An electromagnetic flowmeter for simultaneous measurements of pulmonary arterial and aortic blood flows in the conscious animal

  • Published:
Medical electronics and biological engineering Aims and scope Submit manuscript

Abstract

An electromagnetic flowmeter apparatus for simultaneous measurements of pulmonary arterial and aortic blood flows in the conscious animal is described. The apparatus retains the basic features of the sine wave flowmeter and coreless flow transducers developed byKolin and provides for recordings from two adjacent transducers. The proximity of the implanted transducers results in a constant interaction which alters the techniques for calibration and zero flow determination which have been reported for single-flow measurements. The apparatus has linear frequency and phase shift characteristic to ensure undistorted flow measurement up to 50 c/s.

Résumé

On décrit un appareil électromagnétique permettant de mesurer simultanément le courant sanguin dans l'artère pulmonaire et l'aorte d'animaux non anesthésiés. Cet appareil réunit les propriétés principales d'un débitmètre à ondes sinusoïdes, à celles d'un transducteur de courant sans noyau d'après Kolin, et il permet d'enregistrer les valeurs provenants de deux transducteurs adjacents. Les transducteurs implantés étant proches l'un de l'autre, il en résulte une interaction constante qui influence la méthode d'étalonnage et la détermination du courant zéro, méthodes qui ont été décrites pour la mensuration de courants isolés. L'appareil a des propriétés linéaires de changement de fréquence et de phase ce qui assure un mesurage sans déformation du courant jusqu'à 50 c/s.

Zusammenfassung

Eine Apparatur für elektromagnetische Strömungsmessung zur gleichzeitigen Bestimmung der Strömungsgeschwindigkeiten in der arteria pulmonalis und der Aorta beim nicht narkotisierten Tier wird beschrieben. Der Apparat vereinigt in sich die wesentlichen Eigenschaften eines Sinuswellen Strömungsmessers und des kernlosen Strömungsüberträgers in der Ausführung nach Kolin, und gestattet die Messwerte von zwei nebeneinanderliegenden Überträgern aufzuzeichnen. Die Nachbarschaft zweier implantierter Überträgerführt zu dauernder gegenseitiger Beeinflussung, wodurch sich die Methoden der Tarierung und der Bestimmung des Nullflusses verändern, die für einfache Strömungsmessungen angegeben wurden. Der Apparat arbeitet mit linearer Frequenz und hat einen Phasengang, der eine störungsfreie Strömungsmessung bis hinauf zu 50c/s gestattet.

Резюме

Описывается аппаратура для электромагнитного измерения потока и одновременного определения скорости потока крови в артериях пульмоналис и аорта у ненаркотизированного животного. В этом аппарате совмещены основные свойства синусно-волнового потокомера и бесстержневого передатчика поточноизмерительных данных, в исполнении по Колину, и он разрешает производить регистрацию измеренных величин от двух расположенных рядом передатчиков. Соседство двух имплантированных передатчиков ведет к оказанию непрерывного взаимного влияния между последними, вследствие чего изменяются методы эталонирования и определения нулевого течения, которые задавались при простых измерениях потока. Аппарат работает на линейной частоте, и имеет фазный пропуск, разрешающий бесперебойное измерение потока вплоть до 50 гц.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Blood Flowmeters Symposium (1959)IRE Trans. med. Electron.,6, 207–251.

    Google Scholar 

  • Ferguson, D. J. andWells, H. S. (1959) Frequencies in pulsatile flow and response of magnetic meter.Circulat. Res.,7, 336–341.

    Google Scholar 

  • Gessner, U. (1961) Effects of the vessel wall on electromagnetic flow meters.Biophys. J.,1, 627–637.

    Article  Google Scholar 

  • James, W. G. (1951) An induction flowmeter design suitable for radioactive liquids.Rev. sci. Instrum.,22, 989–1002.

    Article  Google Scholar 

  • Kolin, A. (1936) An electromagnetic flowmeter, principle of the method and its application to blood flow measurement.Proc. Soc. exp. Biol., N.Y.,35, 53–56.

    Google Scholar 

  • Kolin, A. (1938) Flow of blood in the carotid artery of the dog under various circumstances as determined with the electromagnetic flowmeter.Amer. J. Physiol.,22, 788–804.

    Google Scholar 

  • Kolin, A. (1945) An alternating field induction flow meter of high sensitivity.Rev. sci. Instrum.,16, 109–116.

    Article  Google Scholar 

  • Kolin, A. (1959) Electromagnetic blood flow meters.Science,130, 1088–1097.

    Google Scholar 

  • Kolin, A. andKado, R. T. (1959) Miniaturization of the electromagnetic blood flow meter and its use for the recording of circulatory responses of conscious animals to sensory stimuli.Proc. nat. Acad. Sci., Wash.,45, 1312–1321.

    Article  Google Scholar 

  • Olmsted, F. (1959) Measurement of cardiac output in unrestrained dogs by an implanted electromagnetic meter.IRE Trans. med. Electron.,6, 210–213.

    Article  Google Scholar 

  • Shirer, H. W., Shackleford, R. B. andJochim, K. E. (1959) A magnetic flowmeter for recording cardiac output.IRE Trans. med. Electron.,6, 232–234.

    Google Scholar 

  • Weterer, E. (1937) Eine neue Methode zur Registrierung der Blutströmungsgeschwindigkeit am uneröffneten Gefäss.Z. Biol.,98, 26–36.

    Google Scholar 

  • Wyatt, D. G. (1961) Problems in the measurement of blood flow by magnetic induction.Phys. in Med. Biol.,5, 289–320 and 369–399.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

This work was supported by grants HTS-5251 and H-6285 from the United States Public Health Service.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Elliott, S.E., Hoffman, J.I.E. & Guz, A. An electromagnetic flowmeter for simultaneous measurements of pulmonary arterial and aortic blood flows in the conscious animal. Med. Electron. Biol. Engng 1, 323–332 (1963). https://doi.org/10.1007/BF02474416

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02474416

Keywords

Navigation