A 62-year-old patient presented with dizzy spells after her dual chamber pacemaker (Medtronic Enrhythm P1501DR), implanted for complete AV block, had been reprogrammed to deliver antitachycardia therapy (ATP) for paroxysmal atrial tachycardia. Her symptoms were caused by inhibition of ventricular backup pacing during ATP, leading to ventricular asystoles. Inhibition was the result of premature ventricular beats occurring prior to ATP: when ventricular backup pacing is left in the default setting, this pacemaker withholds backup pacing if any of the four preceding events is a sensed event. This case illustrates the possibly hazardous effects of default pacemaker settings, especially in pacemaker-dependent patients. (Neth Heart J 2010;18:323-6.)
Similar content being viewed by others
References
Medtronic EnrhythmTM P1501DR Reference Manual.
Ruiter JH, Barrett MJ, Weteling L, Jansen R. Inhibition of a dual chamber pacemaker by oversensing of far-field atrial depolarisation in a patient after His-bundle ablation. Neth Heart J. 2004;12:543-6.
Herweg B, Barold SS. Ventricular asystole during atrial antitachycardia pacing by implanted cardioverter-defibrillator. Pacing Clin Electrophysiol. 2005;28:432-3.
Navarrete A. Failure to pace: pacemaker malfunction? J Cardiovasc Electrophysiol. 2008;19:100-1.
Author information
Authors and Affiliations
Corresponding author
Additional information
Department of Cardiology, Medical Center Alkmaar, Alkmaar, the Netherlands
Rights and permissions
About this article
Cite this article
van de Sandt, F.M., Jansen, R., Kimman, G.P. et al. Ventricular asystole due to inhibition of ventricular backup pacing by a dual chamber pacemaker: possible pitfalls of default settings . NHJL 18, 323–326 (2010). https://doi.org/10.1007/BF03091784
Issue Date:
DOI: https://doi.org/10.1007/BF03091784