Ventricular asystole due to inhibition of ventricular backup pacing by a dual chamber pacemaker: possible pitfalls of default settings

  • Case report
  • Published:
Netherlands Heart Journal Aims and scope Submit manuscript

Abstract

A 62-year-old patient presented with dizzy spells after her dual chamber pacemaker (Medtronic Enrhythm P1501DR), implanted for complete AV block, had been reprogrammed to deliver antitachycardia therapy (ATP) for paroxysmal atrial tachycardia. Her symptoms were caused by inhibition of ventricular backup pacing during ATP, leading to ventricular asystoles. Inhibition was the result of premature ventricular beats occurring prior to ATP: when ventricular backup pacing is left in the default setting, this pacemaker withholds backup pacing if any of the four preceding events is a sensed event. This case illustrates the possibly hazardous effects of default pacemaker settings, especially in pacemaker-dependent patients. (Neth Heart J 2010;18:323-6.)

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Figure 1
Figure 2
Figure 3

Similar content being viewed by others

References

  1. Medtronic EnrhythmTM P1501DR Reference Manual.

  2. Ruiter JH, Barrett MJ, Weteling L, Jansen R. Inhibition of a dual chamber pacemaker by oversensing of far-field atrial depolarisation in a patient after His-bundle ablation. Neth Heart J. 2004;12:543-6.

    Google Scholar 

  3. Herweg B, Barold SS. Ventricular asystole during atrial antitachycardia pacing by implanted cardioverter-defibrillator. Pacing Clin Electrophysiol. 2005;28:432-3.

    Google Scholar 

  4. Navarrete A. Failure to pace: pacemaker malfunction? J Cardiovasc Electrophysiol. 2008;19:100-1.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to F. M. van de Sandt.

Additional information

Department of Cardiology, Medical Center Alkmaar, Alkmaar, the Netherlands

Rights and permissions

Reprints and permissions

About this article

Cite this article

van de Sandt, F.M., Jansen, R., Kimman, G.P. et al. Ventricular asystole due to inhibition of ventricular backup pacing by a dual chamber pacemaker: possible pitfalls of default settings . NHJL 18, 323–326 (2010). https://doi.org/10.1007/BF03091784

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF03091784

Navigation