Abstract
Corning archeological reference glasses A, B, C, and D have been made to simulate different historic technologies of glass production and are used as standards in historic glass investigations. In this work, nanoseconds (193, 266 nm) and femtosecond (800 nm) laser ablation were used to study the elemental composition of Corning glasses using laser ablation inductively coupled plasma mass spectrometry. The determined concentrations of 26 oxides (Li2O, B2O3, Na2O, MgO, Al2O3, SiO2, P2O5, K2O, CaO, TiO2, V2O5, Cr2O3, MnO, Fe2O3, CoO, NiO, CuO, ZnO, Rb2O, SrO, ZrO2, SnO2, Sb2O5, BaO, PbO, Bi2O3) are compared with values reported in the literature. Results show variable discrepancies between the data, with the largest differences found for Cr2O3 in Corning A; Li2O, B2O3, and Cr2O3 in Corning B; and MnO, Sb2O5, Cr2O3, and Bi2O3 in Corning C. The best agreement between the measured and literature values was found for Corning D. However, even for this reference, glass re-evaluation of the data was necessary and new values for PbO, BaO, and Bi2O3 are proposed.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The knowledge about the elemental composition of historic glasses is crucial for establishing their age, provenance, or the technology used for their production [1–6]. The required chemical information can be obtained from instrumental techniques, which offer variable sensitivity as well as limits of detection [7–10] exploiting various reference materials to support method validation and evaluation of the measurement uncertainty. The limited collection of available solid reference materials limits the quality assurance and quality control when analyzing solids. Archeometrical analyses are additionally complicated by the uniqueness of the analyzed historic objects, and measurements need to be performed non-destructively or pseudo non-destructively using micro-sampling [7, 11–21].
For historic glass analysis, laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) is a good choice, because it offers direct micro-sampling by focused laser beam and fast multielemental determination with low detection limits [7, 22–25]. This method has been widely and successfully used for elemental analysis of historic glasses [5, 11, 21, 26–29], although the lack of suitable matrix-matched solid standards is still recognized as an important limitation and prevents insights into the accuracy achievable with this technique.
Glass standards in analysis of historic objects
Taking into account the widely varying composition of historic glass, the selection of appropriate matrix-matched solid reference materials is difficult [2, 11]. The most popular glass reference materials are those which are available from NIST (National Institute of Standards and Technology) in form of wafers. These have been widely used as calibration materials for LA-ICP-MS analysis of historic glass [21, 27–30]. The silicate matrix (Si-Na-Ca-Al) of NIST SRM 610 to 617 is doped with the well-known concentrations of 61 elements. However, the composition of historic glasses usually differs considerably from the major element composition of these reference materials. To meet these analytical needs, Corning archeological reference glasses: A, B, C, and D have been produced by the Corning Museum of Glass (USA). These materials reflect the different technologies of historic glass production. Glasses A and B mimic the composition typical for Egyptian, Mesopotamian, Roman, Byzantine, and Islamic glasses; C is similar to glasses from East Asia, and D reflects glasses, which have been produced from the seventeenth to the nineteenth centuries [2, 31, 32].
High purity synthetic oxides and carbonates, NaCl, NH4H2PO3, alumina hydrate, natural ZrSiO4, and SiO2 have been used as precursor materials to produce the Corning glasses [31]. Trace elements have been mixed in two groups (Ti, Sn, B, Ba, Sr, Li, Rb and V, K, Ag, Zr, Ni, Zn, Bi) and ball-milled prior to incorporation. The precursor mixtures were melted, stirred, and quenched in deionized water, then crushed, and re-melted again. The rehomogenized melt was poured into 1-cm-thick sheets. Two groups of elements can be distinguished in the elemental composition of Corning glasses [31, 32]. The first group includes the elements which oxides have been determined by a number of techniques, i.e., gravimetry, polarography, flame photometry, atomic absorption, X-ray fluorescence, or neutron activation (SiO2, Al2O3, Fe2O3, MgO, CaO, Na2O, K2O, MnO, P2O5, TiO2, Sb2O5, CuO, PbO, CoO, BaO, SnO2, SrO, and ZnO). The second group includes the oxides for which nominal compositions were calculated from mass fractions of the precursor constituents added to the batch (B2O3, Li2O, Rb2O, V2O5, Cr2O3, NiO, ZrO2, Bi2O3). It can be assumed that the results obtained by several laboratories should be more reliable then indicative information about the contents of the oxides calculated for the second group, although Corning archeological reference glasses are not characterized by a metrologically valid procedure. These glasses have been mainly used for estimation of the precision and accuracy of measurements in several investigations of historic glass objects [5, 11, 17, 27, 29, 30, 33–40], because large amounts of elemental concentration data have been published for them [11, 27, 31, 32, 37–40 ]. The contents of the literature values of the oxides used in this study are from Vicenzi [32] who referred to Brill [2] (Table 1).
Brief revision of the literature
Brill [2, 31] coordinated the first interlaboratory investigations of elemental compositions of the Corning archeological reference glasses and published recommended and nominal values. Tentative recommended values were given in 1971 [31], with emphasis that these preliminary values would need further evaluation. The final compositions of Corning glasses were published by Brill [2] in 1999, and they were then used as the recommended ones [32]. These values have been used for calibration in quantitative analyses of historic glasses and to check the accuracy of the methods used, although the principles of quality control were not always sufficiently documented [33, 40–43]. The deficiency of information about the obtained precision and accuracy can be noticed in some reports [33, 40, 43] as well as the lack of identifying the Corning glass used [42]. In some cases, even when the accuracy and precision were given and discussed in the text, the detailed results of these control measurements were omitted [17, 34, 35].
Corning C and D were used by Kuisma-Kursula [36, 37] to determine the accuracy of the measurements. Bronk and Freestone [38] validated the usefulness of the scanning electron microscopy/energy-dispersive X-ray analysis in analyzing glass objects with the use of Corning A and B. They described the influence of the procedure used for the preparation of the Corning glasses to the final results and reported problems with quantification of some oxides (Sb2O5, PbO in Corning B). Other authors [39] reported inconvenience of quantification of SnO2, in Corning C. The most comprehensive and multitechnique (electron probe micro-analysis (EPMA), LA-ICP-MS, secondary ion mass spectrometry) investigations of Corning archeological reference glasses were described by Vicenzi et al. [32]. This study reports the determination of minor and trace elements and their distribution in the Corning glasses. The agreement of the presented EPMA results with recommended values is highly variable, while the general agreement between LA-ICP-MS and previously published data is within 5% to 20%. EPMA measurements validated heterogeneous distribution of some elements (SrO, ZnO in all glasses, and BaO, SnO2 in Corning B), while LA-ICP-MS indicated relatively high degree of the compositional uniformity of all the Corning glasses which were examined using 193 nm laser with beam diameter of 23 μm. The details of the LA-ICP-MS results have been discussed for Corning A, B, and D. A value of the MnO content in Corning C has been reported here for the first time [32].
LA-ICP-MS has also been applied by Shortland et al. [30] for the determination of trace elements in glass objects using Corning A as a quality control for the accuracy of the measurements. The authors concluded that the agreement of their results with the values recommended by Vicenzi et al. [32] was rather poor. It is important to note that the most significant discrepancies have been observed for Cr2O3 (161%), PbO, SnO2, and SrO (about 30%). The quantitative results of MnO in Corning C with a mean content of 0.001 wt.% obtained in our laboratory [11] or published by Dussubieux [26, 27] varied strongly from the value given by Vicenzi [32]. Dussubieux [27] reported for few elements, not only Mn in Corning C, that the concentrations provided by Brill [2] might not be accurate.
Aim of the work
The evaluation of the published data revealed various inconsistencies between the recommended and determined values of several constituents of Corning archeological reference glasses, which are of interest to both archeologists and art historians. Corning archeological reference glasses are widely used as calibration or validation materials for the analysis of historic glass. Therefore, the reliability and usefulness of the available data requires a careful re-examination.
The aim of this work was focused on a detailed investigation of the composition of Corning archeological reference glasses A, B, C, and D using LA-ICP-MS. Therefore, various nanosecond (λ = 193, 266 nm) and femtosecond (λ = 800 nm) laser wavelengths were used for sampling. The quantification scheme is based on NIST 610, and the results of each individual sampling using commonly applied wavelengths for archeological studies are summarized and discussed in comparison to literature data.
Experimental
Samples and standard
Three different types of archeological reference glasses, which were fabricated to mimic historic glass recipes, were investigated: Corning glass A and Corning glass B are Na-rich/Ca-bearing silicates; Corning Glass C is rich in Pb and Ba while Corning Glass D is K- and Ca-rich silicate [31, 32] (Table 1).
Standard glass NIST SRM 610 was used as the external standard. The similarity of the bulk composition of NIST 610 to the composition of the Corning glasses [32] was considered to be close to the matrix composition of the Corning glass matrix. Therefore, reduced influence of element fractionation effects on the calculated results was assumed. The comparison of the calibration materials and samples with the correlation coefficient indicating the relationship between the oxides concentration in calibration and unknown samples is given in Fig. 1. The matrix-matching is acceptable for three of the four investigated glasses (R 2 = 0.9864 for Corning A; R 2 = 0.9880 for Corning B, and R 2 = 0.9150 for Corning D), and it is most significantly different for the Corning C (R 2 = 0.4034), when compared with the other glasses.
Instrumentation
LA-ICP-MS was used in this study with the involvement of three different laser systems (Table 2):
-
1.
The laser ablation system LSX-200+ (CETAC, USA) combining a stable, environmentally sealed 266-nm UV laser (Nd–YAG, solid-state, Q-switched) was coupled to inductively coupled plasma mass spectrometer ELAN 9000 (Perkin Elmer SCIEX, Canada). All experiments were performed using Ar as the carrier gas.
-
2.
The 193-nm ArF-Excimer laser (GeoLas Compex, MicroLas, Goettingen, Germany) was coupled to ELAN 6100 (Perkin Elmer, Norwalk, CT, USA).
-
3.
The Ti/sapphire femtosecond laser system (Legend, Coherent Inc., Santa Clara, CA, USA) was coupled to Elan DRC II (Perkin Elmer SCIEX, Thornhill, Canada). All ablations which were carried out using the 193 and 800 nm lasers were performed using He as the sample carrier gas, which was mixed with Ar before entering the ICP.
For all measurements, the samples were placed inside the ablation cell with NIST SRM 610. The calibration material was measured twice at the beginning and twice at the end of each run to correct for instrumental drift using the algorithm proposed by Longerich et al. [44]. Seven replicate single point ablations at locations randomly selected on the glass surface were carried out on each sample. Transient signals were recorded and evaluated for subsequent elemental quantification. The LA-ICP-MS signals were background-corrected and integrated using LAMTRACE program developed by Jackson [45]. Table 3 summarizes the average limits of detection for the different types of laser wavelengths used for sampling (for crater diameter, see operating conditions, Table 2).
The results for all samples were calculated using SiO2 as the internal standard, applying the data displayed in Table 1. Normalization was performed to the total content reported by Vicenzi [32] as 99.79 wt.% for Corning A; 99.33 wt.% for Corning B; 99.80 wt.% for Corning C; and 99.97 wt.% for Corning D. Therefore, although the initial contents of SiO2 as the internal standard were given according to the literature [34], the final contents differed from these recommended values after normalization.
Results and discussion
Analysis of corning archeological reference glasses by means of LA-ICP-MS
The results of Corning archeological reference glasses [31, 32] measured using various laser wavelengths and pulse width for sampling by LA-ICP-MS are given in Table 4. It was found that the results acquired with different laser wavelengths varied significantly and that the 193 nm ns or 800 nm fs lasers provided the comparable results, while the 266 nm ns laser showed most significant deviations.
Homogeneity of Corning glasses
The homogeneity of the Corning glasses has already been reported by Vicenzi et al. [32]. The majority of the concentrations of the elements reported in Table 4 were determined with relative standard deviations less than 5%. The mean Sb2O5 in Corning C is the only exception, with the high relative standard deviations obtained for the all laser wavelengths used in this study (RSD = 19% for 193 ns-laser, RSD = 12% for 266 ns-laser, and RSD = 26% for 800 fs-laser ablation). The mean Sb2O5 in Corning C was calculated consequently as equal to the c = 0.0001 wt.%. irrespective of the laser wavelength. However, these results suggest an inhomogeneous distribution of Sb in Corning C.
Comparison of the results obtained after ablation by means of 193 ns, 266 ns, and 800 Fs lasers
The ratios of the measured to the literature [32] values for each oxide element of the Corning glasses are displayed in Fig. 2.
The deviations of the results acquired within this study to the recommended values caused variable shapes of radar-charts for the overall chemical composition of particular Corning glasses (Fig. 2). Analogous shapes of the charts for different laser wavelengths could be obtained if the interaction of these lasers with sample of identical composition was similar. However, it is known that both laser wavelength and pulse duration affect the degree of fractionation, therefore, some differences of the results were expected. Fractionation effects often are reduced by the use of shorter wavelengths (e.g., ns-193 nm) and shorter pulse widths (e.g., fs-800 nm) [46, 47], and the results obtained with these laser wavelengths for the most of elements were comparable.
The analyzed Corning glass results obtained using a 266 nm laser differed to some extent from the results obtained using ablation with 193- and 800-nm wavelength lasers. The most pronounced difference was observed for the soda–lime–silica glass (Fig. 2, Corning B) but can also be seen in the pattern for the high-lead-and-barium glass (Fig. 2, Corning C). The differences of the respective results for potash–lime–silica glass (Fig. 2, Corning D) are low. The highest differences in all samples were observed for Al2O3, CaO, ZrO2, Bi2O3, SrO, and K2O when ablating at 266 nm when compared with 193 or 800 nm lasers. Inconsistent results for PbO were obtained for the high-lead-and-barium glass (Corning C).
Apart from the mentioned inconsistencies, some systematic variations from the recommended values for each Corning glass can be noticed, and these values will be discussed in more detail. Independent of the lasers used for ablation, all data showed an underestimation of SnO2 and overestimation of Li2O and Cr2O3 in Corning B as well as underestimation of MnO and Sb2O2 and overestimation of Cr2O3 and Bi2O3 in Corning C with respect to the recommended values [32]. The data for Corning D are in good agreement with the data reported by Brill [31], which indicates and accurate characterization. However, Bi2O3, PbO, and BaO differ significantly and indicate an overestimation in the literature.
The 266 nm shows the largest and non-coherent, with the other lasers (193 ns and 800 fs lasers), deviations from the expected values, therefore, the quantitative results obtained using this laser wavelength are excluded from the further comparisons. The use of 193-nm laser wavelength creates smaller particles which are more effectively transported and ionized in the plasma and thus exhibit less elemental fractionation compared with 266 nm lasers [48]. Also, the use of femtosecond laser pulses (800 nm) has been reported to be beneficial for LA-ICP-MS measurements [49]. Therefore, the data acquired using these two types of lasers will be discussed in more detail.
Recommended and determined values for 800 and 193 nm lasers
The recommended contents of the oxides were compared with data acquired using 800 and 193 nm. Discrepancies between these values were evaluated using the z-scores for each individual result. The target standard deviation was assigned as 10% of the recommended value for each element, because the information about the relative standard deviations could only be extracted from the report describing individual results of four repetitive inductively coupled plasma-optical emission spectrometry measurements of Corning glasses from Brill [2]. The calculated values of RSDs exceeded 10% for three oxides: PbO in Corning A (X rec = 0.120 wt.%, SDt = 0.043 wt.%, RSD = 35.5%), Al2O3 in Corning C (X rec = 0.870 wt.%, SDt = 0.115 wt.%., RSD = 13.2%), and Sb2O5 in Corning C (X rec = 0.025 wt.%, SDt = 0.006 wt.%, RSD = 23.1%). The high RSD of the mean content was confirmed by our LA-ICP-MS study only for Sb2O5 in Corning C. Therefore, the SDt applied was taken from Brill [2].
Generally, the z-scores are acceptable (within ±2; 95%) and unacceptable when the score is outside ±3 (greater than 99%), and questionable with intermediated values. The calculated z-score values for the oxides in Corning glasses were used to indicate discrepancies between the measured and recommended data (Fig. 3). The values which were outside the calculated threshold for both wavelengths (193 and 800 nm) indicated the need of a re-evaluation of the recommended concentrations in the Corning glass. Most of the results acquired in this study agree with the published and recommended values [2, 31, 32], but some discrepancies were found for different oxides in all Corning glasses (Fig. 3). The results of P2O5, PbO, BaO, and Cr2O3 were found unacceptable in Corning A; BaO, SnO2, B2O5, Li2O, Cr2O3, and BiO3 in Corning B; MnO, P2O5, Sb2O5, Cr2O3, and Bi2O3 in Corning C; and PbO, BaO, and Bi2O3 in Corning D.
Re-evaluation of the elemental composition of Corning archeological reference glasses
Although the use of fs laser typically decreases laser-induced elemental fractionation, increased errors due to mass-load-induced matrix effects in the ICP can influence the quantification capabilities and figures of merit such as accuracy [50–53]. Furthermore, a detailed study about the influence of the wavelength on the accuracy needs further validation and has not been studied as extensively as for a 193 nm ns laser ablation system.
Therefore, the re-evaluation of the recommended values is proposed using the mean of seven LA-ICP-MS measurements at 193 nm. These values are summarized in Table 5 with their standard deviations (SD) and compared with values from literature [2, 27, 30–32].
Conclusions
Three types of archeological reference glasses, reflecting different historical technologies of glass production, were investigated using LA-ICP-MS. The results obtained using different laser wavelengths and pulse durations (nanoseconds, 193 and 266 nm, and femtoseconds, 800 nm) were compared among each other and to recommended values from the literature. Comparison of the re-evaluated values which were proposed here with the data recommended by Brill [2, 31] indicated that some tentative values [31] were closer to the data set generated in this study using a 193 and an 800 nm laser for sampling then the values published later on [2]. The recommended quantification especially of PbO in Corning A, BaO in Corning B, P2O5 or Sb2O5 in Corning C, and PbO or BaO in Corning D were overestimated in the final report [2], comparing to the published preliminary results [31].
Based on this LA-ICP-MS study, some significant discrepancies were found. The data indicate that the 266 nm laser ablation used for sampling is not suitable for quantification of these glasses using NIST 610 for calibration. Therefore, the data set generated in this study using a 193 and an 800 nm laser for sampling was compared with the recommended glass composition. The concentrations of P2O5, PbO, BaO, and Cr2O3 in Corning A, as well as BaO, SnO2, B2O5, Li2O, Cr2O3, and BiO3 in Corning B differ most significant, and some new values are proposed. This confirms reports in the literature where the determination of some trace elements, including Sn, have been reported to be difficult (Corning B). Similar reports on Cr, Rb, and Bi in Corning C have been found. Therefore, corrected values for MnO, P2O5, Sb2O5, Cr2O3, and Bi2O3 in Corning C and PbO, BaO, and Bi2O3 in Corning D are also reported. All other oxides contents in the various Corning glasses were quantified by LA-ICP-MS with good agreement to the data reported in the literature [2, 31, 32].
References
Henderson J (2000) The science and archaeology of materials. An investigation of inorganic materials. Routledge, London & New York
Brill RH (1999) Chemical analyses of early glasses. The Corning Museum of Glass, vol 2. Corning, New York
Shortland AJ (2002) Archaeom 44:517–530
Dussubieux L, Gratuze B, Blet-Lemarquand M (2010) J Archaeol Sci 37:1646–1655
Van Der Linden V, Cosysns P, Schalm O, Cagno S, Nys K, Janssens K, Nowak A, Wagner B, Bulska E (2009) Archaeom 51:822–844
Rehren Th (2008) J Archaeo Sci 35:1345–1354
Mass JL (1999) The conservation of glass and ceramics. In: Tannent NH (ed) Research, practice and training. James & James, London
Pollard AM, Heron C (1996) Archaeological chemistry. The Royal Society Of Chemistry, Cambridge
Adriaens A (2005) Spectrochim Acta Part B 60:1503–1516
Hogg SET, McPhail DS, Oakley VL, Rogers PS (1999) The conservation of glass and ceramics. In: Tannent NH (ed) Research, practice and training. James & James, London
Wagner B, Nowak A, Bulska E, Kunicki-Goldfinger J, Schalm O, Janssens K (2008) Microchim Acta 162:415–424
De Raedt I, Janssens K, Veeckman J, Vincze L, Vekemans E, Jeffries T (2001) J Anal At Spectrom 16:1012–1017
Falcone R, Sommariva G, Verità M (2006) Microchim Acta 155:137–140
Mass J, Ammerman A, Hunt JA (2001) Amer Lab 4:52–55
Hall ME, Yablonsky L (1998) J Archaeol Sci 25:1239–1245
Arletti R, Maiorano C, Ferrari D, Vezzalini G, Quartieri S (2010) J Archaeol Sci 37:703–712
Kunicki-Goldfinger J, Pańczyk E, Dzierżanowski P, Waliś L (2008) J Radioanal Nucl Chem 278:307–311
Šmit Ž, Pelicon P, Vidmar G, Zorko B, Budnar M, Demortier G, Gratuze B, Sturm S, Necemer M, Kump P, Kos M (2000) Nucl Instrum Methods Phys Res Sect B 161–163:718–723
Mäder M, Jembrih-Simbürger D, Neelmeijer C, Schreiner M (2005) Nucl Instrum Methods Phys Res Sect B 239:107–113
Calligaro T (2008) X-Ray Spectrom 37:169–177
Šmit Ž, Janssens K, Bulska E, Wagner B, Kos M, Lazar I (2005) Nucl Instrum Methods Phys Res Sect B 239:94–99
Günther D (2002) Anal Bioanal Chem 327:31–32
Gratuze B, Blett-Lemarquand M, Barrandon J-N (2001) J Radioanal Nucl Chem 247:645–656
Giussani B, Monticelli D, Rampazzi L (2009) Anal Chim Acta 635:6–21
Resano M, Garcia-Ruiz E, Vanhaecke F (2010) Mass Spectrom Rev 29:55–78
Dussubieux L, Kusimba CM, Gogte V, Kusimba SB, Gratuje B, Oka R (2008) Archaeometry 50:797–821
Dussubieux L, Robertshaw P, Glascock MD (2009) Int J Mass Spectrom 284:152–161
van Elteren JT, Tennent NH, Selih VS (2009) Anal Chim Acta 644:1–9
Rbertshaw P, Benco N, Wood M, Dussubieux L, Melchiorre E, Ettahiri A (2010) Archaeometry 52:355–279
Shortland A, Rogers N, Eremin K (2007) J Archaeol Sci 34:781–789
Brill RH (1972) A chemical–analytical round-robin on four synthetic ancient glasses. The Corning Museum of Glass. Corning, New York
Vicenzi EP, Eggins S, Logan A, Wysoczanski R (2002) J Res Natl Inst Stand Technol 107:719–727
Jackson CM, Nicholson PT (2010) J Archaeol Sci 37:295–301
Aerts A, Velde B, Janssens K, Dijkman W (2003) Spectrochim Acta Part B 58:659–667
Schalm O, Caluwe D, Wouters H, Janssens K, Verhaeghe F, Pieters M (2004) Spectrochim Acta Part B 59:1647–1656
Kuisma-Kursula P (2000) X-Ray Spectrom 29:111–118
Kuisma-Kursula P, Raisanen J (1999) Archaeometry 41:71–79
Bronk H, Freestone IC (2001) Archaeometry 43:517–527
Falcone R, Renier A, Verità M (2002) Archaeometry 44:531–542
Mass JL, Wypyski MT, Stone RE (2002) Archaeometry 44:67–82
Wolf S, Kessler CM, Stern WB, Gerber Y (2005) Archaeometry 47:361–380
Robertshaw P, Wood M, Melchiorre E, Popelka-Filcoff RS, Glascock MD (2010) J Archaeol Sci 37:1898–1912
Shortland AJ, Eremin K (2006) Archaeometry 48:581–603
Longerich HP, Jackson SE, Günther D (1996) J Anal At Spectrom 11:899–904
Jackson SE (2008) In: Sylvester P (ed) Laser ablation ICP–MS in the earth sciences: current practices and outstanding issues. Mineralogical Association of Canada, Vancouver
Horn I, Guillong M, Günther D (2001) Appl Surf Sci 182:91–102
Koch J, Lindner H, von Bohlen A, Hergenroder R, Niemax K (2005) J Anal At Spectrom 20:901–906
Wang Z, Hattendorf B, Günther D (2006) J Am Soc Mass Spectrom 17:641–651
Shaheen M, Gagnon JE, Yanga Z, Fryer BJ (2008) J Anal At Spectrom 23:1610–1621
Kuhn H-R, Günther D (2004) J Anal At Spectrom 19:1158–1164
Guillong M, Horn I, Günther D (2003) J Anal At Spectrom 18:1224–1230
Glaus R, Kaegi R, Krumeich F, Günther D (2010) Spectrochim Acta B 65:812–822
Kroslakova I, Günther D (2007) J Anal At Spectrom 22:51–62
Acknowledgments
Financial support for the research project N N204 241734, from the Ministry of Science and High Education of Poland, is gratefully acknowledged.
Open Access
This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://creativecommons.org/licenses/by-nc/2.0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.