Skip to main content
Log in

New method of determining the Q-factor of quartz-crystal resonators

  • High and Ultrahigh Frequency Measurements
  • Published:
Measurement Techniques Aims and scope

Conclusions

The measurement error of the above method does not depend on the Q-factor of the quartz resonator nor on its frequency, which makes it applicable in any frequency or Q-factor range, whereas other methods are restricted in their application either by their minimum decay time or by their minimum bandwidth.

Knowing the Q-factor, the equivalent resistance and the frequency of the quartz resonator, it is possible to calculate its equivalent inductance and dynamic capacitance. Thus, the above circuit makes it possible to determine all the parameters of a quartz resonator equivalent circuit.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Authors

Additional information

Translated from Izmeritel'naya Tekhnika, No. 11, pp. 50–51, November, 1961

Rights and permissions

Reprints and permissions

About this article

Cite this article

Novgorodov, E.D. New method of determining the Q-factor of quartz-crystal resonators. Meas Tech 4, 914–916 (1961). https://doi.org/10.1007/BF02614128