Abstract
Purpose
Dual-chamber rate-modulated pacing provides haemodynamic benefits compared with ventricular pacing at rest, but it is unclear whether this also holds true during physical exercise in patients with heart failure. This study assessed the haemodynamic response to a walk test during dual-chamber pacing and ventricular pacing in patients with depressed or normal left ventricular (LV) function.
Methods
Twelve patients with an LV ejection fraction <50% and 11 patients with an LV ejection fraction ≥50% underwent two randomised 6-min walk tests under dual-chamber rate-modulated pacing and ventricular pacing at a fixed rate of 70 beats/min. All patients had a dual-chamber pacemaker implanted for complete heart block. LV function was monitored by a radionuclide ambulatory system.
Results
In patients with depressed LV function, the change from dual-chamber pacing to ventricular pacing induced a decrease in end-systolic volume at the peak of the walk test (P<0.05), with no difference in end-diastolic volume. As a consequence, higher increases in LV ejection fraction (P<0.0001) and stroke volume (P<0.01) were observed during ventricular pacing. No difference in cardiac output was found between the two pacing modes. In patients with normal LV function, the change from dual-chamber pacing to ventricular pacing induced a significant decrease in cardiac output (P<0.005 at rest and P<0.05 at the peak of the walk test).
Conclusion
Compared with dual-chamber rate-modulated pacing, ventricular pacing improves cardiac function and does not affect cardiac output during physical activity in patients with depressed LV function, whereas it impairs cardiac output in those with normal function.
Similar content being viewed by others
References
Lamas GA, Orav EJ, Stambler BS, Ellenbogen KA, Sgarbossa EB, Huang SK, et al. Quality of life and clinical outcomes in elderly patients treated with ventricular pacing as compared with dual-chamber pacing. Pacemaker selection in the elderly investigators. N Engl J Med 1998;338:1097–104
Lamas GA, Lee KL, Sweeney MO, Silverman R, Leon A, Yee R, et al. Ventricular pacing or dual-chamber pacing for sinus-node dysfunction. N Engl J Med 2002;346:1854–62
Janosik DL, Labovitz AJ. Basic physiology of cardiac pacing. In: Ellenbogen KA, Kay GN, Wilkoff BL, editors. Clinical cardiac pacing. Philadelphia PA: WB Saunders; 1995. p. 367–98
Sulke N, Chambers J, Dritsas A, Sowton E. A randomised double-blind crossover comparison of four rate-responsive pacing modes. J Am Coll Cardiol 1991;17:696–706
Sulke N, Dritsas A, Bostock J, Wells A, Morris R, Sowton E. “Subclinical” pacemaker syndrome: a randomised study of symptom-free patients with ventricular demand (VVI) pacemakers upgraded to dual-chamber devices. Br Heart J 1992;67:57–64
Kass DA. Pathophysiology of physiologic cardiac pacing. JAMA 2002;288:3159–61
Kristensson BE, Arnman K, Ryden L. The haemodynamic importance of atrioventricular synchrony and rate increase at rest and during exercise. Eur Heart J 1985;6:773–78
Ferro A, Duilio C, Santomauro M, Cuocolo A. Walk test at increased levels of heart rate in patients with dual-chamber pacemaker and with normal or depressed left ventricular function. Eur Heart J 2003;24:2123–32
Wilkoff BL, Cook JR, Epstein AE, Greene HL, Hallstrom AP, Hsia H, et al. Dual-chamber pacing or ventricular backup pacing in patients with an implanted defibrillator. The dual chamber and VVI implantable defibrillator (DAVID) trial. JAMA 2002;288:3115–23
Connolly SJ, Kerr CR, Gent M, Roberts RS, Yusuf S, Gillis AM, et al. Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes. N Engl J Med 2000;342:1385–91
Zaret BL, Jain D. Continuous monitoring of left ventricular function with miniaturized non-imaging detectors. In: Zaret BL, Beller GA, editors. Nuclear cardiology: state of the art and future directions. St. Louis: Mosby; 1999. p. 191–200
Tamaki N, Gill JB, Moore RH, Yasuda T, Boucher CA, Strauss HW. Cardiac response to daily activities and exercise in normal subjects assessed by an ambulatory ventricular function monitor. Am J Cardiol 1987;59:1164–69
Tamaki N, Yasuda T, Moore RH, Gill JB, Boucher CA, Hutter AM Jr, et al. Continuous monitoring of left ventricular function by ambulatory radionuclide detector in patients with coronary artery disease. J Am Coll Cardiol 1988;12:669–79
Ferro A, Salvatore M, Cuocolo A. Radionuclide monitoring of left ventricular function after sublingual nifedipine administration at rest and during moderate physical activity. J Nucl Cardiol 2001;8:669–76
Poole-Wilson PA. The 6-minute walk. A simple test with clinical application. Eur Heart J 2000;21:507–8
Pace L, Cuocolo A, Nappi A, Nicolai E, Trimarco B, Salvatore M. Accuracy and repeatability of left ventricular systolic and diastolic function measurements using an ambulatory radionuclide monitor. Eur J Nucl Med 1992;19:800–6
Pace L, Cuocolo A, Stefano ML, Nappi A, Nicolai E, Imbriaco M, et al. Left ventricular systolic and diastolic function measurements using an ambulatory radionuclide monitor: effects of different time averaging on accuracy. J Nucl Med 1993;34:1602–6
Imbriaco M, Cuocolo A, Pace L, Nappi A, Nicolai E, Maurea S, et al. Repeatability of haemodynamic responses to cardiac stimulations by ambulatory monitoring of left ventricular function. J Nucl Biol Med 1993;37:238–44
Heyndrickx GR, Vantrimpont PJ, Rousseau MF, Pouleur H. Effects of asynchrony on myocardial relaxation at rest and during exercise in conscious dogs. Am J Physiol 1988;254:H817–22
Sweeney MO, Hellkamp AS, Ellenbogen KA, Greenspon AJ, Freedman RA, Lee KL, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation 2003;107:2932–7
Gallagher KP, Osakada G, Matsuzaki M, Kemper WS, Ross J Jr. Myocardial blood flow and function with critical coronary stenosis in exercising dogs. Am J Physiol 1982;12:H698–707
Ross J Jr. Myocardial perfusion-contraction matching. Implications for coronary heart disease and hibernation. Circulation 1991;83:1076–83
Scholl JM, Chaitman BR, David PR, Dupras G, Brevers G, Val PG, et al. Exercise electrocardiography and myocardial scintigraphy in the serial evaluation of the results of percutaneous transluminal coronary angioplasty. Circulation 1982;66:380–90
Broderick T, Sawada S, Armstrong WF, Ryan T, Dillon JC, Bourdillon PD, et al. Improvement in rest and exercise-induced wall motion abnormalities after coronary angioplasty: an exercise echocardiographic study. J Am Coll Cardiol 1990;15:591–9
Lamas GA, Ellebogen KA. Evidence base for pacemaker mode selection. Circulation 2004;109:443–51
Rostagno C, Olivo G, Comeglio M, Boddi V, Banchelli M, Galanti G, et al. Prognostic value of 6-minute walk corridor test in patients with mild to moderate heart failure: comparison with other methods of functional evaluation. Eur J Heart Failure 2003;5:247–52
Bittner V, Weiner DH, Yusuf S, Rogers WJ, McIntyre KM, Bangdiwala SI, et al. Prediction of mortality and morbidity with a 6-minute walk test in patients with left ventricular dysfunction. JAMA 1993;270:1702–7
Demers C, McKelvie RS, Negassa A, Yusuf S. Reliability, validity, and responsiveness of the six-minute walk test in patients with heart failure. Am Heart J 2001;142:698–703
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Ferro, A., Duilio, C., Santomauro, M. et al. Haemodynamic effects of dual-chamber pacing versus ventricular pacing during a walk test in patients with depressed or normal left ventricular function. Eur J Nucl Med Mol Imaging 32, 1075–1080 (2005). https://doi.org/10.1007/s00259-005-1790-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00259-005-1790-3