Abstract
Extrapolated surface dose measurements have been performed using radiographic film to measure 2-Dimensional maps of skin and surface dose with and without a magnetic deflector device aimed at reducing surface dose. Experiments are also performed using an Attix parallel plate ionisation chamber for comparison to radiographic film extrapolation surface dose analysis. Extrapolated percentage surface dose assessments from radiographic film at the central axis of a 6 MV x-ray beam with magnetic deflector for field size 10 × 10 cm2, 15 × 15 cm2 and 20 × 20 cm2 are 9±3 %, 13±3 % and 16±3 %, these compared to 14±3 %, 19±3 %, and 27±3 % for open fields, respectively. Results from Attix chamber for the same field size are 12±1 %, 15±1 % and 18±1 %, these compared to 16±1 %, 21±1 % and 27±1 % for open fields, respectively. Results are also shown for profiles measured in-plane and cross-plane to the magnetic deflector and compared to open field data. Results have shown that the surface dose is reduced at all sites within the treatment field with larger reductions seen on one side of the field due to the sweeping nature of the designed magnetic field. Radiographic film extrapolation provides an advanced surface dose assessment and has matched well with Attix chamber results. Film measurement allows for easy 2 dimensional dose assessments.
Similar content being viewed by others
References
Nilsson, B. and Brahme, A.,Absorbed dose from secondary electrons in high energy photon beams, Phys. Med. Biol. 24(5): 901–912, 1979.
Petti, P. L., Goodman, M. S., Sisterson, J. M., Biggs, P.J., Gabriel, T.A. and Mohan, R.,Sources of electron contamination for the Clinac-35 25-MV photon beam, Med. Phys. 10(6): 856–861, 1983.
Sjogren, R. and Karlsson, M.,Electron contamination in clinical high energy photon beams, Med. Phys. 23(11):1873–1881, 1996.
Rao, B. M., Prasad, S. G., Parthasaradhi, K., Lee, Y., Ruparel, R. and Garces, R.,Investigations on the near surface dose for three 10 -MV x-ray beam accelerators with emphasis on the reduction of electron contamination, Med. Phys.15 (2):246–249, 1988.
Butson, M. J.,Yu, P., Kan, M., Carolan, M., Young, E., Mathur, J. N. and Metcalfe, P. E.,Skin dose reduction by a clinically viable magnetic deflector, Australasian Physics & Engineering Sciences in Medicine. 20(2):107–110, 1997.
Bose, L. J., Danciu, C., Cheng, C. W., Brugmans, M. J., van der Horst, A., Minken, A. and Mijnheer, B. J.,Interinstitutional variations of sensitometric curves of radiographic dosimetric films, Med. Phys.29(8):1772–1780, 2002.
Constantinou, C., Attix, F. H. and Paliwal, B. R.,A solid water phantom material for radiotherapy X-ray and gamma ray beam ray calculations, Med. Phys. 9(3): 436–441, 1982.
Butson, M. J., Cheung, T., Yu, P. and Currie, M.,Surface dose extrapolation measurements with radiographic film, Phys. Med. Biol. 49: N197–201, 2004.
Butson, M. J., Yu, P. and Metcalfe, P. E.,Extrapolated surface dose measurements with radiochromic film, Med. Phys. 26(3): 485–488, 1999.
Rawlinson, J. A.,Design of parallel plate ion chambers for build up measurements in megavoltage photon beams, Med Phys. 19(3): 641–648, 1992.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Damrongkijudom, N., Butson, M. & Rosenfeld, A. Extrapolated surface dose measurements using a NdFeB magnetic deflector for 6 MV x-ray beams. Australas. Phys. Eng. Sci. Med. 30, 46–51 (2007). https://doi.org/10.1007/BF03178409
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF03178409