Abstract
Online monitoring of dynamic chemical processes involving a wide volatility range of hydrocarbon species is challenging due to long chromatographic measurement times. Mass spectrometry (MS) overcomes chromatographic delays. However, the analysis of n-alkane mixtures by MS is difficult because many fragment ions are formed, which leads to overlapping signals of the homologous series. Atmospheric pressure chemical ionization (APCI) is suitable for the analysis of saturated hydrocarbons and is the subject of current research. Still, although APCI is a “soft ionization” technique, fragmentation is typically inevitable. Moreover, it is usually applied for liquid samples, while an application for online gas-phase monitoring is widely unexplored. Here, we present an automated APCI-MS method for an online gas-phase analysis of volatile and semi-volatile n-alkanes. Mass spectra for n-heptane and n-decane reveal [M-H]+, [M-3H]+ and [M-3H+H2O]+ as abundant ions. While [M-H]+ and [M-3H]+ show an excessive fragmentation pattern to smaller CnH2n+1+ and CnH2n-1+ cations, [M-3H+H2O]+ is the only relevant signal within the CnH2n+1O+ ion group, i.e., no chain cleavage is observed. This makes [M-3H+H2O]+ an analyte-specific ion that is suitable for the quantification of n-alkane mixtures. A calibration confirms the linearity of C7 and C10 signals up to concentrations of ~1000–1500 ppm. Moreover, validated concentration profiles are measured for a binary C7/C10 mixture and a five-alkane C7/C10/C12/C14/C20 mixture. Compared to the 40-min sampling interval of the reference gas chromatograph, MS sampling is performed within 5 min and allows dynamic changes to be monitored.
Graphical Abstract
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
In the chemical industry, a reliable analytical instrumentation is indispensable for enabling suitable process control strategies and ensuring good product qualities, from economic and/or ecological perspectives [1,2,3,4,5]. Modern processes make use of automated, online, real-time, or even in-situ/operando analytical measurements that are capable of monitoring the product’s composition over time and provide the opportunity for direct interventions [6]. Moreover, rapid analytical procedures allow for investigating dynamic process conditions and thus reveal temporal phenomena, e.g., in chemical reactors. Dynamic conditions may originate from the ongoing process itself (e.g., changing reaction kinetics, residence time distributions, transient phase changes) or be externally forced by intentional dynamic operation. For instance, dynamic operation of catalytic reactors is gaining an increased interest, as it allows to handle a varying feedstock supply (e.g., in power-to-X technologies with renewable energy fluctuations [7,8,9,10]) or provides opportunities of process intensification [11,12,13].
Processes that are relevant in the fuel and chemical production industry, such as petrochemical processes, biomass/waste pyrolysis or Fischer-Tropsch synthesis, can produce a wide range of volatile organic compounds (VOCs) semi-volatile organic compounds (SVOCs) in the gas phase [14,15,16]. The most common analytical strategy for an online gas-phase analysis of VOCs and SVOCs implies gas chromatography (GC), where even complex hydrocarbon mixtures can be separated and analyzed species-wise. However, GC requires a rather long measuring time, especially when the sample contains a wide range of hydrocarbons.
To speed up online gas-phase analysis, analytical devices with high scan rates can be used, for instance via infrared spectroscopy (IR) or mass spectrometry (MS). However, for hydrocarbon mixtures with lacking functional groups, the resulting spectra consist of overlaying signals, which are difficult to interpret. Especially mass spectrometry has been regarded as a promising tool for hydrocarbon analysis, because its versatility provides a full repository of analytical techniques [17]. Many options for sample ionization exist, including “soft ionization” techniques that aim to preserve the molecular identity of the analyte. Moreover, nowadays available ultrahigh-resolution devices (Fourier transform ion cyclotron resonance (FT-ICR) or Orbitrap) are capable of resolving structure and elemental composition even of complex petroleum mixtures [18, 19]. For hydrocarbon gas-phase monitoring, particularly proton transfer reaction (PTR)-MS [20,21,22] as well as photon ionization (PI)-MS [23, 24] has gained a lot of interest. However, many online gas-phase MS studies focus on aromatic, unsaturated, and/or heteroatomic molecules, since saturated straight-chain hydrocarbons, i.e., n-alkanes, are generally more reluctant to ionization and are therefore difficult to analyze [22, 25, 26]. Some MS studies successfully monitor gas-phase n-alkanes, e.g., by carefully choosing a suitable reagent molecule during chemical ionization (CI)-MS [26,27,28] or by applying single photon ionization (SPI)-MS using vacuum ultra-violet radiation (VUV) [29,30,31,32]. However, finding suitable, widely applicable and robust online methods for n-alkane characterization is still an ongoing research topic.
Atmospheric pressure chemical ionization, utilized in positive ionization mode (APCI(+)), is regarded as suitable ionization technique to measure saturated hydrocarbons, by forming [M-H]+, [M-3H]+, or M+• ions as well as nitrogen and/or oxygen-containing adducts [33, 34]. During APCI, a corona discharge needle is used to form reagent ions, which further ionize the analyte molecules [35]. Typical reagent ions are O2+•, NO+, N2+•, N2H+ as well as H3O+; the latter occurs due to trace amounts of water vapor in the air [36, 37]. In gas-phase or aerosol monitoring applications, APCI(+)-MS is regarded as an alternative option for VOC analysis, again usually applied for unsaturated, aromatic, and/or heteroatomic compounds [38,39,40].
With respect to n-alkanes, most APCI(+)-MS investigations deal with liquid analytes, ranging from VOCs up to heavy petroleum species. Here, a solvent is used, which—after vaporization and ionization—usually serves as the source of the subsequent reagent ion itself [41]. Generally, any formed ion may be part of subsequent ionization processes. Unfortunately, in most cases, many fragment ions are formed, which may lead to overlapping signals of different analytes [42, 43]. Thus, the APCI ionization process is highly complex and there is ongoing research in observing and controlling APCI spectra of saturated hydrocarbons. For instance, Marotta and Paradisi studied the analysis of C5–C8 alkanes and identified several routes of how [M-H]+ ions may be formed in air plasma, including the reagent gases O2+•, NO+, H3O+, and hydrocarbon fragments [34]. Owen et al. measured lignin model compounds as well as n-hentriacontane (C31H64) and showed that ion distribution can be simplified by using CS2 as solvent and nitrogen as source gas, which led to abundant M+• ions and only minor fragmentation [44]. Gao et al. reported no or only little fragmentation and [M-H]+ as the only significant ion for n-hentriacontane when using pentane or hexane as solvent and nitrogen as source gas [41]. Hourani and Kuhnert investigated a direct infusion analysis of light shredder waste samples in heptane using APCI with N2 as source gas and enabled a quantification in the range of C7–C40 using [M-H]+ [25]. Tose et al. investigated the influence of different solvents for analyzing saturated, unsaturated, and oxygen-containing compounds in paraffin samples and reported iso-octane to be favorable for the first class and pentane for the latter ones with nitrogen as ion source gas [45]. Nyadong et al. coupled APCI with laser-induced acoustic desorption (LIAD) and applied both N2 and O2 as reagent gas. A C21–C40 alkane mixture as well as a paraffin mixture were analyzed. With N2, predominantly M+• ions were formed including significant fragmentation, while using O2 led to mainly [M-H]+ ions with only minor fragmentation [46]. A benefit of using O2 as ion source gas was confirmed by Jin et al. and Manheim et al. who analyzed lubricant oils by APCI-MS and determined stable [M-H]+ signals [35, 47]. An influence of the three ion source gases N2, He, and synthetic air on the ionization of saturated hydrocarbons (C5–C102) was evaluated by Souza et al., where helium and synthetic air showed better ionization efficiencies than nitrogen [48]. Again, [M-H]+ was the dominant ion. Recently, Li et al. characterized heavy base oils by APCI-MS and compared results with GC×GC/EI-MS evaluation. [M-H]+ ions were again produced for alkane species, but analysis of smaller alkanes failed due to an overlap with fragment ions. Interestingly, O2 and N2 as ion source gases gave identical results [43].
While most of the publications suggested hydride abstraction as the responsible mechanism for forming [M-H]+ ions, Manheim et al. determined this assumption to be wrong [42]. Instead, the authors concluded that primarily formed strong Brønsted acids, e.g., N2H+ and H3O+, transfer a proton to saturated hydrocarbons, which form [M-H]+ ions via subsequent H2 elimination. As the proton transfer is highly exothermic, the authors regarded it as the reason for fragmentation.
The literature studies highlight that there are exemplary successful applications of APCI(+) for analyzing saturated hydrocarbons (single compounds as well as hydrocarbon mixtures), often without chromatographic pre-separation. However, detailed ionization and fragmentation mechanisms and hence a reliable spectrum interpretation are still under discussion.
Although APCI-MS of gaseous hydrocarbons is not new, the method is usually applied for liquid samples. An online gas-phase monitoring application of volatile and semi-volatile n-alkanes via APCI(+) has, to the best of our knowledge, not been addressed in literature so far. As liquid samples are vaporized anyway, a direct gas-phase application can be assumed to be feasible and faster than conventional GC measurements.
In this work, we present a fully automated APCI(+)-MS method as a tool for a rapid online gas-phase analysis of n-alkanes. Measured spectra, most abundant ion groups, and fragmentation patterns are systematically investigated for n-heptane and n-decane as exemplary alkanes, to identify the most promising ion type that is suitable for a quantification in n-alkane mixtures. We further show that species calibration is feasible and applicable to both a binary and a five-component alkane mixture. By comparing temporal concentration profiles to a state-of-the-art GC system, the potential of increasing the analysis frequency via APCI(+)-MS is demonstrated.
Materials and methods
Experimental setup
The experiments were performed in a lab-scale evaporation set-up according to Fig. 1. n-heptane (VWR chemicals, 99.9% purity) and n-decane (Thermo Scientific, 99% purity) as well as a mixture of those two served as alkane standards. In a subsequent five-component experiment, n-dodecane (VWR chemicals, ≥99% purity), n-tetradecane (Alfa Aesar, 99% purity), and n-eicosane (SigmaAldrich, ≥99.8% purity) were added as well. The liquid is injected via a syringe pump, equipped with a pressure-tight stainless-steel syringe (both KD Scientific), into a heated tube, where the liquid is vaporized. A glass wool filling at the injector tip served as a disperser for more homogenous evaporation. Vaporized gases were transported by a diluted syngas mixture (CO/H2/N2=1:2:1, 200 mlSTP/min) as well as an internal GC standard (Ar/cyclopropane=99:1, 25 mlSTP/min) through a separator to an analytical set-up consisting of a Q Exactive Plus Orbitrap (Thermo Fisher) mass spectrometer and a three-channel gas chromatograph (Agilent 8860) equipped with one TCD and two FIDs. The separator, all pipes, and fittings were heated to at least 200 °C so that no condensation occurred within the periphery.
Alkane injections were performed by a programmed syringe pump program including five flow rate steps (3/2/1/0.5/0.25 µL/min), holding each flow rate for 4 h. To overcome initial evaporation inertia, especially at low syringe pump flow rates, it was advantageous to go from high to low injection speeds. The calibration experiments using n-heptane or n-decane only were repeated three times.
It is important to emphasize that both analytical devices were not combined as found in conventional GC/MS applications, but installed in parallel, i.e., they worked completely independently. While the GC device already had pre-installed injection valves and two sample loops (500 µl) for TCD and FIDs respectively, an additional temperature-stable 6-way injection valve (Vici Valco) including a sample loop (500 µl) was mounted upstream of the mass spectrometer. Hence, in contrast to often found real-time analysis applications [37, 39], the gas flow was not guided directly into the ionization chamber, but sampled via a valve switch and a constant argon carrier gas flow. In this way, a reproducible and process-independent sample injection was ensured. With no GC or MS method running, all sample loops were permanently purged by the gas flow. Two needle valves ensured an equal mass flow towards both instruments. The sample loop outflow was guided to individual separators at room temperature to collect condensing hydrocarbons. Pressure was controlled by two regulators and the end of the set-up to a relative pressure of 0.3 bar to facilitate sampling of the condensate.
The gas chromatograph served as a reference instrument to validate and calibrate the MS results. The GC method was a state-of-the-art analytical procedure to detect hydrocarbon species in the range from C1 to C20. For quantification, the calibrated internal standard cyclopropane (C3H6) was used. As the weight-specific FID response factor of all n-alkanes is nearly equal to one [49], we calculated the mole fraction of any n-alkane by
where \({y}_{{{\text{C}}}_{3}{{\text{H}}}_{6}}\) is the known mole fraction of cyclopropane, \({A}_{{{\text{C}}}_{i}}\) and \({M}_{{{\text{C}}}_{i}}\) are the peak area and the molar mass of the n-alkane analyte with the chain length i, and \({y}_{{{\text{C}}}_{3}{{\text{H}}}_{6}}\), \({A}_{{{\text{C}}}_{3}{{\text{H}}}_{6}}\), and \({M}_{{{\text{C}}}_{3}{{\text{H}}}_{6}}\) are the known mole fraction, the peak area, and the molar mass of cyclopropane. The GC measurement time was 33.5 min, allowing to capture alkanes up to C20. When only n-heptane or n-decane were investigated, the method duration could be reduced to 10 min or 17 min, respectively. More details about the GC configuration are given in the Supplementary Information.
APCI-MS method
The mass spectrometer was equipped with its APCI ion source, utilized in positive ionization mode (APCI(+)) over a mass-to-charge range of m/z 50–350. Mass-to-charge ratios below 50 could not be detected by the device, as it is usually applied for analyzing liquid samples. Hence, the gas components CO, N2, H2, and Ar as well as low-mass hydrocarbon fragments could not be analyzed. The presented method is therefore limited in the detection of low-mass VOCs and focuses on a liquid fuel range (~C5–C20 alkanes). To prevent species condensation, the vaporizer temperature was set to 400 °C. The capillary temperature was found to have a significant influence on the magnitude of fragmentation and was therefore set to only 250 °C. All MS parameters are listed in Table 1.
The sample loop injection was fully automated and executed by the MS method itself (via external circuit output). With a mass resolution of 70,000 and three microscans, the scan rate was about 1 scan/s. The signals of the ionized hydrocarbon gases were integrated to a peak area, determining the species concentration. The total measurement duration was selected to be 289.8 s (=4.83 min) in order to capture peak tailing. To enable an online monitoring, we did not evaluate the data using the standard Xcalibur (Thermo Fisher) software, but implemented an automatic workflow for processing the .raw files using a Python script. In this workflow, a list of all mass-to-charge ratios and its intensities of every scan is produced and anticipated ions are looked up within a defined m/z tolerance (measured m/z values differ slightly from theoretic values by Δm/z ±0.002-0.008). This way, all relevant alkane peak areas are calculated fully automated and plotted online. More details on the injection procedure and the data evaluation workflow are given in the Supplementary Information.
Results and discussion
APCI spectra
Every MS sample injection period consisted of almost 300 spectra. Despite the fact that cyclopropane has a molecular mass below 50 u, it showed a characteristic signal at m/z 74.0362, which corresponds to C3H6O2+• (Figure S5 in Supplementary Information). All other abundant signals were related to the two analytes n-heptane or n-decane. Exemplary mass spectra of the gas mixture containing n-heptane and n-decane, respectively, are shown in Fig. 2a and b. The signals refer to a measurement time \({t}_{{\text{meas}}}=10 ~ {\text{s}}\), i.e., 4 s after sample injection. A background spectrum that evolves before sample injection (\({t}_{{\text{meas}}}=4 ~ {\text{s}}\)), i.e., during a pure argon stream, is added to the graphs as well. Although we measured up to m/z 350, only m/z 50–175 is depicted here, as n-heptane and n-decane did not produce significant signals in the upper mass range.
Generally, we observe a high amount of fragment ions. Most abundant signals occur for C4, C5, and C6 hydrocarbon chain lengths. Smaller fragments are not visible (m/z < 50). Among the fragments, alkyl cations, i.e., CnH2n+1+, as well as alkenyl cations (alkyl cations with additional H2 elimination), i.e., CnH2n-1+, are dominant, which is typical for APCI(+) ionization of saturated hydrocarbons [25, 33, 34]. Ions without C-C bond cleavage, i.e., [M-H]+ (m/z 99.1168 for n-heptane; m/z 141.1638 for n-decane) and [M-3H]+ (m/z 97.1012 for n-heptane; m/z 139.1481 for n-decane) were detected as well, but with smaller intensities compared to short-chained fragments. Moreover, [M-H+O]+ cations were measured (m/z 115.1117 for n-heptane; m/z 157.1587 for n-decane). These oxygen-containing ions are often denoted as monohydrated [M-3H+H2O]+ ions, assuming a water cluster formation based on atmospheric humidity in the ionization chamber [25, 33, 46, 50]. Although we did not verify a water-based mechanism in this study, we assumed the same formation pathway and adopted this formulation in the following evaluation.
Peak shapes
As outlined in the “APCI-MS method” section, the abundance of an ion is evaluated by integrating its signal over the measurement time, which results in a peak area. Exemplary peaks of three ions [M-H]+, [M-3H]+ and [M-3H+H2O]+ are depicted for n-heptane and for n-decane in Fig. 3a and b, respectively (at \({y}_{{{\text{C}}}_{7}{{\text{H}}}_{16}}\approx 1500 ~ {\text{ppm}},{y}_{{{\text{C}}}_{10}{{\text{H}}}_{22}}\approx 900 ~ {\text{ppm}}\)). To make the most interesting part of the peaks more visible, they are only shown in the range 0 < \({t}_{{\text{meas}}}\) < 50 s.
It can be seen that [M-H]+ and [M-3H]+ had rather similar peak shapes, whereas [M-3H+H2O]+ formed a very broad peak. While the baseline value was reached within the 50 s window for [M-H]+ and [M-3H]+, tailing was more pronounced for [M-3H+H2O]+, especially with n-decane as analyte. This means that if only [M-H]+ and [M-3H]+ ions were to be evaluated, the measured duration could be shortened to about 1 min. However, we wanted to include the oxygen-containing ions into the evaluation and therefore chose a rather long measurement duration of about 5 min. The observation suggests that water addition occurred not immediately, but via a secondary reaction step. As metal surfaces are known to be covered by water molecules [51, 52], this secondary reaction might have occurred on the walls of the ionization chamber and hence cause a temporal delay.
Although having a lower concentration level, the n-decane peak areas were larger than n-heptane peak areas, except for [M-3H]+. This was expected, since the higher proton affinity of longer n-alkane chains should cause a higher ionization probability by proton transfer [53, 54].
Fragmentation pattern of CnH2n+1 +, CnH2n-1 +, and CnH2n+1O+ ion groups
The fragmentation pattern is crucial for the ability of analyzing not only pure analytes but also alkane mixtures. Thus, we did not only investigate [M-H]+, [M-3H]+ and [M-3H+H2O]+ ions but all smaller fragment ions as well. We defined the three chain-length independent ion groups CnH2n+1+, CnH2n-1+, and CnH2n+1O+ for a detailed evaluation.
The temporal development of calculated peak areas from a five-step syringe injection experiment of n-decane is shown for all ion groups and different chain lengths in Fig. 4.
To highlight the general data processing workflow of measured MS signals, the inset in Fig. 4a presents one exemplary peak (CnH2n+1+ with n=6), measured from one sample injection. The general signal trend is similar for all three ion groups. As fragment allocation is trivial with a single n-alkane analyte, each ion group is able to qualitatively depict the five concentration levels of n-decane. However, apparently, the fragmentation pattern is significantly different for each ion group: CnH2n+1+ and CnH2n-1+ chains are highly fragmented; for CnH2n+1+, especially n=4, n=5, and n=6 are dominant, while for CnH2n-1+, n=6, n=7, and n=8 are abundant. The ions without chain cleavage, i.e., [M-H]+ and [M-3H]+ (n=10), are visible but with rather low abundance. In contrast, the abundance of [M-3H+H2O]+ is not only high, but within its ion group CnH2n+1O+, it is the only directly observable signal. By magnifying the time period of the start of the syringe pump injection (inset in Fig. 4c), a little increment of the short fragment peak areas is indeed existent, but hardly recognizable. This fragmentation behavior, i.e., a high probability for C-C bond cleavage in case of CnH2n+1+ and CnH2n-1+, but an insignificant one for CnH2n+1O+ was not only observed with n-decane as analyte, but with n-heptane as well (Figure S6 in Supplementary Information). If the assumption of a secondary surface reaction with adsorbed water is correct, the low fragmentation tendency might be explained by a release of internal energy to the wall. A mechanistic investigation on the formation of [M-3H+H2O]+ is currently in progress and will be dealt with in an upcoming study. Regardless of the mechanism of formation of these ions, we were able to show that they are analyte-specific signals, which enable substance identification in n-alkane mixtures. Therefore, we focused on this ion type in the following evaluation and — in contrast to most previous APCI studies dealing with n-alkanes — neglected all other ions, such as [M-H]+ or [M-3H]+.
Calibration of [M-3H+H2O]+ ions
The [M-3H+H2O]+ peak area data were calibrated by scaling them to the mole fractions measured by the GC reference instrument. All small amounts of fragment ratios within the CnH2n+1O+ ion group, which were determined in the previous chapter, were neglected. Parity plots between the measured mole fractions \({y}_{i}\) by GC and by MS are shown for n-heptane and n-decane in Fig. 5a and b, respectively. The calibration functions for the MS peak areas \({A}_{{{\text{C}}}_{7}{{\text{H}}}_{16}}\) and \({A}_{{{\text{C}}}_{10}{{\text{H}}}_{22}}\) are added to the graphs as well. All data points were averaged in each concentration step for both GC as well as MS analysis. Error bars indicate the standard deviation for both instruments. The black dashed line presents a perfect agreement; additionally, an error range of 10 % and 30 % is indicated.
The peak areas of [M-3H+H2O]+ signals go along quite linearly with the GC reference concentrations for both n-heptane and n-decane. Most data points lie within the 10 % error region; the rest is valid with a maximum error of 30 %. Moreover, standard deviations (error bars) are quite similar for the MS signals compared to the GC data, which means that the uncertainty of the MS method is in the same range as the uncertainty of the reference method or of the entire syringe injection procedure. As can be seen from the calibration functions added to the panels, an offset had to be subtracted in case of n-heptane, while it could be neglected for n-decane. This might be due to trace impurities in the system. Although one would assume a stronger memory effect for n-decane, the impact of an offset signal in case of n-heptane could be more pronounced because the signal response itself is lower (calibration factor is about 2.5-fold compared to n-decane).
Binary mixture of n-heptane/n-decane
We used a binary mixture of n-heptane/n-decane (v/v = 1:1) to check whether the calibration functions are affected by the presence of the other hydrocarbon species. It is important to mention that the rest of the gas-phase matrix (CO/H2/N2/Ar/cyclopropane) remained the same, i.e., strictly speaking, the final sample was not a binary mixture. This designation only refers to the injected liquid alkane standard. To investigate more dynamic concentration profiles, we programmed a syringe pump injection procedure with multiple step changes in varying intervals. Since GC data still served as reference, syringe pump changes were not implemented too fast, so that they could be still observed by the longer GC measurements. To enable a partial species separation of n-heptane and n-decane, the temperature of the evaporation zone and the separator were set down to 50 °C.
The mole fractions according to the syringe pump program (color-filled area) as well as the development of actual mole fractions as measured by GC and MS are presented in Fig. 6. It can be seen that the lower temperature actually led to a slower evaporation speed for n-decane compared to n-heptane. Still, both mole fractions eventually reached their expected gas-phase concentrations based on the syringe injection profile, indicating no significant remaining liquid hold-up. This behavior is consistent for all concentration changes.
The concentration profiles were well captured by both GC and MS analysis, emphasizing the suitability of the MS method to correctly measure n-alkane concentrations in gas-phase mixtures. It has to be noted that the MS scaling function, required for matching the GC data, slightly changed to \({y}_{{{\text{C}}}_{7}{{\text{H}}}_{16}}=2.3\times {10}^{-5}\bullet {A}_{{{\text{C}}}_{7}{{\text{H}}}_{16}}\) ; \({y}_{{{\text{C}}}_{10}{{\text{H}}}_{22}}=8.93\times {10}^{-6}\bullet {A}_{{{\text{C}}}_{10}{{\text{H}}}_{22}}\)(before: \({y}_{{{\text{C}}}_{7}{{\text{H}}}_{16}}=2.55\times {10}^{-5}\bullet ({A}_{{{\text{C}}}_{7}{{\text{H}}}_{16}}-1.85\times {10}^{-6})\); \({y}_{{{\text{C}}}_{10}{{\text{H}}}_{22}}=9.95\times {10}^{-6}\bullet {A}_{{{\text{C}}}_{10}{{\text{H}}}_{22}}\)). This small deviation was not due to the presence of another alkane species. The mixture experiment was not performed right after calibration and other measurements were conducted in between. A fine-tuned recalibration of the system was thus to be expected. An offset for n-heptane is no longer visible, which indicates that this had been an artifact due to impurities before. Although the GC measuring time was generally sufficient to capture all applied concentrations changes, mole fraction dynamics were captured in more detail by MS. Moreover, at t = 60 min, heating of the evaporation zone was turned off, which temporally decreased \({y}_{{{\text{C}}}_{10}{{\text{H}}}_{22}}\). A corresponding short-term increase of \({y}_{{{\text{C}}}_{7}{{\text{H}}}_{16}}\) was noticed by MS, but not by GC. This behavior cannot be verified, but it seems reasonable.
Five-component mixture
So far, only n-heptane and n-decane were investigated as hydrocarbon analytes. To evaluate the MS method for more complex n-alkane mixtures, we tested a five-component n-alkane standard mixture consisting of C7/C10/C12/C14/C20 (\({y}_{i}\) = 0.624/0.214/0.105/0.051/0.006). The same dynamic injection profile (Fig. 6) was used to deliver the alkane mixture into the setup. The temperature of the evaporation zone and of the separator were set to 60 °C at the start of the experiment and increased step-wise to enable some dynamic condensation/evaporation processes of the different species. The GC measurement time was increased to 33.5 min to measure hydrocarbons up to n-eicosane. Including the subsequent oven cooling, the interval between two GC data points was about 40 min.
The concentration profiles measured by GC and MS are shown are Fig. 7a. Again, the theoretical syringe-injected mole fraction profiles — valid if all molecules are in the gas phase only — are added as color-filled areas. The mass-to-charge ratios (m/z) and the determined scaling parameters of [M-3H+H2O]+ ions are listed for all five components in Table 2. Here, the scaling factors for n-heptane and n-decane remained the same compared to the experiment using the binary alkane mixture. Thus, calibration is indeed independent of the other hydrocarbon analytes, which means that a simultaneous calibration of n-alkanes is possible.
Compared to the binary experiment, alkane species were not detected right after injection but an initial delay time is visible. We observed such a phenomenon in several experiments, indicating a saturation process of periphery surfaces. Afterwards, a different dynamic behavior of the individual species is apparent, which is due to their different vapor pressures. At the time of t = 200 min, n-heptane and n-decane had reached a plateau, and the signal of n-dodecane was slowly increasing. C14 and C20 were not yet detected. At t = 218 min, the temperatures \({T}_{{\text{Evap}}}\) and \({T}_{{\text{Sep}}}\) were increased to 100 °C, causing an immediate increase of \({y}_{{{\text{C}}}_{10}}\), \({y}_{{{\text{C}}}_{12}},\) and \({y}_{{{\text{C}}}_{14}}\). After the second temperature increment (\({T}_{{\text{Evap}}}\) = \({T}_{{\text{Sep}}}\) = 150 °C at t = 282 min), \({y}_{{{\text{C}}}_{12}}\) and \({y}_{{{\text{C}}}_{14}}\) increased strongly, followed by a slow increase of \({y}_{{{\text{C}}}_{20}}\). A high mole fraction peak for n-eicosane was observed immediately after the third temperature increment (\({T}_{{\text{Evap}}}\) = 250 °C, \({T}_{{\text{Sep}}}\) = 150 °C at t = 326 min). These temperature-driven concentration changes were overlapped by the dynamics caused by the syringe pump program. As expected, with increasing chain length the signal response to any applied change was slower and less intense, due to a decreasing vapor pressure.
For all five n-alkanes, we see a quite good agreement of GC and MS data. Hence, although not every hydrocarbon species was calibrated previously, it does not seem to be a problem to scale any n-alkane to the GC reference. Moreover, the results not only confirm again the suitability of the MS method to characterize hydrocarbons in gas streams, but also illustrate the benefit of an increased temporal resolution. This is especially visible in the marked magnification window shown in Fig. 7b. The short-term increase of syringe pump volume flow was not recognized by GC, but captured quite well by MS. With an MS data point interval of about 5 min, temporal resolution is increased by a factor of 8 compared to the GC measurements and is thus better suited for measuring the system dynamics.
A closer look to the concentration profile of C20 (second magnification window in Fig. 7c) reveals two striking aspects. Firstly, the measured mole fractions of n-eicosane were higher than their theoretical values as derived from the syringe pump profile, while C12 and C14 were lower (for better illustration, data of C7 and C10 are not shown here). As most C20 molecules accumulated as liquid within the setup after starting the experiment and evaporated only after increasing the temperature, this observation is reasonable and demonstrates the transient separation process. Secondly, although generally matching the GC data quite well, the MS data of C20 shows more signal fluctuations compared to the other species. Within the setup, less volatile species usually respond more slowly to temperature changes, so that the oscillations for C20 do not really make sense. An explanation could be a the poor C20 peak quality due to significant molecule transport limitations based on low volatility. Either within the sample loop or in the ionization chamber C20 molecules probably remained adsorbed to the surfaces for quite a long time, so that the ion signal did not fit into the 5-min measurement window (Figure S7 in the Supplementary Information). Multiple adsorption/desorption processes likely occurred on the way to the mass analyzer. This is clearly a disadvantage compared to conventional liquid MS applications where heavy compounds are dissolved in a suitable solvent and do not suffer from major transport limitations. In this gas-phase application, improvements might be possible by increasing the temperature of sample loop, vaporizer, and/or capillary. The broad peak further explains the relatively large scaling factor for n-eicosane listed in Table 2, where a generally decreasing factor with increased chain length is visible. With a full-range peak, calculated C20 peak areas would be larger, resulting in a lower response factor. Nevertheless, despite the low peak quality, the MS data of C20 follow the trend of GC measurements quite well, i.e., a clear correlation with the mole fraction is visible.
Conclusion
We presented a new method for a rapid online gas-phase analysis of volatile and semi-volatile n-alkanes using APCI(+)-MS using [M-3H+H2O]+ ions for quantification. In contrast to [M-H]+ and [M-3H]+ ions, [M-3H+H2O]+ ions showed negligible chain cleavage, which allowed for an explicit analyte identification and a good match of concentration profiles with GC reference data for a one-, two-, and five-component alkane mixture.
In this context, [M-3H+H2O]+ ions are assumed to be formed via secondary reactions with surface adsorbed water molecules. A corresponding mechanistic study is currently being carried out. Unfortunately, [M-3H+H2O]+ ion transport to the analyzer is slow, which prolongs the measuring time to capture the long-tailing peaks. Still, the APCI-MS method has proven to be valid and faster than a conventional GC method. The temporal benefit depends on the range of hydrocarbons to be analyzed. In this study, temporal resolution for a characterization up to n-eicosane was improved by a factor of 8. This way, dynamic concentration changes, which were invisible to the GC system, were captured by the MS instrument.
However, with a minimum m/z 50 of the detector used, a mixture characterization below alkane chain lengths of n=3 is impossible. Short-chained alkanes might be problematic anyway, as their proton affinity might be too low for ionization. For long-chained molecules, such as n-eicosane, significant transport limitations due to low volatility became apparent, which have to be further examined. Furthermore, matrix effects have to be assessed, by other hydrocarbons (e.g., branched, unsaturated, aromatic), heteroatomic compounds (especially oxygen-containing) and by changes of the permanent gases (e.g., CO or H2). The latter aspect is especially important when applying the method to reaction systems such as Fischer-Tropsch synthesis, where CO and H2 are consumed, i.e., the matrix changes over time. Hence, there is still a range of uncertainties that need to be clarified, which will be topic of further investigations. Nevertheless, the method is already usable to identify system dynamics in transient n-alkane gas-liquid systems, such as residence time experiments or separation processes.
Data availability
All data, the automated monitoring workflow, and the evaluation scripts that were used for this study are available in an online repository [55].
References
Callis JB, Illman DL, Kowalski BR. Process analytical chemistry. Anal Chem. 1987;59:624A-637A. https://doi.org/10.1021/ac00136a001.
Koch KH. Process analytical chemistry. Berlin, Heidelberg: Springer-Verlag GmbH; 1999.
Koel M, Kaljurand M. Application of the principles of green chemistry in analytical chemistry. Pure Appl Chem. 2006;78:1993–2002. https://doi.org/10.1351/pac200678111993.
Bakeev KA. Process analytical technology: spectroscopic tools and implementation strategies for the chemical and pharmaceutical industries. John Wiley & Sons; 2010.
Ge Z, Song Z, Gao F. Review of recent research on data-based process monitoring. Ind Eng Chem Res. 2013;52:3543–62. https://doi.org/10.1021/ie302069q.
Workman JJr, Lavine B, Chrisman R, Koch M (2011) Process analytical chemistry. Anal Chem 83:4557–4578. https://doi.org/10.1021/ac200974w.
Kalz KF, Kraehnert R, Dvoyashkin M, Dittmeyer R, Gläser R, Krewer U, Reuter K, Grunwaldt J-D. Future challenges in heterogeneous catalysis: understanding catalysts under dynamic reaction conditions. ChemCatChem. 2017;9:17–29. https://doi.org/10.1002/cctc.201600996.
Giglio E, Pirone R, Bensaid S. Dynamic modelling of methanation reactors during start-up and regulation in intermittent power-to-gas applications. Renew Energy. 2021;170:1040–51. https://doi.org/10.1016/j.renene.2021.01.153.
Palys MJ, Daoutidis P. Power-to-X: a review and perspective. Comput Chem Eng. 2022;165:107948. https://doi.org/10.1016/j.compchemeng.2022.107948.
Wentrup J, Pesch GR, Thöming J. Dynamic operation of Fischer-Tropsch reactors for power-to-liquid concepts: a review. Renew Sustain Energy Rev. 2022;162:112454. https://doi.org/10.1016/j.rser.2022.112454.
Nikačević N, Todić B, Mandić M, Petkovska M, Bukur DB. Optimization of forced periodic operations in milli-scale fixed bed reactor for Fischer-Tropsch synthesis. Catal Today. 2020;343:156–64. https://doi.org/10.1016/j.cattod.2018.12.032.
Duerksen A, Thiessen J, Kern C, Jess A. Fischer-Tropsch synthesis with periodical draining of a liquid-filled catalyst by hydrogenolysis. Sustain Ener Fuels. 2020;4:2055–64. https://doi.org/10.1039/C9SE01269A.
Tafete GA, Habtu NG. Reactor configuration, operations and structural catalyst design in process intensification of catalytic reactors: a review. Chem Eng Proc - Process Intensif. 2023;184:109290. https://doi.org/10.1016/j.cep.2023.109290.
Hsu CS, Robinson PR. Natural gas and petroleum products. In: Hsu CS, Robinson PR, editors. Petroleum Science and Technology. Cham: Springer International Publishing; 2019. p. 301–42.
Sophonrat N, Sandström L, Zaini IN, Yang W. Stepwise pyrolysis of mixed plastics and paper for separation of oxygenated and hydrocarbon condensates. Appl Energy. 2018;229:314–25. https://doi.org/10.1016/j.apenergy.2018.08.006.
Hodala JL, Moon DJ, Reddy KR, Reddy CV, Kumar TN, Ahamed MI, Raghu AV. Catalyst design for maximizing C5+ yields during Fischer-Tropsch synthesis. Int J Hydrog Energy. 2021;46:3289–301. https://doi.org/10.1016/j.ijhydene.2019.12.021.
Rüger CP, Tiemann O, Neumann A, Streibel T, Zimmermann R. Review on evolved gas analysis mass spectrometry with soft photoionization for the chemical description of petroleum, petroleum-derived materials, and alternative feedstocks. Energy Fuels. 2021;35:18308–32. https://doi.org/10.1021/acs.energyfuels.1c02720.
Marshall AG, Rodgers RP. Petroleomics: the next grand challenge for chemical analysis. Acc Chem Res. 2004;37:53–9. https://doi.org/10.1021/ar020177t.
Bojkovic A, Vermeire FH, Kuzmanović M, Dao Thi H, Van Geem KM. Analytics driving kinetics: advanced mass spectrometric characterization of petroleum products. Energy Fuels. 2022;36:6–59. https://doi.org/10.1021/acs.energyfuels.1c02355.
N. Hewitt C, Hayward S, Tani A (2003) The application of proton transfer reaction- mass spectrometry ( PTR-MS ) to the monitoring and analysis of volatile organic compounds in the atmosphere. J Environ Monit 5:1–7. https://doi.org/10.1039/B204712H.
Jordan A, Haidacher S, Hanel G, Hartungen E, Märk L, Seehauser H, Schottkowsky R, Sulzer P, Märk TD. A high resolution and high sensitivity proton-transfer-reaction time-of-flight mass spectrometer (PTR-TOF-MS). Int J Mass Spectrom. 2009;286:122–8. https://doi.org/10.1016/j.ijms.2009.07.005.
Yu Z, Liu C, Niu H, Wu M, Gao W, Zhou Z, Huang Z, Li X. Real time analysis of trace volatile organic compounds in ambient air: a comparison between membrane inlet single photon ionization mass spectrometry and proton transfer reaction mass spectrometry. Anal Meth. 2020;12:4343–50. https://doi.org/10.1039/D0AY01102A.
Zhou Z, Chen X, Ma H, Liu C, Zhou C, Qi F. Real-time monitoring biomass pyrolysis via on-line photoionization ultrahigh-resolution mass spectrometry. Fuel. 2019;235:962–71. https://doi.org/10.1016/j.fuel.2018.08.098.
Liu P, Huang J, Yang K, Zhuang H, Chen L, Pan Y, Yang J, Jia L. Exploring the reaction chemistry of biomass upgrading over HZSM-5 catalyst through model compounds. Fuel. 2022;312:122874. https://doi.org/10.1016/j.fuel.2021.122874.
Hourani N, Kuhnert N. Development of a novel direct-infusion atmospheric pressure chemical ionization mass spectrometry method for the analysis of heavy hydrocarbons in light shredder waste. Anal Meth. 2012;4:730–5. https://doi.org/10.1039/C2AY05249K.
Zhang T, Li Z-Y, Zhang M-Q, He S-G. Gas-phase reactions of atomic gold cations with linear alkanes (C2–C9). J Phys Chem A. 2016;120:4285–93. https://doi.org/10.1021/acs.jpca.6b03836.
Dehon C, Lemaire J, Heninger M, Chaput A, Mestdagh H. Chemical ionization using CF3+: efficient detection of small alkanes and fluorocarbons. Int J Mass Spectrom. 2011;299:113–9. https://doi.org/10.1016/j.ijms.2010.09.033.
Inomata S, Tanimoto H, Yamada H. Mass spectrometric detection of alkanes using NO+ chemical ionization in proton-transfer-reaction plus switchable reagent ion mass spectrometry. Chem Lett. 2014;43:538–40. https://doi.org/10.1246/cl.131105.
Hanley L, Zimmermann R. Light and molecular ions: the emergence of vacuum UV single-photon ionization in MS. Anal Chem. 2009;81:4174–82. https://doi.org/10.1021/ac8013675.
Adam T, Zimmermann R. Determination of single photon ionization cross sections for quantitative analysis of complex organic mixtures. Anal Bioanal Chem. 2007;389:1941–51. https://doi.org/10.1007/s00216-007-1571-x.
Dang M, Liu R, Dong F, Liu B, Hou K. Vacuum ultraviolet photoionization on-line mass spectrometry: instrumentation developments and applications. TrAC Trends Anal Chem. 2022;149:116542. https://doi.org/10.1016/j.trac.2022.116542.
He S, Cui H, Lai Y, Sun C, Luo S, Li H, Seshan K. A temperature-programmed reaction/single-photon ionization time-of-flight mass spectrometry system for rapid investigation of gas–solid heterogeneous catalytic reactions under realistic reaction conditions. Catal Sci Technol. 2015;5:4959–63. https://doi.org/10.1039/C5CY01550B.
Bell SE, Ewing RG, Eiceman GA, Karpas Z. Atmospheric pressure chemical ionization of alkanes, alkenes, and cycloalkanes. J Am Soc Mass Spectrom. 1994;5:177–85. https://doi.org/10.1016/1044-0305(94)85031-3.
Marotta E, Paradisi C. A mass spectrometry study of alkanes in air plasma at atmospheric pressure. J Am Soc Mass Spectrom. 2009;20:697–707. https://doi.org/10.1016/j.jasms.2008.12.005.
Jin C, Viidanoja J, Li M, Zhang Y, Ikonen E, Root A, Romanczyk M, Manheim J, Dziekonski E, Kenttämaa HI. Comparison of atmospheric pressure chemical ionization and field ionization mass spectrometry for the analysis of large saturated hydrocarbons. Anal Chem. 2016;88:10592–8. https://doi.org/10.1021/acs.analchem.6b02789.
Kolakowski BM, Grossert JS, Ramaley L. Studies on the positive-ion mass spectra from atmospheric pressure chemical ionization of gases and solvents used in liquid chromatography and direct liquid injection. J Am Soc Mass Spectrom. 2004;15:311–24. https://doi.org/10.1016/j.jasms.2003.10.019.
Badjagbo K, Sauvé S, Moore S. Real-time continuous monitoring methods for airborne VOCs. TrAC Trends Anal Chem. 2007;26:931–40. https://doi.org/10.1016/j.trac.2007.07.004.
Badjagbo K, Picard P, Moore S, Sauvé S. Direct atmospheric pressure chemical ionization-tandem mass spectrometry for the continuous real-time trace analysis of benzene, toluene, ethylbenzene, and xylenes in ambient air. J Am Soc Mass Spectrom. 2009;20:829–36. https://doi.org/10.1016/j.jasms.2008.12.021.
Warscheid B, Kückelmann U, Hoffmann T. Direct quantitative analysis of organic compounds in the gas and particle phase using a modified atmospheric pressure chemical ionization source in combination with ion trap mass spectrometry. Anal Chem. 2003;75:1410–7. https://doi.org/10.1021/ac025788d.
Zuth C, Vogel AL, Ockenfeld S, Huesmann R, Hoffmann T. Ultrahigh-resolution mass spectrometry in real time: atmospheric pressure chemical ionization Orbitrap mass spectrometry of atmospheric organic aerosol. Anal Chem. 2018;90:8816–23. https://doi.org/10.1021/acs.analchem.8b00671.
Gao J, Owen BC, Borton DJ, Jin Z, Kenttämaa HI. HPLC/APCI mass spectrometry of saturated and unsaturated hydrocarbons by using hydrocarbon solvents as the APCI reagent and HPLC mobile phase. J Am Soc Mass Spectrom. 2012;23:816–22. https://doi.org/10.1007/s13361-012-0347-5.
Manheim JM, Milton JR, Zhang Y, Kenttämaa HI. Fragmentation of saturated hydrocarbons upon atmospheric pressure chemical ionization is caused by proton-transfer reactions. Anal Chem. 2020;92:8883–92. https://doi.org/10.1021/acs.analchem.0c00681.
Li W, Manheim JM, Fu Y, Laaksonen T, Kilaz G, Kenttämaa HI. Comparison of APCI orbitrap MS and GCxGC/EI TOF MS for the hydrocarbon analysis of heavy base oils. Fuel. 2023;343:127993. https://doi.org/10.1016/j.fuel.2023.127993.
Owen BC, Gao J, Borton DJ II, Amundson LM, Archibold EF, Tan X, Azyat K, Tykwinski R, Gray M, Kenttämaa HI. Carbon disulfide reagent allows the characterization of nonpolar analytes by atmospheric pressure chemical ionization mass spectrometry. Rapid Commun Mass Spectrom. 2011;25:1924–8. https://doi.org/10.1002/rcm.5063.
Tose LV, Cardoso FMR, Fleming FP, Vicente MA, Silva SRC, Aquije GMFV, Vaz BG, Romão W. Analyzes of hydrocarbons by atmosphere pressure chemical ionization FT-ICR mass spectrometry using isooctane as ionizing reagent. Fuel. 2015;153:346–54. https://doi.org/10.1016/j.fuel.2015.03.004.
Nyadong L, Quinn JP, Hsu CS, Hendrickson CL, Rodgers RP, Marshall AG. Atmospheric pressure laser-induced acoustic desorption chemical ionization mass spectrometry for analysis of saturated hydrocarbons. Anal Chem. 2012;84:7131–7. https://doi.org/10.1021/ac301307p.
Manheim J, Zhang Y, Viidanoja J, Kenttämaa HI. An automated method for chemical composition analysis of lubricant base oils by using atmospheric pressure chemical ionization mass spectrometry. J Am Soc Mass Spectrom. 2019;30:2014–21. https://doi.org/10.1007/s13361-019-02284-6.
Souza LM, Tose LV, Cardoso FMR, Fleming FP, Pinto FE, Kuster RM, Filgueiras PR, Vaz BG, Romão W. Evaluating the effect of ion source gas (N2, He, and synthetic air) on the ionization of hydrocarbon, condensed aromatic standards, and paraffin fractions by APCI(+)FT-ICR MS. Fuel. 2018;225:632–45. https://doi.org/10.1016/j.fuel.2018.03.180.
Dietz WA. Response factors for gas chromatographic analyses. J Chromatogr Sci. 1967;5:68–71. https://doi.org/10.1093/chromsci/5.2.68.
Borsdorf H, Nazarov EG, Eiceman GA. Atmospheric pressure chemical ionization studies of non-polar isomeric hydrocarbons using ion mobility spectrometry and mass spectrometry with different ionization techniques. J Am Soc Mass Spectrom. 2002;13:1078–87. https://doi.org/10.1016/S1044-0305(02)00429-4.
Dobrozemsky R, Menhart S, Buchtela K. Residence times of water molecules on stainless steel and aluminum surfaces in vacuum and atmosphere. J Vac Sci Technol Vac Surf Films. 2007;25:551–6. https://doi.org/10.1116/1.2718958.
Hodgson A, Haq S. Water adsorption and the wetting of metal surfaces. Surf Sci Rep. 2009;64:381–451. https://doi.org/10.1016/j.surfrep.2009.07.001.
Arnold ST, Viggiano AA, Morris RA. Rate constants and product branching fractions for the reactions of H3O+ and NO+ with C2–C12 alkanes. J Phys Chem A. 1998;102:8881–7. https://doi.org/10.1021/jp9815457.
Hunter KC, East ALL. Properties of C−C bonds in n-alkanes: relevance to cracking mechanisms. J Phys Chem A. 2002;106:1346–56. https://doi.org/10.1021/jp0129030.
Wentrup J (2023) Supporting files for “rapid online analysis of n-alkanes in gaseous streams via APCI mass spectrometry.” https://doi.org/10.5281/zenodo.10203915.
Acknowledgements
The authors like to thank Robert Güttel (University of Ulm) and Alica Rother (Uni Bremen, CVT) for valuable discussions.
Funding
Open Access funding enabled and organized by Projekt DEAL. The authors appreciate the funding by the German Federal Ministry of Education and Research (BMBF) and the Federal Ministry for Economic Affairs and Energy (BMWi) through the QUARREE100 project (Grant 03SBE113B).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Wentrup, J., Bösing, I., Dülcks, T. et al. Rapid online analysis of n-alkanes in gaseous streams via APCI mass spectrometry. Anal Bioanal Chem 416, 1843–1855 (2024). https://doi.org/10.1007/s00216-024-05182-3
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-024-05182-3