 6-minute walk test in patients with left ventricular dysfunction. JAMA 1993;270:1702–7

    Article  CAS  PubMed  Google Scholar 

  28. Demers C, McKelvie RS, Negassa A, Yusuf S. Reliability, validity, and responsiveness of the six-minute walk test in patients with heart failure. Am Heart J 2001;142:698–703

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Alberto Cuocolo.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ferro, A., Duilio, C., Santomauro, M. et al. Haemodynamic effects of dual-chamber pacing versus ventricular pacing during a walk test in patients with depressed or normal left ventricular function. Eur J Nucl Med Mol Imaging 32, 1075–1080 (2005). https://doi.org/10.1007/s00259-005-1790-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00259-005-1790-3

Keywords

Navigation