Introduction

Polycaprolactone diol (PCLD) is an aliphatic polyester in which two polycaprolactone (PCL) chains are linked via a diethylene glycol molecule. The low molecular weight PCLD is widely used in the manufacture of biodegradable polyurethanes [1, 2]. From the point of view of stability, PCLD is known be to biodegradable making it attractive for specific applications in biomedicine, e.g., drug delivery, in medical devices, and as degradable sutures [3,4,5]. Recently, PCL-related polymers have been proved useful as carbon source and biofilm carrier for denitrifying microorganisms in tertiary wastewater treatments [6]. Thus, they can directly contribute to the release of synthetic polymers in wastewater treatment plants (WWTP). WWTPs essentially consist of a sequence of treatment steps designed to remove both organic matter and nutrients. Typically, they include a biological treatment under aerobic conditions designed to reduce organic load, followed by a tertiary treatment with nitrification/denitrification process to remove ammonium and nitrate. Each treatment step is characterized by specific conditions (aerobic or anoxic), which result in the development of adapted microorganisms consortia [7]. It is well known that the degradation pathways and kinetics of synthetic organic compounds largely depend on the conditions (aerobic or anaerobic) under which the wastewater bacteria are striving. For instance, synthetic polyesters have been reported to display higher degradation rates under aerobic conditions than under anaerobic ones [6,7,8,9].

Table 1 High mass range (MW > 800 Da) transformation product structures identified by in the different polymer samples studied. Structures on the right column are those of the starting polycaprolactone diol (PCLD) polymer, while those in the left column correspond to polycaprolactone (PCL) polymer formed as transformation product under denitrifying conditions
Table 2 Low mass transformation product structures identified by either LC-HRMS, MALDI MSI or both in the different polycaprolactone diol samples studied

A plethora of mass spectrometric analytical techniques have been used for evaluating the degradation of synthetic polymers [10]. The main drawback of the techniques used to date (gas chromatography coupled to mass spectrometry (MS), liquid chromatography-MS, and matrix-assisted laser desorption/ionization time of flight MS (MALDI-ToF-MS) is the loss of spatial information that occurs in the preparation of a liquid sample (extract). This loss is due to the homogenization and dissolution of the sample, a step that can be obviated by employing MALDI-MS imaging (MALDI MSI) on the intact solid sample. By two-dimensional scanning of the sample surface with the laser beam following deposition of a suitable matrix, analyte ions are being desorbed from the surface and extracted into the ToF analyzer for mass spectral recording. Due to the large m/z operating range of the ToF-MS, high molecular weight components of the polymer under investigation can easily be studied. Characterization of polymers [11], polymeric dialyzer membranes [12], and polymers used for joint implants [13] have been performed by MALDI-ToF-MSI. However, the accuracy of MALDI average molecular weights has often been discussed. When MALDI-ToF-MSI is compared to high-resolution MS (HRMS) instruments commonly used in environmental laboratories for identification of small molecules, e.g., Orbitrap-MS [14], it presents limited MS/MS capability relative to its lower mass resolving power and mass accuracy [15].

Several studies have evaluated the biodegradability of polymers in soil, compost [6], wastewaters [16], and marine environments [17]; however, less attention has been paid to the biodegradation of polyesters in wastewaters. Recently, we performed a comparative study on the degradation of a polyester, PCLD, in two aquatic environments, a river and a laboratory mesocosm, applying MALDI-ToF-MS [18]. In a follow-up study [19], we explored the capability of MALDI-MSI as an analytical tool for investigating modifications in the chemical composition at the polymer surface after exposure to either sterile, aerobic, or denitrifying conditions, using PCLD as a probe polymer. Whereas the acquisition of the MALDI mass spectral data and the subsequent image processing proved a powerful technique for studying 2D spatial degradation patterns of the polymer surface, its mass resolution capability turned out to be insufficient for assigning chemical structures to the mass peaks attributed to a series of emerging transformation products [20]. This limitation prompted us to undertake a third study of PCLD degradation exploiting in parallel the spatial capabilities of MALDI-MSI and the identification power of LC-HRMS [11].

The objective of the present study was to investigate the degradation of PCLD following exposure to wastewater bacteria in the aerobic and denitrifying reactor tanks of a municipal WWTP. We set out to identify the chemical changes in the polymer structure leading to degradation products and to determine their spatial distribution on the surface. This was accomplished by making use of LC-HRMS and MALDI-MSI, respectively.

Materials and methods

Chemical reagents and materials

Polycaprolactone diol homopolymer (average MW = 1250 Da) used in the degradation experiments was purchased from Polysciences (Warrington, PA, USA). The matrix trans-2-[3-(4-tert-butylphenyl)-2-methyl-2-propenylidene] malononitrile (DCTB) was purchased from Aldrich (Steinheim, Germany). PCLD was placed in the capsule units (Lancaster, UK) used as polymer holders in the biodegradation experiments; they were purchased from Exposmeter AB (Tavelsjo, Sweden). Each capsule unit consisted of a plastic piston and a plastic cap with a window 2 cm in diameter (see Electronic Supplementary Material (ESM), Fig. S1). Acetonitrile and water were purchased from Fisher Scientific (Geel, Belgium). Formic acid (98–100%) was ACS grade and purchased from Sigma-Aldrich (Schnelldorf, Germany).

Degradation experiments

Wastewater treatment plant description

The wastewater generated by the southern district of Barcelona reaches the WWTP located in Prat de Llobregat (see ESM, Fig. S2) through several pumping stations after a pretreatment for the elimination of sand and grease it passes through an activated sludge biological treatment plant. About 60% of the treated water is discharged into the sea through 3.2 km long emissary. The remaining 40% is subjected to tertiary treatment for nutrient removal and subsequent reuse (environmental flow and irrigation). Main technical characteristics of the plant are reported in the ESM, Table S1.

Degradation experiments in WWTP

Degradation experiments were carried out by placing three capsules filled with PCLD (3–6 g) in the secondary (aerobic) and denitrifying reactors. Samples were collected after 7 days of exposure and analyzed as explained below. Moreover, the following parameters were routinely monitored as part of the regular in-process operational WWTP control: suspended solid, BOD, COD, nitrates, nitrites, phosphor, nitrogen, conductivity, and pH. Data corresponding to the days in which the experiment took place are given (see ESM, Table S1).

Control experiments at lab scale

For control and comparison purposes, a parallel blank experiment was carried out in the lab, in which capsules containing the polymer were exposed to sterilized water at room temperature during the same period of time (7 days). Furthermore, a polymer sample standard (i.e., non-exposed) was analyzed as explained below.

MALDI MSI

Sample preparation

For MALDI-TOF-MS measurements, the polymer was removed from the capsules and dried in a dessicator for 15 min. Slices of 12–16 μm were cut in a cryo-microtome (Leica CM3050S, Nußloch, Germany) and placed onto indium-tin oxide (ITO)-coated glass slides (Sigma-Aldrich). The samples were placed in a desiccator for 15 min before matrix deposition. The section was coated with 30 mg of DCTB using a sublimation device (Fisher Scientific, Madrid, Spain). Sublimation conditions were 100 mTorr pressure at 140 °C during 15 min. The sample slice was externally cooled with ice.

Mass spectrometry

All experiments were performed using an AuToFlex III MALDI-TOF/TOF instrument (Bruker Daltonik GmbH, Bremen, Germany) equipped with a Smartbeam laser at 200 Hz laser at the “medium focus” setting and were controlled using FlexControl 3.0 (Bruker Daltonik GmbH). The analytical conditions for the MALDI MSI analysis were set as follows: positive reflector ion mode; ion source 1 voltage, 19 kV; ion source 2 voltage, 16 kV; lens voltage, 8.5 kV; and mass range m/z 100–3000. Laser irradiance was maintained slightly above threshold. Mass spectra were acquired with spatial resolution set at 200 μm. For each pixel, 250 laser shots were measured per position as a sum of 25 consecutive laser shots in 10 random walk shot steps.

Image processing

Raw spectra were analyzed with Flex Imaging 4.0 software (Bruker Daltonik GmbH, Bremen, Germany) after baseline subtraction and root mean square (RMS) normalization. Ion density maps were created for ions observed from the skyline projection spectrum. Masses were manually selected with a mass accuracy set to ±0.1% (manual peak picking). Regions of interest were manually defined by using both the sample image and MSI data. Image processing and data analysis were done using SCiLS Lab software (Bremen, Germany) allowing identification and quantification of the altered areas.

LC-HRMS

Clean cuts of the polymer exposed in wastewater tanks at the WWTP were dissolved in methanol and then diluted with water (95:5 water:methanol) to achieve the initial concentration of 1 mgL−1 of the polymer. The samples were injected directly (10 μL) into LC-HRMS. A generic LC-HRMS method was applied to detect and identify TPs of the PCLD using an Acquity UPLC coupled to Q-Exactive (Thermo Scientific, Bremen, Germany). Electrospray ionization (ESI) interface was operated in both positive and negative ionization mode to obtain the exact masses of the TPs. The chromatographic separation was performed using an Acquity UPLC C18 column (100 × 2.1 mm, 1.7 μm) proceeded by a pre-column of the same packing material (5 × 2.1 mm, 1.7 μm). The mobile phases for the chromatographic run were as follows: (a) acetonitrile with 0.1% formic acid and (b) water (0.1% formic acid). The total chromatographic run was 15 min with a flow rate of 300 μL min−1. The elution was accomplished with the following solvent gradient: 0 min (10% A)–1 min (10% A)–11 min (95% A) and 13 min (95% A)—13.5 min (10% A) and stabilized until 15 min. The column temperature was set to 40 °C. For the detection method, a data dependent scan was applied with the following parameters: full scan and data-dependant MS2 were set to resolution of 35,000 [full width at half maximum (FWHM)], with the full scan range of m/z 100–1000. In the data-dependent scan, the isolation window was m/z 2 and the normalized collision energy was (NCE) either 15 or 35. Both positive and negative ionization mode runs were used and then compared to a blank sample (containing only the parent homopolymer dissolved in pure water and methanol (95:5)) with the degraded samples using SIEVE 2.2 (Thermo Fisher).

Results and discussion

MALDI mass spectra of polymer sections

Representative average MALDI-TOF mass spectra (mass range up to m/z 2600) of sections of the starting material, the sterile control, and the two samples exposed to aerobic and denitrifying wastewater are shown in Fig. 1A–D at different attenuation factors. Highest mass range (m/z 700–2600) is consistent with the starting polymer structure with increasing number of monomer (caprolactone) units (peak spacing m/z 114). Overall, there is a general decrease of the relative peak intensities in the exposed samples (Fig. 1B–D) as well as differences in the peak distribution, thus suggesting a different distribution of the chain lengths.

Fig. 1
figure 1

Average MALDI-ToF mass spectra recorded in the positive reflectron mode, showing a polycaprolactone diol series [M+Na]+ in the range from m/z 23 to 2600. The differences in the intensity scales: (A) standard polycaprolactone diol; (B) after 7 days of exposure in sterile water. (C) After 7 days of exposure in water under aerobic conditions (WWTP secondary treatment tank); (D) after 7 days of exposure in water under denitrifying conditions (WWTP tertiary treatment tank).

This trend is more pronounced in the samples exposed to aerobic and denitrifying conditions (Fig. 1C, D). In addition, in the spectrum of the latter sample (Fig. 1D), the presence of a second series of oligomers, likewise, spaced by m/z 114 (i.e., m/z 1051.6 and 1165.6) and separated from the homolog peaks of the main series by m/z 104. It was attributed to PCL (Table 1) formally corresponding to a loss of the diethylenglycol unit from PCLD polymer and they were identified as PCL cyclic oligomers [21,22,23,24], previously reported as a possible degradation pathway of PCL (Fig. 1) [25].

As regards the lower mass range series (m/z < 800), with the peak at m/z 500 corresponding to the matrix DCTB, there is also evidence in the degraded polymer samples (Fig. 1C–D) of peaks attributable to degradation of the standard material as well. Again, this is detected in the polymer exposed to aerobic and denitrifying conditions. The structural identification of these peaks was further investigated in detail using LC-HRMS (see “Identification of transformation products using LC-HRMS”). Overall, the above findings are indicative of degradation processes affecting the exposed samples. This is consistent with the fact that polyesters and specifically polycaprolactones are reported to undergo degradation under both aerobic and anaerobic conditions [6].

Identification of transformation products using LC-HRMS

Since there was no qualitative difference between the runs in the positive and negative ionization mode, transformation products (TPs) were identified using the positive ESI. For the identification of polymer TPs, the product ion spectra of the TPs were recorded to determine the most likely elemental compositions and to identify structures of the fragment ions.

All compounds detected were formed as the result of ester hydrolysis of the parent homopolymer (Fig. 2). In total, ten TPs were detected and identified in LC-HRMS (see ESM, Tables S1 and S2). Five were the result of ester hydrolysis forming caprolactone oligomers (see ESM, Table S1 in the homologous series TPs 220, 334, 448, 562, and 676) while the other series detected corresponded to formation of PCL chain with a terminal diethylene glycol (see ESM, Table S2), via (again) ester hydrolysis (TPs 246, 360, 474, 588, and 702). In all cases, the difference between the polymer series was the repeating unit of m/z 114.0681 (6-hydroxyhexanecarbonyl, HOCH2-(CH2)5-CO-) (see ESM, Fig. S3). In general, all TPs were predominantly detected as ionized sodium adducts and [M+H]+.

Fig. 2
figure 2

Proposed degradation pathway of polycaprolactone diol: (A) formation of polycaprolactone (PCL) linear oligomers; (B) formation of cyclic oligomers and (C) formation of polycaprolactone-diethylene glycol monoester

The fragmentation of TP 220 started with a water loss followed by loss of m/z 44.0262 (C2H4O). Its homolog series followed the same pattern (see ESM, Table S1). As for the other series, their fragmentation pattern again was the water loss in combination with loss of C2H4O, after the initial water loss from the acid itself. Detailed fragmentation pattern and mass spectra can be found in the ESM on Tables S1 and S2. A summary of the main peaks identified in “MALDI mass spectra of polymer sections” and “Identification of transformation products using LC-HRMS,” together with their proposed structure is given in Tables 1 and 2 for low mass range (MW < 800 Da) and high mass range (MW > 800 Da) TPs, respectively.

UPLC-HRMS chromatograms corresponding to the low mass range TPs of the denitrifying exposed sample are shown in Fig. 3. Peaks corresponding to polycaprolactone (PCL) fragments (Fig. 3A; TPs 246, 360, 474, 588, and 702) are single as expected from the hydrolysis pattern, while those of polycaprolactone-diethylene glycol monoesters (Fig. 3B: TPs 220, 334, 448, 562, and 676) appear resolved totally or partially in more than one peak (Table 2), tentatively pointing to the presence of isomer mixtures. However, their HRMS-(ESI)-mass spectra (Fig. 4) cannot confirm this hypothesis.

Fig. 3
figure 3

UPLC-HRMS chromatogram of the TPs in the low mass range (m/z < 800); (A) formed by ester hydrolysis and (B) formation of polycaprolactone-diethylene glycol monoester in degradation sample at t 7 days (see also Fig. 2)

Fig. 4
figure 4

HRMS-(ESI)-mass spectra of (A) TP220, (B) TP334, (C) TP448, (D) TP562, and (E) TP676

Spatial distribution of polymer degradation patterns by MALDI MSI

Biotic-mediated processes taking place in the surface of polymer samples exposed in the WWTP environment were expected to exhibit large spatial heterogeneity, especially if they are associated to the formation of biofilm colonies in the surface of the exposed material. This fact should be correspondingly reflected in the surface composition pattern of the exposed polymer probes. To investigate it, MALDI MSI appears as a suitable technique since it allows generating 2D images across the probe surface using selected ions. In Fig. 5A–D, MALDI MSI images generated from polymer cuts of the four samples analyzed using the main ions characteristic of the starting polymer (m/z 1155.7, 1269.7, 1383.8, and 1497.9) are shown as an example. They exhibit notable differences in the spatial distribution on both intensity and uniformity (homogeneity). Whereas the standard polymer (Fig. 5A) shows a highly uniform and intense distribution, exposed samples (Fig. 5B–D) deploy increasingly heterogeneous spatial patterns with a decreasing intensity in the order sterile->aerobic->denitrifying. In the latter case (Fig. 5D), the disappearance of the peak monitored is almost complete in some parts (deep blue colored areas). These findings are fully consistent with those derived from the average MALDI MS exposed in “MALDI mass spectra of polymer sections.”

Fig. 5
figure 5

Comparison of MALDI-ToF images corresponding to sections of (A) PCDL standard, (B) control sample kept in sterilized water, (C) sample exposed to aerobic wastewater, and (D) sample exposed to denitrifying wastewater. Images were obtained by data processing with Flex Imaging 4.0 software. Spatial differences among pixels using the different ions are visible across the surface of the samples. Ions shown are sodium adducts [M + Na]+

The spatial pattern of second oligomer series predominantly formed in the denitrifying exposed sample was investigated using MALDI MSI as well. The images corresponding to the main parent oligomer and the second series formed by hydrolytic loss of the diethylene glycol moiety are shown in Fig. 5. In each compartment, the transformed polymer peak is shown on the right side while the formally parent peak is on the left side. Their MW differs on m/z 104, which corresponds to the diethylene glycol fragment lost. The images corresponding to the main parent oligomer and the second series formed by internal “back-biting” cyclization are shown in Fig. 6. In each compartment, the cycled analog ion distribution is shown on the left side while the formally parent peak ion distribution is on the right side. It is worth noting that the transformation polymer is highly unevenly distributed across the sample, appearing mostly located in the upper part and it is roughly complementary to the starting polymer. Such pattern is approximately repeated for all peaks shown in Fig. 6.

Fig. 6
figure 6

Comparison of MALDI-ToF images corresponding to sections of samples exposed to denitrifying wastewater. Images were obtained by data processing with Flex Imaging 4.0 software. Each pair of images shown corresponds (right) to the peaks of the starting polycaprolactone diol polymer PCDL (M) and (left) the cyclic polycaprolactone analog (M-104). Ions shown are sodium adducts [M + Na]+

We finally studied the spatial distribution of low MW range ions (i.e., below m/e 700). Their structures are reported in Table 2. Most of them were confirmed by LC-HRMS as explained above. Comparing the aerobic and denitrifying exposed sample images, some differences are perceptible. Whereas ions m/z 221.1, 247.1, and 449.2 seem to predominate in the aerobic exposed sample, those at m/z 475.5 and 677.4 are characteristic for the denitrifying exposed sample. Remarkably, while the surface distribution of the latter one (m/z 677.4) follows a similar spatial pattern than those observed for the aforementioned transformation polymer (Fig. Fig. 7, right side images, upper part), the former (m/z 475.5) fits the opposite and complementary distribution (Fig. Fig. 7, right side images, lower part).

Fig. 7
figure 7

Comparison of MALDI-ToF images corresponding to sections of samples exposed to aerobic (left) and denitrifying (right) wastewater using characteristic ions (m/z < 800). Images were obtained by data processing with Flex Imaging 4.0 software.

Conclusions

This study has been demonstrated the complementarity of the MSI and UPLC-HRMS techniques for the determination of the biodegradation products of polycaprolactone diol. Whereas UPLC-HRMS has higher sensitivity thus allowing the unequivocal identification of degradation products through their exact masses, MALDI MSI technique shows how these degradation products are distributed across the surface of the exposed sample.

In addition, different transformation products of the starting polycaprolactone diol oligomer were identified in both the low and high molecular weight mass ranges. On the other hand, there is evidence of three different degradation processes as well, namely (a) the hydrolytic cleavage of the diethylenglycol-polycaproic ester bond, (b) another one taking place on the polyester bonds of the polycaprolactone units, and (c) the cyclization of polycaprolactone oligomers via internal ester formation. The former one seems specific of the denitrifying exposed samples. On the other hand, there is evidence of two different ester hydrolysis processes as well, namely the cleavage of the diethylenglycol-polycaproic ester bond and another one taking place on the polyester bonds of the polycaprolactone moiety. The former one seems specific of the denitrifying exposed samples. Both processes were traced through the corresponding degradation products.

Finally, the study of MALDI MSI images reveal the spatial complexity and specificity of the degradation processes taking place in the exposed polymer surface, which allow clearly differentiation between aerobic and denitrifying environments as found in conventional engineered systems. Such differential behavior is better highlighted in comparison with a sample kept under sterile conditions. Whereas abiotic processes seem to be present as well, it clearly supports the contribution of biotic processes acting on the aerobic and denitrifying exposed samples.