Introduction

In recent decades, as a rising star in the field of chromatography, two-dimensional liquid chromatography (2D-LC) has attracted more and more attention by rapid developments in research of complex systems, such as natural products, environmental pollutants, food and pharmacy, proteomics, and metabolomics [1]. The largest benefit of 2D-LC is the dramatic increase in peak capacity that may be achieved, which is reflected in the component overlapping reduction [2].

In online comprehensive 2D-LC (LCxLC), where the sample is completely and continuously transferred from the first dimension (1D) to the second dimension (2D), the combination of orthogonal dimensions is usually hard to achieve because incompatible mobile phases are often used in both dimensions. For example, one of the most orthogonal combinations that can be used in 2D-LC is the combination of RPLC and HILIC because these two separation mechanisms present completely different selectivity for the retention of the analytes. Besides, the mobile phases used in RPLC and HILIC are miscible and compatible with detectors such as MS. Therefore, a 2D-LC system based on the coupling of RPLC and HILIC (HILICxRPLC or RPLCxHILIC) should theoretically be a promising system [1]. However, the compatibility of the mobile phases in such two dimensions is always a problematic and an important issue. The main reason is that the weak solvent in the 1D consists of a strong eluent solvent for the 2D, which can lead to a problem of mobile phase mismatch and poor or no separation in the 2D.

In particular, the analysis of polar compounds by an RPxHILIC coupling has been substantially less employed than the HILICxRP combination in the literature so far. The difficulty arises from the high HILIC sensitivity to the injection solvent, fractions coming from the 1D-RP solved in a high aqueous concentration are hardly focused in the 2D-HILIC causing a breakthrough phenomenon due to the non-retention of the transfer compounds in the 2D column. This effect is even more negative when the injection volume exceeds the 9% of the column dead volume [3]. However, HILIC is able to provide fast analysis of compounds with a low loss of resolution [4] that can be a very interesting characteristic for the separation carried out in the 2D. So, the fast analysis provided by HILIC could be extrapolated to the analysis of polar compounds, which 2DLC analysis was mainly limited to the coupling of HILIC in the 1D and RP in the 2D up to now.

A big effort should be done in this kind of orthogonal coupling to avoid the mismatch solvent effect as far as possible. As a core part of a 2D-LC system, the modulator is responsible for the 1D effluent cutting, collecting, and transferring to the 2D column. The incompatibility of two dimensions will negatively affect the modulation process, resulting in analytes unfocussed on the head of the 2D column. The tradition standard modulation (TS) is the most common and simple way to transfer the fractions from the 1D to the 2D, and consists on the use of two identical injection loops installed in a two-position switching valve. These two loops continuously alternate their functions collecting the fractions from the 1D and injecting them in the 2D column. This modulation mechanism can be considered as a passive modulation since the fraction injected in the 2D column is going to depend only on the 1D flow rate and mobile phases. Due to this passive modulation, usually, the TS modulation produces a low focusing effect in the head of the 2D column, because it is not able to reduce the strength of the fraction solvent before reaching the 2D column.

Recently, several kinds of modulators have been developed to overcome the incompatibility problem [5]. Some of them are based on the modification of the modulator configuration like the Fixed Solvent Modulator (FSM) developed by Petersson et al. [6] or the Active Solvent Modulation (ASM) proposed by Stoll et al. [7, 8].

Besides the solvent incompatibility, LCxLC presents other limitations. For instance, due to the realistic limitation of the transferred volume from the 1D to the 2D, 1D column diameter and flow rate are usually low in LCxLC and a small fraction is collected in the modulator which will be diluted by the 2D high flow rate. Therefore, these limitations result in a hard detection of low abundant targets, i.e., low sensitivity. Based on the FSM modulation strategy, Chen’s group made a new development to improve the 2D detection sensitivity by replacing the sample loops with short C18 trapping columns to construct an RPLC×HILIC system [9], which is shown in Fig. S1 in the Electronic Supplementary Material (ESM). However, in all three described modulators, the fraction dilution factor depends on the length and inner diameter of the sample loop channel and the bypass channel or the length of the trapping and the splitting column. This fact involves precise control and optimization of the dilution factor during the analysis of a complex sample containing analytes with a wide range of polarity which can limit the method development.

As a further improvement, a novel modulator, called at-column dilution (ACD) modulator, was proposed in our previous work to achieve free adjustment of the online dilution factor [10]. The RPLCxHILIC configuration equipped with an ACD modulator is shown in Fig. 1. The ACD modulator was modified from the TS modulation (Fig. 1) by adding an independent transfer pump, which is responsible for the injection of the fractions into the 2D. In this modulation strategy, the 2D mobile phase is no longer flowing through the sample loops, but the additional transfer pump is connected to the valve, and therefore, it is the responsible for eluting the 1D fraction storage in the loop with a weaker solvent for the 2D (transfer flow rate). Afterward, the transfer flow and the 2D gradient flow meet and are mixed in a mixer before reaching the 2D column. Thereby, the high water content fraction from the 1D-RP is diluted with acetonitrile and gradually enters the 2D column, and the analytes can be focussed on the head of the 2D column. More importantly, the dilution factor could be controlled by adjusting the transfer flow and the 2D flow rates, since the transfer flow can be optimized regardless of the 1D and 2D conditions.

Fig. 1
figure 1

Configuration of the traditional standard (TS) modulator and at-column dilution (ACD) modulator. In this scheme, a 2-position 8-port switching valve is shown to simplify the visualization of the new configuration

Herbal medicines are naturally occurring plant-derived substances, which are used for medical purposes [11]. According to the World Health Organization (WHO), about 80% of the world’s population uses herbal medicines as its primary healthcare [12]. There is no doubt that herbal medicines have great medical potential and economical importance. However, the difficulty of obtaining a clear description of the complex chemical constituents limits the in-depth study of the efficacy principles and the quality control of these herbal medicines in practice. Therefore, the development of analytical methods and tools that allow the comprehensive analysis of the herbal medicines is highly demanded. However, due to the high complexity of the chemical constituents in herbal medicines, conventional one-dimensional chromatographic techniques usually do not provide enough separation power to comprehensively analyze the chemical profile of such complex plants. Therefore, there is a motivation to seek more advanced analytical tools with higher separation performance.

Buddleja davidii is a multi-stemmed shrub or small tree that is native to China and has been introduced nowadays as an ornamental plant worldwide, both in sub-oceanic climates and sub-Mediterranean zones. In Germany, Buddleja davidii, commonly known as the butterfly bush, has very wide distribution due to its small wind-dispersed seeds rapidly colonize and barely disturb the area. It also grows widely due to its potential invasive power and adaptability.

Similar to other species of Buddleja, Buddleja davidii has significant pharmaceutical activities related to its metabolites. For example, terpenes (ESM Fig. S2a), phenylethanoids as verbascoside, shown in ESM Fig. S2b, which is known for its antioxidant, anti-inflammatory and photoprotective actions [13], and phenylpropanoid glycosides (ESM Fig. S2c), with antioxidant activities, also isolated and identified in this plant [14,15,16]. A flavone glycoside called linarin, shown in ESM Fig. S2d, was also isolated, and this compound revealed analgesic, antipyretic, anti-inflammatory, and neuroprotective activities [17]. Furthermore, a systematic survey of the genus Buddleja has indicated that some isolated compounds have anti-inflammatory and inhibitory activity against eicosanoid synthesis by inhibiting cyclo-oxygenase (COX) or 5-lipoxygenase (5-LOX) [15, 18]. Therefore, a comprehensive study about its chemical constituents is highly required due to its rich and diverse potential medicinal value.

In this study, the applicability of ACD modulator is tested for the coupling of RPLCxHILIC-MS system. For this propose different parts of the herbal medicine Buddleja davidii were used as an example for the method development. As a starting point, careful optimization of 1D and 2D gradient separation was carried out for the root extract. Then, with the optimized 2D-LC separation conditions, four parts of Buddleja davidii were also analyzed. Finally, the 2D-LC-UV method was coupled to MS, and MS/MS experiments were carried out to identify the complex profile of the root sample.

Experimental

Chemicals

Acetonitrile (ACN, MS grade) was purchased from VWR International (USA). Formic acid (HPLC/SPECTRO grade) was from Merck KGaA (Germany) and ethanol (Analytical reagent grade) was purchased from Fisher Chemical (Fair Lawn, New Jersey, USA). Distilled water was filtered through a Sartorius Stedim Biotech system (Germany).

Sample collection and preparation

The whole Buddleja davidii plant was collected from the field around the University of Duisburg-Essen, Campus Essen, Germany. The whole plant was divided into five parts: roots, stems, leaves, flowers, and fruits. The roots were washed with distilled water and dried with a dryer at 45 °C during 12 h. The other four parts were directly dried in a dryer at 45 °C. Then, they were smashed into powder and preserved in brown glass bottles at room temperature. 5.0 g of each sample were extracted with 50 mL 60% ethanol for 1 h at room temperature in glass flasks, and then the samples were extracted in an ultrasonic bath (VWR, USA) for another 1 h. After that, the samples were centrifuged (5840R Eppendorf, Germany) at 4000 rpm/min for 15 min at room temperature. The supernatant of the extract was moved to an ultracentrifuge (Spin plus from Eppendorf, Germany) and centrifuged at 14000 rpm/min for 10 min at room temperature. Afterward, 1 mL volume of the supernatants was diluted with 4 mL distilled water and the sample solution was filtered through a 0.2-μm filter prior to 2D-LC analysis.

Two-dimensional liquid chromatography (2D-LC) system

An Agilent 1260 Infinity (1D) combined with Agilent 1290 Infinity (2D) was utilized as a 2D liquid chromatography system (Agilent Technologies, USA) in this study. In particular, the 1D was built by a 1260 Infinity Flexible Cube solvent management module, a 1260 Infinity Degasser, a 1260 Infinity Capillary Pump, a 1260 Infinity Micro Autosampler, a 1260 Infinity Thermostat Column Compartment, a 1260 Infinity Diode Array Detector. The 2D consisted of a 1290 Infinity II Binary Pump and a 1290 Infinity II Diode Array Detector. The interface or modulator consisted of a 1290 Infinity Valve Drive and a 2-position/4-port duo-valve equipped with two 80 μL loops. The third pump, used for the ACD modulation was a 1290 Infinity binary pump. The 2DLC system was controlled by the OpenLab CDS software. The data of the diode array detector were collected during the running of the system and then imported to LC Image 2.7r LC × LC (64-bit) to generate 2D contour plots. The detection wavelength was set at 280 nm during the whole separation process.

The configuration of the RPLCxHILIC-MS system with the ACD modulator is shown on the right side of Fig. 1. At first, the mobile phase from the 1D pump transferred the injected sample (2 μL) to the RP-1D column. The 1D effluent was alternatively stored in the sampling-loops of the electronically controlled 2-positon/4-port duo-valve, which hyphenated the 1D column and the 2D column. The switching time of the valve was set at 2 min (modulation time). Once the sample loop was filled with 1D effluent, the valve changed the position and the fraction was connected in the line of 2D flow, and then the collected fraction was pushed out of the sample loop by the transfer flow from the additional pump (transfer pump). At downstream of the modulator, the transfer flow—with a relative low flow rate—and the 2D eluent—with a high flow rate—met and mixed in a mixer and were transferred into the 2D column. Finally, after the detection in the 2D UV-detector, the 2D effluent was split to enter into the MS. In this work, the optimized transfer flow rate was 0.2 mL/min and the 2D flow rate was established at 2.4 of mL/min, which lead to a total 2D flow rate of 2.6 mL/min and a dilution factor of 13 (total flow rate/transfer flow rate).

1D RP separation was performed at room temperature on a C18 column (150 × 1.0 mm) with 2.6 μm core-shell particles (Phenomenex, CA, USA). The mobile phase contained water (A) and acetonitrile with 0.1% formic acid (B). The optimized gradient time was 70 min at a flow rate of 15 μL/min. The gradient consisted of: 0 min, 10%B; 3 min, 20%B; 20 min, 20%B; 36 min, 45%, 50 min, 100% B; 64 min, 100% B; 65 min, 10% B (Fig. 2).

Fig. 2
figure 2

Optimized 2D-LC gradient for 1D and 2D separation

2D HILIC separation was performed at room temperature on a Cys-HILIC column (150 × 3.0 mm) packed with 5 μm particles (Acchrom, Inc., Beijing, China). The mobile phase contained water with 0.1% formic acid (A) and acetonitrile (B). The modulation time was 2 min, which leads to 35 cycles presented in the 2D separation. As shown in Fig. 2, the lowest point of the 2D shifting gradient in each cycle increases in 70 min from 65 to 75% of acetonitrile (B). The gradient was established as follow: 0 min, 100% B; 0.1 min, 0.1% B; 0.2 min, 95% B; 1.5 min, 75% B (at the beginning) or 65% B (at the end); 1.6 min, 75% B (at the beginning) or 65% B (at the end), 1.7 min, 100% B; 100% was maintained until 2 min for the equilibration of the column before the start of the next measurement.

Mass spectrometry

Mass spectrometry was performed on an Agilent 6545 Q-TOF LC/MS (Agilent Technologies, USA) using an electrospray ionization source (ESI) operating in negative ion mode. A flow of 0.4 mL/min was introduced into the ion source of the MS by splitting the 2D flow rate with a T-valve. Nitrogen was utilized as sheath gas and drying gas. The sheath gas and the drying gas were set to 11 L/min at 300 °C and 10 L/min at 350 °C, respectively. The nebulizer was set at 35 psig and the capillary voltage and nozzle voltage were set at 3500 V and 1000 V, respectively. The fragmentor, skimmer, and Oct 1 RF Vpp were set at 220 V, 65 V, and 750 V, respectively. The acquisition mode was performed on full MS and auto MS/MS. The acquisition was performed in full scan and automatic MS/MS, at 1 spectra/s, in the m/z range 100–1500 (MS) or 100–1300 (MS/MS). The collision energy was 40 eV. The MS system was controlled by the software Agilent Mass Hunter Workstation Data Acquisition. Then the data could be analyzed by Agilent Mass Hunter Qualitative Analysis Navigator and imported into LC Image 2.7r LC × LC (64-bit) to generate MS contour plots.

Results and discussion

Optimization of the gradient for the 1D (RPLC) separation

The different parts of Buddleja dividii were expected to present similar metabolite families, therefore, the root was chosen to optimize the 1D RP separation gradient. Several gradients were tested, the detailed information about these gradients is listed in ESM Table S1. Figure 3 shows the chromatograms of the root extract separation obtained with the listed gradients. At first, to measure the polarity distribution of all constituents, a full gradient with ACN as organic solvent from 5 to 90% in 55 min was performed as shown in Fig. 3a. Most of the peaks were overlapped without being well separated in the time interval between 20 and 25 min. To perform a better separation of all compounds, a de novo optimization strategy was applied. In this strategy, the optimization was consequently performed segment by segment from the beginning. When the separation of analytes in a segment was optimized, the analytes in the next segment were eluted directly with a high content of organic phase to save time. By analogy, the optimized segment was fixed and the next segment continued to be optimized until the separation optimization of all analytes was completed (Fig. 3a–e). As shown in Fig. 3e, after only four steps, an evenly distributed separation was achieved with a final analysis time of 70 min. Therefore, the gradient showed in Fig. 3e was applied in the following experiments as an optimized 1D gradient.

Fig. 3
figure 3

Optimization of the gradient for the 1D RP separation

Optimization of the gradient for the 2D (HILIC) separation

The composition of mobile phase in the early stage of 2D gradient cycle is critical for focusing the transfer analytes on the head of the 2D column. Any modification of the 2D gradient in this stage might change the separation in 2D. Therefore, the optimization of the 2D HILIC gradient will focus on the early stage of the gradient cycle.

In this term, as a first and common strategy, we installed 80 μL loops in the interface. Considering the 1D flow rate (15 μL/min) and the modulation time (2 min), loops of 30 μL volume would be enough for the collection of the whole 1D fraction; however, with the use of larger loops (80 μL) connected to the interface, 50 μL of the loops was filled with 2D mobile phase, which allowed diluting the 1D fraction in a more adequate solvent, as demonstrated by Montero et al. [19]. This fact contributed to reduce the mismatch solvent incompatibility together with the ACD modulation, helping the focusing effect in the head of the 2D column.

In this regard, the effect of the injection volume in relation to the column void volume of the 2D column is a critical point in online LCxLC, mainly due to the combination of the 1D solvent strength and the very fast flow rates typically used in the 2D may hamper the performance of the separation of the analytes solved in the injected fraction, producing a band-broadening effect. Typically, the injection volume in the 2D column (loop size) should not exceed the 15% of the column void volume to reduce the band-broadening negative effects [20]; however, in the case of HILIC, the recommendation is much stricter considering the important effect of the injection solvent on the HILIC separation. Some results recommend that the injection volume should not exceed the 9% of the column void volume [3] but others reduce this percentage until the 1% [21]. In this work, when the TS modulator was employed, the injection volume (80 μL) implied the 11% of the 2D column volume, considering the dimensions of the 2D column (150 × 3 mm, 5 μm) and the interstitial column volume (68% of the total volume), which determine a dead volume of the 2D column of 720.8 μL. However, with ACD modulation, there is not a real injection volume as it happens in the TS modulation, since in ACD the transfer pump elutes the fraction from the loop with a controlled flow rate (slower than the 2D pump flow rate used in the TS modulator configuration for the elution of the fraction collected in the loop) and after that, this flow rate is mixed with the high 2D flow rate, so in this case, we can talk about a transfer fraction diluted by two different flow rates more than an injection volume, which improves even more the focusing effect in the head of the 2D column.

Moreover, the use of shifting gradients along the LCxLC analysis not only helps to achieve more orthogonality but also to reduce the breakthrough since a modification of the 2D initial mobile phase conditions can be set accordingly to the 1D eluent. Fig. 4b, d, f, and h show four different 2D shift gradients, in which α and β represent the first gradient cycle and the last gradient cycle, respectively in the 2D shift gradients. In Fig. 4a, c, e, and g, the corresponding separations are shown. At the beginning and the end of the 4b and 4d gradients, the organic content of ACN was set at 95% and 100%, respectively. The separations obtained using these two gradients are shown in Fig. 4a and c. Although theoretically, under HILIC mode separation a small aqueous percentage is needed to create a water-layer around the polar particles, comparing both conditions, using 100% ACN as initial conditions the retention of the compounds was significantly increased (Fig. 4d). In addition, the peak width in zone c (Fig. 4c) was much narrower than that in zone a (Fig. 4a). This indicates that a 100% ACN used at the beginning and at the end of the fast 2D gradient leads to a better separation. Applying higher percentage of ACN at the end of the gradient would accelerate the equilibrium of the 2D HILIC column and prepare the conditions for the injection of the next cycle. Furthermore, applying higher content of ACN at the beginning of the gradient would improve the dilution effect of transferred fractions under ACD modulation mode, which results in better focus of analytes on the head of the 2D HILIC column. However, as can be observed in Fig. 4c, the separation with the gradient 4d was not optimal since the main part of the peaks were separated within a narrow small 2D area. The adjustment of the gradient was performed based on the gradient shown in Fig. 4d as shown in Fig. 4f and h. A rapid decrease of ACN to 95% (Fig. 4f) or 90% (Fig. 4h) was set in each cycle between 0.2 and 0.25 min. The chromatogram shown in Fig. 4f indicates that 5% decrease of ACN resulted in a slightly higher coverage of the 2D space in comparison with the distribution achieved using gradient 4h.

Fig. 4
figure 4

Optimization of the 2D shift gradient (α, first gradient cycle and β, last gradient cycle in 2D shift gradient). a 2D plot of the root separation using the gradient shown in b (95% of mobile phase B as initial conditions in the 2D gradients). c 2D plot of the root separation using the gradient shown in d (100% of mobile phase B as initial conditions in the 2D gradients). e 2D plot of the root separation using the gradient shown in f (100% of mobile phase B followed by a second step reducing the mobile phase B as initial conditions in the 2D gradients). g 2D plot of the root separation using the gradient shown in h (100% of mobile phase B followed by a second step with a faster reduction of mobile phase B as initial conditions in the 2D gradients)

It is worth to point out that serious breakthrough was observed in the 2D contour plot. This might be attributed to a large amount of low polar compounds in the roots, which were hardly retained on the 2D HILIC column. Nevertheless, any combination would have limitations. In this case, the combination of RPLC/HILIC is appropriated for the separation of amphiphilic compounds but is not suitable enough for the separation of homophilic compounds. The target of this work would focus on the amphiphilic compounds rather than homophilic compounds. Therefore, gradient 4f was determined as the optimal 2D gradient used for the LCxLC system for the following measurements.

RPLCxHILIC for different parts of Buddleja davidii

With the optimized 1D RP gradient and 2D HILIC shift gradient, the separation of the other four parts of Buddleja davidii including flower, leaf, stem, and fruit were performed on the RPLCxHILIC system with ACD modulation. The results are shown in Fig. 5a, b, e, and f. Due to the high-efficiency dilution of the transferred fraction, the constituents of the Buddleja davidii extracts were well separated into narrow and symmetric peaks over the 2D analysis. Compared to Fig. 5a, b, and f, a serious breakthrough could be observed in Fig. 5e. This result indicates that there are many homophilic compounds existing in the stem, which is similar to the result obtained in root separation. The results shown in Fig. 5a and b also indicate that more polar compounds with high concentration could be observed in the flowers and leaves than in other parts, which also implies that the metabolic profiles and, therefore, the metabolic activities of flowers and leaves are more vigorous than in other parts.

Fig. 5
figure 5

Contour plots of RPLCxHILIC with ACD modulation or with TS modulation for different parts of Buddleja davidii. Flower with ACD modulation (a), leaf with ACD modulation (b), leaf with TS modulation (c), flower with TS modulation (d), stem with ACD modulation (e), fruit with ACD modulation (f)

To demonstrate the advantage of the ACD modulation, the separation of the flower and leaf extract was also performed with TS modulation under the optimized gradient, the separation results are shown in Fig. 5c and d. Compared to the separation shown in Fig. 5a and b, without ACD modulation much less peaks could be separated and they presented lower peak intensities. This negative effect occurred because of the no-dilution of the fraction in the TS modulation which resulted in a considerable breakthrough in the dead volume. The solvent in the transferred fraction (high content of water) was a very strong elution solvent for the 2D HILIC separation, which prevents a focussing effect of the analytes on the head of the 2D HILIC column. In conclusion, the ACD modulation provides much better flexibility than the TS modulation for the dilution of the transferred fraction and the achievement of the re-focusing and 2D separation.

Constituents analysis for the root of Buddleja davidii by RPLCxHILIC-Q-TOF MS

The advantage of RPLCxHILIC coupled to MS with ACD modulation

For further evaluation of the established RPLCxHILIC method, two 2D-LC configurations with and without ACD modulation were coupled to DAD and Q-TOF MS detectors to study the chemical constituents in the Buddleja davidii root. The high effective separation power provided by the ACD modulation facilitates the online identification of the constituents in root by coupling 2D-LC to high-resolution mass spectrometry. The results are shown in ESM Fig. S3. Figs. S3A and S3E show the UV contour plots for both measurements. As previously discussed, serious breakthrough could be observed at the dead volume without ACD modulation (ESM Fig. S3A), and much more improvement was achieved with ACD modulation (ESM Fig. S3E). The measurement without ACD modulation showed that only a few peaks could be well retained and observed in the 2D contour plot, which means that most constituents were eluted with the transferred fraction solvent in the 2D system without retention on the 2D HILIC stationary phase.

The same result could be obtained from the extracted ion chromatogram (EIC) contour plots in ESM Fig. S3. As observed from the EIC contour plots, with ACD modulation (ESM Figs. S3F, G and H), the spots show very good shapes and a clear background without any breakthrough, which can be attributed to the sufficient dilution of the transferred fraction. This leads to highly focus of amphiphilic analytes on the head of the 2D HILIC column. In contrast, with TS modulation, a significant peak splitting and fronting were found in the EIC contour plots (Figures S3B, C and D). It indicated that the amphiphilic compounds were divided in the 2D analysis into a non-retained portion and a retained portion. The compounds from the non-retained portion were quickly eluted together with the high water content solvent from the transferred fraction, forming a breakthrough peak in the dead volume due to a scarce dilution during the modulation. However, in the retained portion, part of the amphiphilic compounds presented relative separation in the 2D analysis probably due to an inter-diffusion effect in the loop between the transferred fraction and the 2D mobile phase, that allowed a partial focusing in the 2D column.

Nevertheless, the insufficient dilution produced by the inter-diffusion in the loop still resulted in serious peak fronting. As a serious consequence, the existence of the splitting peaks at dead volume might lead to a misunderstanding on the number of isomers, as shown in the EIC contour plot in ESM Figs. S3B, C and D, which could be considered as double when TS modulation is applied. Those results indicate that enough dilution of the transferred faction in high orthogonality 2D-LC is necessary to achieve proper separation and true MS chromatogram, avoiding misinformation.

The results of the EIC contour plots with ACD modulation demonstrate that the coupling of the proposed 2D-LC system with high-resolution mass spectrometry could provide high-quality MS information for the online identification of chemical components from complex samples.

Qualitative analysis of constituents in the roots of Buddleja davidii

The MS and MS/MS measurements of the root extract were performed with the optimized 2D-LC-UV-MS method. The general method for the identification of compounds in the sample mixture was carried out by comparing the detected molecular ion and MS/MS fragment ions with the reported information in the literature. Therefore, the reported researches about Buddleja davidii were carefully reviewed. Firstly, the comparison of the theoretically calculated molecular ion masses of the reported compounds with the detected molecular ion masses obtained by Q-TOF was done. Once the detected mass value was the same as the calculated ion mass of the reported compounds, it was postulated that the same compound was detected in the prepared sample. However, without MS/MS information and only with the sum formula is hardly possible to carry out an identification due to a multitude of possible isobaric compounds. Therefore, MS/MS analysis was performed. Even with MS/MS data, uncharacteristic fragmentation is often difficult to interpret and, besides, the differentiation between isomeric compounds is not possible. Here, the combination of ion mobility with MS could be helpful [22].

Due to the lack of previous information about the chemical composition of this plant, only a few compounds with defined structures were found in the reported work. To reduce this problem, another strategy was developed to characterize as much Buddleja davidii compounds separated by the 2D-LC-MS analysis as possible. According to the principles of biology, the plants with a genetic relationship usually have similar metabolic networks, resulting in similar chemical constituents. Based on this, in addition to the reported information of the compounds in Buddleja davidii, the components with defined structure in various species of the genus Buddleja were collected to extend the candidate compounds. As shown in ESM Fig. S1, in the literature several compounds of the genus Buddleja were described and these were considered possible candidates for Buddleja davidii [13,14,15,16,17,18, 23,24,25,26,27,28,29]. The accurate molecular ion masses [M-H] of each compound collected from the information about the Buddleja genus were calculated with Agilent MassHunter Qualitative Analysis Navigator software. Then the data were imported to the software LC Image. With this software, the contour plots can be easily generated, which allows the determination of a given molecular ion masses [M-H] existing in the huge total ion current (TIC) mass information. According to the calculated molecular ion masses [M-H] of the Buddleja described compounds, the EIC contour plots were extracted from the TIC contour plot.

After that, the compounds that were not found in the analysis were excluded. For example, two typical compounds present in the genus Buddleja have an m/z 283.0612 and 423.3632, which correspond to acacetin and δ-amyrone, respectively. The EIC of both compounds in the RPxHILIC analysis of the root is shown in ESM Fig. S4. In the EIC contour plot of m/z 283.0612, a clear spot was observed, which means that this compound is presented with a relative high concentration. In contrast, in the contour plot of the m/z 423.3632, there is no clear spot, but a complex background was observed, which means that the ion with m/z 423.3632 was not present in Buddleja davidii root or its intensity was too low to be detected. Thus, this ion would be ignored in this study.

Obviously, in the TIC, many more ions existed than those we screened with the above method. In order to find out more ions with relatively high concentrations, further effort was done with the LC image software by comparing the UV and TIC contour plots. On one hand, the positions of the signals in the TIC contour plot were compared with the spots in the UV contour plot, and carefully examined to find out new ions, and then the EIC contour plot was generated and saved in LC image software. On the other hand, the spots observed in the TIC contour plot, but not in the UV contour plot, were also checked. After that, the ions in the EIC contour plot were further selected and excluded according to the method shown in ESM Fig. S4. Considering the molecular mass and the MS/MS fragments, these compounds were also tentatively identified.

Finally, EIC contour plots, which were screened by the previous three methods, were merged. As shown in ESM Fig. S5, some EIC contour plots contained one spot, for example in the m/z 905.2727 (ESM Fig. S5 left panel). Other EIC contour plots might contain several spots for the same mass, which indicated the existence of multiple isomers such as m/z 637.2148 (ESM Fig. S5 right panel). For both situations, the molecular ions and corresponding MS/MS fragments, once detected, were collected and then listed in Table 1. In total, 45 compounds including isomers have been found in the Buddleja davidii root sample. However, due to scarce available information about the chemical composition of this plant, some ions could not be identified accurately by comparing to the detected ions with the reported works.

Table 1 MS identification of the constituents found in the RPxHILIC-DAD-QTOF MS analysis of the root sample of Buddleja davidii

Conclusion

In this study, a recent developed RPLCxHILIC system with at-column dilution (ACD) modulation was applied for a comprehensive analysis of constituents in Buddleja davidii. With ACD modulation, the incompatibility of high orthogonal 2D-LC systems could be successfully overcome. Meanwhile, the flexible adjustment of the dilution of the transferred fraction supported the advantages of the RPLCxHILIC coupling for an excellent separation of amphiphilic compounds with a wide range of polarity. The results demonstrated the potential of a 2D-LC system with ACD modulation for wide applications in the herbal medicine analysis. With this method, a broad separation of the metabolites present in Buddleja davidii was achieved. However, more information about the chemical composition of the separated compounds should be done to obtain comprehensive information about the metabolic profile of this herbal medicinal. Nevertheless, the detailed MS and MS/MS data of some detected ions with relative high intensities can be considered as a good reference for further studies.