Does mechanical pressure on radiochromic film affect optical absorption and dosimetry?

  • Technical Note
  • Published:
Australasian Physics & Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

EBT Gafchromic film, a new high sensitivity radiochromic film has been tested to evaluate if external pressure on the film can affect absorption spectra analysis and thus radiation dosimetry. This question arises from the fact that Gafchromic film is often cut into smaller pieces or to certain shapes for dosimetric analysis using scissors which can apply significant pressure to the sides of the film and small film pieces are placed within a solid phantom at depth which can produce significant pressure on the film if appropriate weight distribution procedures are not performed. As expected, results have shown that films cut by scissors can produce a large increase in OD near the film edge up to 5–10 mm away due to physical damage to the EBT film layers. Films placed within a solid phantom receiving up to 39.5 kg/cm2 pressure showed negligible differences in measured absorption spectra compared with control films subject to no external pressure. This equates to negligible external pressure effects for as much as 44 cm of 30 cm × 30 cm solid water placed on a 1 cm2 area film piece. As such, we recommend based on results herein, that film analysis should be performed with a boundary around every film edge, which can be defined visually based in the film. Also film dosimetry in a phantom can be performed with weights up to 39.5 kg/cm2 (or 44 cm of 30 cm × 30 cm solid water or equivalent) placed on the film without effecting the absorption spectra and thus dosimetry of radiation beams.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Butson, M.J., Yu, K.N., Cheung, T., Metcalfe, P.E., “Radiochromic film for Medical Radiation Dosimetry”, Materials Science & Engineering R: Reports, 41, 61–120, 2003.

    Article  CAS  Google Scholar 

  2. Butson, M.J., Cheung, T., Yu, K.N., “Radiochromic film: the new x-ray dosimetry and imaging tool”, Australasian Physical & Engineering Sciences in Medicine, 27, 230, 2004.

    Article  Google Scholar 

  3. Cheung, T., Butson, M.J., Yu, K.N., “Multilayer Gafchromic film detectors for breast skin dose determinationin vivo”, Physics in Medicine and Biology, 47, N31–37, 2002.

    Article  PubMed  Google Scholar 

  4. Chiu-Tsao ST, Duckworth T, Zhang C, Patel NS, Hsiung CY, Wang L, Shih JA, Harrison LB.Dose response characteristics of new models of GAFCHROMIC films: dependence on densitometer light source and radiation energy. Med Phys. Sep;31(9):2501–8, 2004.

    Article  CAS  Google Scholar 

  5. Mack A, Mack G, Weltz D, Scheib SG, Bottcher HD, Seifert V. High precision film dosimetry with GAFCHROMIC films for quality assurance especially when using small fields. Med Phys. 30(9):2399–409, 2003.

    Article  PubMed  Google Scholar 

  6. Butson, M.J., Cheung, T., Yu, K.N., “Spatial resolution of a stacked radiochromic film dosimeter”, 2001, Radiotherapy and Oncology, 61, 211–213, 2001.

    Article  CAS  PubMed  Google Scholar 

  7. Klassen N., Zwan L., Cygler J., Gafchromic MD-55: investigated as a precision dosimeter.Med phys 24 1924–1934, 1997.

    Article  CAS  PubMed  Google Scholar 

  8. Butson, M.J., Cheung, T., Yu, K.N., “XR type R radiochromic film x-ray energy response”, Physics in Medicine and Biology, 50, N195-N199, 2005.

    Article  PubMed  Google Scholar 

  9. Cheung, T., Butson, M.J., Yu, K.N., “Weak energy dependence of EBT Gafchromic film dose response in the 50 kVp-10 MVp X-ray range”, Applied Radiation and Isotopes, 64, 60–62, 2006.

    Article  PubMed  CAS  Google Scholar 

  10. Butson M, Yu P, Metcalfe P, Effects of readout light sources and ambient light on radiochromic film Phys Med Biol 43 2407–2412, 1998.

    Article  CAS  PubMed  Google Scholar 

  11. Niroomand-Rad A, Blackwell C, Coursey B, Gall K, Galvin J, McLaughlin W, Meigooni A, Nath R, Rodgers J, Soares C Radiochromic film dosimetry: Recommendation of AAPM radiation therapy task group 55.Med Phys 25 (11) 2093–2115, 1998.

    Article  CAS  PubMed  Google Scholar 

  12. ISP Corp web site, www.ispcorp.com, (2005).

  13. Butson, M.J., Cheung, T., Yu, K.N., “Absorption spectra variations of EBT radiochromic film from radiation exposure”, Physics in Medicine and Biology, 50, N135-N140, 2005.

    Article  CAS  PubMed  Google Scholar 

  14. Butson, M.J., Yu, K.N., Cheung, T., Metcalfe, P., “High sensitivity radiochromic film dose comparisons”, Physics in Medicine and Biology, 47, N291-N295, 2002.

    Article  PubMed  Google Scholar 

  15. Cheung, T., Butson, M.J., Yu, K.N., “Post irradiation coloration of Gafchromic EBT radiochromic film”, Physics in Medicine and Biology, 50, N281-N285, 2005.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to M. Butson.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Yu, P.K.N., Butson, M. & Cheung, T. Does mechanical pressure on radiochromic film affect optical absorption and dosimetry?. Australas. Phys. Eng. Sci. Med. 29, 285–287 (2006). https://doi.org/10.1007/BF03178580

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF03178580

Key words

Navigation