Acceptance testing and commissioning of Kodak Directview CR-850 digital radiography system

  • Technical Note
  • Published:
Australasian Physics & Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

This Technical Paper describes Acceptance Testing and Commissioning of the Kodak DirectView CR-850 digital radiography system installed at the Royal Adelaide Hospital. The first of its type installed in Australia, the system is a “dry” image processor, for which no chemicals are required to develop images. Rather, latent radiographic images are stored on photostimulable phosphor screens, which are scanned and displayed by a reader unit. The image can be digitally processed and enhanced before it is forwarded to a storage device, printer or workstation display, thereby alleviating the need to re-expose patients to achieve satisfactory quality images. The phosphor screens are automatically erased, ready for re-use. Results are reported of tests carried out using the optional “Total Quality Tool” quality assurance package installed with the system. This package includes analysis and reporting software which provides for simple testing and reporting of many important characteristics of the system, such as field uniformity, aspect ratio, line and pixel positions, image and system noise, exposure response, scan linearity, modulation transfer function (MTF) and image artefacts. Acceptance Tests were performed for kV and MV exposures. Resolution for MV exposures was at least 0.8 l/mm, and measured phantom dimensions were within 1.05% of expected magnification. Reproducibility between cassettes was within 1.6%. The mean pixel values on the central axis were close to linear for MV exposures from 3 to 10 MU and reached saturation level at around 20 MU for 6 MV and around 30 MV for 23 MV beams. Noise levels were below 0.2 %.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Seibert, J.A.,Computed Radiography Overview, 84th Scientific Assembly and Annual Meeting, Radiological Society of North America, Chicago, November, 1998.

  2. Samei E.et al, Performance Evaluation of computed radiography systems, Medical Physics 28 No. 3, 2001, 361–371, 2001.

    Article  CAS  PubMed  Google Scholar 

  3. Kodak publication no. 3E4043 Rev C, 11 Nov 03:Kodak DirectView Total Quality Tool — Theory of Operation. 4. Kodak publication no. 3E4025 Rev C, 11 Nov 03:Kodak DirectView Total Quality Tool User Guide.

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to E. Bezak.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Bezak, E., Nelligan, R.A. Acceptance testing and commissioning of Kodak Directview CR-850 digital radiography system. Australas. Phys. Eng. Sci. Med. 29, 30 (2006). https://doi.org/10.1007/BF03178825

Download citation

  • Received:

  • Accepted:

  • DOI: https://doi.org/10.1007/BF03178825

Key words

Navigation