Abstract
Lipidomics analysis for large-scale studies aiming at the identification and quantification of natural lipidomes is often performed using LC–MS-based data acquisition. However, the choice of suitable LC–MS method for accurate lipid quantification remains a matter of debate. Here, we performed the systematic comparison between two HRAM-MS-based quantification workflows based on HILIC and RPLC MS by quantifying 191 lipids from five lipid classes in human blood plasma using deuterated standards in the “one ISTD-per-lipid class” approach. Lipid quantification was performed considering all necessary isotopic corrections, and obtained correction factors are illustrated. Concentrations of lipids in NIST® SRM® 1950 human blood plasma determined by the two methods were comparable for most of the studied lipid species except for highly unsaturated phosphatidylcholines (PC). A comparison of lipid concentrations to consensus values determined in a previously published multi-laboratory study illustrated possible “overestimation” of concentrations for these highly unsaturated lipids by HILIC MS. We evaluated the influence of lipid loading amounts as well as the difference between quantified lipid and internal standard concentrations on the HILIC MS quantification results. We conclude that both HILIC and RPLC HRAM-MS workflows can be equally used for accurate lysophosphatidylcholine (LPC), lysophosphatidylethanolamine (LPE), phosphatidylcholine (PC), phosphatidylethanolamine (PE), and sphingomyelin (SM) lipid quantification, despite significant differences in the concentration of highly unsaturated PC lipids which need to be addressed by establishing response factors to account for the differences in degree of lipid unsaturation.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Lipids play crucial roles in a plethora of physiological functions ranging from energy storage, cellular compartmentalization, regulation of protein function to signaling [1]. In order to unravel lipid function, it is of utmost importance to identify and quantify single lipid molecular species in complex biological mixtures. Lipidomics studies based on the application of liquid chromatography (LC) coupled online to mass spectrometry (MS) are often used to identify and quantify lipids in a complex biological sample. The lipidomics community nowadays underlines the importance of reporting quantitative values for lipid species as the only way to facilitate inter-laboratory and inter-study comparison of results at least for most often studied lipidomes such as human blood plasma [2].
The clear need for protocol standardization to ensure quality of reported lipid concentrations is significantly challenged by the fact that there are more than one MS-based method used for lipid analysis. Several quantification approaches have been developed implementing MS without (shotgun) or with prior chromatographic separation (LC–MS) in which MS quantification is based on adding compound-specific internal standards (ISTD) [3,4,5]. Quantitative mass spectrometry data may be acquired in a targeted or untargeted fashion whereas each acquisition method has certain advantages and disadvantages. Targeted methods (e.g., single ion monitoring (SIM), multiple/parallel reaction monitoring (MRM/PRM)) need prior knowledge of the sample matrix and specific LC and MS properties of the analytes accompanied by a limited number of quantification targets [6,7,8]. Untargeted data acquisition modes (e.g., full-MS, data-dependent (DDA-MS), and data-independent acquisition MS (DIA-MS)) [9,10,11] allow for the quantification of theoretically all analytes present in a sample and do not require prior knowledge of the sample matrix; thus, they are well suited for Discovery Lipidomics approaches. Furthermore, feature quantification may be performed on intact precursor ions enabling excellent chromatographic peak integration and sensitivity at the cost of decreased specificity (Full-MS, SIM) or on analyte-specific fragment ions increasing specificity (SWATH-MS, PRM/MRM) but complicating quantification due to a limited number of data points over chromatographic peak.
Accuracy of quantification also relies on the close similarity between physicochemical properties of added ISTD and the native lipid that is ought to be quantified. For that purpose, a chemically identical, isotopically labeled ISTD can be added to allow for the absolute quantification of a single lipid molecular species. It is experimentally evident that up to several thousand lipids can be present in a biological sample, and it is therefore not possible so far to obtain an ISTD for each single molecular lipid species. Usually, non-native lipid class-specific ISTDs (isotopically labeled, odd-numbered, or the one absent in the sample) [12] are added in a similar concentration as the lipids to be quantified which allow for accurate quantification in the so-called one ISTD-per-lipid class approach. However, it should be considered that native lipids and added ISTD are structurally not identical, and thus, several correction algorithms have to be implemented during the data processing workflow to make up for those structural differences [13,14,15]. Additionally, other factors are influencing the analytical response in MS experiments and therefore the analytical accuracy: (i) chromatographic matrix effects originating from different elution times; (ii) molecular species-dependent ionizability, surface activity, and adducts formation during the ESI process; and (iii) molecular species-dependent in-source fragmentation [13].
In LC–MS-based lipidomics, the two most widely applied separation modes are reversed-phase liquid chromatography (RPLC) and hydrophilic interaction chromatography (HILIC) [16] which have fundamentally different separation modes. RPLC is separating lipids mainly based on their fatty acyl/alkyl chain hydrophobicity whereas HILIC is separating lipids based on their headgroup polarity leading to coelution of all lipids of a specific lipid class [17, 18]. MS response of a lipid is mainly determined by its lipid class, the fatty acid chain length, and unsaturation as well as the solvent in which it is ionized [19, 20]. Therefore, HILIC-based MS quantification is capable of diminishing elution-dependent matrix effects due to coelution of the ISTD with the lipids of the same class. On the other hand, due to the coelution of all species within a defined lipid class, possible ionization suppression can occur favoring detection of high abundant lipid species over the low abundant ones. RPLC-based MS methods allow for quantification of isomeric lipids of the same class. However, since RPLC-separated lipids and the corresponding ISTD are distributed over a broad retention time range with quite different solvent compositions, increased matrix effects might decrease the accuracy of quantification results. Surprisingly, yet there has been no direct comparison of HILIC- and RPLC-based MS methods for their accuracy to quantifying lipids. To close this gap, we compare quantitative values for 191 lipids from five different lipid classes (LPC, LPE, PC, PE, and SM) in NIST® SRM® 1950 Metabolites in Frozen Plasma determined by RPLC and HILIC MS quantitative workflows.
Materials and methods
Chemicals
Acetonitrile (MeCN), 2-propanol (i-PrOH), methanol (MeOH), and formic acid (all ULC/MS-CC/SFC grade) were purchased from Biosolve (Valkenswaard, Netherlands). Chloroform Emsure®, tert-butyl methyl ether (MTBE) (≥ 99%), butylated hydroxytoluene (≥ 99%), and the NIST® SRM® 1950 Metabolites in Frozen Plasma were purchased from Sigma-Aldrich (Taufkirchen, Germany). Ammonium formate (NH4HCO2) and ammonium acetate (NH4OAc) (both MS grade) were purchased from Fluka Analytical (München, Germany). SPLASH® LIPIDOMIX® was purchased from Avanti Polar Lipids Inc. (Alabaster, AL, USA). Water was ultrapurified by an ELGA PURELAB Ultra Analytic (Berlin, Germany) instrument delivering water quality ≥ 18.2 MΩ-cm.
Preanalytics
NIST plasma was delivered frozen in 1-mL tubes and stored at − 80 °C until further processing. For lipid extraction, NIST plasma was thawed on ice for 1 h with subsequent vortexing. Plasma was aliquoted in 50-μL aliquots and stored at − 80 °C.
Lipid extraction
Five aliquots of NIST plasma were thawed by incubating tubes containing 50 μL plasma on ice for 1 h. SPLASH® LIPIDOMIX® was added (ratio 1:10, SPLASH:plasma, v/v) and incubated on ice for 15 min. Lipids were extracted as described before [14] by adding ice-cold MeOH (375 μL) and MTBE (1250 μL) with subsequent vortexing. Homogenates were incubated for 1 h at 4 °C (orbital shaker, 32 rpm). Phase separation was induced by addition of H2O (375 μL), vortexed and incubated for 10 min at 4 °C (orbital shaker, 32 rpm). Afterwards, the sample was centrifuged to separate the organic and aqueous phases (10 min, 4 °C, 1000×g). The organic phase was collected into a new tube. Re-extraction of the remaining aqueous phase was performed by addition of MTBE/MeOH/H2O (4/1.2/1, v/v; 500 μL) with subsequent vortexing. The samples were centrifuged (10 min, 4 °C, 1000×g), organic phases were combined, and the solvent was removed in vacuo (Eppendorf concentrator 5301, 1 ppm). A quality control (QC) sample was prepared by mixing obtained lipid extracts in an equivolumetric manner and was used for method development and monitoring of analytical accuracy.
Liquid chromatography
Subsequently, the lipid extract was reconstituted in pure i-PrOH (200 μL) by vigorous vortexing. Lipids were diluted with i-PrOH to a final concentration of 0.03 μLplasma/μLi-PrOH of which 5 μL was injected onto the column corresponding to 0.15 μLplasma.
Reversed-phase liquid chromatography (RPLC) was carried out on a Vanquish focused+ (Thermo Fisher Scientific, Bremen, Germany) equipped with an Accucore C18 column (150 × 2.1 mm; 2.6 μm, 150 Å; Thermo Fisher Scientific, Bremen, Germany). Lipids were separated by gradient elution with solvent A (MeCN/H2O, 1:1, v/v) and B (i-PrOH/MeCN/H2O, 85:10:5, v/v), both containing 5 mM NH4HCO2 and 0.1% (v/v) formic acid. Separation was performed at 50 °C with a flow rate of 0.3 mL/min using the following gradient: 0–20 min—10 to 86% B (curve 4), 20–22 min—86 to 95% B (curve 5), 22–26 min—95% isocratic, and 26–26.1 min—95 to 10% B (curve 5) followed by 5 min re-equilibration at 10% B.
Hydrophilic interaction chromatography (HILIC) was carried out on a Vanquish focused+ (Thermo Fisher Scientific) equipped with an Acquity UPLC BEH HILIC Si column (100 × 1.0 mm, 1.7 μm, 130 Å; Waters Corp.). Lipids were separated as described previously [3] by gradient elution with solvent A (MeCN/H2O, 96:4, v/v) and B (H2O) both containing 7 mM NH4OAc. Separation was performed at 40 °C with a flow rate of 0.15 mL/min using the following gradient: 0–10 min—0 to 10% B (curve 5) and 10–10.1 min—10 to 0% B (curve 5) followed by 5 min re-equilibration at 0% B.
Mass spectrometry
Both UHPLC experiments were performed using a Q Exactive Plus Hybrid Quadrupole-Orbitrap mass spectrometer (Thermo Fisher Scientific, Bremen, Germany) equipped with a HESI probe. Mass spectra were acquired in the mass range of m/z 100–1500 in positive and negative modes with the following ESI parameters: sheath gas—40 L/min, auxiliary gas—10 L/min, sweep gas—1 L/min, spray voltage—2.5 kV, spray current—10 μA, capillary temperature—300 °C, S-lens RF level—35, and aux gas heater temperature—370 °C.
For lipid identification, parallel reaction monitoring (PRM) using m/z of 191 previously reported blood plasma lipids (LPC, PC, LPE, PE, and SM) as precursors was used in negative mode for phospholipids or positive mode for SM at the resolution of 17,500 at m/z 200, AGC target of 2e5, and a maximum injection time of 40 ms. The isolation window for precursor selection was 1.2 m/z, and normalized stepped collision energy (10, 20, and 30 eV) was used for HCD. Data were acquired in profile mode.
For quantification, data was acquired in full MS mode only in positive polarity at the resolution of 140,000 at m/z 200, AGC target of 1e6, and maximum injection time of 100 ms in the mass range from m/z 100 to 1500. Data were acquired in profile mode.
Data processing
Lipids were identified based on MS2 fragmentation patterns. A lipid identification was accepted when the mass difference between expected and measured m/z was below 5 ppm and characteristic fragments for specific lipids could be detected.
For quantification, raw data sets of full MS measurements were processed using TraceFinder™ 4.1 (Thermo Fisher Scientific, Bremen, Germany). For all investigated lipid classes, only the monoisotopic mass peaks of protonated lipid adducts [M + H]+ were detectable and area under the curve (AUC; also referred to as peak integral or an extracted ion chromatogram) was determined using the following settings: mass tolerance—5 ppm, area noise factor—5, peak noise factor—10, baseline window—150, and S/N ≥ 3 using ICIS detection algorithm.
A detailed description of the quantitative workflow is presented in Fig. S1 in the Electronic Supplementary Material (ESM). Type I isotopic correction for 13C-abundance was performed with an excel macro as described previously [3] using the following correlation [15]:
AUCn(k) total = total ion area under curve, AUCn(k) = quantified area under curve of monoisotopic mass, n = no. of C atoms, k = no. of double bonds
Type II isotopic correction for overlapping, coeluting isotopologues was performed as previously described using the following equation [15]:
AUCn(k) total = total ion area under curve, AUCn(k) = quantified area under curve of monoisotopic mass, n = no. of C atoms, k = no. of double bonds
Accurate quantification of lipid species was performed by relating the corrected AUC values of the lipid species to the corrected AUC of lipid class ISTD.
Clipid = concentration of lipid species, CISTD = concentration of ISTD, AUClipid = corrected area under curve for lipid species, AUCISTD = area under curve for ISTD
The obtained lipid concentrations were compared to the established consensus values using LipidQC [21] software as previously described.
Quantitative reliability of obtained lipid concentrations was assessed by determination of the relative standard deviation (RSD) in between replicates. A lipid concentration was chosen to be reliably quantified if RSD was ≤ 20% in accordance with the definition of the upper and lower limits of quantification provided by the FDA [22].
Results and discussion
Evaluation of quantitative lipidomic workflow
The aim of this work was to compare HILIC and RPLC-MS workflows in their ability to quantify lipids from biological matrices and to assess systematic differences between those methods arising during the quantitation. Therefore, the NIST® SRM® 1950 Metabolites in Frozen Plasma was chosen as the reference matrix since consensus values for 339 different lipids were already established and validated by the community [2]. The SPLASH® LIPIDOMIX® was used as a ready, commercially available mixture of lipid class-specific internal standards (ISTD) as it was designed to match lipid concentrations in human plasma. Lipid extracts were separated by either HILIC or RPLC and measured under identical high-resolution accurate-mass (HRAM) mass spectrometry settings with a standardized data processing workflow.
Here, we focused on quantification of five lipid classes including lysophosphatidylcholines (LPC), lysophosphatidylethanolamines (LPE), phosphatidylcholines (PC), phosphatidylethanolamines (PE), and sphingomyelins (SM) as the most abundant lipids for which consensus values have been established and that show good chromatographic retention by both, RPLC and HILIC MS methods, used in our study (ESM Table S1) [23, 24]. Lipid identities were confirmed in the PRM experiment, and corresponding retention times were determined for lipids with reported consensus values [2]. Unambiguous identification was based on accurate mass and the characteristic fragmentation behavior upon HCD as reported previously [25]. Subsequently, lipid extracts were separated by both methods using full-MS scans at the resolution of 140,000 (at m/z 200) to ensure MS separation of isobaric lipid species. Quantification workflow was subdivided into several steps:
(1). Peak area integration of extracted ion chromatograms of the monoisotopic precursor
The quantitative data processing workflow starts with generating extracted ion chromatograms (XIC) for the monoisotopic mass ([M]) of identified lipids with a mass accuracy of 5 ppm within a determined retention time window. Monoisotopic XIC area integration can be routinely performed by different software tools, and here, TraceFinder v4.1 was used. Lipid concentrations could be potentially determined by relating the peak area of the native lipid to the peak area of the corresponding class-specific ISTD. However, several important correction factors have to be implemented to increase the accuracy of lipid quantification as illustrated below (ESM Table S2).
(2). Type I isotopic correction
The intensity of [M] relative to the total ion intensity (summed intensity of all isotopologues [M], [M + 1], [M + 2]) differs depending on the number of carbon atoms in the molecule due to isotopic effects caused by the natural 13C abundance. Since ISTD and native lipids are not identical on the molecular level (different number of carbon atoms), using area under the curve (AUC) for only [M] will impact accuracy of the quantification as it does not reflect the total ion abundance. Consequently, the total ion abundance of both the ISTD and the native lipid has to be taken into account to ensure accurate quantification.
Type I isotopic correction can be performed by extrapolating the peak area intensities for the corresponding isotopologues based on the intensity of the monoisotopic mass, the natural abundance of 13C, and the number of carbons in the molecule [15]. Thus, the magnitude of the correction factor will depend on the number of carbon atoms and will increase with the size of the lipid (Fig. 1, red trace). For instance, the type I isotopic correction factor for PC lipids results in differences relative to uncorrected values ranging from 39.9% for PC(30:0) with 38 C atoms to 51.3% for PC(42:6) with 50 C atoms (ESM Table S2).
(3). Type II isotopic correction
Furthermore, accurate lipid quantification is also affected by the overlap of the monoisotopic signal of the quantified lipid and the second isotopologue ([M + 2]) of a lipid with one more double bond. Those signals cannot be distinguished at a resolution of 140,000 at m/z 200 or lower. Therefore, type II isotopic correction has to be applied. First, the peak area of the monoisotopic signal of the lipid of interest (AUCn(k)) and the lipid with one more double bond (AUCn(k + 1)) are determined. Next, the AUC for the second isotopologue ([M + 2]) of the lipid with one more double bond is calculated and subtracted from AUCn(k). The type II isotopic correction is only necessary when lipid n(k) and lipid n(k + 1) cannot be chromatographically resolved which is usually the case for HILIC separations but not RPLC. Moreover, the type II correction factor will be strongly influenced by the relative concentrations of coeluting lipids that differed by one double bond (Fig. 1, blue trace), resulting in the higher percent of the difference relative to the uncorrected values for lipid pairs in which the signal intensity of lipid n(k + 1) is much higher than that of lipid n(k) (Fig. 1, blue trace, ESM Table S2). Type II corrected values are further subjected to type I isotopic correction discussed above. Therefore, AUC values for monoisotopic signal acquired in HILIC mode were corrected with type II and type I isotopic corrections whereas those acquired in RPLC mode were corrected only with type I (Fig. 1, black trace).
(4). Correction for all-ion abundance of deuterated standards
The SPLASH® LIPIDOMIX® contains a non-naturally occurring, deuterated representative for each main lipid class in blood plasma and can therefore be used for quantification of those classes. Determination of the total ion abundance of those deuterated lipids is complicated by the incomplete deuteration of the ISTD which has to be considered in the data processing workflow. Usually, only the integrated peak area of the monoisotopic mass [M] of a certain lipid ion is used because it can be determined with the highest sensitivity and accuracy. Isotopologue intensities are then mathematically extrapolated and summed up to determine total ion abundance (similar to type I isotopic correction described above). However, isotopically labeled molecules can have incomplete isotopic enrichment and purity; therefore, non-negligible ion intensities are found also for [M−1] and [M−2] depending on the number of introduced deuterium atoms and the compounds’ purity. Each of those isotopologues itself has a carbon-dependent 13C abundance profile which makes the calculation of all isotopologues using just the monoisotopic mass rather complicated. Additionally, for those calculations, the isotopic enrichment and purity of each ISTD have to be quantified accurately and can vary in between batches. Due to those disadvantages, the total ion abundance is determined by integrating and summing up peak areas of each isotopologue of the deuterated ISTD (ESM Fig. S1), i.e., [M−2], [M−2], [M], [M + 1], [M + 2], and [M + 3].
Eventually, corrected abundances of the native species are multiplied with the concentration of ISTD and divided by the total ion abundance of the ISTD to get the concentration of the native lipid.
Type I isotopic correction depends only on the total carbon number nC and, if not performed, can introduce a quantification error of 26% in the case of lipids with nC = 20 up to 51% for nC = 50. Type II isotopic correction depends on the presence of a lipid signal with one more double bond relative the quantified lipid as well as the signal intensities for both of those signals. For instance, in the case of the 88 PC lipids quantified here with the HILIC MS method, 64 required type II correction. The impact of lipid relative concentrations on type II correction can be exemplified on the PC(38:x) series. For instance, PC(38:6) (76.2 nmol/mL) was corrected for the second isotopologue of PC(38:7) (62.6 nmol/mL) leading to the decrease of the PC(38:6) concentration by only ≈ 10% since both of the lipids were present at similar concentrations. Whereas type II correction for PC(38:3) (15.4 nmol/mL) using the isotopologue of PC(38:4) (99.6 nmol/mL) changed the obtained concentration value by ≈ 57%. Thus, we would like to highlight one more time the importance of correction factors to ensure high accuracy of the quantification results.
Comparison of lipid quantification using RPLC and HILIC MS
HILIC and RPLC are complementary separation methods [17]. HILIC is separating lipids based on their headgroup polarity leading to coelution of all lipid molecular species of a corresponding subclass and its ISTD. RPLC is separating lipids mainly based on hydrophobicity of their fatty acid chains leading to separation of lipid molecular species over a relatively broad retention time range. Compared to HILIC-based separations, RPLC is capable of separating isomeric lipids within one class. However, here, in order to allow for direct comparison of both methods, the quantities of isomeric lipids from RPLC were summed up.
Only lipids that show good chromatographic retention for both methods were compared. HILIC is well suited for the separation of polar lipids whereas for unpolar lipids such as glycerolipids (GL), ceramides (Cer), and cholesteryl esters (CE), no retention takes place and those lipids elute in the void volume. Furthermore, acidic lipid classes such as phosphatidylserines (PS), phosphatidic acids (PA), and phosphatidylinositols (PI) show a bad retention behavior in HILIC separations due to the presence of several different ionization states in solution [23]. RPLC is well suited for the separation of GL, phospholipids (PL), sphingolipids (SL), and CEs. Therefore, in this study, we compared the capability of HILIC and RPLC to quantify five lipid classes including lysophosphatidylcholines (LPC; 23 species), lysophosphatidylethanolamines (LPE; 7 species), phosphatidylcholines (PC; 88 species), phosphatidylethanolamines (PE; 37 species), and sphingomyelins (SM; 36 species) as the most abundant lipids well separated both by RPLC and HILIC MS methods used in our study (ESM Table S1).
Out of 204 lipids from five lipid classes considered in this study (LPC, LPE, PC, PE, and SM) with established consensus values, we were able to quantify 184 and 171 lipids by HILIC and RPLC MS workflows, respectively. Lipid concentrations determined using all correction factors mentioned above were compared to the reported consensus values using LipidQC tool [21]. For 36 SM lipids, a good correlation between concentrations determined by HILIC and RPLC MS workflows was observed for nearly all molecular species independent of molecular mass, fatty acid composition, and concentration (Fig. 2). The quantities obtained by both methods are largely consistent with each other with only two lipid molecular species that vary by more than 50% in between both methods. Moreover, SM concentrations measured both by HILIC and RPLC MS were consistent with the consensus values for nearly all molecular species. A similar trend was observed for 7 LPE species quantified in the study (ESM Fig. S2).
Concentrations obtained by RPLC MS for 88 PC and 23 LPC lipids were in a good agreement with the consensus values for the majority of lipid species (Fig. 3 and ESM Fig. S3). However, quantities determined by HILIC MS showed that some lipid species were vastly overestimated in comparison to the consensus values (marked by k-eq ≥ 10). By plotting lipid quantities determined by HILIC vs RPLC MS, one can see that almost all “overestimated” lipids were characterized by a high unsaturation degree (n(DB) ≥ 4). A similar trend but to a smaller extent was observed for PE lipids as well (ESM Fig. S4).
Effect of retention time distribution between lipid and corresponding standard on the quantification results
Quantification by HILIC MS-derived methods has been stated to provide superior results relative to RPLC MS due to close elution of ISTDs and corresponding lipid molecular species, thus diminishing differential matrix effects [12], whereas in RPLC, lipids of the same class are separated over a broad retention time range exposing ISTD and lipid molecular species to different matrix effects. We therefore addressed the influence of retention time differences between the ISTD and the lipids to be quantified in HILIC and RPLC MS. Three SM species were chosen to display the effect of earlier, later, or coelution with the ISTD in RPLC (Fig. 4).
Separation by HILIC and RPLC is exemplified for the lipids SM(d32:1), SM(d34:1), SM(d38:1), and the corresponding ISTD SM(d36:2)-d9. In RPLC, SM(d32:1) is eluting before (ΔtR = 1.8 min) and SM(d38:1) is eluting after (ΔtR = 1.6 min) ISTD SM(d36:2)-d9 whereas in HILIC, close elution with a maximum retention time difference of 0.3 min to the ISTD can be observed. Quantification of those signals with all applied correction factors yields similar concentrations between HILIC and RPLC well suited within the 95% confidence interval calculated by LipidQC. Furthermore, those results were compared to SM(d34:1) that is coeluting with the ISTD in both applied separation modes. Slightly higher concentrations are determined in the case of RPLC but as shown previously (Figs. 2 and 3) remains within the 95% confidence interval. In conclusion, coelution of ISTD and lipids is not obligatory to obtain accurate quantification for lipid species addressed in this study.
Effect of lipid concentrations on the HILIC MS quantification results
Previous direct injection experiments on equimolar mixtures of lipids with different degrees of unsaturation and fatty acyl chain lengths demonstrated that lipids with a higher number of double bonds as well as lipids with decreasing chain length show higher intensities [26, 27]. This effect was dependent on the total lipid concentration during ionization and could be diminished by diluting lipid mixtures prior to injection. The difference in the surface activity during ionization was proposed to be the main factor explaining higher signal intensities for unsaturated lipids. Additionally, it was proposed that double bonds in lipids could weaken the intermolecular interactions on the droplet surface [26]. Since all lipid species for a certain lipid class are coeluting in HILIC, it might be compared to direct injection experiments in which no separation prior to MS analysis is performed. Moreover, PC lipids for which “overestimated” concentrations were measured have the highest total plasma concentrations followed by SM, LPC, PE, and LPE lipid classes. For instance, total SM concentration in plasma is around three times lower than that of PC. Furthermore, all measured SM lipids had n(DB) ≤ 3, whereas PC lipid species contained up to eight double bonds.
Another point to be considered is the contribution of added ISTD lipids to electrospray saturation effects. However, here, the concentration of PC ISTD in plasma was ≈ 19 μM (corresponding to 3.2 pmol on the column) and thus was comparable to the medium abundant PC lipids. Thus, it is possible that electrospray saturation and its effect on MS response are unlikely to be attributed to PC ISTD when compared to abundant PC species (e.g., PC 34:2 with 240 μM/67.7-pmol on the column or PC 36:4 with 150 μM/27.3 pmol on the column) with potentially higher ESI-MS response due to increased surface activity determined by a higher number of double bonds [28,29,30].
Here, to test the effect of total PC concentration on the “overestimation” of unsaturated molecular species, lipid extract was quantified by HILIC MS using several consecutive dilutions with the total amounts of lipid extracts equivalent to 0.15, 0.1, 0.05, 0.01, and 0.005 μL of original human blood plasma.
Effect of different loading amounts on lipid quantification by HILIC MS exemplified here for two PC series (Fig. 5). For PC lipids with 38 carbon atoms, good fit to the consensus values as well as RPLC MS quantitative results was observed for species with nDB ≤ 4 independent of the dilution factor. However, concentrations of PC(38:5) until PC(38:7) were “overestimated” using the highest loading amount (0.15 μL of blood plasma). This effect was slightly reduced for PC(38:5) and PC(38:6), but was still evident for PC(38:7) (Fig. 5a). A similar trend can be observed for the series PC(40:x) where the effect of “overestimation” was reduced by using diluted lipid extract for PC(40:6) but remained unchanged for PC(40:7) and PC(40:8) (Fig. 5b). Furthermore, decreasing sample loading led to higher standard deviations due to the integration of signals of very low intensity.
Effect of different ratios of quantified lipids and internal standard
Diluting lipid extracts before HILIC analysis did not significantly improve the accuracy of quantification for highly unsaturated PC. We then evaluated if the difference in intensity between the quantified lipid and corresponding ISTD could explain the observed “overestimation.” Previously, using the direct injection approach, it was proposed that a certain lipid can be quantified only if its intensity lies within the range of 20–500% of the intensity of the corresponding ISTD signal [15].
Using corresponding deuterated PC lipid present in SPLASH® LIPIDOMIX®, we determined those ratios for all PC lipids. Intensities of blood plasma PC lipids were within the range from 1 to 600% of the ISTD (Fig. 6). However, no correlation between “overestimation” and intensity values relative to ISTD was observed. In fact, majority of highly unsaturated PC lipids for which substantial differences between HILIC and RPLC MS methods were measured were within the range of intensity ratios suitable for quantification according to the previous report [15].
It was previously reported that isobaric interference of the [M + Na]+ signal of a lipid with 3 less double bonds and 2 less carbons with the [M + H]+ adduct of the quantified lipid may lead to overestimation in quantitative results [31,32,33]. For example, the [M + H]+ adduct of PC 40:7 (m/z 832.5851) overlaps with the [M + Na]+ adduct of PC 38:4 (m/z 832.5827). Those signals to be accurately differentiated would need a mass accuracy below 3 ppm and a resolving power of ≈ 350,000 that even modern high-resolution accurate mass MS cannot routinely deliver. Those artifacts occur only if lipids with nDB ≥ 3 are present in the studied biological matrix which was the case for PC, LPC, and PE but not SM and LPE lipids in human plasma studied here and therefore could potentially explain the observed “overestimation” effect.
Assessing those isobaric interferences by HILIC MS was not possible at the analytical conditions used here (mass accuracy of 5 ppm and MS resolving power of 73,300 at m/z 832.5827), so we estimated the contribution of isobaric overlap by calculating the sum of [M + Na]+ and [M + H]+ of isobarically overlapping lipid signals in RPLC MS (ESM Table S3). For instance, concentrations determined for PC 40:7 [M + H]+ by RPLC MS and HILIC MS corresponded to 3.49 μM and 43.97 μM, respectively. To mimic the HILIC MS situation where signals of PC 40:7 [M + H]+ and PC 38:4 [M + Na]+ are indistinguishable and thus quantified together, those two signals, well separated by RPLC, were quantified and summed, providing the value of 17.59 μM, which could explain only 35% of “overestimated” values obtained by HILIC MS. Overall, those values varied between 5 and 94% for 14 analyzed PC lipids (ESM Table S3). However, it should be considered that it is rather rough approximation, since different solvent systems, different elution times, and isomer separation in RPLC could lead to a different relative intensity of the [M + Na]+ to [M + H]+ adducts for different lipid species.
Conclusion
The lipidomics community nowadays aims to establish guidelines “to develop common standards for minimum acceptable data quality and reporting for lipidomics data” with a focus on providing reliable and accurate quantitative readouts for lipid species in natural lipidomes [34]. To aid the question of selecting the optimal LC–MS method for lipid quantification, we performed the systematic comparison between two HRAM MS-based quantification workflows based on HILIC and RPLC MS by quantifying 191 lipids from five lipid classes (LPC, LPE, PC, PE, and SM) in human blood plasma using deuterated standards in the “one ISTD-per-lipid class” approach. Quantification workflow followed the recommendation of the Lipidomics Standard Initiative (https://lipidomics-standards-initiative.org/guidelines/lipid-species-quantification) by considering two types of isotopic corrections as well as correction for the isotopic purity of deuterated standards used for the quantification. We provide a detailed comparison of quantification results with and without each type of correction and highlight the necessity of utilizing these correction factors to ensure high accuracy of the results.
Lipid concentrations determined using HILIC and RPLC MS were compared to each other as well as to the consensus values established for the same lipid species in NIST SRM 1950 human blood in a previously published multi-laboratory study [2]. It has been stated that different separation modes (HILIC vs RPLC) have varying suitability for the quantification of lipids from biological matrices. HILIC is believed to be better suited for lipid quantification since native lipids coelute with corresponding ISTD, thus reducing matrix effects originating from different solvent compositions during ionization [12, 13]. Following this logic, RPLC is not suited for lipid quantification since ISTD usually elute in a quite different solvent composition and therefore suffer from increasing solvent-dependent matrix effects. We have shown that HILIC and RPLC, despite the obvious difference in matrix effects, yield similar values for PE, LPE, and SM consistent with the reported consensus values. Yet, quantification of (L)PC lipids showed higher quantities in the HILIC method compared to RPLC MS and consensus values, especially in the case of highly unsaturated PC lipids.
Neither reduced lipid loading nor the differences between intensities of the ISTD and quantified lipid signals, or isotopic overlap with [M + Na]+ adducts of lipids with 3 less double bonds and 2 less carbons could fully explain observed “overestimation.” As it has been stated before, the differences in lipid MS response are dependent on distinct molecular features such as unsaturation level, acyl/alkyl chain length, and bond types [19, 26, 27, 35,36,37] and thus can only be properly addressed by application of a response factor approach or the use of several different ISTD per class (i.e., with different chain lengths and number of double bounds) to make up for differential response. However, so far, almost no equimolar mixtures of isotopically labeled molecular species with different degrees of unsaturation and fatty acyl/alkyl chain length for a given specific lipid class are commercially available and thus generation of response factors is tedious. One possible solution is provided by Lipidyzer™ Platform kits including 50 ISTD covering 13 lipid classes.
Furthermore, comparison of the measured concentrations to the reported consensus values might be biased as there is no information available on which separation modes, mass analyzers, or ISTD have been used for establishing them. If it is the case that mostly RPLC MS was used for generation of the consensus values then those quantities have to be reevaluated by incorporating also HILIC-based quantification approaches.
Abbreviations
- AUC:
-
Area under the curve
- CE:
-
Cholesteryl ester
- Cer:
-
Ceramides
- GL:
-
Glycerolipids
- HILIC:
-
Hydrophilic interaction chromatography
- HRAM:
-
High-resolution accurate mass
- i-PrOH:
-
Isopropanol
- ISTD:
-
Internal standard
- LC-MS:
-
Liquid chromatography–mass spectrometry
- LPC:
-
Lysophosphatidylcholine
- LPE:
-
Lysophosphatidylethanolamine
- m/z :
-
Mass to charge ratio
- MeCN:
-
Acetonitrile
- MeOH:
-
Methanol
- MS:
-
Mass spectrometry
- MS2 :
-
Tandem mass spectrometry
- MTBE:
-
Methyl-tert-butyl ether
- NH4HCO2 :
-
Ammonium formate
- NH4OAc:
-
Ammonium acetate
- PA:
-
Phosphatidic acid
- PC:
-
Phosphatidylcholine
- PE:
-
Phosphatidylethanolamine
- PI:
-
Phosphatidylinositol
- PRM:
-
Parallel reaction monitoring
- PS:
-
Phosphatidylserine
- RPLC:
-
Reversed-phase liquid chromatography
- RSD:
-
Relative standard deviation
- SL:
-
Sphingolipids
- SM:
-
Sphingomyelin
References
Casares D, Escribá PV, Rosselló CA. Membrane lipid composition: effect on membrane and organelle structure, function and compartmentalization and therapeutic avenues. Int J Mol Sci. 2019;20:2167. https://doi.org/10.3390/ijms20092167.
Bowden JA, Heckert A, Ulmer CZ, Jones CM, Koelmel JP, Abdullah L, et al. Harmonizing lipidomics: NIST interlaboratory comparison exercise for lipidomics using SRM 1950–metabolites in frozen human plasma. J Lipid Res. 2017;58:2275–88. https://doi.org/10.1194/jlr.M079012.
Lísa M, Cífková E, Khalikova M, Ovčačíková M, Holčapek M. Lipidomic analysis of biological samples: comparison of liquid chromatography, supercritical fluid chromatography and direct infusion mass spectrometry methods. J Chromatogr A. 2017;1525:96–108. https://doi.org/10.1016/j.chroma.2017.10.022.
Sadowski T, Klose C, Gerl MJ, Wójcik-Maciejewicz A, Herzog R, Simons K, et al. Large-scale human skin lipidomics by quantitative, high-throughput shotgun mass spectrometry. Sci Rep. 2017;7:43761. https://doi.org/10.1038/srep43761.
Schweizer S, Liebisch G, Oeckl J, Hoering M, Seeliger C, Schiebel C, et al. The lipidome of primary murine white, brite, and brown adipocytes—impact of beta-adrenergic stimulation. PLoS Biol. 2019;17:e3000412. https://doi.org/10.1371/journal.pbio.3000412.
Ribbenstedt A, Ziarrusta H, Benskin JP. Development, characterization and comparisons of targeted and non-targeted metabolomics methods. PLoS One. 2018;13:e0207082. https://doi.org/10.1371/journal.pone.0207082.
Aretz I, Meierhofer D. Advantages and pitfalls of mass spectrometry based metabolome profiling in systems biology. Int J Mol Sci. 2016;17.
Zhou J, Liu H, Liu Y, Liu J, Zhao X, Yin Y. Development and evaluation of a parallel reaction monitoring strategy for large-scale targeted metabolomics quantification. Anal Chem. 2016;88:4478–86. https://doi.org/10.1021/acs.analchem.6b00355.
Raetz M, Duchoslav E, Bonner R, Hopfgartner G. Hybrid SWATH/MS and HR-SRM/MS acquisition for phospholipidomics using QUAL/QUANT data processing. Anal Bioanal Chem. 2019;411:5681–90. https://doi.org/10.1007/s00216-019-01946-4.
Triebl A, Burla B, Selvalatchmanan J, Oh J, Tan SH, Chan MY, et al. Shared reference materials harmonize lipidomics across MS-based detection platforms and laboratories. J Lipid Res. 2019:jlr.D119000393. https://doi.org/10.1194/jlr.D119000393.
Cajka T, Smilowitz JT, Fiehn O. Validating quantitative untargeted Lipidomics across nine liquid chromatography–high-resolution mass spectrometry platforms. Anal Chem. 2017;89:12360–8. https://doi.org/10.1021/acs.analchem.7b03404.
Holčapek M, Liebisch G, Ekroos K. Lipidomic analysis. Anal Chem. 2018;90:4249–57. https://doi.org/10.1021/acs.analchem.7b05395.
Burla B, Arita M, Arita M, Bendt AK, Cazenave-Gassiot A, Dennis EA, et al. MS-based lipidomics of human blood plasma: a community-initiated position paper to develop accepted guidelines. J Lipid Res. 2018;59:2001–17. https://doi.org/10.1194/jlr.S087163.
Matyash V, Liebisch G, Kurzchalia TV, Shevchenko A, Schwudke D. Lipid extraction by methyl-tert-butyl ether for high-throughput lipidomics. J Lipid Res. 2008;49:1137–46. https://doi.org/10.1194/jlr.D700041-JLR200.
Wang M, Wang C, Han X. Selection of internal standards for accurate quantification of complex lipid species in biological extracts by electrospray ionization mass spectrometry-what, how and why? Mass Spectrom Rev. 2017;36:693–714. https://doi.org/10.1002/mas.21492.
Cajka T, Fiehn O. Comprehensive analysis of lipids in biological systems by liquid chromatography-mass spectrometry. TrAC Trends Anal Chem. 2014;61:192–206. https://doi.org/10.1016/J.TRAC.2014.04.017.
Lange M, Ni Z, Criscuolo A, Fedorova M. Liquid chromatography techniques in lipidomics research. Chromatographia. 2019;82:77–100. https://doi.org/10.1007/s10337-018-3656-4.
Buré C, Ayciriex S, Testet E, Schmitter JM. A single run LC-MS/MS method for phospholipidomics. Anal Bioanal Chem. 2013;405:203–13. https://doi.org/10.1007/s00216-012-6466-9.
Höring M, Ejsing CS, Hermansson M, Liebisch G. Quantification of cholesterol and cholesteryl ester by direct flow injection high-resolution Fourier transform mass spectrometry utilizing species-specific response factors. Anal Chem. 2019;91:3459–66. https://doi.org/10.1021/acs.analchem.8b05013.
Weir JM, Wong G, Barlow CK, Greeve MA, Kowalczyk A, Almasy L, et al. Plasma lipid profiling in a large population-based cohort. J Lipid Res. 2013;54:2898–908. https://doi.org/10.1194/jlr.P035808.
Ulmer CZ, Ragland JM, Koelmel JP, Heckert A, Jones CM, Garrett TJ, et al. LipidQC: method validation tool for visual comparison to SRM 1950 using NIST interlaboratory comparison exercise lipid consensus mean estimate values. Anal Chem. 2017;89:13069–73. https://doi.org/10.1021/acs.analchem.7b04042.
Rogatsky E, Stein D, Toxicology F, Workplace F, Testing D, Toxicology PF, Fachi MM, Leonart LP, Cerqueira LB, et al. Bioanalytical Method Validation Guidance. J Chromatogr B Anal Technol Biomed Life Sci. 2017;1043:25 http://www.ich.org/fileadmin/Public_Web_Site/ICH_Products/Guidelines/Quality/Q2_R1/Step4/Q2_R1__Guideline.pdf.
Cífková E, Hájek R, Lísa M, Holčapek M. Hydrophilic interaction liquid chromatography-mass spectrometry of (lyso)phosphatidic acids, (lyso)phosphatidylserines and other lipid classes. J Chromatogr A. 2016;1439:65–73. https://doi.org/10.1016/j.chroma.2016.01.064.
Ovčačíková M, Lísa M, Cífková E, Holčapek M. Retention behavior of lipids in reversed-phase ultrahigh-performance liquid chromatography-electrospray ionization mass spectrometry. J Chromatogr A. 2016;1450:76–85. https://doi.org/10.1016/j.chroma.2016.04.082.
Pi J, Wu X, Feng Y. Fragmentation patterns of five types of phospholipids by ultra-high-performance liquid chromatography electrospray ionization quadrupole time-of-flight tandem mass spectrometry. Anal Methods. 2016;8:1319–32. https://doi.org/10.1039/C5AY00776C.
Koivusalo M, Haimi P, Heikinheimo L, Kostiainen R, Somerharju P. Quantitative determination of phospholipid compositions by ESI-MS: effects of acyl chain length, unsaturation, and lipid concentration on instrument response. J Lipid Res. 2001;42:663–72.
Hofmann T, Schmidt C. Instrument response of phosphatidylglycerol lipids with varying fatty acyl chain length in nano-ESI shotgun experiments. Chem Phys Lipids. 2019;223:104782. https://doi.org/10.1016/J.CHEMPHYSLIP.2019.05.007.
Wilm M. Principles of electrospray ionization. Mol Cell Proteomics. 2011;10:1–8. https://doi.org/10.1074/mcp.M111.009407.
Cech NB, Enke CG. Relating electrospray ionization response to nonpolar character of small peptides. Anal Chem. 2000;72:2717–23. https://doi.org/10.1021/ac9914869.
Schuett BS, Millar TJ. Lipid component contributions to the surface activity of meibomian lipids. Investig Ophthalmol Vis Sci. 2012;53:7208–19. https://doi.org/10.1167/iovs.12-10471.
Cífková E, Holčapek M, Lísa M. Nontargeted lipidomic characterization of porcine organs using hydrophilic interaction liquid chromatography and off-line two-dimensional liquid chromatography-electrospray ionization mass spectrometry. Lipids. 2013;48:915–28. https://doi.org/10.1007/s11745-013-3820-4.
Cífková E, Holčapek M, Lísa M, Vrána D, Gatěk J, Melichar B. Determination of lipidomic differences between human breast cancer and surrounding normal tissues using HILIC-HPLC/ESI-MS and multivariate data analysis. Anal Bioanal Chem. 2015;407:991–1002. https://doi.org/10.1007/s00216-014-8272-z.
Cífková E, Holčapek M, Lísa M, Ovčačíková M, Lyčka A, Lynen F, et al. Nontargeted quantitation of lipid classes using hydrophilic interaction liquid chromatography–electrospray ionization mass spectrometry with single internal standard and response factor approach. Anal Chem. 2012;84:10064–70. https://doi.org/10.1021/ac3024476.
(2019) Lipidomics needs more standardization. Nat Metab 1:745–747. https://doi.org/10.1038/s42255-019-0094-z
Holcapek M, Lísa M, Jandera P, Kabátová N. Quantitation of triacylglycerols in plant oils using HPLC with APCI-MS, evaporative light-scattering, and UV detection. J Sep Sci. 2005;28:1315–33. https://doi.org/10.1002/jssc.200500088.
Li M, Butka E, Wang X. Comprehensive quantification of triacylglycerols in soybean seeds by electrospray ionization mass spectrometry with multiple neutral loss scans. Sci Rep. 2014;4:1–11. https://doi.org/10.1038/srep06581.
Han X, Gross RW. Quantitative analysis and molecular species fingerprinting of triacylglyceride molecular species directly from lipid extracts of biological samples by electrospray ionization tandem mass spectrometry. Anal Biochem. 2001;295:88–100. https://doi.org/10.1006/abio.2001.5178.
Acknowledgments
Financial support from the German Federal Ministry of Education and Research (BMBF) within the framework of the e:Med research and funding concept for SysMedOS project (to MF) are gratefully acknowledged. We thank Prof. Ralf Hoffmann (Institute of Bioanalytical Chemistry, University of Leipzig) for providing access to his laboratory.
Funding
Open Access funding provided by Projekt DEAL.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interests
The authors declare that they have no conflict of interests.
Additional information
Topical collection Current Progress in Lipidomics with guest editors Michal Holčapek, Gerhard Liebisch, and Kim Ekroos.
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lange, M., Fedorova, M. Evaluation of lipid quantification accuracy using HILIC and RPLC MS on the example of NIST® SRM® 1950 metabolites in human plasma. Anal Bioanal Chem 412, 3573–3584 (2020). https://doi.org/10.1007/s00216-020-02576-x
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-020-02576-x