Polarity effect on surface dose measurement for an attix parallel plate ionisation chamber

  • Technical Note
  • Published:
Australasian Physics & Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

The effects of chamber polarity have been investigated for the measurement of 6MV and 18MV x-ray surface dose using a parallel plate ionization chamber. Results have shown that a significant difference in measured ionization is recorded between polarities at 6MV and 18MV at the phantom surface. A polarity ratio ranging from 1.062 to 1.005 is seen for 6MV x-rays at the phantom surface for field sizes 5cm × 5cm to 40cm × 40cm when comparing positive to negative polarity. These ratios range from 1.024 to 1.004 for 18MV x-rays with the same field sizes. When these charge reading are compared to the Dmax readings of the same polarity it is found that these polarity effects are minimal for the calculation of percentage dose results with variations being less than 1% of maximum.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Fiorino, C., Mangili, P., Cattaneo, G. M., Calandrino, R.,Polarity effects of ionization chambers used in tbi dosimetry due to cable irradiation, Med Dosim. 25: 121–6, 2000.

    Article  CAS  PubMed  Google Scholar 

  2. Fiorino, C., Cattaneo, G. M., del Vecchio, A., Fusca, M., Longobardi, B., Signorotto, P., Calandrino, R.,Cable-induced effects on plane — parallel ionization chambe measurements in large clinical electron beams, Med Dosim 19: 73–4, 1994

    CAS  PubMed  Google Scholar 

  3. Wickman, G., Holmstrom, T.,Polarity effect in plane-parallel ionization chambers using air or a dielectric liquid as ionization medium, Med Phys 19: 637–40, 1992.

    Article  CAS  PubMed  Google Scholar 

  4. Ramsey, C. R., Spencer, K. M., Oliver, A. L.,Ionization chamber, electrometer, linear accelerator, field size, and energy dependence of the polarity effect in electron dosimetry, Med Phys 26: 214–9, 1999.

    Article  CAS  PubMed  Google Scholar 

  5. Gerbi, B. J.,The response characteristics of a newly designed plane-parallel ionization chamber in high-energy photon and electron beams, Med Phys 20: 1411–5, 1993.

    Article  CAS  PubMed  Google Scholar 

  6. Rawlinson, J. A., Arlen, D., Newcombe, D.,Design of parallel plate ion chambers for build up measurements in megavoltage photon beams, Med Phys 19: 641–8, 1992.

    Article  CAS  PubMed  Google Scholar 

  7. Constantinou, C., Attix, F. H., Paliwal, B. R.,A solid water phantom material for radiotherapy X-ray and gamma ray beam ray calculations, Med. Phys. 9: 436–441 1982.

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to M. J. Butson.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Butson, M.J., Yu, P.K.N. & Cheung, T. Polarity effect on surface dose measurement for an attix parallel plate ionisation chamber. Australas. Phys. Eng. Sci. Med. 26, 84–86 (2003). https://doi.org/10.1007/BF03178463

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF03178463

Key words

Navigation