Abstract
This study examines how the height and width of peaks exiting the secondary column of a comprehensive two-dimensional gas chromatography (GC × GC) separation are affected by the width of the pulse introduced to the secondary column. A flow-modulated GC × GC apparatus was assembled that allowed input pulse widths to be controlled precisely over a range of 10 to 70 ms. GC × GC chromatograms were obtained using secondary columns containing a polyethylene glycol stationary phase with internal diameters of 0.25 and 0.32 mm. The area, height, and width of peaks emerging from the secondary column were found to be accurately modeled by the convolution of a rectangular function with a Gaussian distribution. The rectangular function represents the input pulse, and the Gaussian distribution represents the broadening that occurs in the secondary column. The minimum peak width that could be produced by the secondary column was determined for a wide range of compounds. Injection pulse widths that matched a compound’s minimum peak width produced peaks that were 25% wider than the minimum width and had heights that were 76% of the maximum possible peak height. Increasing the injection width significantly above the minimum width yielded substantially broader peaks with only a modest increase (< 25%) in peak height.
Graphical abstract
Similar content being viewed by others
References
Bahaghighat HD, Freye CE, Synovec RE. Recent advances in modulator technology for comprehensive two dimensional gas chromatography. Trends Anal Chem. 2019;113:379–91. https://doi.org/10.1016/j.trac.2018.04.016.
Tranchida PQ. Comprehensive two-dimensional gas chromatography: a perspective on processes of modulation. J Chromatogr A. 2018;1536:2–5. https://doi.org/10.1016/j.chroma.2017.04.039.
Amaral MSS, Nolvachai Y, Marriott PJ. Comprehensive two-dimensional gas chromatography advances in technology and applications: biennial update. Anal Chem. 2020;92:85–104. https://doi.org/10.1021/acs.analchem.9b05412.
Liu Z, Phillips JB. Comprehensive two-dimensional gas chromatography using an on-column thermal modulator interface. J Chromatogr Sci. 1991;29:227–31. https://doi.org/10.1093/chromsci/29.6.227.
Luong J, Guan X, Xu S, Gras R, Shellie RA. Thermal independent modulator for comprehensive two-dimensional gas chromatography. Anal Chem. 2016;88:8428–32. https://doi.org/10.1021/acs.analchem.6b02525.
Panić O, Górecki T, McNeish C, Goldstein AH, Williams BJ, Worton DR, et al. Development of a new consumable-free thermal modulator for comprehensive two-dimensional gas chromatography. J Chromatogr A. 2011;1218:3070–9. https://doi.org/10.1016/j.chroma.2011.03.024.
Zoccali M, Giocastro B, Tranchida PQ, Mondello L. Use of a recently developed thermal modulator within the context of comprehensive two-dimensional gas chromatography combined with time-of-flight mass spectrometry: gas flow optimization aspects. J Sep Sci. 2019;42:691–7. https://doi.org/10.1002/jssc.201801095.
Bruckner CA, Prazen BJ, Synovec RE. Comprehensive two dimensional high-speed gas chromatography with chemometric analysis. Anal Chem. 1998;70:2796–804.
Ghosh A, Bates CT, Seeley SK, Seeley JV. High speed Deans switch for low duty cycle comprehensive two-dimensional gas chromatography. J Chromatogr A. 2013;1291:146–54. https://doi.org/10.1016/j.chroma.2013.04.003.
Seeley JV, Kramp F, Hicks CJ. Comprehensive two-dimensional gas chromatography via differential flow modulation. Anal Chem. 2000;72:4346–52.
Seeley JV, Micyus NJ, McCurry JD, Seeley SK. Comprehensive two-dimensional gas chromatography with a simple fluidic modulator. Am Lab News. 2006;38:24.
Griffith JF, Winniford WL, Sun K, Edam R, Luong JC. A reversed-flow differential flow modulator for comprehensive two-dimensional gas chromatography. J Chromatogr A. 2012;1226:116–23. https://doi.org/10.1016/j.chroma.2011.11.036.
Trinklein TJ, Gough DV, Warren CG, Ochoa GS, Synovec RE. Dynamic pressure gradient modulation for comprehensive two-dimensional gas chromatography. J Chromatogr A. 2020;1609:460488. https://doi.org/10.1016/j.chroma.2019.460488.
Cai H, Stearns SD. Comprehensive two-dimensional gas chromatography using direct flow modulation to extend the secondary dimension separation time. J Chromatogr A. 2022;1669:462930. https://doi.org/10.1016/j.chroma.2022.462930.
Tranchida PQ, Maimone M, Franchina FA, Bjerk TR, Zini CA, Purcaro G, et al. Four-stage (low-)flow modulation comprehensive gas chromatography-quadrupole mass spectrometry for the determination of recently-highlighted cosmetic allergens. J Chromatogr A. 2016;1439:144–51. https://doi.org/10.1016/j.chroma.2015.12.002.
Seeley JV, Schimmel NE, Seeley SK. The multi-mode modulator: a versatile fluidic device for two-dimensional gas chromatography. J Chromatogr A. 2018;1536:6–15. https://doi.org/10.1016/j.chroma.2017.06.030.
Franchina FA, Maimone M, Tranchida PQ, Mondello L. Flow modulation comprehensive two-dimensional gas chromatography–mass spectrometry using ≈4mLmin−1 gas flows. J Chromatogr A. 2016;1441:134–9. https://doi.org/10.1016/j.chroma.2016.02.041.
Tranchida PQ, Franchina FA, Dugo P, Mondello L. Use of greatly-reduced gas flows in flow-modulated comprehensive two-dimensional gas chromatography-mass spectrometry. J Chromatogr A. 2014;1359:271–6. https://doi.org/10.1016/j.chroma.2014.07.054.
Virtanen P, Gommers R, Oliphant TE, Haberland M, Reddy T, Cournapeau D, et al. SciPy 1.0: fundamental algorithms for scientific computing in Python. Nat Methods. 2020;17:261–72. https://doi.org/10.1038/s41592-019-0686-2.
Funding
The submitted work was supported by Oakland University. No external funds or grants were received for conducting this study.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Published in the topical collection Comprehensive 2D Chromatography with guest editors Peter Q. Tranchida and Luigi Mondello.
Supplementary information
Below is the link to the electronic supplementary material.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Seeley, J.V., Schimmel, N.E. & Seeley, S.K. Influence of modulator injection width on comprehensive two-dimensional gas chromatography peak dimensions. Anal Bioanal Chem 415, 2399–2409 (2023). https://doi.org/10.1007/s00216-022-04387-8
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-022-04387-8