CT-ED conversion on a GE Lightspeed-RT scanner: influence of scanner settings

  • Technical Report
  • Published:
Australasian Physics & Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

The influence of tube voltage (kV) and current (mA) on the resulting relationship of computed tomography number to electron density (CT-ED) was investigated for a wide-bore GE scanner. The influence of kV and mA scan settings were examined in combination with a 16-bit image reconstruction algorithm made available via the scanner software and which allowed resolution of CT numbers for high density materials. By using titanium and stainless steel inserts in an electron density phantom, mA variation was found to have minimal impact on the CT-ED relationship, whereas variation in kV led to significant differences in CT number for the high density materials. The scanner is also equipped with automatic tubecurrent modulation capabilities. The influence of automatic tube-current modulation on CT number was investigated for a range of materials in a phantom geometry. It was found that tube current modulation has negligible effect on CT number, though the changing dimension of the phantom did influence CT number of an aluminium insert for scans undertaken with both fixed and modulated tube currents. In light of evidence from other studies examining the influence of CT number on dose calculation, it is recommended that scanner settings and specific CT-ED look-up tables be considered when calculations will be required with high-density materials present.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Thomas, S.J.,Relative electron density calibration of CT scanners for radiotherapy treatment planning, The British Journal Of Radiology, 72(860): 781–786, 1999.

    CAS  PubMed  Google Scholar 

  2. Metcalfe, P., Kron, T. and Hoban, P.,The Physics of Radiotherapy X-rays from Linear Accelerators, Medical Physics Publishing, Madison WI, 1997.

    Google Scholar 

  3. Schneider, U., Pedroni, E. and Lomax, A.,The calibration of CT Hounsfield units for radiotherapy treatment planning, Physics In Medicine And Biology, 41(1): 111–124, 1996.

    Article  CAS  PubMed  Google Scholar 

  4. Coolens, C. and Childs, P.J.,Calibration of CT Hounsfield units for radiotherapy treatment planning of patients with metallic hip prostheses: the use of the extended CT-scale, Physics in Medicine & Biology, 48(11): 1591–603, 2003.

    Article  CAS  Google Scholar 

  5. Keall, P.J., Chock, L.B., Jeraj, R., Siebers, J.V. and Mohan, R.,Image reconstruction and the effect on dose calculation for hip prostheses, Medical Dosimetry, 28(2): 113–7, 2003.

    Article  PubMed  Google Scholar 

  6. Yazdi, M., Gingras, L. and Beaulieu, L.,An adaptive approach to metal artifact reduction in helical computed tomography for radiation therapy treatment planning: experimental and clinical studies.[erratum appears in Int J Radiat Oncol Biol Phys, 2005 Oct 1;63(2):651 Note: Yazdia, Mehran [corrected to Yazdi, Mehran]], International Journal of Radiation Oncology, Biology, Physics, 62(4): 1224–31, 2005.

    PubMed  Google Scholar 

  7. Gies, M., Kalender, W.A., Wolf, H. and Suess, C.,Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies, Medical Physics, 26(11): 2235–2247, 1999.

    Article  CAS  PubMed  Google Scholar 

  8. Greess, H., Wolf, H., Baum, U., Lell, M., Pirkl, M., Kalender, W.A. and Bautz, W.A.,Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions, European Radiology, 10(2): 391–394, 2000.

    Article  CAS  PubMed  Google Scholar 

  9. Kalender, W.A., Wolf, H. and Suess, C.,Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements, Medical Physics, 26: 2248–2253, 1999.

    Article  CAS  PubMed  Google Scholar 

  10. Kalra, M.K., Maher, M.M., Toth, T.L., Kamath, R.S., Halpern, E.F. and Saini, S.,Comparison of Z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis, Radiology, 232(2): 347–353, 2004.

    Article  PubMed  Google Scholar 

  11. Kalra, M.K., Maher, M.M., Toth, T.L., Kamath, R. S., Halpern, E.F. and Saini, S.,Radiation from “extra” images acquired with abdominal and/or pelvic CT: effect of automatic tube current modulation, Radiology, 232(2): 409–414, 2004.

    Article  PubMed  Google Scholar 

  12. Kalra, M.K., Maher, M.M., Toth, T.L., Schmidt, B., Westerman, B.L., Morgan, H.T. and Saini, S.,Techniques and applications of automatic tube current modulation for CT, Radiology, 233(3): 649–657, 2004.

    Article  PubMed  Google Scholar 

  13. Chu, J.C., Ni, B., Kriz, R. and Amod Saxena, V.,Applications of simulator computed tomography number for photon dose calculations during radiotherapy treatment planning, Radiotherapy & Oncology, 55(1): 65–73, 2000.

    Article  CAS  Google Scholar 

  14. Purdy, J.A.,Photon dose calculations for three-dimensional radiation therapy, Seminars in Radiation Oncology, 2: 235–245, 1992.

    Article  PubMed  Google Scholar 

  15. Biggs, P.J. and Russell, M.D.,Effect of a femoral head prosthesis on megavoltage beam radiotherapy, International Journal of Radiation Oncology, Biology, Physics, 14(3): 581–6, 1988.

    CAS  PubMed  Google Scholar 

  16. Keall, P.J., Siebers, J.V., Jeraj, R. and Mohan, R.,Radiotherapy dose calculations in the presence of hip prostheses, Medical Dosimetry, 28(2): 107–12, 2003.

    Article  PubMed  Google Scholar 

  17. Carolan, M., Dao, P., Fox, C. and Metcalfe, P.,Effect of hip prostheses on radiotherapy dose, Australasian Radiology, 44(3): 290–5, 2000.

    Article  CAS  PubMed  Google Scholar 

  18. Reft, C., Alecu, R., Das, I.J., Gerbi, B.J., Keall, P., Lief, E., Mijnheer, B.J., Papanikolaou, N., Sibata, C. and Van Dyk, J.,Dosimetric considerations for patients with HIP prostheses undergoing pelvic irradiation. Report of the AAPM Radiation Therapy Committee Task Group 63, Medical Physics, 30(6): 1162–1182, 2003.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to M. A. Ebert.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ebert, M.A., Lambert, J. & Greer, P.B. CT-ED conversion on a GE Lightspeed-RT scanner: influence of scanner settings. Australas. Phys. Eng. Sci. Med. 31, 154–159 (2008). https://doi.org/10.1007/BF03178591

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF03178591

Key words

Navigation