The extraction of data from an optical image by scanning

  • Published:
Medical and biological engineering Aims and scope Submit manuscript

Abstract

The Vickers Cytology Scanning Apparatus is described and critically examined to determine how it works and to discover the uses and limitations of scanning methods. Studies were made on cervical smears of the appearance of the scan, the relationship between the pulse width and the dimension of the nucleus, the range of density levels of the nucleus and cytoplasm, the slope at the edge of the nucleus, and on interference by two cells in proximity. A theoretical study of conditions at the trigger point showed that the pulse width was a function of the densities of both nucleus and cytoplasm.

Sommaire

L'appareil de Vickers pour Identification Cytologique est décrit et examiné d'une manière critique pour déterminer comment il fonctionne et pour découvrir les usages et les limitations de la méthode d'identification. Des études ont été entreprises sur des frottis du col utérin concernant l'apparence du champ de vision, le rapport entre l'envergure du pouls et la dimension du noyau, l'étendue des niveaux de la densité du noyau et du cytoplasme, l'inclinaison au bord du noyau, et sur l'interférence par deux cellules attenantes. Une étude théorique des conditions au point de la détente a montré que l'envergure du pouls était en fonction de la densité du noyau et du cytoplasme.

Zusammenfassung

Beschreibung des Vickers Cytologie-Abtastungsapparates und sorgfältige Untersuchung zur Bestimmung seiner Wirkungsweise und der Begrenzung von Abtastungs-methoden. Es wurden Studien angestellt über: zervikale Abstriche betreffend das Aussehen der Abtastung, die Beziehung zwischen Pulsweite und den Kernschatten, bezw. seine Dimensionen, die Reichweite des Dichteniveaus des Kernschattens und der Cytoplasmen, die Senkung am Rande des Kernschattens und Interferenz zweier nahe bei einander liegenden Zellen. Eine theoretische Studie über Verhältnisse am Auslöserpunkt zeigte, dass die Pulsweite eine Funktion der Dichte sowohl von Kernschatten als auch von Cytoplasmen darstellte.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Diamond, R. A. (1967) Cytological criteria applicable to automatic scanning of cervical smears for malignancy. B.Sc. Thesis, Oxford.

  • Hawksley, P. G. W., Blackett, J. H., Meyer E. W. andFitzsimmons, A. E. (1954) The design and construction of a photoelectric scanning machine for sizing microscopic particles.Br. J. appl. Phys. 5, 165–173.

    Article  Google Scholar 

  • Mansberg, H. P. andOhringer, P. (1969) Design considerations for electronic and electromechanical flying spot scanners.Ann. N.Y. Acad. Sci. 157, 1–37.

    Google Scholar 

  • Papanicolaou, G. N. (1954)Atlas of Exfoliative Cytology. Harvard University Press, Cambridge, Mass.

    Google Scholar 

  • Roman, T. N. andLatour, J. P. A. (1967) The effect of early diagnosis on survival statistics in carcinoma of the uterine cervix.Am. J. Obstet. Gynaec. 97, 739–749.

    Google Scholar 

  • Tolles, W. E., Horvath W. J. andBostrom, R. C. (1961) A study of the quantitative characteristics of exfoliated cells from the female genital tract—I. Measurements, methods and results.Cancer 14, 437–454.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Burn, G.P. The extraction of data from an optical image by scanning. Med. & biol. Engng. 9, 87–95 (1971). https://doi.org/10.1007/BF02474737

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02474737

Keywords

Navigation