Abstract

Problem

To design a gantry-type beam delivery system for heavy-ion therapy beams. In order to allow for installation of the gantry in hospital environment, small overall gantry size and low weight are required. The isocentric configuration is to be preferred.

Sulution

A new gantry concept with oblique output beam is introduced. Transformations between the treatment angle, angles of gantry and patient table rotations are derived.

Results and Conclusions

The isocentric gantry with the overall radius of 2.8 m is feasible using the 60° output beam. The oblique output beam imposes some restrictions on gantry capabilities, which has to be carefully evaluated by radiation oncology community.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Carli Ch, Rocher Ch, Fietier N, et al. EULIMA — Preliminary design of the gantry for the EULIMA Project. March 1992; CERN: CH-1211 Geneva 23, Switzerland.

  2. Janik J, Müller M. Some remarks on the ion optics of beam delivery systems for tumor treatment. Nucl Instr and Meth 1994;B84:117–9.

    Article  CAS  Google Scholar 

  3. Kats MM, Onosovski KK. A planar magnetooptical system for the irradiation of a lying patient with a proton beam from various directions. Instruments and Experimental Techniques 1996;39/1:127–31.

    Google Scholar 

  4. Kats MM, Onosovski KK. A simple, compact, flat system for the irradiation of a lying patient with a proton beam from different directions. Instruments and Experimental Techniques 1996;39/1:132–4.

    Google Scholar 

  5. Pavlovic M. Beam-optics study of the gantry beam delivery system for light-ion cancer therapy. Nucl Instr and Meth 1997;A399:439–54.

    Article  CAS  Google Scholar 

  6. Pavlovic M. Gantries for light-ion cancer therapy. In: Kraft G, Langbein K (eds.). Book of Abstracts 6th Workshop on Heavy-Charged Particles in Biology and Medicine; September 29, 30–October 1, 1997; Baveno, Italy:I5.

  7. Pavlovic M. Recent development of gantry design activities at GSI Darmstadt. Proc. 5th Int. European Particle Accelerator Conference EPAC96 June 10–14 1996; Vol. 3; Sitges (Barcelona), Spain: 2665–2667.

  8. Pavlovic M. An alternative solution of a gantry for light-ion cancer therapy. GSI-Preprint-97-50; September 1997.

  9. Vorobiev LG, Wollnik H, Winkler M. Ion optical design for a gantry for heavy ion tumor treatment. GSI-Report-97-06; April 1997.

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Márius Pavlovic.

Additional information

on leave from the Slovak Institute of Metrology

Rights and permissions

Reprints and permissions

About this article

Cite this article

Pavlovic, M. Oblique gantry — An alternative for heavy-ion cancer therapy. Strahlenther Onkol 175 (Suppl 2), 24–26 (1999). https://doi.org/10.1007/BF03038881

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF03038881

Key Words

Navigation