Abstract
A new type of oxygen analyzer was constructed utilizing the electrochemical oxygen pump: Pt, air/ZrO2 (+ CaO)/Ar + H2 + H20 gas mixture, Pt. Copper, indium, or tin samples were melted in a circulating Ar + H2 gas mixture; the oxygen content was then determined within 3 minutes from the change in the electrical current passing through the oxygen pump. The analytical values were highly consistent and reliable despite the fairly low oxygen concentrations in the samples. Neither standard samples nor the establishment of blank conditions was necessary. This type of oxygen analyzer appears to be useful for determining oxygen concentrations less than 1 wt ppm.
Similar content being viewed by others
References
Y. Inoue:Collected Abstracts of the 1983 Fall Meeting of Jpn. Inst. Met., 1983, p. 193.
H. Kimura:Collected Abstracts of the 1983 Fall Meeting of Jpn. Inst. Met., 1983, p. 195.
Y. Nakamura:Kinzokuzairyo to Chobiryobunseki, Jpn. Inst. Met., Sendai, Japan, 1983, pp. 39–43.
S. Otsuka and Z. Kozuka:Trans. Jpn. Inst. Met., 1984, vol. 25, pp. 639–48.
C. Wagner:Z. Eleclrochem., 1956, vol. 60, pp. 4–7.
C. Wagner:Z. Electrochem., 1959, vol. 63, pp. 1027–30.
J. W. Patterson, E. C. Bogren, and R. A. Rapp:J. Electrochem. Soc, 1967, vol. 114, pp. 752–58.
S. Otsuka: Master Thesis, Osaka Univ., Japan, 1972.
S. Otsuka and Z. Kozuka:Metall. Trans. B, 1979, vol. 10B, pp. 565–74.
S. Otsuka, T. Sano, and Z. Kozuka:Metall. Trans. B, 1980, vol. 11B, pp. 313–19.
S. Otsuka, Y. Matsumura, and Z. Kozuka:Solid State Ionics, 1981, vol. 3/4, pp. 495–98.
K. Kiukkola and C. Wagner:J. Electrochem. Soc, 1957, vol. 104, pp. 379–87.
O. Kubaschewski and C. B. Alcock:Metallurgical Thermo-chemistry, 5th ed., Pergamon Press, Oxford, England, 1979, p. 380.
S. Otsuka and Z. Kozuka:Trans. Jpn. Inst. Met., 1981, vol. 22, pp. 558–66.
M. Iwase, E. Ichise, M. Takeuchi, and T. Yamasaki:Trans. Jpn. Inst, Met., 1984, vol. 25, pp. 43–52.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Otsuka, S., Kozuka, Z. A new type of oxygen analyzer utilizing a potentiostatic coulometric titration technique. Metall Trans B 16, 113–119 (1985). https://doi.org/10.1007/BF02657496
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02657496