Introduction

Osteoporosis is a bone disorder that affects millions of people worldwide. It is characterized by an imbalance in bone resorption and formation rates [1, 2], resulting in low bone mass and increased fracture risk. Approximately 50% of age-related vertebral fractures are believed to be spontaneous fractures, resulting from daily activities or from cyclic loading, rather than from trauma [3, 4].

Bisphosphonates are often used to treat osteoporotic patients. They inhibit bone resorption and thereby slow down the process of bone loss, maintaining bone mass, microstructure and strength in relevant anatomical sites like the femur and vertebra, in animals as well as in humans [57]. Importantly, fracture risk is significantly reduced in osteoporotic patients treated with bisphosphonates [6, 810]. Zoledronic acid is a potent, relatively new bisphosphonate that recently has been shown to significantly reduce fracture risk in osteoporotic patients who received once-yearly doses [11]. We have previously shown that zoledronic acid both inhibited and recovered loss of bone structure and strength in ovariectomized rat tibiae and vertebrae resulting in values that were similar to control levels [12, 13].

Bisphosphonates have also been shown to influence the degree of mineralization of bone tissue due to decreased bone turnover rates and the subsequent prolongation of secondary mineralization [14, 15], which may lead to more brittle mechanical behavior [1619]. Crystallinity of bone tissue has been shown to influence monotonic and fatigue mechanical properties in human cortical bone [20]. Microcracks and diffuse damage are commonly seen in human bone [2123] and may act as a stimulus for bone remodeling [24]. Studies in dogs have shown that low resorption rates induced by bisphosphonates lead to accumulation of microcracks and diffuse damage [25]. It is unknown whether these increases in mineralization and microdamage resulting from bisphosphonates influence the mechanical properties of bone when cyclically loaded.

Compressive and tensile fatigue behavior has been well documented for cortical bone from humans as well as animals [2629]. More recently, the fatigue behavior of trabecular bone in animals and humans has been found to exhibit similar characteristics as cortical bone [3033]. Although these studies have provided fundamental information regarding bone fatigue behavior, the integral function of cortical and trabecular bone, i.e., the way they act together, which plays an important role in the vertebra, has not yet been determined. Moreover, drug efficacy studies in rats generally focus on changes in bone mass, structure, and static mechanical strength, whereas fatigue behavior, which may play an important role in vertebral fractures, may respond differently to pharmacologic intervention than other statically determined mechanical parameters.

Our primary aim was to develop an experimental approach to determine compressive fatigue mechanical properties in whole rat vertebra. We then used this method to compare fatigue properties in ovariectomized rats treated with zoledronic acid to SHAM, ovariectomized controls, which exhibited similar structural and static, compressive properties.

Materials and methods

Seventeen female, 35-week-old, Wistar rats were used from a previous study described elsewhere [12]. At week 0, eight rats were ovariectomized (OVX-ZOL), and nine rats were SHAM-ovariectomized (SHAM-OVX). Zoledronic acid was kindly provided as the disodium salt hydrate by Novartis Pharma AG (Basel, Switzerland) and was dissolved in a saline vehicle prior to injection. It was administered at a single dose of 20 μg/kg body weight s.c. at the time of OVX to all rats of the OVX-ZOL group. The dose was chosen based on a dose–response study in rats, in which 20 μg/kg body weight was found to be most effective [34]. Rats were humanely sacrificed 16 weeks later, and whole L4 vertebrae were dissected, soaked in 0.9% saline solution gauze, and frozen at −20°C. This study was approved by the Animals Ethics Committee of the University of Maastricht, the Netherlands.

Micro-CT scanning

Vertebrae were thawed to room temperature and scanned with a desktop micro-CT system (microCT40, Scanco Medical AG, Bruettisellen, Switzerland) at an isotropic resolution of 16 μm (55 kV, 145 μA, 500 projections per 180°, 200 ms integration time). After scanning, samples were frozen again until mechanical testing. Images were Gaussian filtered (sigma = 0.8, support = 1 voxel) and binarized to separate bone from background using a global thresholding procedure [35].

From the CT scans, the trabecular region was manually selected starting ten slices below the cranial growth plate and ending ten slices above the caudal growth plate, resulting in a trabecular region of approximately 5 mm in axial direction. From this region, six bone structural parameters (bone volume fraction (BV/TV), connectivity density (Conn.D), structure model index (SMI), trabecular number (Tb.N), trabecular thickness (Tb.Th), and separation (Tb.Sp)) were automatically determined. Cortical bone was semi-automatically delineated from the CT scans by drawing contour lines, using the same set of slices as used for trabecular bone measurements.

Specimen preparation

To achieve plano-parallel ends, vertebrae were fixed in a custom-made jig. A double-blade, wafering, low-speed diamond saw (Isomet, Buehler, Lake Bluff, IL, USA) was used under constant saline irrigation to remove cranial and caudal ends including the growth plate. After sawing, the exact vertebral height was measured using a caliper and found to be 4.06 ± 0.09 mm (mean ± SD). An example of a processed vertebra can be seen in Fig. 1. A single-blade, wafering, low-speed diamond saw was used under constant saline irrigation to remove all posterior pedicles and processes. Anterior elements were clipped off using a rounger, resulting in a separated vertebral body. CT scans taken for pilot samples had shown no splintering resulted from sawing and clipping. Vertebrae were kept frozen in a 0.9% saline solution until fatigue testing.

Fig. 1
figure 1

Schematic of fatigue loading test. The lower platen, designed as a cup, contained the vertebra. The top platen, smaller in diameter than the cup, was lowered onto the vertebra to a compressive preload of 5 N, at which point the displacement was set at zero. A 0.9% saline solution containing protease inhibitors was added to the cup to prevent the vertebra from dehydrating and to inhibit microorganism growth

Fatigue compression tests

Vertebrae were thawed to room temperature prior to mechanical testing. In total, all samples were frozen and thawed for three cycles. Previously, five cycles of freezing and thawing has been found not to affect mechanical properties determined in a static, compression [36], and indentation test [37]. Therefore, we assumed that fatigue properties determined in our study were not affected by the freezing and thawing. L4-vertebral bodies were used for fatigue compression tests, which were conducted in load-control at 25°C. A custom-made, steel lower plate was designed in the form of a cup (Fig. 1). The vertebral body was placed with the cranial end facing upward in the lower cup of the testing machine (Instron 8874, Instron Corp. Canton, MA, USA). The top platen, smaller in diameter than the cup, was lowered onto the vertebra to a compressive preload of 5 N, at which point the displacement was set at zero. Displacement was measured from the actuator displacement transducer of the testing machine. A 0.9% saline solution containing protease inhibitors was added to the cup to prevent the vertebra from dehydrating and to inhibit microorganism growth. Since bone is known to fail at a certain strain rather than at a certain load or stress [38, 39] and since our aim was to compare the fatigue properties at the tissue rather than structural level between the two groups, all tests were started at the same apparent strain. In a pilot study, the relation between initial apparent strain and number of cycles to failure was studied. Thirteen samples were tested between 0.6% and 0.94% initial apparent strain. A significant correlation (r 2 = 0.48, p = 0.009) was found between the log of strain and log of number of cycles, which corresponds to typical fatigue behavior [27, 3033]. It was found that 0.75% initial apparent strain resulted in a reasonable number of cycles to failure (average number of cycles ∼40,000), and therefore, this value was used in all tests. Since the stiffness varied per sample and the test was run in load-control, the load needed to reach the initial apparent strain criterion varied as well. Therefore, prior to testing, each sample was cyclically loaded for about 400 cycles with increasing load until the load was reached at which the desired apparent strain was met. The maximum load ranged from 63 to 97 N.

During the test, load cycled between 5 N and the determined maximum load in a sinusoidal shape at a frequency of 2 Hz. Tests were ended after the sample failed, which was characterized firstly by an increasing displacement range per cycle, increased hysteresis, and increased total apparent strain. Then, the sample could not bear the loads anymore and was crushed. A full load–displacement cycle could not be reached anymore, after which the test was stopped. Tests were stopped after 120,000 cycles if failure had not occurred. Every fourth cycle, force and displacement were acquired during one cycle at a sampling frequency of 100 Hz.

For each sample, creep characteristics exhibited three classical phases: an initial phase of high creep rate, a phase of a lower creep rate, and a phase in which creep rate was high again, finally resulting in failure (Figs. 2 and 3) [33, 40]. From each apparent strain versus time curve, the steady-state creep rate of the secondary phase was determined by fitting a linear line through the central part of the curve. According to the method of Bowman et al. [33], a line parallel to this line was drawn at 0.5% higher offset. The intersection of this line with the apparent strain curve was defined as the time to failure and the apparent strain at failure. Young’s modulus could not be properly calculated, as the effective area in a vertebra that contains trabecular and cortical bone varies going from cranial to caudal ends. Therefore, secant stiffness was calculated for each recorded cycle by dividing the load range by the displacement range of that cycle. Initial secant stiffness was determined at the start of the experiment, and final secant stiffness was determined at the time of failure. For each sample, time to failure, apparent strain at failure, steady-state creep rate, initial stiffness, and percent loss of stiffness at failure were calculated.

Fig. 2
figure 2

Three representative force–displacement cycles throughout the testing period: 20, 55, and 10,620 cycles for a typical sample. Force–displacement cycles display typical fatigue behavior characterized by decreasing secant stiffness, increasing hysteresis, and increasing nonlinearity. Displacement increases over time due to mostly creep and to a lower extent, a decreasing secant stiffness

Fig. 3
figure 3

Typical sample for which creep characteristics exhibit three typical phases of fatigue: an initial phase of high creep rate, a phase of a steady-state lower creep rate, and a phase in which creep rate is high again, finally resulting in failure [33, 40]. From each apparent strain against time curve, the creep rate of the secondary phase is determined by fitting a linear line. According to the method of Bowman et al. [33], a line parallel to this line is drawn at 0.5% higher offset. The intersection of this line with the apparent strain curve is defined as the time to failure and the strain at failure

Data analysis

Pearson correlation coefficients were used to determine the relation between trabecular bone microarchitecture, cortical thickness, and compressive fatigue properties. For this, all structural properties were correlated with fatigue properties as well as with log-transformed values of the fatigue properties. Also, all structural and fatigue parameters were compared between the two groups using a Student’s t test. p values below 0.05 were considered significant.

Results

During fatigue testing, 12 samples failed between 10 min and 14.7 h (1,200 and 106,000 cycles), and five samples did not fail within the studied period of time. The latter samples showed a decreasing, rather than an increasing, apparent strain range per cycle during the test, accompanied by an increasing secant stiffness, suggesting that artifacts were present in these tests [41]. These samples were subsequently removed from all analyses in the study, resulting in seven samples in the SHAM-OVX and five in the OVX-ZOL group.

Trabecular and cortical microarchitecture

No significant differences were found in trabecular bone microarchitecture and cortical thickness between the SHAM-OVX- and the OVX-ZOL-treated group except for Tb.Th, which was significantly higher in the ZOL-treated group (Table 1). These results indicate that ZOL treatment inhibited bone loss and trabecular deterioration that has previously been shown to occur after ovariectomy [13].

Table 1 Cortical thickness, trabecular bone volume, and trabecular microarchitecture as determined by micro-CT in L4 vertebrae (mean ± SD) from SHAM-OVX and OVX-ZOL rats

Fatigue compression tests

For all failed samples, force–displacement cycles displayed typical fatigue behavior characterized by decreasing secant stiffness, increasing hysteresis, and increasing nonlinearity (Fig. 2). Displacement increased over time due to mostly creep and to a lower extent, decreasing secant stiffness. For each sample, the steady-state creep rate was determined from the apparent strain versus time curve, as well as the time to failure and apparent strain at failure (Fig. 3).

Time to failure, apparent strain at failure, steady-state creep rate, initial stiffness, and percent loss of stiffness at failure were not significantly different between the two groups (Table 2). Steady-state creep rate and log of the time to failure have shown to be inversely linearly correlated in compressive fatigue studies on bovine trabecular bone [32, 33]. Here, we also found a strong inverse correlation between log of the steady-state creep rate and log of the time to failure of all samples taken together (r 2 = 0.86, p < 0.001, Fig. 4). The relationship between steady-state creep rate and time to failure was similar between SHAM-OVX and OVX-ZOL.

Fig. 4
figure 4

Steady-state creep rate plotted against time to failure for all samples on a log–log scale. A significant inverse linear correlation was found between log of the time to failure and log of the steady-state creep rate (r 2 = 0.84, p < 0.001)

Table 2 Compressive fatigue properties determined in L4 vertebrae (mean ± SD) from SHAM-OVX and OVX-ZOL rats

Relation between morphology and fatigue properties

BV/TV, Conn.D, Tb.N, and Tb.Sp each correlated with apparent strain at failure as well as with log of the apparent strain at failure (0.31 < r 2 < 0.50, p < 0.05). All other correlations between morphologic parameters and fatigue properties were not significant (Fig. 5).

Fig. 5
figure 5

A nonsignificant and a significant correlation between, respectively, cortical thickness and apparent strain at failure and BV/TV and apparent strain at failure

Discussion

In this study, we developed a method for assessing the fatigue properties of whole rat vertebrae. Typical fatigue behavior was seen in SHAM-ovariectomized as well as in ZOL-treated, ovariectomized rats. Fatigue properties, trabecular microarchitecture, and cortical thickness were similar in both groups. Previously, we showed that static compressive behavior was also similar in L3 vertebrae of the same groups of rats [13]. Altogether, this suggests that ZOL treatment of ovariectomized rats results in the same vertebral bone mass and structure as SHAM, ovariectomized rats, as well as the same vertebral static and fatigue properties.

For all vertebrae, force–displacement curves displayed typical fatigue behavior characterized by decreasing secant stiffness, increasing hysteresis, and increasing nonlinearity. This agrees with compressive, fatigue behavior previously reported for cortical and trabecular bone specimens [27, 3133]. Also, the strong linear correlation between the log steady-state creep rate and the log time to failure agrees with the literature [32, 33], which indicates the validity of the test. This also indicates that the integral fatigue behavior of cortical and trabecular bone in rats is similar to the two bone compartments assessed separately. We found an average apparent strain at failure of about 4% for both groups, which is just slightly higher than the 3.4% and 2.8% reported for, respectively, human and bovine trabecular bone [31, 33].

Samples that did not fail during the test were removed from further analysis and showed a decreasing rather than an increasing apparent strain range per cycle during the test accompanied by an increasing secant stiffness. This behavior suggests that artifacts were present in these tests [41, 42], possibly due to vertebral ends that were not perfectly parallel. In this case, when the force range, leading to 0.75% apparent strain, was determined at the start of the test, the actual area bearing the load would be smaller than the total bone area. During the test, the area bearing the load would then be compressed, resulting in the same load being born by the area of the whole vertebra and thus in lower strains. Improving the sawing procedure and specimen fixation in the loading device could possibly reduce the rate of exclusion of samples.

The fatigue behavior in these whole vertebrae was comparable to the fatigue behavior found in studies on cortical and trabecular bone, though no fatigue data on rat bone are available. Although not determined in our study, it would be interesting to study whether failure starts in the cortical or trabecular bone.

Most of the fatigue properties were unrelated to cortical or trabecular bone morphology, with the exception of weak relationships between trabecular bone morphology and apparent strain at failure. Vertebral morphology and microarchitecture were similar in both groups, and any intrinsic variance in bone mass and stiffness was accounted for in the fatigue test by starting all tests at the same apparent strain level. One might therefore expect that trabecular microarchitecture would not be well correlated with fatigue properties in this test protocol. However, it is possible that despite our normalized test, some types of structure are more favorable over time in a fatigue test than others, which could result in a correlation between a structural parameter and a fatigue property. Additional studies need to be conducted to further delineate the possible relationship between bone microarchitecture and fatigue behavior. Notably, in human trabecular bone, bone volume fraction is weakly correlated with strain at failure, which agrees with our findings [30].

Rather than applying the same load, which will result in low bone mass samples failing earlier than high bone mass samples, we applied the same apparent strain in each test. By developing this normalized fatigue test, we aimed at determining changes in fatigue properties due to differences at the tissue rather than the structural level. The fact that no difference in fatigue behavior was found between both groups indicates that either no changes occurred in the bone tissue fatigue properties or that we were unable to detect them. Increased mineralization that may have taken place in the ZOL group due to lower turnover rate apparently did not lead to detectable changes in fatigue properties of the bone tissue. It may be, however, that a longer treatment period would have led to noticeable changes. Also, no untreated OVX group was included in this study, and therefore, the effects of OVX versus those associated with ZOL treatment cannot be distinguished. Theoretically, it could be that OVX would lead to altered fatigue properties, which could have then been reversed by ZOL resulting in no differences between SHAM-OVX- and ZOL-treated OVX rats. This will need to be tested by additional studies.

In our study, several samples did not fail during the test, which reduced the sample size. Also, between-subject variation was found to be high, which, combined with the small sample size, reduced the power to detect differences between the groups. A power analysis revealed that a scientifically relevant difference of 30% between the two groups in apparent strain at failure would have been detectable if the sample size would have been 22. Therefore, large sample sizes would have been needed to detect any scientifically relevant differences, which were not noted in this study. Also, after starting the test, all samples needed to “settle in”. Thus, strains sometimes decreased or increased slightly directly after starting the test, and this may have affected the time to failure. However, this phenomenon occurred in both groups and, therefore, would not be expected to contribute to group-related differences. The sufficient quality of the fatigue test, however, was reflected in the fact that typical fatigue behavior was seen in terms of force–displacement curves and a strong correlation between steady-state creep rate and log of the time to failure. No observation of substantial differences between groups was found, and this study presents therefore the first indication that ZOL-treated, ovariectomized rats have similar fatigue properties as control rats. More studies are needed to further elucidate the effects of bisphosphonates on fatigue properties. The gained insight in testing limitations will allow for future study designs to be optimized.

In this study, we developed a method to determine compressive fatigue mechanical behavior of whole vertebrae in rats. Fatigue properties of whole rat vertebra exhibited similar characteristics as isolated cortical and trabecular bone specimens. Vertebral morphology, as well as fatigue properties of ZOL-treated ovariectomized rats, were similar to SHAM-OVX rats. These findings indicate that ZOL treatment does not have a pronounced negative influence on cyclic mechanical properties, as might be expected if ZOL-treated bone tissue were more brittle or contained excessive microdamage. The development of this methodology will allow further investigation of the effects of osteoporosis treatments on vertebral compressive fatigue behavior.