Permanent cardiac pacing in pediatric patients presents many challenges related to small patient size, complex anatomy, electrophysiologic abnormalities, and limited access to cardiac chambers [1, 7, 10]. Although many adolescent patients are well served by transvenous pacemakers, epicardial pacing currently remains the conventional technique for infants and those with complex congenital heart disease [9, 10]. Epicardial pacemaker leads must be affixed to the outer surface of the heart by direct application in the operating room. Because of the complex cardiac anatomy, lack of pediatric-specific devices, and variable surgeon training and experience, a range of surgical techniques have evolved [14]. These techniques include different surgical approaches (partial or full sternotomy, subxyphoid, thoracotomy), generator pocket location (retrocostal, subrectus, subxyphoid), and various lead fixation methods.

Findings have shown that epicardial lead placement is a risk factor for lead failure compared with transvenously placed leads [6, 11]. Historically, the pacemaker generator pocket has been formed in the abdominal wall within the rectus abdominus muscle sheath, typically at the level of the umbilicus. At the institution where this study was conducted, this approach is termed subrectus.

In 1999, surgeons at Rady Children’s Hospital, San Diego, California began implanting generators closer to the heart, using either a subxiphoid pocket location or positioning the generator pocket in an intrathoracic location, which we term the retrocostal approach. With the retrocostal approach, the generator and coils of redundant leads are positioned posterior to the rectus muscle at a location behind the lower margin of the rib cage.

The rationale for the retrocostal approach was to avoid the exposed nature of the epicardial leads as they traverse from the abdomen superiorly into the mediastinum and to leave the leads and device in a “protected” pocket behind the rib cage. We hypothesized that the retrocostal approach would reduce the rate of lead failure due to fracture. This study aimed to assess lead survival characteristics among our epicardial pacemaker patients to determine whether the retrocostal pocket location was associated with increased lead survival compared with the subrectus and subxiphoid pocket locations and to examine risk factors for epicardial lead failure in this patient population.

Methods

Design and Subjects

We undertook a retrospective chart review of all patients with epicardial pacemaker systems implanted or followed at Rady Children’s Hospital–San Diego between January 1980 and May 2007. Institutional review board approval was obtained before data collection. Patients were identified by review of medical records and our surgical database. Transvenous pacemaker systems and leads placed prophylactically were excluded from the analysis. Patients whose devices were implanted outside our institution were included in the study when adequate follow-up evaluation, pocket location, and variables of their implanted devices were documented.

Data Collection

We reviewed all medical records in hard and soft copy for the purposes of our study. These included electronic databases, paper charts, and surgical and electrophysiological records. The demographic and clinical variables collected were date of birth, gender, height and weight at the time of the implantation and the latest follow-up evaluation, anatomic and electrophysiologic diagnosis, and number of prior operations. When patient height or weight were not available for a device implantation procedure, we used two to three points on the patient’s plotted growth curve (National Center for Health Statistics 2000, 3rd to 97th percentile charts) and extrapolated the curve to determine the patient’s approximate implant dimensions. The implantation procedure-related variables collected were pocket location (derived from the surgeon’s operative notes and confirmed using radiography and follow-up records whenever possible), the implantation date, and the implanting surgeon. The lead-specific variables collected were the make and model of the lead (and associated manufacturing details such polarity, insulation material, steroid-eluting properties), lead tip location (epicardial or transmural-endocardial), lead chamber (atria or ventricle), lead functional characteristics (implant and latest follow-up capture threshold, sensing, and impedance), lead failure (yes/no), and lead failure type described in the following discussion.

Definitions

We defined a lead’s lifetime from the implantation date to the date of documented lead failure, removal, or latest follow-up date within the study period. If the pocket location was changed and an existing lead reused, the lead was censored at the time of pocket relocation. When an apparently normally functioning lead was removed or replaced (elective replacement), the latest follow-up date was recorded, and the lead was censored. Other outcomes recorded included patient death, heart transplantation, abandonment of pacing system, patient lost to follow-up evaluation, or completion of the study period. Failure type was noted for all failed leads whenever possible and categorized as proven lead fracture by direct inspection or radiograph, pacing problems such as failure to capture or unacceptably high capture threshold, clinically significant sensing problems, or suspected lead fracture due to lead impedance or cause otherwise not specified.

Surgical Technique

The standard techniques for creation of subrectus and subxiphoid pacemaker pockets have been reviewed elsewhere [8]. The retrocostal approach is performed as follows. An entry into the left posterior rectus sheath is made near the lower margin of the sternotomy incision, and a pocket is developed within the rectus sheath. The upper margin of the rectus sheath is delimited by its attachment to the subcostal margin. This attachment is carefully dissected off the ribs to permit extension of the pocket cephalad, beyond and posterior to the subcostal margin. Care is taken to remain extrapleural in the cephalad extension of the pocket.

The pacemaker generator is positioned within the pocket so as to be almost entirely retrocostal. An anchoring suture secures the generator to the periosteum of the ribs to prevent it from migrating outside the protection of the overlying rib. The header of the generator, with lead attachments, is entirely posterior to the ribs, so the course of the leads never emerges from the protection of the overlying rib cage. This method is in contradistinction to the prior method of the rectus sheath pocket, in which the leads pass over the abdominal wall to course into the pericardial envelope (Fig. 1).

Fig. 1
figure 1

Chest radiographs demonstrating pocket locations. The typical sites of epicardial pacemaker implantation are demonstrated from left to right: retrocostal, subxiphoid, and subrectus. Notably, the retrocostal location keeps all hardware within an intrathoracic location, whereas the subrectus and subxiphoid pacemaker pockets require that the pacemaker leads traverse the abdominal musculature

Statistical Analysis

For the primary outcome analysis, the patient cohort was divided into three groups based on the pocket location used (retrocostal, subxiphoid, or subrectus). Kruskal–Wallis one-way analysis of variance by ranks was used for testing equality of population medians among the three groups. Lead survival as a function of pocket location was plotted using Kaplan–Meier analysis. The resulting survival curves were assessed for statistical difference using the log-rank test. The chi-square test was used to test for an association between lead failure type and pocket location.

For our secondary outcome analysis, we tested candidate variables for correlation with lead failure in the entire cohort. Both patient- and lead-specific variables were examined. For other patient-specific variables, Cox proportional hazard analysis was performed using the patient as the unit of analysis and modeling on time to the first lead failure. Each patient was entered only once. For lead-specific variables, the lead was considered the primary unit of analysis. A patient could have more than one lead failure.

Multivariate analyses were performed using Cox proportional hazards, with an entry criteria of univariate p value less than 2. Confidence intervals (CI) at 95% are presented for hazard ratios in Cox regression. The statistical programming package of Stata/IC for Windows version 10 (StataCorp, College Station, TX, USA) was used for uni- and multivariate Cox proportional hazards analysis. All other statistical analyses were performed using SPSS, version 16 (SPSS Inc., Chicago, IL, USA).

A power analysis was conducted using Power and Precision 3.0 (Biostat Inc, 2001, Englewood, NJ, USA). For the described study design, sample size (number of leads), attrition rate (0), alpha (p ≤ 0.05) and tails (2), and population effect size described earlier, the study had a power of 83% to yield a statistically significant result.

Results

Patient Characteristics

Our cohort consisted of 219 patients with a total of 620 leads that comprised the study sample. Of these patients, 107 were women, and 84% (183/219) had structural congenital heart disease. The majority of the patients were young, with 45% younger than 3 years, and only 8% of the patients 18 years or older at the time of the first implantation. The predominant indication for pacing was third-degree heart block.

The cohort was divided into three groups by pocket location. The majority of the patients had either a subrectus or retrocostal pocket location. Demographic data for our patient population were assessed for equivalence between groups. Age, gender, weight, height, implant body surface area, cardiac anatomy, pacemaker indication, and number of prior surgeries were not significantly different between the three groups. The follow-up period was longer in the subrectus pocket location group (Table 1).

Table 1 Patient demographics

Lead Characteristics

The median duration of lead follow-up assessment was 3.1 years (n = 620). The majority of leads were manufactured by Medtronic (Minneapolis, MIN) (442/620, 71%), and 55% of the leads used were steroid eluting. Most of the leads were silicone (77%), and slightly more than half of the leads used a “screw-in” fixation mechanism. As outlined in Table 2, manufacturer, insulation material, steroid elution, polarity, and fixation methods differed between the groups by pocket location (p = 0.01).

Table 2 Lead characteristics by pocket location

Procedural Characteristics

A total of 284 leads were placed in 99 patients using the subrectus pocket location. The retrocostal approach was used for 274 leads in 96 patients. For the remainder, a subxiphoid pocket location was used (Tables 1, 2). The implanting surgeon was recorded for 583 of 620 leads (94%). A total of 11 surgeons implanted one or more of the 620 leads analyzed. Of these 11 surgeons, 3 accounted for 83% (518/620) of the procedures. Most of the retrocostal implants (84%, 230/274) were performed by a single surgeon.

Lead Failure

Overall, 43% (95/219) of the patients experienced one or more lead failures during a median follow-up period of 3.7 years, and 28% (173/620) of the leads failed during a mean follow-up period of 3.1 years. The estimated lead survival was 93% at 2 years and 83% at 5 years. The majority of the leads failed due to pacing problems (54%, 93/173) (e.g., high pacing thresholds), followed by lead fracture (31%, 53/173) and sensing problems (14%, 24/173). When divided into groups by pocket location, the majority of lead failures were seen in the subrectus group (116/173). Overall, 40% of the leads using this pocket location failed compared with 16% and 17% of the leads placed in subxiphoid and retrocostal locations, respectively (p = 0.001). However, when adjustment was made for length of follow-up evaluation, the rates of failure by implant location did not differ significantly (Fig. 2, Table 3). In general, chi-square analysis showed no significant differences of lead failure cause by implant site (Fig. 3). There were no statistical differences between lead failure rates and the implanting surgeon.

Fig. 2
figure 2

Kaplan–Meyer survival curve for epicardial pacemaker leads by pocket location. Cumulative lead longevity, defined as freedom from lead failure for any reason, does not differ between retrocostal, subrectus, and subxiphoid pacemaker pocket locations (p = 0.491, Mantel-Cox)

Table 3 Lead failure by pocket location
Fig. 3
figure 3

Bar graph showing the cause of lead failure by pocket location expressed as the proportion of total lead failures. When lead failure is stratified by cause, groups do show statistical differences based on their epicardial pacemaker pocket location (retrocostal vs subrectus vs subxiphoid; p = 0.194, Pearson chi-square)

Risk Factors for Lead Failure

Consistent with the Kaplan–Meier analysis, a univariate Cox proportional hazard analysis failed to show any differences between implant location and risk of failure (Table 4). This variable was tested using both a patient- and lead-specific analysis. Other candidate variables were tested, with none proving a significant relation to risk of failure. The patient-dependent variables included number of prior surgeries, age at first implantation, size, and presence of congenital heart disease. The lead-specific variables included brand, steroid elution, insulation material, fixation mechanism, and polarity.

Table 4 Risk factors for lead failure (univariate analysis)

Discussion

Previous studies have demonstrated an increased risk for failure of epicardial leads versus transvenous pacing systems. However, epicardial lead systems remain the technique of choice for patients who are either too small or possess contraindications to standard transvenous lead placement.

This study, the largest epicardial lead survival study to date, aimed to determine whether the risk of lead failure could be modified by pocket location, specifically by comparing lower rectus versus subxiphoid versus implantation behind the lower costal margin (retrocostal). Despite analysis of 620 leads in 219 patients over a median follow-up period longer then 3 years, we were unable to show any statistically significant benefit of one pocket location over another. Secondary analysis of a number of both patient- and lead-specific variables also failed to identify any predictors of lead failure.

We were unable to demonstrate a statistically significant advantage of the retrocostal pocket location over other pocket locations in terms of lead fracture prevention. Qualitatively, however, the data (Fig. 3) suggest that lead failure by this mechanism appears to be less frequent with the retrocostal implantation technique. Although admittedly not rigorously supported by the data in this report, it is the authors’ practice for patients with a retrocostal implant who are not pacemaker dependent to liberate them from the contact sports restrictions that typically accompany the alternative implant locations.

The relative surgical risk of the various implantation techniques was not addressed in this study. The experience of the surgeon performing the majority of the retrocostal implantations indicates that the various implant locations are comparable in terms of associated risks. At the primary study institution, all epicardial generator changes were performed by cardiothoracic surgeons, although this is an institutional bias and does not reflect the suitability of these procedures for an appropriately trained electrophysiologist. Similar to the subxiphoid pocket location, generator change or removal of a device in the retrocostal position can be performed by reopening the subxiphoid incision and dissecting through the largely avascular planes defined by the pseudointima surrounding the generator. This does not require a sternotomy. Generator changes or removals, unless performed in combination with larger cardiac procedures, typically do not require an inpatient hospital stay.

Among leads, there were significant differences in manufacturer, steroid elution, polarity, insulation material, and fixation method when compared by group. We speculate that the differences between groups result from their date of implantation. Our subrectus implantations were the earliest epicardial surgeries performed at our center, whereas the retrocostal and subxiphoid implantations were the most recent.

During the study period, pacemaker leads were engineered with newer insulation materials, bipolarity, and steroid elution in attempts to improve their longevity [3]. Yet with all this, we still were impressed with the consistency in the rate of lead failure across groups regardless of lead characteristic or pocket location.

Previous lead survival studies have demonstrated remarkable consistency in estimated lead survival despite a broad period and range of techniques covered. For example, Fortescue et al. [6] undertook a retrospective analysis of all pediatric pacemaker placements between 1980 and 2002 at a single pediatric hospital. They identified 1,007 leads in 497 subjects, 523 of which were epicardial leads. Consistent with the practices at other centers, epicardial leads were more often used in younger children and those with congenital heart disease. These two factors were found in both single and multivariate analysis to be independent correlates of reduced lead longevity. Similarly, epicardial placement also was an independent correlate with lead failure. As in our study, the steroid-eluting versus noneluting leads showed no difference in lead longevity for epicardial devices (Fig. 4). Although the 5-year survival rate for their epicardial leads approached 85%, compared with approximately 75% in our study, the median age at implantation was 9 years compared with 3.62 years in our study.

Fig. 4
figure 4

Kaplan–Meyer survival curve for epicardial pacemaker leads by (non)steroid elution. For all leads for all pacemaker pocket locations, comparison of steroid-eluting leads with non–steroid-eluting leads shows no significant difference in lead longevity (p = 0.244, Mantel-Cox)

Smerup et al. [12] undertook a prospective study attempting to quantify risk among 61 patients. Germane to this discussion, they too found that the epicardial leads, low age at implantation, and high risk-adjusted congenital heart surgery (RACHS) score were significant predictors for early device mortality. Approximately 50% of all their patients had their first device 5 years after implantation.

Cohen et al. [2] provides another comparison for our study. Between 1983 and 2000, these authors had 123 patients with 207 epicardial leads. The median age at implantation was 4.1 years, and 84% of the patients had underlying structural congenital heart disease. The 5-year lead survival rate among this population was 74%. Similar failure rates were seen in Epstein et al. [5], with 33 epicardial patients and a 60% lead survival at 6 years; Dodge-Khatami et al. [4], with 41 children and an 86% lead survival at 5 years; and Thomson et al. [13], with 326 patients and a lead survival of 50% to 76% at 5 years.

Study Limitations

Because our study was retrospective, our analysis was limited in scope. We were not able to control for surgeon technique or device preference. Similarly, due to the technological evolution of pacemakers and leads, lead characteristics significantly differed by implant location. For example, the subrectus group included more non–steroid-eluting leads than the subxiphoid or retrocostal group. Nevertheless, we were surprised to see that not only was the null hypothesis not disproven, but our risk model also was statistically not significant. It would appear that our pediatric epicardial pacemaker population, with its low median age, preponderance of congenital structural heart disease, and multiple surgeries, does not accommodate itself to an internal risk model because the vast majority of patients already possess the known risk factors for early lead failure.

Conclusions

Based on the outcomes of this study and a review of the current literature, we conclude that the retrocostal approach is neither superior nor inferior to the subrectus or subxiphoid approach in terms of lead longevity and freedom from failure. Identifying modifiable risk factors that may have an impact on epicardial lead survival remains challenging.