Navigated total knee arthroplasty: is it error-free?

  • Knee
  • Published:
Knee Surgery, Sports Traumatology, Arthroscopy Aims and scope

Abstract

Purpose

The aim of this study was to determine whether errors do occur in navigated total knee arthroplasty (TKAs) and to study whether errors in bone resection or implantation contribute to these errors.

Methods

A series of 20 TKAs was studied using computer navigation. The coronal and sagittal alignments of the femoral and tibial cutting guides, the coronal and sagittal alignments of the final tibial implant and the coronal alignment of the final femoral implant were compared with that of the respective bone resections. To determine the post-implantation mechanical alignment of the limb, the coronal alignment of the femoral and tibial implants was combined.

Results

The median deviation between the femoral cutting guide and bone resection was 0° (range −0.5° to +0.5°) in the coronal plane and 1.0° (range −2.0° to +1.0°) in the sagittal plane. The median deviation between the tibial cutting guide and bone resection was 0.5° (range −1.0° to +1.5°) in the coronal plane and 1.0° (range −1.0° to +3.5°) in the sagittal plane. The median deviation between the femoral bone resection and the final implant was 0.25° (range −2.0° to 3.0°) in the coronal plane. The median deviation between the tibial bone resection and the final implant was 0.75° (range −3.0° to +1.5°) in the coronal plane and 1.75° (range −4.0° to +2.0°) in the sagittal plane. The median post-implantation mechanical alignment of the limb was 0.25° (range −3.0° to +2.0°).

Conclusions

When navigation is used only to guide the positioning of the cutting jig, errors may arise in the manual, non-navigated steps of the procedure. Our study showed increased cutting errors in the sagittal plane for both the femur and the tibia, and following implantation, the greatest error was seen in the sagittal alignment of the tibial component. Computer navigation should be used not only to guide the positioning of the cutting jig, but also to check the bone resection and implant position during TKA.

Level of evidence

IV.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

References

  1. Anderson KC, Buehler KC, Markel DC (2005) Computer assisted navigation in total knee arthroplasty: comparison with conventional methods. J Arthroplasty 20(7 Suppl 3):132–138

    Article  PubMed  Google Scholar 

  2. Bathis H, Perlick L, Tingart M, Luring C, Zurakowski D, Grifka J (2004) Alignment in total knee arthroplasty. A comparison of computer-assisted surgery with the conventional technique. J Bone Joint Surg Br 86(5):682–687

    Article  CAS  PubMed  Google Scholar 

  3. Bathis H, Perlick L, Tingart M, Perlick C, Luring C, Grifka J (2005) Intraoperative cutting errors in total knee arthroplasty. Arch Orthop Trauma Surg 125(1):16–20

    Article  CAS  PubMed  Google Scholar 

  4. Biant LC, Yeoh K, Walker PM, Bruce WJ, Walsh WR (2008) The accuracy of bone resections made during computer navigated total knee replacement. Do we resect what the computer plans we resect? Knee 15(3):238–241

    Article  PubMed  Google Scholar 

  5. Bolognesi M, Hofmann A (2005) Computer navigation versus standard instrumentation for TKA: a single-surgeon experience. Clin Orthop Relat Res 440:162–169

    Article  PubMed  Google Scholar 

  6. Bove JC (2010) Computer-assisted total knee arthroplasty: does the tibial component remain at malposition risk? Orthop Traumatol Surg Res 96(5):536–542

    Article  PubMed  Google Scholar 

  7. Catani F, Biasca N, Ensini A, Leardini A, Bianchi L, Digennaro V, Giannini S (2008) Alignment deviation between bone resection and final implant positioning in computer-navigated total knee arthroplasty. J Bone Joint Surg Am 90(4):765–771

    Article  PubMed  Google Scholar 

  8. Chauhan SK, Clark GW, Lloyd S, Scott RG, Breidahl W, Sikorski JM (2004) Computer-assisted total knee replacement. A controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT Protocol). J Bone Joint Surg Br 86(6):818–823

    Article  CAS  PubMed  Google Scholar 

  9. Chauhan SK, Scott RG, Breidahl W, Beaver RJ (2004) Computer-assisted knee arthroplasty versus a conventional jig-based technique: a randomised, prospective trial. J Bone Joint Surg Br 86(3):372–377

    Article  CAS  PubMed  Google Scholar 

  10. Dalury DF, Dennis DA (2005) Mini-incision total knee arthroplasty can increase risk of component malalignment. Clin Orthop Relat Res 440:77–81

    Article  PubMed  Google Scholar 

  11. Ensini A, Catani F, Leardini A, Romagnoli M, Giannini S (2007) Alignments and clinical results in conventional and navigated total knee arthroplasty. Clin Orthop Relat Res 457:156–162

    CAS  PubMed  Google Scholar 

  12. Fehring TK, Odum S, Griffin WL, Mason JB, Nadaud M (2001) Early failures in total knee arthroplasty. Clin Orthop Relat Res 392:315–318

    Article  PubMed  Google Scholar 

  13. Fu Y, Wang M, Liu Y, Fu Q (2012) Alignment outcomes in navigated total knee arthroplasty: a meta-analysis. Knee Surg Sports Traumatol Arthrosc 20(6):1075–1082

    Article  PubMed  Google Scholar 

  14. Hasegawa M, Yoshida K, Wakabayashi H, Sudo A (2013) Cutting and implanting errors in minimally invasive total knee arthroplasty using a navigation system. Int Orthop 37(1):27–30

    Article  PubMed Central  PubMed  Google Scholar 

  15. Hepinstall MS, Ranawat AS, Ranawat CS (2010) High-flexion total knee replacement: functional outcome at one year. HSS J 6(2):138–144

    Article  PubMed Central  PubMed  Google Scholar 

  16. Hufner T, Kendoff D, Citak M, Geerling J, Krettek C (2006) Precision in orthopaedic computer navigation. Orthopade 35(10):1043–1055

    Article  CAS  PubMed  Google Scholar 

  17. Jenny JY, Clemens U, Kohler S, Kiefer H, Konermann W, Miehlke RK (2005) Consistency of implantation of a total knee arthroplasty with a non-image-based navigation system: a case-control study of 235 cases compared with 235 conventionally implanted prostheses. J Arthroplasty 20(7):832–839

    Article  PubMed  Google Scholar 

  18. Kim TK, Chang CB, Kang YG, Chung BJ, Cho HJ, Seong SC (2010) Execution accuracy of bone resection and implant fixation in computer assisted minimally invasive total knee arthroplasty. Knee 17(1):23–28

    Article  PubMed  Google Scholar 

  19. Koyonos L, Stulberg SD, Moen TC, Bart G, Granieri M (2009) Sources of error in total knee arthroplasty. Orthopedics 32(5):317

    Article  PubMed  Google Scholar 

  20. Laskin RS, Beksac B (2006) Computer-assisted navigation in TKA: where we are and where we are going. Clin Orthop Relat Res 452:127–131

    Article  PubMed  Google Scholar 

  21. Lewold S, Knutson K, Lidgren L (1993) Reduced failure rate in knee prosthetic surgery with improved implantation technique. Clin Orthop Relat Res 287:94–97

    PubMed  Google Scholar 

  22. Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R (2009) Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplasty 24(4):570–578

    Article  PubMed  Google Scholar 

  23. Lotke PA, Ecker ML (1977) Influence of positioning of prosthesis in total knee replacement. J Bone Joint Surg Am 59(1):77–79

    CAS  PubMed  Google Scholar 

  24. Manzotti A, Cerveri P, De Momi E, Pullen C, Confalonieri N (2010) Relationship between cutting errors and learning curve in computer-assisted total knee replacement. Int Orthop 34(5):655–662

    Article  PubMed Central  PubMed  Google Scholar 

  25. Nakahara H, Matsuda S, Moro-oka TA, Okazaki K, Tashiro Y, Iwamoto Y (2012) Cutting error of the distal femur in total knee arthroplasty by use of a navigation system. J Arthroplasty 27(6):1119–1122

    Article  PubMed  Google Scholar 

  26. Otani T, Whiteside LA, White SE (1993) Cutting errors in preparation of femoral components in total knee arthroplasty. J Arthroplasty 8(5):503–510

    Article  CAS  PubMed  Google Scholar 

  27. Plaskos C, Hodgson AJ, Inkpen K, McGraw RW (2002) Bone cutting errors in total knee arthroplasty. J Arthroplasty 17(6):698–705

    Article  PubMed  Google Scholar 

  28. Ritter MA, Davis KE, Meding JB, Pierson JL, Berend ME, Malinzak RA (2011) The effect of alignment and BMI on failure of total knee replacement. J Bone Joint Surg Am 93(17):1588–1596

    Article  PubMed  Google Scholar 

  29. Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM (2002) Insall Award paper. Why are total knee arthroplasties failing today? Clin Orthop Relat Res 404:7–13

    Article  PubMed  Google Scholar 

  30. Slover JD, Tosteson AN, Bozic KJ, Rubash HE, Malchau H (2008) Impact of hospital volume on the economic value of computer navigation for total knee replacement. J Bone Joint Surg Am 90(7):1492–1500

    Article  PubMed  Google Scholar 

  31. Sparmann M, Wolke B, Czupalla H, Banzer D, Zink A (2003) Positioning of total knee arthroplasty with and without navigation support. A prospective, randomised study. J Bone Joint Surg Br 85(6):830–835

    CAS  PubMed  Google Scholar 

  32. Yau WP, Chiu KY (2008) Cutting errors in total knee replacement: assessment by computer assisted surgery. Knee Surg Sports Traumatol Arthrosc 16(7):670–673

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgments

We would like to thank Dr Khin Lay Wai, Saw Swee Hock School of Public Health, National University of Singapore for assisting us in our statistical analysis.

Conflict of interest

No authors have signed any agreement with a commercial interest related to this study which would in any way limit publication of any and all data generated for the study or to delay publication for any reason. One of the authors (LK) was previously a consultant for Stryker Orthopaedics.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Krishna Lingaraj.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Chua, K.H.Z., Chen, Y. & Lingaraj, K. Navigated total knee arthroplasty: is it error-free?. Knee Surg Sports Traumatol Arthrosc 22, 643–649 (2014). https://doi.org/10.1007/s00167-013-2641-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00167-013-2641-6

Keywords

Navigation