Abstract
Problem
To design a gantry-type beam delivery system for heavy-ion therapy beams. In order to allow for installation of the gantry in hospital environment, small overall gantry size and low weight are required. The isocentric configuration is to be preferred.
Sulution
A new gantry concept with oblique output beam is introduced. Transformations between the treatment angle, angles of gantry and patient table rotations are derived.
Results and Conclusions
The isocentric gantry with the overall radius of 2.8 m is feasible using the 60° output beam. The oblique output beam imposes some restrictions on gantry capabilities, which has to be carefully evaluated by radiation oncology community.
Similar content being viewed by others
References
Carli Ch, Rocher Ch, Fietier N, et al. EULIMA — Preliminary design of the gantry for the EULIMA Project. March 1992; CERN: CH-1211 Geneva 23, Switzerland.
Janik J, Müller M. Some remarks on the ion optics of beam delivery systems for tumor treatment. Nucl Instr and Meth 1994;B84:117–9.
Kats MM, Onosovski KK. A planar magnetooptical system for the irradiation of a lying patient with a proton beam from various directions. Instruments and Experimental Techniques 1996;39/1:127–31.
Kats MM, Onosovski KK. A simple, compact, flat system for the irradiation of a lying patient with a proton beam from different directions. Instruments and Experimental Techniques 1996;39/1:132–4.
Pavlovic M. Beam-optics study of the gantry beam delivery system for light-ion cancer therapy. Nucl Instr and Meth 1997;A399:439–54.
Pavlovic M. Gantries for light-ion cancer therapy. In: Kraft G, Langbein K (eds.). Book of Abstracts 6th Workshop on Heavy-Charged Particles in Biology and Medicine; September 29, 30–October 1, 1997; Baveno, Italy:I5.
Pavlovic M. Recent development of gantry design activities at GSI Darmstadt. Proc. 5th Int. European Particle Accelerator Conference EPAC96 June 10–14 1996; Vol. 3; Sitges (Barcelona), Spain: 2665–2667.
Pavlovic M. An alternative solution of a gantry for light-ion cancer therapy. GSI-Preprint-97-50; September 1997.
Vorobiev LG, Wollnik H, Winkler M. Ion optical design for a gantry for heavy ion tumor treatment. GSI-Report-97-06; April 1997.
Author information
Authors and Affiliations
Corresponding author
Additional information
on leave from the Slovak Institute of Metrology
Rights and permissions
About this article
Cite this article
Pavlovic, M. Oblique gantry — An alternative for heavy-ion cancer therapy. Strahlenther Onkol 175 (Suppl 2), 24–26 (1999). https://doi.org/10.1007/BF03038881
Issue Date: