Abstract
Objective: To determine whether infusion line compliance contributes to irregular drug delivery during vertical displacement of syringe pumps.¶Design: Five different commercially available infusion lines were studied at infusion rates of 0.5, 1.0, and 1.5 ml/h. Zero drug delivery time was measured after acute line loop formation (70 cm) using an electronic balance. Compliance of each infusion line was calculated using a pressure transducer and measurement of the occlusion release bolus at 300 mmHg occlusion pressure. Finally, the influence of infusion line compliance on drug delivery during acute lowering of the syringe pump was studied using low- and high-compliance infusion lines.¶Results: Acute line loop formation resulted in zero drug delivery time from 5.1 ± 1.5 to 44.0 ± 6.8 s at flow rates of 0.5 ml/h. Increased flow rates significantly reduced loop-induced flow variability. A close correlation was found between zero drug delivery time and calculated infusion line compliance at 0.5 ml/h (linear regression R 2 = 0.79). Lowering of the syringe pump 50 cm prolonged zero drug delivery time from 295.8 ± 20.7 s with the low-compliance tube to 463.3 ± 24.0 s with the high-compliance infusion line.¶Conclusions: Infusion line compliance contributes to irregular drug delivery associated with vertical displacement of syringe pumps. Siphoning of the infusion line during patient care should be avoided, and flow rates of 1 ml/h or higher are recommended. Low-compliance infusion lines are indicated whenever highly short-acting vasoactive drugs at low delivery rates are administered.
Similar content being viewed by others
Author information
Authors and Affiliations
Additional information
Received: 8 September 1999/Final revision received: 3 January 2000/Accepted: 3 January 2000
Rights and permissions
About this article
Cite this article
Weiss, M., Bänziger, O., Neff, T. et al. Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med 26, 776–779 (2000). https://doi.org/10.1007/s001340051246
Issue Date:
DOI: https://doi.org/10.1007/s001340051246