Tin foil as bolus material for therapeutic electron beams from the Varian Clinac 2100C/D

  • Scientific Papers
  • Published:
Australasian Physics & Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

Tin foils of sub-millimetre thickness have been investigated as bolus material for therapeutic electron beams from the Varian Clinac 2100C/D linear accelerator. Measurements with ionisation chamber and radiographic film in Plastic Water or water were performed under tin foil bolus to determine surface dose, therapeutic ranges, output factor correction, penumbra and dose outside the field edge. Appropriate thicknesses of tin foil for 90% dose at the surface were found to be approximately 0.3 mm for 6 MeV, and 0.4 mm for 9 MeV and 12 MeV. Enhanced therapeutic interval with tin foil bolus over water-equivalent bolus has previously been reported, but was found not to be evident for 12 MeV and for a small (4 × 4 cm2) 9 MeV field. The penumbra width of fields with tin foil and water-equivalent bolus were found to be within 2 mm, while the doses at 1 cm outside the field edge were within 1.5% of peak dose. Output factor corrections for fields with tin foil were measured as within 2% of unity. Air gaps between the tin foil and phantom surface up to 5 mm were observed to have minimal effect on output correction factor, relative surface dose, and therapeutic range.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Tung, S. S., Shiu, A. S., Starkschall, G., Morrison, W. H. and Hogstrom, K. R.,Dosimetric evaluation of total scalp irradiation using a lateral electron-photon technique, Int. J. Radiat. Oncol. Biol. Phys. 27:153–60, 1993.

    CAS  PubMed  Google Scholar 

  2. Humphries, S. M., Boyd, K., Cornish, P. and Newman, F. D.,Comparison of Super stuff and Paraffin wax bolus in radiation therapy of irregular surfaces, Medical Dosimetry. 21: 155–157, 1996.

    Article  CAS  PubMed  Google Scholar 

  3. Chang, F., Chang, P., Benson, K. and Share, F.,Study of elasto-gel pads used as surface bolus material in high energy photon and electron therapy, Int. J. Radiat. Oncol. Biol. Phys. 22:191–3, 1992.

    CAS  PubMed  Google Scholar 

  4. Alasti, H. and Galbraith, D. M.,Depth dose flattening of electron beams using a wire mesh bolus, Med. Phys. 22:1675–83, 1995.

    Article  CAS  PubMed  Google Scholar 

  5. Cederbaum, A., Rarkin, A., Rosenblatt, E. and Gez, E.,Implementing a Tantalum wire mesh to increase the skin dose in low energy irradiation of the chest wall, Medical Dosimetry 26: 275–279, 2001.

    Article  CAS  PubMed  Google Scholar 

  6. Moyer, R. F., King, G. A. and Hauser, J. F.,Lead as surface bolus for high-energy photon and electron therapy, Med. Phys. 13:263–269, 1986.

    Article  CAS  PubMed  Google Scholar 

  7. Lambert, G. D., Richmond, N. D., Kermode, R. H. and Porter, D. J. T.,The use of high density metal foils in low-energy clinical electron beams, Radiotherapy and Oncology 53:161–166, 1999.

    Article  CAS  PubMed  Google Scholar 

  8. Tello, V. M., Tailor, R. C. and Hanson, W. F.,How water equivalent are water-equivalent solid materials for output calibration of photon and electron beams? Med. Phys. 22:1177–1189, 1995.

    Article  CAS  PubMed  Google Scholar 

  9. IAEA TRS 398,Absorbed dose determination in external beam radiotherapy, International Atomic Energy Agency, Vienna, 2000.

    Google Scholar 

  10. IAEA TRS 381,The use of plane parallel ionization chambers in high energy electron and photon beams, International Atomic Energy Agency, Vienna, 1997.

    Google Scholar 

  11. ICRU Report 35,Radiation dosimetry: electron beams with energies between 1 and 50 MeV, International commission on radiation units and measurements, Bethesda, 1984.

    Google Scholar 

  12. Khan, F. M. et al.,Clinical electron-beam dosimetry: report of the AAPM Radiation Therapy Committee Task Group No. 25, Med. Phys. 18:73–109, 1991.

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to B. J. Healy.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Healy, B.J., Padmanabhan, P. & Nitschke, K.N. Tin foil as bolus material for therapeutic electron beams from the Varian Clinac 2100C/D. Australas. Phys. Eng. Sci. Med. 28, 8 (2005). https://doi.org/10.1007/BF03178858

Download citation

  • Received:

  • Accepted:

  • DOI: https://doi.org/10.1007/BF03178858

Key words

Navigation