Skip to main content
Log in

Method for recording electrocardiograms with dry electrodes applied to unprepared skin

  • Published:
Medical and biological engineering Aims and scope Submit manuscript

Abstract

A plate electrode placed on a limb, while making a high-resistive contact, has a considerable capacitative contact with the skin. An electrode is described which contains its own preamplifier and which utilises this capacitance contact to allow the recording of electrocardiograms when applied to unprepared skin. The electrode has been integrated with a commercial electrocardiograph and evaluated during extensive clinical trials. A comparison is made with conventional recording techniques.

Sommaire

Une électrode en forme de plaque placée sur un membre, a un contact capacitatif considérable avec la peau tout en formant un contact à résistivité élevée. L'article décrit une électrode qui contient son propre pré-amplificateur et utilise le contact capacitatif pour permettre l'enregistrement des électrocardiogrammes lorsqu'elle est appliquée à la peau non préparée. L'électrode a été incorporée dans un électrocardiographe commercial et a été évaluée au cours d'essais cliniques rigoureux. On fait la comparaison avec les techniques conventionnelles d'enregistrement.

Zusammenfassung

Eine Scheibenelektrode, die auf einem Glied angebracht wird, hat einen wesentlichen kapazitativen Kontakt mit der Haut und stellt gleichzeitig einen stark ohmschen Kontakt her. Es wird eine Elektrode beschrieben, die einen eigenen Vorverstärker enthält und diesen kapazitativen Kontakt dazu verwendet, die Aufzeichnung des Elektrokardiogramms vorzunehmen, wenn sie auf die unvorbereitete Haut angewandt wird. Die Elektrode gehöhrt zu einem handelsüblichen Elektro kardiographen und wurde während ausgedehnter klinischer Erprobungen beurteilt. Es erfolgt ein Vergleich mit konventionellen Aufzeichnungsverfahren.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Bergey, G. E., Squires, R. D. andSipple, W. C. (1971) Electrocardiogram recording with pasteless electrodes.IEEE Trans. BME-18, 206–211.

    Google Scholar 

  • Geddes, L. A., Bourland, J. D., Smalling, R. W. andSteinberg, R. B. (1975) The use of the same pair of dry electrodes to record skin resistance and beat-by-beat heart rate.Med. & Biol. Engng. 12, 89–96.

    Article  Google Scholar 

  • Lewes, D. (1965) Multipoint electrocardiography without skin preparation.Lancet, 3rd July, 17–18.

  • Lopez, A. andRichardson, P. C. (1969) Capacitive electrocardiographic and bioelectric electrodes.IEEE Trans. BME-16, 99.

    Google Scholar 

  • Stevens, W. G. S. (1963) The current voltage relationship in human skin.Med. Electron. Biol. Eng. 1, 389–399.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Betts, R.P., Brown, B.H. Method for recording electrocardiograms with dry electrodes applied to unprepared skin. Med. & biol. Engng. 14, 313–315 (1976). https://doi.org/10.1007/BF02478127

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02478127

Keywords

Navigation