It is common opinion that the elastic lung is enclosed within another elastic structure, the chest wall, and that both require pressure to be expanded. However, this is not true. An isolated lung will collapse to airlessness, but a lung inside the chest will not. It will stay open at functional residual capacity, FRC, which is the volume at which the inward elastic forces of the lung are balanced by the outward elastic forces of the chest wall [1]. Thus, the chest wall acts to expand the lung. This balance between lung and chest wall results in a negative pleural pressure (negative in relation to atmospheric pressure) that is more negative in the upper, non-dependent region than in lower, dependent ones, because of the weight of the lung and possible shape differences between lung and rib cage [2]. Lung elastance decreases with increasing age because of loss of elastic tissue, and pleural pressure may become positive in the dependent part of the pleural space. Airways will close and alveoli may eventually be airless. During anesthesia, FRC is reduced by half a liter, presumably by reduced outward force by the rib cage because of loss of respiratory muscle tone [3]. This also promotes airway closure and atelectasis formation. However, if the lung or part of it is aerated at resting condition, i.e., at an alveolar pressure of 0 cm H2O (FRC), then extra-luminal pressure, assumed to be reflected by pleural pressure at the level of the aerated lung, must be negative. And the other way around, with a higher extra-luminal than intra-luminal pressure, alveoli will collapse. Airways in the same region will also close and may delay the collapse but atelectasis will eventually occur [3]. This holds true whether the subject is breathing spontaneously or is connected to a ventilator, and irrespective of body position. In other words, to stay open, a lung region must be exposed to a positive transpulmonary pressure (airway pressure minus pleural pressure). Pleural pressure must be negative if the subject is breathing at atmospheric pressure, while it may be positive to a varying degree if the subject is breathing or ventilated at positive airway pressure.
Will the reasoning above also apply to a sick lung? Let us consider acute respiratory distress syndrome (ARDS), where lung volume is further reduced (“baby lung”). Lung elastance is increased and this should make pleural pressure more negative, but overall volume including tissue may instead increase or remain unaltered, and the net effect on pleural pressure will therefore be less predictable. The chest wall may also be affected in ARDS, becoming heavier, with potential impact on lung volume and pleural pressure (“extrapulmonary ARDS”) [4]. Still, if part of the lung is aerated at FRC (resting condition with zero airway pressure), then pleural pressure at the level of the aerated lung must be negative. Similar reasoning has been presented earlier [5].
During spontaneous breathing, pleural pressure reflects the force applied to expand the lung, and during mechanical ventilation, the same pressure reflects any force applied to expand the chest wall. The chest wall is made up by the rib cage and the diaphragm. At FRC, the rib cage acts as an outward force, and the resting position of the rib cage is well above FRC [1, 3]. This force generates the negative pleural pressure. Thus, at end-expiration, the diaphragm prevents the abdominal organs from encroaching on the thoracic space and influencing the lung in the supine/prone position. It is exposed to the pressure generated by the weight of the abdominal organs, a weight that does not change to any substantial degree when lung volume changes [6]. The abdomen can be considered a partly fluid-filled container, and its content creates a pressure that increases down the abdomen in the gravitational direction, from close to zero in the anterior part to 10–30 cm H2O in the posterior part, depending on vertical distance and water content. A passive (mechanically controlled) inflation of the lung will move the thoracic wall outwards and the diaphragm caudally, changing the shape of the abdomen and its content. At end-inspiration, when abdominal movement has ceased, any additional pressure to maintain the new lung volume will depend (disregarding the lung) on abdominal tissue characteristics and vertical tissue displacement. There is no reason why any abdominal organ would behave as an elastic tissue; that would require that it is distended during passive lung inflation and causes an increased recoil pressure. The possibility remains that the abdominal wall is distended and recoils but increasing PEEP per se need not increase abdominal pressure [6]. Vertical displacement of abdominal tissue is more likely, but the amount and weight of the displaced tissue is difficult to predict since displacement will initially occur at an iso-gravitational level and even downwards without any increase in end-inspiratory pressure. We erroneously translate the additional pressure needed to store “energy” at a higher vertical level to reflect chest wall elastance. We thus hypothesize that, in the absence of a recoiling abdominal wall and vertical tissue displacement, pleural pressure need not increase with increasing static (no gas flow) airway pressure [7].
The statement of an unaltered pleural pressure may appear provocative and controversial, not least when it is a frequent finding that pleural pressure (or its substitute, esophageal pressure; see below) increases by a passive tidal inflation of the lung [8–10]. One possible answer to the differing observations is that the mechanical behavior of the abdomen (perhaps also the rib cage) is time-dependent. During passive tidal lung inflation, the chest wall is moving and this requires energy (pressure). This adds to the inspiratory airway pressure and to a calculated respiratory resistance. This resistive component remains for a second or more at end-inspiration, keeping pleural pressure elevated, and is traditionally interpreted as reflecting a chest wall elastance. Lu and co-workers found chest wall elastance to be lower the slower the inspiratory flow [11]. We elaborated on their observation by increasing the PEEP step by step in an ARDS patient, staying at each PEEP level for a couple of minutes until the lung volume stabilized at a new level. We measured airway and esophageal pressures and lung volume changes by electric impedance tomography (see Fig. 1). As can be seen, a step change in PEEP caused an immediate and persisting increase in airway pressure. Esophageal pressure, on the other hand, rose initially but then decreased partly or completely to the initial level, while the lung volume increased, somewhat gradually. On the decrease of PEEP, all changes were reversed. Thus, the response by the chest wall to a pressure change is time-dependent. Such an influence by time with decreasing end-inspiratory pressure until a plateau is reached after seconds was noticed long ago and nick-named “creeping” [12]. The time course of PEEP inflation has also been described by Katz et al. [13] and shown to last for minutes. Putensen and co-workers even suggested that it may take up to an hour to obtain a new steady state after increasing PEEP [14].
For practical and safety reasons, it has become common practice to use esophageal pressure as a substitute for pleural pressure to separate lung and chest wall elastance (or its inverse, “compliance”). However, it has been developed and refined in upright subjects and is less reliable in supine subjects [15] with the balloon exposed to many forces that generate pressure [16]. Thus, esophageal pressure is frequently reported as positive at FRC in supine intensive care patients [17, 18]. However, to be representative of pleural pressure at mid-thoracic level, where the lung is normally aerated in most subjects, it should be negative if the subject is breathing at atmospheric pressure, while it may be positive during positive pressure breathing. What always matters is the transpulmonary pressure. Correction for the influence of gravity by subtracting 5 cm H2O has been suggested [18] (see Fig. 2), but it may well be that additional forces on the esophageal balloon raise the baseline level [16]. It may also be that the slope of the pressure–volume curve is correct so that the calculated lung compliance is also correct [19]. This conclusion does not preclude any of our previous discussion regarding pleural pressure and lung volume.
To summarize, in the mechanically ventilated patient:
-
1.
The lung is not enclosed by an elastic chest wall except when the abdomen is severely expanded; vertical shift of abdominal organs may affect pleural pressure but in the absence of such shift, static pleural pressure need not increase with increase in lung volume;
-
2.
Opposing observations on chest wall elastance may be explained by time-dependent behavior of chest wall mechanics;
-
3.
Alveolar pressure must be higher than extra-alveolar pressure to keep the alveoli open; thus at zero alveolar pressure, relative to atmospheric pressure (FRC), extra-alveolar pressure, assumed to be reflected by pleural and esophageal pressure, must be negative at the same vertical level as there is aerated lung, otherwise the lung would be collapsed.
References
Rahn H, Otis AB et al (1946) The pressure-volume diagram of the thorax and lung. Am J Physiol 146(2):161–178
Cherniack RM, Farhi LE, Armstrong BW, Proctor DF (1955) A comparison of esophageal and intrapleural pressure in man. J Appl Physiol 8(2):203–211
Hedenstierna G, Rothen HU (2012) Respiratory function during anesthesia: effects on gas exchange. Compr Physiol 2(1):69–96
Gattinoni L, Pelosi P, Suter PM, Pedoto A, Vercesi P, Lissoni A (1998) Acute respiratory distress syndrome caused by pulmonary and extrapulmonary disease. Different syndromes? Am J Respir Crit Care Med 158(1):3–11
Behazin N, Jones SB, Cohen RI, Loring SH (2010) Respiratory restriction and elevated pleural and esophageal pressures in morbid obesity. J Appl Physiol 108(1):212–218
Jakob SM, Knuesel R, Tenhunen JJ, Pradl R, Takala J (2010) Increasing abdominal pressure with and without PEEP: effects on intra-peritoneal, intra-organ and intra-vascular pressures. BMC Gastroenterol 10:70
Lundin S, Grivans C, Stenqvist O (2015) Transpulmonary pressure and lung elastance can be estimated by a PEEP-step manoeuvre. Acta Anaesthesiol Scand 59(2):185–196
Cortes-Puentes GA, Gard KE, Adams AB, Faltesek KA, Anderson CP, Dries DJ et al (2013) Value and limitations of transpulmonary pressure calculations during intra-abdominal hypertension. Crit Care Med 41(8):1870–1877
Valenza F, Chevallard G, Porro GA, Gattinoni L (2007) Static and dynamic components of esophageal and central venous pressure during intra-abdominal hypertension. Crit Care Med 35(6):1575–1581
Stenqvist O, Grivans C, Andersson B, Lundin S (2012) Lung elastance and transpulmonary pressure can be determined without using oesophageal pressure measurements. Acta Anaesthesiol Scand 56(6):738–747
Lu Q, Vieira SR, Richecoeur J, Puybasset L, Kalfon P, Coriat P et al (1999) A simple automated method for measuring pressure-volume curves during mechanical ventilation. Am J Respir Crit Care Med 159(1):275–282
Van de Woestijne KP (1967) Influence of forced inflations on the creep of lungs and thorax in the dog. Respir Physiol 3(1):78–89
Katz JA, Ozanne GM, Zinn SE, Fairley HB (1981) Time course and mechanisms of lung-volume increase with PEEP in acute pulmonary failure. Anesthesiology 54(1):9–16
Putensen C, Baum M, Hormann C (1993) Selecting ventilator settings according to variables derived from the quasi-static pressure/volume relationship in patients with acute lung injury. Anesth Analg 77(3):436–447
Mead J, Gaensler EA (1959) Esophageal and pleural pressures in man, upright and supine. J Appl Physiol 14(1):81–83
Hedenstierna G (2012) Esophageal pressure: benefit and limitations. Minerva Anestesiol 78(8):959–966
Chiumello D, Cressoni M, Colombo A, Babini G, Brioni M, Crimella F et al (2014) The assessment of transpulmonary pressure in mechanically ventilated ARDS patients. Intensive Care Med 40(11):1670–1678
Talmor D, Sarge T, O’Donnell CR, Ritz R, Malhotra A, Lisbon A et al (2006) Esophageal and transpulmonary pressures in acute respiratory failure. Crit Care Med 34(5):1389–1394
Gattinoni L, Chiumello D, Carlesso E, Valenza F (2004) Bench-to-bedside review: chest wall elastance in acute lung injury/acute respiratory distress syndrome patients. Crit Care 8(5):350–355
Conflicts of interest
On behalf of all authors, the corresponding author states that there is no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Stenqvist, O., Gattinoni, L. & Hedenstierna, G. What’s new in respiratory physiology? The expanding chest wall revisited!. Intensive Care Med 41, 1110–1113 (2015). https://doi.org/10.1007/s00134-015-3685-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00134-015-3685-5