Six Hologic QDR-1000 DEXA bone densitometers at different centers across the USA were compared to determine the intermachine variability. Nine scans in succession were acquired on each machine using a single anthropomorphic lumbar spine phantom (manufactured by Hologic). Values for BMC, area, and BMD were recorded for each measurement. Means, standard deviations (SD), and coefficients of variation (CV) were calculated for each machine. All the CVs (BMC, area, BMD) were less than 1% (range 0.3%–0.6%). The CV of the means at the six sites were 0.4%, 0.6%, and 0.5% for BMC, area, and BMD, respectively. Although several significant differences for BMC, area, and BMD were noted by ANOVA between machines at different sites, the difference between the highest and lowest means of the individual machines was only 1.1%, 1.31%, and 1.07% for BMC, area, and BMD. The small variations between the DEXA systems are encouraging for researchers involved in multicenter trials in which data are pooled.
Similar content being viewed by others
References
Lindsay R, Fey C, Haboubi A (1987) Dual photon absorptiometry measurements of bone mineral increase with source life. Calcif Tiss Int 41:293
DaCosta MD, Luckey MM, Meier DE, Mandeli JP, DeLaney ML, Stritzke PH, Goldsmith SJ (1989) Effect of source strength and attenuation on dual photon absorptiometry. J Nucl Med 30:1875–1881
Ott SM (1986) Should women get screening bone mass measurements? Ann Intern Med 104:(6)874–876
Pacifici R, Rupith R, Vered I, Fischer KC, Griffin M, Susman N, Avioli LV (1988) Dual energy radiography (DER): a preliminary comparative study. Calcif Tiss Int 43:189–191
Kelly TL, Slovik DM, Schoenfeld DA, Neer RM (1983) Quantitative digital radiography versus dual photon absorptiometry of the lumbar spine. J Clin Endo Metab 4:839–844
Sartoris DJ, Resnick D (January 1988) Digital radiography may spark renewal of bone densitometry. Diag Imaging: 145–157
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Rencken, M.L., Murano, R., Drinkwater, B.L. et al. In vitro comparability of dual energy X-ray absorptiometry (DEXA) bone densitometers. Calcif Tissue Int 48, 245–248 (1991). https://doi.org/10.1007/BF02556375
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02556375