Abstract
A new model of the telecobalt unit (TCU), Theratron Equinox-80, (MDS Nordion, Canada) equipped with a single 60 degree motorized wedge (MW), four universal wedges (UW) for 15o, 30o, 45o and 60o have been evaluated. MW was commissioned in Eclipse (Varian, Palo Alto, USA) 3D treatment planning system (TPS). The profiles and central axis depth doses (CADD) were measured with Wellhofer blue water phantom for MW and the measured data was commissioned in Eclipse. These profiles and CADD for MW were compared with UW in a homogeneous phantom generated in Eclipse for various field sizes. The dose was also calculated in the same phantom at 10 cm depth. For the particular MW angle and the respective open and MW beam weights, the dose was measured for a field size of 10 cm* 10 cm in a MEDTEC water phantom at 10 cm depth with a 0.13 cc thimble ion chamber (Scanditronix Wellhofer, Uppsala, Sweden) and a NE electrometer (Nuclear Enterprises, UK). Measured dose with ion chamber was compared with the TPS calculated dose. MW angle verification was also done on the Equinox for four angles (15o, 30o, 45o and 60o). The variation in measured and calculated dose at 10 cm depth was within 2%. The measured and the calculated wedge angles were in good agreement within 2o. The motorized wedges were successfully commissioned in Eclipse for four wedge angles.
Similar content being viewed by others
References
Piet-Hein Van Der Giessen,Design, construction, and performance of an asymmetrical collimator for a cobalt unit. Int. J. Radiation Oncology Biol. Phys., 49(1), 265–271, 2001.
Venselaar, J. L. M., Bierhuizen, H. W. J., and Piet-Hein Van Der Giessen,Verification of dose calculations with a treatment planning system for open and wedged fields of a 60 Co unit with a new asymmetric collimator Medical Dosimetry, 26(4), 309–314, 2001.
Wong, S. and Sixel, K. E.An asymmetric half-wedged field technique for the treatment of inferiorly extended head and neck cancers Medical Dosimetry. 20(4), 237–242, 1995.
Phillips Mark H., Parsaei Homayoon and Cho Paul S.Dynamic and omni wedge implementation on an Elekta SL Linac Med. Phys. 27, 1623–1634, 2000.
Large Field Autowedge Functional Description, Elekta Oncology Systems Ltd., Fleming Way, Crawley.
Dai Jianrong, Zhu Yunping and Ji Qing.Optimizing beam weights and wedge filters with the concept of the super-omni wedge Med. Phys. 27, 2757–2762, 2000.
Van Dyk, J, Barnett, R. B, Cygler, J. E., and Shragge, P. C.,Commissioning and quality assurance of treatment planning Computers Int J Radiat Oncol Biol Phys. 26, 261–273, 1993.
Fraass, B, Doppke, K. and Hunt, M. American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53:Quality assurance for clinical radiotherapy treatment planning Med Phys. 25, 1773–1829, 1998.
Venselaar J, Goodeweerd H and Mijnheer B.Tolerances for the accuracy of photon beam dose calculations of treatment planning systems Radiother Oncol. 60, 191–201, 2001.
Theratron Equinox-80, External Beam Therapy System, User Manual, MDS Nordion, Canada.
AAPM Report No. 46,Comprehensive Quality Assurance for Radiation Oncology. Report of Task Group No. 40, Radiation Therapy Committee, AAPM, Med. Phys. 21, 1994.
International Electrotechnical Commission (IEC 976) Document.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Kinhikar, R.A., Sharma, S., Upreti, R. et al. Commissioning of motorized wedge for the first Equinox-80 telecobalt unit and implementation in the Eclipse 3D treatment planning system. Australas. Phys. Eng. Sci. Med. 30, 127–134 (2007). https://doi.org/10.1007/BF03178417
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF03178417