Four field prostate treatments are a standard treatment procedure in radiotherapy. Dose in the prostate and rectum region were calculated for 6MV and 18MV photon beams on an anthropomorphic phantom with a collapsed cone convolution method using a 3-D planning system. Validation has been performed with radiographic film and thermoluminescent dosimeters. Results have shown that the pinnacle planning system has accurately modelled doses delivered to a heterogeneous phantom with calculations and measurements agreeing within ±3% over most areas. When treating clinically, considerations such as the volume of bowel gas should be taken into account when planning. A sample of patient CT scans showed that in the absence of a heterogeneity correction, the error in estimated dose through the rectum could be as high as 8% in the presence of large volumes of rectal gas. Considerations, such as whether the patient undergoes another CT scan, the bowel gas volume ignored or assigned a specific density needs to be taken into account and brought to the attention of the radiation oncologists for accurate treatment.
Similar content being viewed by others
References
Zietman, A. L., Chung, C. S., Coen, J. J.Shipley WU.10-year outcome for men with localized prostate cancer treated with external radiation therapy: results of a cohort study, J Urol. 2004 171(1): 210–4.
Gray, C. L., Powell, C. R., Riffenburgh, R. H.Johnstone PA. 20-year outcome of patients with T1-3N0 surgically staged prostate cancer treated with external beam radiation therapy, J Urol. 2001 166(1):116–8.
Freedman, G. M., Hanlon, A. L., Lee, W. R., Hanks, G. E.Young patients with prostate cancer have an outcome justifying their treatment with external beam radiation, J Radiat Oncol Biol Phys. 1996 35(2):243–50.
Mackie, T., Scriminger, J., Battista, J.A convolution method of calculating dose for 15MV x-rays, Med Phys 12 188–196 1985.
Kleck, J. H., Smathers, J. B., Holly, F. E.Myers LT Anthropomorphic radiation therapy phantoms: a quantitative assessment of tissue substitutes, Med Phys. 1990 Sep-Oct;17(5):800–6.
Kron, T., Elliott, A., Wong, T. et al.X-ray surface dose measurements using TLD extrapolation, Med Phys 1993;20:703–711
Suchowerska, N., Hoban, P., Davison, A., Metcalfe, P.,Perturbation of radiotherapy beams by radiographic film: measurements and Monte Carlo simulations, Phys Med Biol 1999 Jul;44(7): 1755–65.
Mersseman, B., De Wagter, C.Characteristics of a commercially available film digitizer and their significance for film dosimetry, Phys Med Biol. 43 1803–1812 1998
Butson, M. J., Yu, K. N., Cheung, T., Metcalfe, P. E.Radiochromic film for Medical Radiation Dosimetry, 2003, Materials Science & Engineering R: Reports, 41, 61–120.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Yu, P.K.N., Cheung, T. & Butson, M.J. Prostate dosimetry in an anthropomorphic phantom. Australas. Phys. Eng. Sci. Med. 27, 60 (2004). https://doi.org/10.1007/BF03178377
Received:
Accepted:
DOI: https://doi.org/10.1007/BF03178377