Commissioning of motorized wedge for the first Equinox-80 telecobalt unit and implementation in the Eclipse 3D treatment planning system

  • Technical Report
  • Published:
Australasian Physics & Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

A new model of the telecobalt unit (TCU), Theratron Equinox-80, (MDS Nordion, Canada) equipped with a single 60 degree motorized wedge (MW), four universal wedges (UW) for 15o, 30o, 45o and 60o have been evaluated. MW was commissioned in Eclipse (Varian, Palo Alto, USA) 3D treatment planning system (TPS). The profiles and central axis depth doses (CADD) were measured with Wellhofer blue water phantom for MW and the measured data was commissioned in Eclipse. These profiles and CADD for MW were compared with UW in a homogeneous phantom generated in Eclipse for various field sizes. The dose was also calculated in the same phantom at 10 cm depth. For the particular MW angle and the respective open and MW beam weights, the dose was measured for a field size of 10 cm* 10 cm in a MEDTEC water phantom at 10 cm depth with a 0.13 cc thimble ion chamber (Scanditronix Wellhofer, Uppsala, Sweden) and a NE electrometer (Nuclear Enterprises, UK). Measured dose with ion chamber was compared with the TPS calculated dose. MW angle verification was also done on the Equinox for four angles (15o, 30o, 45o and 60o). The variation in measured and calculated dose at 10 cm depth was within 2%. The measured and the calculated wedge angles were in good agreement within 2o. The motorized wedges were successfully commissioned in Eclipse for four wedge angles.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Piet-Hein Van Der Giessen,Design, construction, and performance of an asymmetrical collimator for a cobalt unit. Int. J. Radiation Oncology Biol. Phys., 49(1), 265–271, 2001.

    Google Scholar 

  2. Venselaar, J. L. M., Bierhuizen, H. W. J., and Piet-Hein Van Der Giessen,Verification of dose calculations with a treatment planning system for open and wedged fields of a 60 Co unit with a new asymmetric collimator Medical Dosimetry, 26(4), 309–314, 2001.

    Article  CAS  PubMed  Google Scholar 

  3. Wong, S. and Sixel, K. E.An asymmetric half-wedged field technique for the treatment of inferiorly extended head and neck cancers Medical Dosimetry. 20(4), 237–242, 1995.

    Article  CAS  PubMed  Google Scholar 

  4. Phillips Mark H., Parsaei Homayoon and Cho Paul S.Dynamic and omni wedge implementation on an Elekta SL Linac Med. Phys. 27, 1623–1634, 2000.

    Article  CAS  PubMed  Google Scholar 

  5. Large Field Autowedge Functional Description, Elekta Oncology Systems Ltd., Fleming Way, Crawley.

  6. Dai Jianrong, Zhu Yunping and Ji Qing.Optimizing beam weights and wedge filters with the concept of the super-omni wedge Med. Phys. 27, 2757–2762, 2000.

    Article  Google Scholar 

  7. Van Dyk, J, Barnett, R. B, Cygler, J. E., and Shragge, P. C.,Commissioning and quality assurance of treatment planning Computers Int J Radiat Oncol Biol Phys. 26, 261–273, 1993.

    PubMed  Google Scholar 

  8. Fraass, B, Doppke, K. and Hunt, M. American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53:Quality assurance for clinical radiotherapy treatment planning Med Phys. 25, 1773–1829, 1998.

    Article  CAS  PubMed  Google Scholar 

  9. Venselaar J, Goodeweerd H and Mijnheer B.Tolerances for the accuracy of photon beam dose calculations of treatment planning systems Radiother Oncol. 60, 191–201, 2001.

    Article  CAS  PubMed  Google Scholar 

  10. Theratron Equinox-80, External Beam Therapy System, User Manual, MDS Nordion, Canada.

  11. AAPM Report No. 46,Comprehensive Quality Assurance for Radiation Oncology. Report of Task Group No. 40, Radiation Therapy Committee, AAPM, Med. Phys. 21, 1994.

  12. International Electrotechnical Commission (IEC 976) Document.

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to R. A. Kinhikar.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kinhikar, R.A., Sharma, S., Upreti, R. et al. Commissioning of motorized wedge for the first Equinox-80 telecobalt unit and implementation in the Eclipse 3D treatment planning system. Australas. Phys. Eng. Sci. Med. 30, 127–134 (2007). https://doi.org/10.1007/BF03178417

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF03178417

Key words

Navigation