Abstract
To investigate the wide range of pollutants occurring in sewage sludge, an analytical method for comprehensive non-target screening is needed. To the best of our knowledge, no procedures currently exist for the full screening of organic contaminants in sewage sludge, which is the ultimate goal of this project. We developed non-discriminating sample preparation methods for gas chromatography-mass spectrometry (GC-MS) analysis. Pressurized liquid extraction (PLE) was used for extraction, with in-line (silica gel selective PLE, SPLE) or off-line clean-up (gel permeation chromatography, GPC). This combination allowed the analysis of non-polar compounds of all sizes and small semi-polar and non-polar compounds. The results show that the combination of SPLE and PLE with GPC is suitable for analysis of established as well as new contaminants. Both methods were validated for 99 compounds with different properties. For all GC suitable analytes, either one of the methods produced acceptable recoveries (64 to 136%). As a test, the two methods were used for non-target screening of Swedish sewage sludge. A tiered approach was used to tentatively identify the sludge contaminants. In total, 1865 and 1593 compounds were found of which 321 and 192 compounds were tentatively identified for the PLE and SPLE method, respectively. For a comprehensive coverage of contaminants, the two methods should be used together, with the PLE method covering a wider polarity range and the SPLE method a wider size range. In addition, polar substances will require liquid chromatography-mass spectrometry analysis, the method for which will be developed soon.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Globally, there are more than 100,000 chemicals currently used every day [1]. Many of these chemicals, among them potential pollutants, are disposed of in wastewater and hence enter sewage treatment plants (STPs). STPs are used to remove nutrients, but also some metals and organic chemicals, from urban water to create a less contaminated effluent. Consequently, STPs form a link between the technosphere and the environment. A by-product of the sewage treatment process is sewage sludge—a solid product that contains nutrients as well as pollutants. These nutrients make the sewage sludge attractive for applications such as fertilizer for agriculture, provided that the contaminant levels are not too high.
Statistics have shown that 54% of the sewage sludge in Europe and 61% of the North American sewage sludge are used in land applications, whereas 10 and 17%, respectively, are placed in landfills [2, 3]. The rest of the sludge is either combusted, disposed of, or reused in other ways.
In order to safely dispose of or reuse the sewage sludge in agriculture the European Union directive 86/278/EEC forces member states to monitor heavy metal concentrations in sludge and soil on a regular basis when sewage sludge is used as fertilizer [4]. Similarly, the US government has defined maximum pollutant loadings for sewage sludge when used as fertilizer on agricultural crops. Again, heavy metals were the only regulated pollutants [5]. For EU member states, national requirements apply as well. Some of these are much stricter than the EU laws (e.g., in Denmark, Finland, Sweden, and the Netherlands) and some of them also include organic contaminants in sewage sludge [6]. In Sweden, for example, maximum loadings for polychlorinated biphenyls (PCBs), nonylphenol ethoxylates (NPEs), polycyclic aromatic hydrocarbons (PAHs), and toluene are defined and those are measured before sludge is spread on crops [6]. Nevertheless, sewage sludge might still pose a risk when used on arable land as in all cases only target compounds are monitored and some crops are known to take up pollutants from soil [7,8,9]. This study aims to develop a method that enables a comprehensive screening of sewage sludge and thereby allowing detection and monitoring of currently unknown organic contaminants present in sewage sludge. Although the literature contains many examples that deal with the analysis of sewage sludge, no study so far has involved non-target screening of sewage sludge, which is the scope of this study. Traditionally, sludge has been extracted using Soxhlet and ultrasound extraction, but nowadays, these methods are often replaced with pressurized liquid extraction (PLE), as highlighted by Zuloaga et al. [10]. By reducing solvent consumption and process time, providing improved extraction rates, and enabling extraction of polar as well as non-polar compounds [10, 11], PLE would appear to be a suitable method for the comprehensive extraction of sewage sludge. In addition, PLE provides the opportunity of an in-cell clean-up using, for example, Florisil, silica gel, alumina [12], or combinations of them [13]. Such procedures, known as selective PLE (SPLE), decrease the amount of co-extracted interfering matrix compounds from solid samples, such as lipids or humic and fulvic acids [14] and may, therefore, enable direct analysis after extraction [13]. SPLE has been applied to various matrices such as soil and sediment [15,16,17], food and feed samples [18], and sewage sludge [13, 19] and dates back to 1996 when the use of alumina was suggested in a Dionex application note to retain fat [12].
In SPLE, there is a balance between the polarity range of chemicals extracted and the purity of the extracts. Polar solvents or solvent mixtures will extract a wide range of chemicals but will also extract more of the matrix. Although in some cases, only filtering and/or derivatization is required before analysis by gas chromatography-mass spectrometry (GC-MS) [13, 19], some analysts apply further clean-up prior to analysis [20, 21] in order to reduce interference and improve the limit of detection (LOD) [22]. Conventional PLE will, in most cases, require further clean-up. For a non-target screening, a non-destructive clean-up is generally used, such as gel permeation chromatography (GPC) [22], liquid-liquid partitioning, or adsorption chromatography.
The combination of PLE and GPC has previously been used for target and non-target analysis of solid matrices [23, 24]. While PLE can be employed with various different types of solvent, a binary mixture of a non-polar solvent (such as n-hexane) and a more polar solvent (such as dichloromethane (DCM)) is often used [25]. For example, Kim et al. [26] extracted polychlorinated dibenzo-p-dioxins, dibenzofurans, and biphenyls from animal feed using n-hexane/DCM (1:3) and Muscalu et al. [27] extracted halogenated organic compounds from soil, sediment, and sludge using n-hexane/DCM (3:1). However, no studies have focused on a comprehensive non-target screening of organic contaminants in sewage sludge.
The current study aimed to develop procedures for non-target screening of sewage sludge. For this purpose, PLE with off-line GPC clean-up was compared to SPLE with in-cell silica clean-up. The extraction efficiency and the amount of co-extracted matrix were assessed for several solvents and solvent mixtures. As a test of utility, the two methods were used for non-target screening of Swedish sewage sludge samples. The number of sample constituents captured and the spectra quality (percentage of peaks that could be tentatively identified) obtained with the two non-target screening techniques were compared. There was also an assessment of whether a combination of the two approaches would enlarge the chemical domain covered. Finally, the potential need for complementary LC-MS analyses was discussed.
Materials and methods
Experiment overview
The method development and evaluation included two methods for extraction, PLE and SPLE, two methods for solvent evaporation, Turbovap and Rotavap, and two methods for sulfur removal, using acid-activated copper and tetra butyl ammonium sulfite (TBA) reagent, respectively. In addition, a method validation study was carried out using spiked and unspiked sewage sludge.
Materials
The 8270 MegaMix® standard (see Electronic Supplementary Material (ESM) Table S1 for compound information) was bought from Restek (Bellefonte, PA, USA). Deuterated PAHs (see ESM Table S2 for more information) were obtained from Cambridge Isotope Laboratories (Tewksbury, MA, USA). Sand (Fontainebleau PROLABO®) and 2-propanol (HiPerSolv Chromanorm 100%, PROLABO) were purchased from VWR (Leuven, Belgium), whereas concentrated hydrochloric acid was obtained from VWR, Fontenay-sous-Bois (France). Sodium sulfate, silica gel 60, acetone (≥99.8%), n-hexane (≥98.0%), and cyclohexane were obtained from Merck KGaA (Darmstadt, Germany). DCM (99.99% purity), isooctane (99.94% purity), ethyl acetate (99.96% purity), and methanol (99.99% purity) were purchased from Fisher Scientific (Loughborough, UK). Copper of mesh size 10–40 (≥99.9% purity) and sodium sulfite (≥98%) were acquired from Sigma-Aldrich (St. Louis, MO, USA). TBA hydrogen sulfate was purchased from Molekula (Shaftesbury, UK). Glass fiber filter papers (GFFs) with a diameter of 27 mm were acquired from Dionex (Sunnyvale, CA, USA). An Omnifit glass column (L 50 cm, i.d. 25 mm) from Diba Industries Ltd. (Cambridge, United Kingdom) and SX-3 Bio-Beads from Bio-Rad Laboratories AB (Hercules, CA, USA) were used for GPC.
Sludge sampling and sample pre-treatment
Digested, dewatered sludge (15 days in digester) was obtained from the STP in Umeå, Sweden. Samples were taken in the morning and frozen immediately until further use. Prior to extraction, the samples were freeze-dried using a Lyovac GT 2 (SRK System Technik GmbH, Riedstadt, Germany) equipped with an Edwards High Vacuum Pump E2M2, and the dry weight was determined (~33.6%). Afterwards, the sludge was homogenized using a mortar and pestle. As a third step directly before extraction, a filling material, either pre-cleaned (PLE with acetone) sand or pre-baked (550 °C) sodium sulfate, was mixed with the dried sludge (approximately 3:2, w/w) to create a homogeneous mixture that filled the extraction cells evenly. The detailed procedures are described below.
Extraction equipment and conditions
Sample extraction was carried out using a Dionex™ ASE™ 350 system equipped with 22-mL stainless steel extraction cells under the following conditions: 120 °C, 5 min static extraction, 3 extraction cycles, 100% flush volume, and 60 s nitrogen purge. To reduce the risk of contamination, high purity solvents were used. During the method development, different solvents and solvent combinations were tested. The solvent volume used for the extraction resulted in approximately 50 mL under the specified conditions. More information on the solvents can be found in the respective sections below. In addition, the extraction cells, sand, and GFFs were pre-cleaned using the PLE system with acetone under the following conditions: 100 °C, 1 min static extraction, 3 extraction cycles, 100% flush volume, and 60 s nitrogen purge.
PLE method development experiments
Non-polar solvents such as n-hexane are expected to release less co-extracted matrix but may not exhaustively extract contaminants. Binary solvent mixtures generally offer better extraction efficiencies. The method development therefore included the following solvents and solvent mixtures: n-hexane, n-hexane/DCM (80:20, v/v), and n-hexane/DCM (50:50, v/v). The polar modifier selected, DCM, is aprotic and known to efficiently desorb difficult to extract compounds such as PAHs from solid matrices [28]. Moreover, conventional PLE and SPLE with silica were compared. To assess the suitability of the methods, the co-extracted matrix amount and extraction efficiency were determined and compared.
For the evaluation of the co-extracted matrix, 1-g sludge aliquots mixed with sand for homogenization were extracted as described above. After extraction, the solvent was fully evaporated and the residue was determined gravimetrically (d = 0.001 g). For the extraction efficiency evaluation, analytical standards were spiked to sand as follows: the PLE cells were filled with a GFF and pre-cleaned sand and spiked with approximately 1 μg of the 8270 MegaMix standard and the SPLE extraction cells were filled with a GFF, 5 g silica gel 60 (dried at 130 °C for 12 h or overnight), a second GFF on top, and pre-cleaned sand and spiked with approximately 1 µg of the 8270 MegaMix standard. A blank containing one GFF and sand and one GFF, silica gel 60, another GFF, and sand were prepared for the PLE and SPLE method, respectively. After solvent exchange to isooctane and volume reduction to about 1 mL, d10-phenanthrene (approximately 544 ng per sample) was added as the volumetric standard. Analysis was carried out using an Agilent 7890A GC (Agilent Technologies, St. Clara, CA, USA) coupled to a high-resolution (HR) time-of-flight (TOF) MS (HRT; Leco Corp. St. Joseph, MI, USA) with electron impact (EI) ionization. The instrument was equipped with a Gerstel CIS4 inlet, which was operated in pulsed splitless mode. The splitless time was 105 s with an inlet purge flow of 25 mL/min and septum purge flow of 3 mL/min. A 30-m DB-5MS Ultra Inert column (0.25 mm i.d., 0.25 μm film thickness) from Agilent was used. The oven program was as follows: 80 °C (3.8 min), 15 °C/min, and 300 °C (6.5 min). Helium was used as carrier gas with a flow of 1 mL/min. The transfer line was held at 300 °C. The ion source temperature was 250 °C and 12 spectra per second were recorded in the range from m/z 38 to 400.
A calibration curve with ten points ranging from 1 to 1000 ng/mL was prepared. A linear regression curve with a fixed intercept at zero was used for the determination of the analyte concentrations.
Solvent evaporation experiments
The MegaMix standard was used for testing two methods of solvent evaporation. The first method used a Turbovap concentration workstation (Biotage AB, Uppsala, Sweden), operated at 35 °C and 500 mbar, and the second method used a rotary evaporator (Rotavap) from Heidolph (Schwabach, Germany), operated at 50 °C with a pressure below 150 mbar. The Turbovap was used with 60 mL PLE vials and the Rotavap with 100 mL pear-shaped flasks, which were tilted to create a horizontal solvent surface and minimize the deposition of chemical residues on dry walls. A MegaMix aliquot equivalent of 1 μg of each analyte was added to 50 mL isooctane, which was then evaporated to 1 mL using the two techniques. The experiments were carried out in triplicate and a blank containing only solvent was included. A mix of deuterated PAHs was added as a volumetric standard and the samples were analyzed with GC-MS, as described in the “Method development” section.
PLE method validation experiments
A total of 99 analytes, including the MegaMix 8270, PCBs, organophosphates, fragrances, pesticides, and others, were used for validation of the final method. For information about native standards/analytes and labeled standards including their spiking levels, please refer to the ESM Tables S1, S3, and S4. The added amounts of native analytes were higher than of labeled standards in order to sufficiently exceed the intrinsic sludge levels. Three sets of samples were prepared in triplicate for each method: (i) 1 g sewage sludge spiked with native and labeled compounds, (ii) sewage sludge spiked with labeled compounds, and (iii) inert material (pre-baked sodium sulfate) spiked with labeled compounds (blanks).
Each set was extracted using PLE and SPLE with n-hexane/DCM (80:20, v/v), leading to a total of 24 samples. Samples extracted with PLE (not SPLE) were further cleaned by using GPC (mobile phase, cyclohexane/ethyl acetate (3:1); flow, 5 mL/min; fraction, 23–59 min). The combination of cyclohexane and ethyl acetate is commonly used in GPC [29]. The flow was adjusted not to exceed the maximum column pressure while the collection window was determined by injecting the MegaMix and collecting fractions for subsequent GC analysis to determine when the compounds elute. The column was packed in-house with approximately 45 g SX-3 Bio-Beads and was compressed to a bed height of 40 cm. For all samples and blanks, sulfur was removed using TBA sulfite reagent as explained below. The analysis was carried out using the GC-HRT system described above, equipped with a secondary oven and a quad jet two stage thermal (liquid nitrogen) modulator for GC × GC analysis. The first column was a 30-m Rtx-5MS (0.25 mm i.d., 0.25 μm film thickness), and the second column was a 1.1 m Rxi-17Sil MS column (0.25 mm i.d., 0.25 μm film thickness), both from Restek. The oven programs were as follows: 90 °C (2 min), 5 °C/min, and 300 °C (5 min) for the first oven and 105 °C (2 min), 5 °C/min, and 300 °C (8 min) for the second. The modulator had a temperature offset of 15 °C relative to the secondary oven, and the modulation period was 4 s with a hot jet and cold jet duration of 1.2 and 0.8 s, respectively. The transfer line was held at 325 °C. The ion source temperature was 250 °C, and 150 spectra per second were recorded in the range from m/z 38 to 1000.
A six-point calibration curve was prepared of which a linear regression curve (intercept at zero) was created for the quantification of the analytes. Information about the linear range and R 2 can be found in Table S5 (see ESM). Before calculating the recovery, the amount of analyte detected in the unspiked sewage sludge, if present, was subtracted from the amount detected in the spiked sewage sludge.
Sulfur removal experiments
Sulfur removal using activated copper and a TBA sulfite reagent were compared using triplicate treatments for recovery of the following contaminants (all at 1 ng/μL in isooctane): the 8270 MegaMix, an organochlorine pesticide mix (GC multiresidue pesticide standard no. 2), and an organophosphorus pesticide mix (GC multiresidue pesticide standard no. 8) from Restek and diazinon, 2-(methylthio) benzothiazole, and thiabendazole from Dr. Ehrenstorfer GmbH (Augsburg, Germany). The recovery of each analyte was determined using GC × GC-MS, as described in the “Method validation” section.
For sulfur removal with copper, the copper was activated using concentrated hydrochloric acid and then rinsed each three times with Milli-Q water (Merck Millipore), methanol, and DCM. The activated copper was added in small portions (~½ teaspoon) to the samples until freshly added copper no longer discolored. Samples were kept overnight in the fridge and more copper was added if additional discoloring was visible the next day.
For sulfur removal using the TBA sulfite, a reagent mixture was prepared and used as described by Jensen et al. [30]. In brief, TBA sulfite reagent was prepared by mixing 1.695 g TBA hydrogen sulfate with 50 mL Milli-Q water followed by threefold extraction, each with 15 mL n-hexane for removal of impurities. Afterwards, the solution was saturated with 12.5 g sodium sulfite. Samples in 2 mL isooctane were mixed with 1 mL 2-propanol and 1 mL TBA sulfite reagent. The mixture was shaken and sodium sulfite was added in 100-mg portions until a solid residue remained after shaking. Then, 5 mL Milli-Q water was added and the mixture was shaken for another minute. Afterwards, the mixture was centrifuged (10 min, 2000 rpm) and the supernatant was transferred.
Data evaluation
The limit of quantification (LOQ) and LOD were derived from method validation blank values (see section “PLE method validation experiments”) where possible. In all other cases, they were determined using the standard deviation of the triplicate injections of the lowest point of the standard curve. The formulae for the LOD and LOQ are as follows:
where σ is the standard deviation of the response (blank or standard dilution close to the LOQ, respectively) and S being the slope of the standard curve.
The peak finding and library search for the non-target application were carried out using the ChromaTOF software (version 1.90.60) from Leco Corporation in connection with the NIST MS library (2011). For a peak to be accepted, the following criteria had to be fulfilled: (i) the area of the peak in the sample had to be at least three times higher than the area of the same peak in the blank and (ii) the peak had to be found in at least two out of three sample replicates. The stepwise procedure of identifying and classifying peaks in sludge chromatograms was as follows:
-
1.
Peaks occurring in the blank (in high enough concentrations) as well as the sample were removed (as defined above).
-
2.
Features that occurred only in one of the triplicates were removed.
-
3.
Peaks were classified into groups according to the rules in Table 1 in combination with the regions defined in Table 1 and Fig. 2 (see “Results” section). All classification regions followed the upwards trend (increasing second dimension retention time) caused through the isothermal (starting at 41 min) in the end of the oven temperature program.
-
4.
The remaining peaks were identified using the NIST library (similarity and probability), fragmentation patterns, and, where possible, retention indexes. Only hits with a similarity match greater than 500 were displayed. To reduce the amount of peaks to look at, only peaks that had a first hit with either a high similarity (>750) or a high probability (>7000) were considered. For compounds where no retention index was found, a simple linear regression model using retention times of standard analytes and their boiling points was used for giving an approximate retention time. Retention times were used for exclusion purposes rather than confirmation.
-
5.
Chlorine and bromine filters were applied. Firstly, ChromaTOF’s built-in chlorine and bromine filters were used. In addition, our own filter criteria were applied (Table 1).
-
6.
The mass defect was used to identify chlorinated and brominated compounds using 81Br–79Br and 37Cl–35Cl (nominal isotope spacing divided by exact isotope spacing), respectively, as reference for normalization. The mass spectrum was summed over a range of 10 min each. Since the raw chromatograms/spectra were used and peaks were identified manually, peaks that were missed in the peak picking process during the data processing could also be identified.
The in silico fragmentation tool MetFrag [31] was used to identify unknown chlorinated compounds (steps 5 and 6 above). MetFrag uses compound structures stored in databases (e.g., PubChem or Chemspider) to predict the fragmentation of small molecules. Those fragmentation patterns are then compared to a spectrum that is inserted by the user. The similarity of the spectrum inserted by the user to the predicted fragmentation is then given. Originally, MetFrag was developed for tandem MS data but, it can also be applied for EI MS data.
Here, the internet database Chemspider was used as a source for candidate structures matching the neutral mass of the highest m/z present in the spectrum, with a 5 ppm mass tolerance. The electron ionization spectra for the unknown compounds were exported from ChromaTOF and compared to the fragments generated by MetFrag from [M+] using a 5 ppm or 0.001 mDa tolerance. Only compounds including (at least) carbon, hydrogen, and chlorine were considered. The Chemspider data source count and reference count were taken into account in scoring the results. Hereby, the spectral match was weighted with 100%, while the data source count and reference count were weighted with 50% each.
Results
Method development
During the method development, PLE and SPLE extraction efficiencies were compared for different solvents or solvent mixtures. For both methods, there was an improvement in extraction efficiency when changing from n-hexane as the pure extraction solvent to the 20% DCM in n-hexane mixture but almost no improvement when increasing the DCM percentage to 50%. For PLE and SPLE, the median recovery percentages for the 20% DCM in n-hexane mixture were almost identical at 71 and 76%, respectively. The 10-percentile values did, however, differ greatly, with greater than 10-fold higher recovery values for PLE (48%) than for SPLE (3%), which is no surprise as SPLE is a more selective extraction technique. For both methods, early eluting compounds showed lower extraction efficiencies or recoveries than later eluting compounds. An evaluation of evaporation methods was, therefore, carried out prior to the validation study.
The amount of co-extracted matrix increased with the amount of DCM used in the extraction solvent mixture and was higher for PLE than SPLE in all cases. The percentages of co-extracted material using n-hexane, 20% DCM in n-hexane, and 50% DCM in n-hexane were 5, 6, and 7% for PLE and 0.7, 1.6, and 2.5% for SPLE, respectively. As the extraction efficiency was considerably higher for the 20% DCM mixture than for pure n-hexane, without showing a significant increase for the 50% DCM mixture, the less polar solvent mixture was chosen for both PLE and SPLE. This mixture also releases slightly less matrix compared to the more polar mixture.
Solvent evaporation
The graph comparing Rotavap versus Turbovap evaporation for solvent volume reduction (Fig. 1) clearly shows that the ratio is above one predominantly and, thus, that Rotavap gives a better result, i.e., higher recovery of analytes. Only bis(2-ethylhexyl) adipate showed a slightly better recovery using Turbovap. However, the difference is not significant. As expected, the difference between the methods was relatively small for high molecular weight analytes such as large PAHs and larger for low molecular weight analytes such as mono- and di-substituted benzenes, phenols, and anilines. On average, the recoveries can be improved by approximately 20% by using Rotavap instead of Turbovap. Rotary evaporation was therefore used during the method validation.
Method validation
The method validation included two different ways of extraction and clean-up, PLE followed by GPC, and SPLE with silica as the in-cell clean-up sorbent. Both methods used 20% DCM in n-hexane as the extraction solvent. The two methods were able to extract most of the analytes. However, some analytes had to be excluded from the dataset due to poor GC performance (see Table 2, footnote). A few other compounds (2,4-dimethylphenol, 2,4-dichlorophenol, bisphenol A) also suffered from relatively poor chromatography and their results are, therefore, slightly more uncertain, as illustrated by a relatively high variation among replicates. Those data were still kept in the dataset for comparison purposes.
In general, the PLE method worked better for most compounds (Table 2), although there were problems with analyzing a few large PAHs (likely due to the relatively narrow (too short) collection windows used in the GPC). The SPLE also worked well for many compounds. It does, however, show the expected losses of polar analytes, such as the organophosphates (OPs), several phenolic compounds, some phthalates, diazinon, and carbamazepine due to the sorption to the silica sorbent.
Taking both methods into account, the non-polar and moderately polar compounds (left half of Table 2 and pesticides and fragrances) showed recoveries ranging from 64 to 136%, while the LOD values ranged from as low as 0.02 ng/g for bis(2-chloroethoxy) methane to 76 ng/g for benzo(g,h,i)perylene (data from ESM Table S5). Several of the more polar compounds (right half of Table 2) did, however, show relatively low recoveries, including many OPs, some chlorophenols, and 1,2-dinitrobenzene. These, and other more polar compounds, would be better analyzed using a complementary LC-MS method.
Sulfur removal
The data for the validation study (Table 2) were produced using sulfur removal with a TBA sulfite reagent, much because it has been claimed to be a soft method [30, 32]. This method was, however, found to be difficult to work with and the process time-consuming. An alternative technique using copper was therefore tested to improve the method further.
The recoveries of roughly 100 compounds were determined for the two sulfur removal techniques: copper and TBA sulfite reagent. Diethyl phthalate, bis(2-ethylhexyl) phthalate, and bis(2-ethylhexyl) adipate showed high blank values for both procedures and were therefore excluded from the dataset. In addition, endrine ketone was excluded from the dataset due to poor reproducibility, i.e., it had a high standard deviation for both procedures. The recovery values for the remaining compounds using the copper treatment ranged from 42 to 114%, while TBA sulfite reagent treatment resulted in recoveries from 38 to 127%. Median values were 90 and 85% for the copper and TBA sulfite reagent treatments, respectively. The 10-percentile and 90-percentile were calculated as 79 and 98% for copper and 53 and 98% for TBA sulfite reagent, respectively. Thus, the copper treatment was found to be slightly better with both a higher median recovery and a narrower recovery range. The copper treatment was also found to be much more user-friendly and therefore more preferable.
Application
In total, after removing the blanks and peaks that occurred in only one replicate, 1865 and 1593 peaks were obtained from the ChromaTOF peak finding algorithm, using sewage sludge processed with the PLE and SPLE methods, respectively. Some 633 (34%) and 378 (19%) peaks had a spectral match with a similarity equal to, or higher than, 75% for the PLE and SPLE methods, respectively, using the NIST library. Figure 2 shows the complexity of the PLE and SPLE sample extracts, and Table 3 shows how many compounds could be identified or characterized using the different techniques. Since a tiered approach was used, compounds that were identified or characterized at an early stage do not appear in later stages. Therefore, the number of identified compounds generally decreases in the later stages. In total, 174 and 45 compounds were uniquely identified (tentatively) using the PLE and SPLE method, respectively. In addition, 147 compounds were tentatively identified through both methods.
The automatic peak classification detected a large number of alkanes, alkenes, and related compounds such as long-chain ketones, aldehydes, amides, as well as free fatty acids and methyl derivatives thereof. It also detected a large number of phthalates. Figure 3 shows the distribution of the classified compounds, as well as four groups of compounds, alkyl-benzenes, flavor and fragrances, PAHs and derivatives, and steroids, which contain constituents with high structural similarity that are difficult to identify correctly without using reference materials. More information on the alkyl-benzenes, flavor and fragrances (mainly terpenoid and musk compounds), PAHs and derivatives, and steroids can be found in the ESM (Table S6).
Roughly 15% of the compounds detected (17% for PLE and 12% for SPLE) could be assigned CAS numbers. A list, including tentatively identified compounds that were found in the final extracts from the PLE and SPLE methods (tiers 2 and 3), is shown in Table 4. This table also contains the first dimension retention times, the second dimension retention times, the nominal molecular weights, the mass deviation (ppm) from the theoretical molecular ion or major fragment ion mass, and the score and probability produced from reverse searches of the NIST library. The compounds identified were loosely divided into the following groups: alkyl-phenols, extractives (of plant origin), organophosphate esters, pharmaceuticals and personal care products (PPCP), stabilizers (antioxidants and UV absorbers), as well as other halogenated compounds and process chemicals.
Further searches for halogenated chemicals using halogen-specific filters (tier 4) or mass defect plots (tier 5) revealed 17 additional chlorinated compounds, 11 using Cl/Br filters and six using mass defect plots (Table 5). All chemicals that could be tentatively identified were aromatic compounds. Five compounds could be tentatively identified using MetFrag. The final searches, using mass defect filters, captured only chlorinated biphenyls.
Discussion
Method development
In SPLE with silica, the polar sorbent retains polar compounds such as phenols, leading to very low recoveries. Non-polar compounds such as PAHs show no difference in extraction efficiency between PLE and SPLE since they are not affected by the silica. As mentioned before, the amount of co-extracted matrix increases with increasing percentage of DCM. DCM is more polar than n-hexane and, for this reason, it has the ability to elute more polar compounds but also more matrix compounds, e.g., humic and fulvic acids from the sewage sludge. The challenge is to find the proper balance between analyte extraction and matrix retention. For sewage sludge, 20% DCM in n-hexane seems to be the best compromise.
In the PLE method, the clean-up was carried out using size exclusion since common matrix compounds such as lipids and humic and fulvic acids are big molecules. However, there is also a risk of losing other large GC-amenable compounds. The combination of both methods will allow a comprehensive screening of GC-amenable analytes for non-target screening of sewage sludge. The PLE method allows detection of compounds that are non-polar or moderately polar, but rather small, while the SPLE method allows the detection of relatively non-polar compounds of all sizes. The more polar analytes would have to be analyzed by LC-MS, which is currently being evaluated. Thus, using three complementary methods, a highly comprehensive non-target screening of environmentally relevant organic contaminants in sewage sludge and similar matrices (e.g., soil and sediment) could be achieved, see Fig. 4. Admittedly, very large (molecular weight of 2000 and above) non-polar contaminants would still not be possible to analyze by the three proposed methods (Fig. 4), but those are generally not bioavailable (i.e., too large to pass biological membranes).
Solvent evaporation
During the method development, it was realized that the extraction efficiency for the studied analytes correlated with the retention time. Early eluting compounds show, in general, lower extraction efficiencies compared to later eluting compounds. Since a non-polar column was used here, the elution order is determined by the boiling point of the analytes. Early eluting compounds have a lower boiling point, while late eluting compounds have a higher one. Thus, the boiling point of the analytes seems to correlate with the recovery of the PLE and SPLE method. There is a possibility that compounds with a lower boiling point were lost in the subsequent evaporation step, which was therefore further evaluated.
The results of the comparison of Rotavap and Turbovap clearly show that Rotavap is the technique of choice for solvent evaporation for the tested compounds. The advantage of Rotavap is that the walls of the small pear-shaped flasks used in this study are constantly covered with a film of solvent. This prevents analytes from transferring to the gas phase by evaporation or adsorbing strongly to the glass walls. In the Turbovap on the other hand, a constant flow of air creates a vortex that speeds up the evaporation but also leads to the formation of a film of dry sample on the vial walls, which may cause a loss of analyte by evaporation or adsorption. Hence, greater losses and lower recoveries are to be expected for relatively volatile analytes using the Turbovap.
Method validation
The method validation showed that most of the analyzed compounds had a reasonable recovery in at least one of the methods (Table 2). Some compounds, such as anilines, yielded bad recovery percentages using both PLE and SPLE. These compounds are hard to analyze using GC in general. A method for LC-MS analysis is currently being developed that will be suitable for these compounds, as discussed above.
High recoveries (>100%) were observed for several compounds, e.g., low molecular weight PAHs and PCBs. For these compounds, the amount used to spike the samples before the extraction was rather low compared to typical concentrations found in sewage sludge. The natural occurrence of the analytes in the sludge, thus, could contribute to the amount detected. The fact that PCB 118 and PCB 105 yield high recovery percentages supports this hypothesis since they are major congeners in technical mixtures. Additionally, the recoveries of the 13C–labeled PCBs that were added as internal standards were calculated. The tetra- to hepta-CBs had recoveries between 71 and 102% for PLE and 84 and 110% for SPLE, which is fit for purpose for non-target screening methods.
However, for some polar compounds, such as bisphenol A and some other phenolics, recoveries and standard deviations were higher than acceptable for at least one of the methods. These high recoveries and variabilities could result from a matrix enhancement effect. When matrix is present, as in the sample, these matrix compounds can bind to active sites in the GC inlet system, column, or ion source making them unavailable for analytes. During instrument calibration, no such matrix compounds are introduced and therefore analytes can bind to active sites. This effect is called matrix induced chromatographic response enhancement (MICRE) and has been described previously [33, 34]. MICRE can be avoided by adding so-called analyte protectants to the sample and/or standard directly before analysis [35,36,37]. These analyte protectants bind to the active surfaces in the GC system, mainly the inlet, thereby facilitating analyte transfer. Since the use of analyte protectants could improve the analysis of polar compounds, the authors recommend the use of those for future analyses. In addition, more internal standards could be added allowing a better matching of analytes and internal standards.
It should be emphasized, however, that in non-target screening, it will never be possible to match analytes and internal standards perfectly and that reference compounds are lacking generally. Thus, the analyses will be semi-quantitative at best, although the relative concentrations among similar samples can be determined much more accurately. A tiered approach is therefore recommended: first, prioritize between the tentatively identified compounds, then confirm the identity of top-priority compounds, and finally develop quantitative analytical methods for the selected contaminants.
Sulfur removal
Sulfur removal with copper and with TBA sulfite reagent produced similar results for the tested analytes. However, the variability among recoveries for copper treatment is lower and both the median as well as 10-percentile are higher. The US EPA stated in their method 3660B that treatment with copper powder (fine granular) may affect certain pesticides. Some of the mentioned pesticides (heptachlor, malathion, ethion, and diazinon) were included in the current study but no major losses could be seen. The recoveries for these analytes ranged from 82 to 92%. The reason for this may be that we used a more coarse copper granulate than what is recommended in method 3660B. The main advantage of the copper treatment, however, is its ease of use and speed. It will therefore be used in future suspect and non-target screening studies.
Application
Applying the non-target screening methods to a sewage sludge sample revealed that both methods are very good at detecting non-target compounds. Although more compounds were identified using the PLE method (see Table 3), the SPLE method gave additional information. This is to be expected as the two methods cover different parts of the chemical property (polarity and size) space, as shown in Fig. 4.
Figure 3 clearly shows that there is a difference in polarity among compounds detected by the PLE and SPLE method, respectively. The majority of classified compounds that were found using the SPLE method were relatively non-polar compounds, such as alkanes or alkenes, alkyl-benzenes, alkyl-PAHs, and PAHs. In contrast, more polar compounds, such as alkanals, phthalates, fatty acids, and methyl derivatives, and long-chain amides were found using the PLE method. This can easily be explained by the loss of polar compounds due to sorption to the silica gel used in the SPLE method.
Size-dependent differences in compounds detected by the two methods are clearly shown in Fig. 2. The GC × GC region between 25 and 40 min first dimension retention time and 2.5 and 3.0 s second dimension retention time, which corresponds to the unresolved complex mixture (UCM) of crude oil or weathered petroleum, differs greatly for the two methods. The SPLE sample is much richer in UCM than the PLE sample, most likely due to losses of big molecules in the GPC step of the PLE method. A similar observation was made for the last part of the analysis (40 min or later): 26% of the compounds detected using the SPLE method (mainly steroids) eluted in this chromatographic region whereas only 16% of the compounds using the PLE method elute in this region. Vitamins E α and E γ, vitamin E α acetate, and 2,4-bis(2-phenylpropan-2-yl)phenol were also among the big compounds that were lost in the PLE method but found using the SPLE method (Table 4).
The non-target screening of sludge using the SPLE and PLE methods revealed many compounds besides the classified or grouped chemicals. Tables 4 and 5 show the compounds that could be tentatively identified. These compounds were primarily of anthropogenic origin (organophosphates, PPCPs, synthetic antioxidants, UV screens and stabilizers, pesticides and other chlorinated compounds, and process chemicals), which may be of environmental relevance, while some were of biogenic origin (e.g., extractives and vitamins). The tables appear to indicate that the PLE method revealed many more unique compounds than the SPLE method. However, many of the compounds that were grouped (PAHs, alkyl-PAHs, N/S/O-PAHs, steroids, and flavor and fragrances) could be tentatively identified using both methods (ESM, Table S6), and some were only tentatively identified using the SPLE method (e.g., large alkyl-benzenes). Hence, combining both methods increases the amount of information made available. The range of compounds that can be captured using the two non-target screening methods (Tables 4 and 5) is quite wide, including small and large compounds, e.g., m-amino-phenylacetylene (117 g/mol) and vitamin E α acetate (473 g/mol), as well as non-polar and relatively polar compounds (e.g., PCBs and organophosphate esters and phenols).
The results also illustrate how the NIST library can be used to identify compounds effectively by using both the similarity and probability scores. As shown in Table 3, the majority of tentatively identified compounds (83% for PLE and 91% for SPLE) were identified using the NIST library similarity scores, with a cutoff at 75% (tier 2). By using such a cutoff, only a limited number of spectra have to be manually reviewed but at the risk of losing information for low level contaminants affected by instrument noise. Some of these contaminants, those with distinct spectra, may be found by probability sorting the data that have similarity scores <75% (tier 3). In this way, 5 and 11% of the tentatively identified compounds using PLE and SPLE, respectively, could be revealed.
However, the majority of existing organic chemicals do not have an MS spectrum in the NIST library. Therefore, additional techniques are needed to identify compounds of particular concern, such as halogenated compounds. An attempt was made to screen for chlorinated and brominated compounds (tiers 4 and 5), as these compounds are generally environmentally relevant (c.f. PCBs, PBDEs, dioxins). Eleven compounds were found using halogen-specific filters (tier 4) and six compounds were found using mass defect plots (tier 5); all of these were chlorinated biphenyls. Molecular formula information for the tier 4 compounds could be generated using information about accurate mass and isotopic patterns. In some cases, searches in databases such as Chemspider and SciFinder resulted in a plausible candidate structure, e.g., 9,10-di(chloromethyl)-9,10-dihydroanthracene. In other cases, it was possible to extract a spectrum manually that was missed in the original peak picking and find a plausible hit in NIST. This resulted in the discovery of triphenylchloromethane, 9,10-di(chloromethyl)-anthracene, and p-(6-chloro-4-phenyl-2-quinolyl)aniline. Finally, an attempt was made to interpret the remaining spectra manually, resulting in two additional tentative identifications: one DDT metabolite and one chlorinated PAH. The spectra of all unknown chlorinated compounds that could not be assigned a tentative molecular structure were analyzed using MetFrag. This resulted in the identification of five additional compounds: (i) one compound which is likely to be dichloroxylenol, a commercial disinfectant [38], (ii) 4-(3,4-dichlorophenyl)tetralone and one of its isomers that are potential impurities in the pharmaceutical Sertraline [39], and (iii) two compounds belonging to the classes of dichlorophenyl coumarins or dichloroflavones, which are highlighted in patents suggesting their use in tire rubber production [40, 41]. To our knowledge, none of these compounds were detected in environmental samples up to now. Spectra for these compounds and other compounds that are not listed in NIST are given in the ESM (Figs. S1–S7).
Some of the other compounds that were tentatively identified have, to our knowledge, not previously been reported in sludge or in environmental samples either, e.g., the chlorinated PAH derivatives and triphenylchloromethane. However, similar compounds have been reported. Chlorinated PAHs have been found in incinerator flue gas, car exhaust, and urban air [42, 43]. Triphenylchloromethane is quite reactive and is frequently used in organic synthesis. It may be degraded to triphenylmethane that has been found in sediment [44].
Overall, the two proposed methods (PLE with GPC and SPLE, followed by sulfur removal with copper) in combination with a soon-to-be developed LC-MS method will provide a comprehensive methodology for the screening of a large variety of compounds with different properties in sewage sludge. This should also be feasible for other similar environmental matrices such as soil and sediment. It should be realized that there is an overlap between the applicability domains of the three methods. Once the methodology is fully developed, this should be highly advantageous. It should, for instance, be possible to use EI spectra from GC-MS to verify the identity of compounds originally detected by LC-MS and vice versa: it should be possible to use LC-MS data to generate molecular ion information that is often missing in GC-MS spectra. This could lead to the discovery of many more new and emerging chemicals in samples of environmental relevance, which could be subject to targeted measurement campaigns, environmental risk assessments, and STP improvement initiatives.
References
Boxall AB, Sinclair CJ, Fenner K, Kolpin D, Maund SJ. When synthetic chemicals degrade in the environment. Environ Sci Technol. 2004;38:368A–75A. doi:10.1021/es040624v.
Kroiss H, Rechberger H, Egle L. Phosphorus in water quality and waste management. Integr Waste Manag - Vol II. doi:10.5772/18482.
Evans TD. Sewage sludge: operational and environmental issues FR/R0001—review of Current Knowledge. Found Water Res. 2016;4th Edit.:1–46.
Council of the European Union. Council directive of 12 June 1986 on the protection of the environment, and in particular of the soil, when sewage sludge is used in agriculture.
U.S. EPA. e-CFR, Title 40: protection of environment, Part 503: standards for the use or disposal of sewage sludge, subpart B: land application.
European Comission. Disposal and recycling routes for sewage sludge—part 2: regulatory report.
Carter LJ, Harris E, Williams M, Ryan JJ, Kookana RS, Boxall ABA. Fate and uptake of pharmaceuticals in soil-plant systems. J Agric Food Chem. 2014;62:816–25. doi:10.1021/jf404282y.
Wu C, Spongberg AL, Witter JD, Fang M, Czajkowski KP. Uptake of pharmaceutical and personal care products by soybean plants from soils applied with biosolids and irrigated with contaminated water. Environ Sci Technol. 2010;44:6157–61. doi:10.1021/es1011115.
Kumar K, Gupta SC, Baidoo SK, Chander Y, Rosen CJ. Antibiotic uptake by plants from soil fertilized with animal manure. J Env Qual. 2005;34:2082–5. doi:10.2134/jeq2005.0026.
Zuloaga O, Navarro P, Bizkarguenaga E, Iparraguirre A, Vallejo A, Olivares M, et al. Overview of extraction, clean-up and detection techniques for the determination of organic pollutants in sewage sludge: a review. Anal Chim Acta. 2012;736:7–29. doi:10.1016/j.aca.2012.05.016.
Schantz MM. Pressurized liquid extraction in environmental analysis. Anal Bioanal Chem. 2006;386:1043–7. doi:10.1007/s00216-006-0648-2.
Haglund P, Spinnel E. A modular approach to pressurized liquid extraction with in-cell clean-up. LCGC Eur. 2010;23:292–301.
Pena MT, Casais MC, Mejuto MC, Cela R. Development of a sample preparation procedure of sewage sludge samples for the determination of polycyclic aromatic hydrocarbons based on selective pressurized liquid extraction. J Chromatogr a. 2010;1217:425–35. doi:10.1016/j.chroma.2009.11.081.
Réveillé V, Mansuy L, Jardé É, Garnier-Sillam É. Characterisation of sewage sludge-derived organic matter: lipids and humic acids. Org Geochem. 2003;34:615–27. doi:10.1016/S0146-6380(02)00216-4.
Chuang JC, Van Emon JM, Schrock ME. High-throughput screening of dioxins in sediment and soil using selective pressurized liquid extraction with immunochemical detection. Chemosphere. 2009;77:1217–23. doi:10.1016/j.chemosphere.2009.09.010.
Zhang Z, Rhind SM, Kerr C, Osprey M, Kyle CE. Selective pressurized liquid extraction of estrogenic compounds in soil and analysis by gas chromatography-mass spectrometry. Anal Chim Acta. 2011;685:29–35. doi:10.1016/j.aca.2010.11.013.
Do L, Lundstedt S, Haglund P. Optimization of selective pressurized liquid extraction for extraction and in-cell clean-up of PCDD/Fs in soils and sediments. Chemosphere. 2013;90:2414–9. doi:10.1016/j.chemosphere.2012.10.070.
Bjorklund E, Sporring S, Wiberg K, Haglund P, Holst C. New strategies for extraction and clean-up of persistent organic pollutants from food and feed samples using selective pressurized liquid extraction. TrAC Trends Anal Chem. 2006;25:318–25. doi:10.1016/j.trac.2005.10.003.
Martínez-Moral MP, Tena MT. Focused ultrasound solid-liquid extraction and selective pressurised liquid extraction to determine bisphenol A and alkylphenols in sewage sludge by gas chromatography-mass spectrometry. J Sep Sci. 2011;34:2513–22. doi:10.1002/jssc.201100159.
Díaz-Cruz MS, Barceló D. Highly selective sample preparation and gas chromatographic-mass spectrometric analysis of chlorpyrifos, diazinon and their major metabolites in sludge and sludge-fertilized agricultural soils. J Chromatogr a. 2006;1132:21–7. doi:10.1016/j.chroma.2006.07.062.
Vallecillos L, Borrull F, Pocurull E. Determination of musk fragrances in sewage sludge by pressurized liquid extraction coupled to automated ionic liquid-based headspace single-drop microextraction followed by GC-MS/MS. J Sep Sci. 2012;35:2735–42. doi:10.1002/jssc.201200326.
Nieto A, Borrull F, Pocurull E, Marcé RM. Pressurized liquid extraction: a useful technique to extract pharmaceuticals and personal-care products from sewage sludge. TrAC - Trends Anal Chem. 2010;29:752–64. doi:10.1016/j.trac.2010.03.014.
Du G, Xiao Y, Yang HR, Wang L, Song YL, Wang YT. Rapid determination of pesticide residues in herbs using selective pressurized liquid extraction and fast gas chromatography coupled with mass spectrometry. J Sep Sci. 2012;35:1922–32. doi:10.1002/jssc.201200169.
Lacorte S, Ikonomou MG, Fischer M. A comprehensive gas chromatography coupled to high resolution mass spectrometry based method for the determination of polybrominated diphenyl ethers and their hydroxylated and methoxylated metabolites in environmental samples. J Chromatogr A. 2010;1217:337–47. doi:10.1016/j.chroma.2009.11.024.
Carabias-Martínez R, Rodríguez-Gonzalo E, Revilla-Ruiz P, Hernández-Méndez J. Pressurized liquid extraction in the analysis of food and biological samples. J Chromatogr A. 2005;1089:1–17. doi:10.1016/j.chroma.2005.06.072.
Kim BH, Jeong JS, Chang YS. Congener-specific distribution of polychlorinated dibenzo-p-dioxins, dibenzofurans and biphenyls in animal feed. Food Addit Contam. 2003;20:659–67. doi:10.1080/0265203031000109468.
Muscalu AM, Reiner EJ, Liss SN, Chen T, Ladwig G, Morse D. A routine accredited method for the analysis of polychlorinated biphenyls, organochlorine pesticides, chlorobenzenes and screening of other halogenated organics in soil, sediment and sludge by GCxGC-μECD. Anal Bioanal Chem. 2011;401:2403–13. doi:10.1007/s00216-011-5114-0.
Lundstedt S, Haglund P, Öberg L. Simultaneous extraction and fractionation of polycyclic aromatic hydrocarbons and their oxygenated derivatives in soil using selective pressurized liquid extraction. Anal Chem. 2006;78:2993–3000. doi:10.1021/ac052178f.
Sundkvist AM, Olofsson U, Haglund P. Organophosphorus flame retardants and plasticizers in marine and fresh water biota and in human milk. J Environ Monit. 2010;12:943–51. doi:10.1039/b921910b.
Jensen S, Renberg L, Reutergårdh L. Residue analysis of sediment and sewage sludge for organochlorines in the presence of elemental sulfur. Anal Chem. 1977;49:316–8. doi:10.1021/ac50010a033.
Ruttkies C, Schymanski EL, Wolf S, Hollender J, Neumann S. MetFrag relaunched: incorporating strategies beyond in silico fragmentation. J Cheminform. 2016;8:1–16. doi:10.1186/s13321-016-0115-9.
U.S. EPA. Method 3660B.
Bruzzoniti MC, Checchini L, De Carlo RM, Orlandini S, Rivoira L, Del Bubba M. QuEChERS sample preparation for the determination of pesticides and other organic residues in environmental matrices: a critical review. Anal Bioanal Chem. 2014;406:4089–116. doi:10.1007/s00216-014-7798-4.
Rahman MM, Abd El-Aty AM, Shim JH. Matrix enhancement effect: a blessing or a curse for gas chromatography?—a review. Anal Chim Acta. 2013;801:14–21. doi:10.1016/j.aca.2013.09.005.
Anastassiades M, Maštovská K, Lehotay SJ. Evaluation of analyte protectants to improve gas chromatographic analysis of pesticides. J Chromatogr a. 2003;1015:163–84. doi:10.1016/S0021-9673(03)01208-1.
Čajka T, Maštovská K, Lehotay SJ, Hajšlová J. Use of automated direct sample introduction with analyte protectants in the GC–MS analysis of pesticide residues. J Sep Sci. 2005;28:1048–60. doi:10.1002/jssc.200500050.
Kirchner M, Húsková R, Matisová E, Mocák J. Fast gas chromatography for pesticide residues analysis using analyte protectants. J Chromatogr a. 2008;1186:271–80. doi:10.1016/j.chroma.2007.08.089.
Grillo JA, Ojo OY. Evaluation of activity of dichloroxylenol (1.2% w/v) on Staphylococcus aureus Oxoid 701/1 Lot 610254 and clinical isolates of Escherichia coli and Salmonella typhi. African J Microbiol Res. 2013;7:4265–70. doi:10.5897/AJMR12.2187.
Sinco Pharmachem. Sertraline impurity (SRTRC-3). http://www.sincopharmachem.com/SertralineImpuritySRTRC-3.html. Accessed 2 Mar 2017.
Ford BA, Sandstrom PH, Wideman LG. Tire with rubber containing flavone. US Patent 6,872,771 B1.
Sandstrom PH. Tire with rubber containing flavone. US Patent 2014/0135433 A1.
Wang Q, Miyake Y, Amagai T, Horii Y, Nojiri K, Ohtsuka N. Profiles and distribution of halogenated polycyclic aromatic hydrocarbons from waste incinerators. Organohalogen Compd. 2013;75:27–30.
Haglund P, Alsberg T, Bergman A, Jansson B. Analysis of halogenated polycyclic aromatic hydrocarbons in urban air, snow and automobile exhaust. Chemosphere. 1987;16:2441–50. doi:10.1016/0045-6535(87)90303-1.
Kadokami K, Li X, Pan S, Ueda N, Hamada K, Jinya D, et al. Screening analysis of hundreds of sediment pollutants and evaluation of their effects on benthic organisms in Dokai Bay. Japan Chemosphere. 2013;90:721–8. doi:10.1016/j.chemosphere.2012.09.055.
Acknowledgements
The authors would like to thank the Swedish EPA for funding through their Environmental Screening Program (Contract 2219-13-002). They are grateful to Matyas Ripszam for help with instrumentation and Darya Kupryianchyk for help with lab work. The authors thank Per Liljelind and Christine Gallampois for fruitful discussions and helpful comments.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors have no conflict of interests to declare.
Additional information
Parts of this work have been awarded with the ABC Poster Prize at the 23rd Analytical Days of the Analytical Division of the Swedish Chemical Society, Umeå, Sweden, June 14th-17th, 2016.
Electronic supplementary material
ESM 1
(PDF 720 kb)
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Veenaas, C., Haglund, P. Methodology for non-target screening of sewage sludge using comprehensive two-dimensional gas chromatography coupled to high-resolution mass spectrometry. Anal Bioanal Chem 409, 4867–4883 (2017). https://doi.org/10.1007/s00216-017-0429-0
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-017-0429-0